Pelvic pitch and roll during total hip arthroplasty performed through a posterolateral approach. A potential source of error in free-hand cup positioning

  • Original Paper
  • Published:
International Orthopaedics Aims and scope Submit manuscript

Abstract

Purpose

Intraoperative pelvic motion can alter the perceived cup inclination and version during non-navigated THA. We quantified pelvic motion during different phases of primary THA performed in the lateral decubitus through a posterolateral approach.

Methods

Pelvic roll (rotation of the coronal plane) and pitch angles (rotation parallel to the coronal plane) were studied in 75 patients undergoing THA for osteoarthritis by four arthroplasty surgeons. Ten steps of surgery were defined. Angular motion was recorded with a miniature surgical device that utilizes inertial sensors.

Results

The mean absolute roll ranged from 0.03° detected at the end of surgery to 4.13° detected during acetabular exposure. The mean absolute pitch ranged from 0.05° detected at the end of surgery to 2.54° detected during hip dislocation. The maximum pelvic roll and pitch detected during surgery averaged 17.62° (SD: 5.08) and 9.3° (SD: 3.39) respectively. Absolute roll and pitch angles were not affected by patient’s BMI, sex, pre-operative hip motion, or surgeon. Before cup insertion, the greatest mean change in roll was observed during acetabular exposure (10.02° anteriorly), and for pitch was observed during dislocation (1.88° caudally).

Conclusion

During THA performed through a posterolateral approach, there is a progressive anterior pelvic roll that peaks before cup insertion. This can lead to underestimation of cup anteversion during non-navigated THA. The anterior roll does not completely correct, even when all retractors and external forces acting on the pelvis are removed. Pelvic pitch that could affect the perceived cup inclination occurs to a lesser extent than pelvic roll.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA (2005) Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplast 20(4):427–435

    Article  Google Scholar 

  2. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2):217–220

    Article  CAS  Google Scholar 

  3. Wan Z, Boutary M, Dorr LD (2008) The influence of acetabular component position on wear in total hip arthroplasty. J Arthroplast 23(1):51–56. https://doi.org/10.1016/j.arth.2007.06.008

    Article  Google Scholar 

  4. Miettinen S, Mäkinen TJ, Laaksonen I, Mäkelä K, Huhtala H, Kettunen J, Remes V (2017) Early aseptic loosening of cementless monoblock acetabular components. Int Orthop 41:715–722

    Article  PubMed  Google Scholar 

  5. Gonzalez Della Valle A, Sharrock NE, Barlow M, Caceres L, Go G, Salvati EA (2016) The modern, hybrid total hip arthroplasty for primary osteoarthritis at Hospital for Special Surgery. Bone Joint J 98(Suppl A):54–59

    Article  PubMed  Google Scholar 

  6. McArthur B, Vulcano E, Cross M, Gonzalez Della Valle A, Salvati EA (2014) Acetabular component orientation in total hip arthroplasty: the impact of obesity. Hip Int 24(3):263–269

    Article  PubMed  Google Scholar 

  7. Rojas J, Bautista M, Bonilla G, Amado O, Huerfano E, Monsalvo D, Llinas A, Navas J (2018) A retrospective study on the relationship between altered native acetabular angle and vertical implant positioning. Int Orthop 42:769–775

    Article  PubMed  Google Scholar 

  8. Meermans G, Van Doorn WJ, Koenraadt K, Kats J (2014) The use of the transverse acetabular ligament for determining the orientation of the components in total hip replacement a randomized controlled trial. Bone Joint J 96-B(3):312–318. https://doi.org/10.1302/0301-620X.96B3.32989

    Article  CAS  PubMed  Google Scholar 

  9. Padgett DE, González Della Valle A (2006) The uncemented acetabular component. In: Barrack RL, Rosenberg AG (eds) Master techniques in orthopaedic surgery. The hip, 1st edn. Lippincott Williams & Wilkins, Philadelphia, pp 283–294

    Google Scholar 

  10. Schwarzkopf R, Muir JM, Paprosky WG, Seymour S, Cross MB, Vigdorchik JM (2017) Quantifying pelvic motion during total hip arthroplasty using a new surgical navigation device. J Arthroplast 32(10):3056–3060. https://doi.org/10.1016/j.arth.2017.04.046

    Article  Google Scholar 

  11. Cross MB, Schwarzkopf R, Miller TT, Bogner EA, Muir JM, Vigdorchik JM (2018) Improving registration accuracy during total hip arthroplasty: a cadaver study of a new, 3-D mini-optical navigation system. Hip Int 28(1):33–39. https://doi.org/10.5301/hipint.5000533

    Article  PubMed  Google Scholar 

  12. Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 51:737–755

    Article  CAS  Google Scholar 

  13. Najarian BC, Kilgore JE, Markel DC (2009) Evaluation of component positioning in primary total hip arthroplasty using an imageless navigation device compared with traditional methods. J Arthroplast 24(1):15–21. https://doi.org/10.1016/j.arth.2008.01.004

    Article  Google Scholar 

  14. Boylan M, Suchman K, Vigdorchik J, Slover J, Bosco J (2018) Technology-assisted hip and knee arthroplasties: an analysis of utilization trends. J Arthroplast 33(4):1019–1023. https://doi.org/10.1016/j.arth.2017.11.033

    Article  Google Scholar 

  15. Grammatopoulos G, Pandit HG, da Assunção R, Taylor A, McLardy-Smith P, De Smet KA, Murray DW, Gill HS (2014) Pelvic position and movement during hip replacement. Bone Joint J 96-B(7):876–883. https://doi.org/10.1302/0301-620X.96B7.32107

    Article  CAS  PubMed  Google Scholar 

  16. Asayama I, Akiyoshi Y, Naito M, Ezoe M (2004) Intraoperative pelvic motion in total hip arthroplasty. J Arthroplast 19(8):992–997

    Article  Google Scholar 

  17. Ezoe M, Naito M, Asayama I, Ishiko T, Fujisawa M (2005) Pelvic motion during total hip arthroplasty with translateral and posterolateral approaches. J Orthop Sci 10(2):167–172

    Article  PubMed  Google Scholar 

  18. Ji W, Stewart N (2016) Fluoroscopy assessment during anterior minimally invasive hip replacement is more accurate than with the posterior approach. Int Orthop 40:21–27

    Article  PubMed  Google Scholar 

  19. Kawarai Y, Iida S, Nakamura J, Shinada Y, Suzuki C, Ohtori S (2017) Does the surgical approach influence the implant alignment in total hip arthroplasty? Comparative study between the direct anterior and the anterolateral approaches in the supine position. Int Orthop 41:2487–2493

    Article  PubMed  Google Scholar 

  20. Kobayashi H, Homma Y, Baba T, Ochi H, Matsumoto M, Yuasa T, Kaneko K (2016) Surgeons changing the approach for total hip arthroplasty from posterior to direct anterior with fluoroscopy should consider potential excessive cup anteversion and flexion implantation of the stem in their early experience. Int Orthop 40:1813–1819

    Article  PubMed  Google Scholar 

  21. Zhu J, Wan Z, Dorr LD (2010) Quantification of pelvic tilt in total hip arthroplasty. Clin Orthop 468(2):571–575. https://doi.org/10.1007/s11999-009-1064-7

    Article  PubMed  Google Scholar 

  22. Bayraktar V, Weber M, von Kunow F, Zeman F, Craiovan B, Renkawitz T, Grifka J, Woerner M (2017) Accuracy of measuring acetabular cup position after total hip arthroplasty: comparison between a radiographic planning software and three-dimensional computed tomography. Int Orthop 41:731–738

    Article  PubMed  Google Scholar 

  23. Craiovan B, Renkawitz T, Weber M, Grifka J, Nolte L, Zheng G (2014) Is the acetabular cup orientation after total hip arthroplasty on a two dimension or three dimension model accurate? Int Orthop 38:2009–2015

    Article  PubMed  Google Scholar 

  24. Opperer M, DeNally F, Lee YY, Gonzalez Della Valle A (2016) A critical analysis at factors contributing to dislocation following total hip arthroplasty. Int Orthop 40(4):703–708

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

We are grateful to Jeffrey M. Muir, MSc, DC, MSc (Clin Epi) and Luke Becker, BASc from Intellijoint Surgical, Inc. for their assistance in data extraction and graphic preparation.

Funding

This study was partially funded by Intellijoint Surgical, Glenn Bergenfiend and The Sidney Milton and Leoma Simon Foundation.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Alejando Gonzalez Della Valle.

Ethics declarations

Conflict of interest statement

Funding for this study has been received by Intellijoint Surgical, Inc.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gonzalez Della Valle, A., Shanaghan, K., Benson, J.R. et al. Pelvic pitch and roll during total hip arthroplasty performed through a posterolateral approach. A potential source of error in free-hand cup positioning. International Orthopaedics (SICOT) 43, 1823–1829 (2019). https://doi.org/10.1007/s00264-018-4141-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00264-018-4141-2

Keywords

Navigation