Abstract
Our aim is to compare the radiation dose associated with a low-dose CT colonography (CTC) protocol for colorectal cancer screening with that delivered by double-contrast barium enema (DCBE). CTC of twenty asymptomatic individuals (M:F = 10:10) participating to a colorectal cancer screening program and DCBE of fifteen patients (M:F = 6:9) were evaluated. For CTC, absorbed dose was determined by calculating the dose-length product for each CTC examination from measurements on a CT dose phantom equipped with a CT ion chamber. For DCBE, the free-in-air Kerma at the patient’s X-ray entry surface and the Kerma-area product during fluoroscopy and fluorography were measured with a Barracuda system, with fluoroscopy times being recorded blinded to the performing operator. Effective dose at CTC was 2.17 ± 0.12 mSv, with good and excellent image quality in 14/20 (70%) and 6/20 cases (30%), respectively. With DCBE, effective patient dose was 4.12 ± 0.17 mSv, 1.9 times greater than CTC (P < 0.0001). Our results show that effective dose from screening CTC is substantially lower than that from DCBE, suggesting that CTC is the radiological imaging technique of the large bowel with the lowest risk of stochastic radiation effects.
Similar content being viewed by others
References
Jemal A, Murray T, Ward E, et al. (2005) Cancer statistics, 2005. CA Cancer J Clin 55(1):10–30
Nelson NJ (2008) Virtual colonoscopy accepted as primary colon cancer screening test. J Natl Cancer Inst 100(21):1492–1499
Brenner DJ, Georgsson MA (2005) Mass screening with CT colonography: should the radiation exposure be of concern? Gastroenterology 129(1):328–337
International Commission on Radiological Protection (1991) 1990 recommendations of the international commission of radiological protection, publication no. 60. In: Smith H, (ed). Annals of the ICRP, vol 21. Oxford: Pergamon Press, pp 1–3
Hospital Physicists’ Association (Diagnostic Radiology Topic Group) (1977) The physics of radiodiagnostics: scientific report series, vol 6. London
Hirofuji Y, Aoyama T, Koyama S, Kawaura C, Fujii K (2009) Evaluation of patient dose for barium enemas and CT colonography in Japan. Br J Radiol 82:219–227
Kemerink GJ, Borstlap ACW, Frantzen MJ, et al. (2001) Patient and occupational dosimetry in double contrast barium enema examinations. Br J Radiol 74:420–428
Graser A, Wintersperger BJ, Suess C, Reiser MF, Becker CR (2006) Dose reduction and image quality in MDCT colonography using tube current modulation. AJR Am J Roentgenol 187:695–701
Macari M, Bini EJ, Xue X, et al. (2002) Colorectal neoplasms: prospective comparison of thin-section low-dose multi-detector row CT colonography and conventional colonoscopy for detection. Radiology 224:383–392
Schopphoven S, Faulkner K, Busch HP (2008) Assessment of patient organ dose in CT virtual colonoscopy for bowel cancer screening. Radiat Prot Dosimetry 129:179–183
Sugiyama A, Ohashi Y, Gomi A, et al. (2007) Colorectal screening with single scan CT colonography in children. Pediatr Surg Int 23:987–990
Wang J, Wang S, Li L, et al. (2008) Virtual colonoscopy screening with ultra low-dose CT and less-stressful bowel preparation: a computer simulation study. IEEE Trans Nucl Sci 55:2566–2675
Branschofsky M, Vogt C, Aurich V, et al. (2006) Feasibility of ultra-low-dose multi-detector-row CT-colonography: detection of artificial endoluminal lesions in an in vitro-model with optimization of image quality using a noise reduction filter algorithm. Eur J Med Res 11:13–19
Taylor SA, Halligan S, Bartram CI, et al. (2003) Multi-detector row CT colonography: effect of collimation, pitch, and orientation on polyp detection in a human colectomy specimen. Radiology 229:109–118
Laghi A, Iannaccone R, Mangiapane F, et al. (2003) Experimental colonic phantom for the evaluation of the optimal scanning technique for CT colonography using a multidetector spiral CT equipment. Eur Radiol 13:459–466
McCollough CH (2002) Optimization of multidetector array CT acquisition parameters for CT colonography. Abdom Imaging 27:253–259
Dalrymple NC, Prasad SR, El-Merhi FM, Chintapalli KN (2007) Price of isotropy in multidetector CT. Radiographics 27:49–62
Luz O, Buchgeister M, Klabunde M, et al. (2007) Evaluation of dose exposure in 64-slice CT colonography. Eur Radiol 17:2616–2621
Jensch S, van Gelder RE, Venema HW, et al. (2006) Effective radiation doses in CT colonography: results of an inventory among research institutions. Eur Radiol 16:981–987
Liedenbaum MH, Venema HW, Stoker J (2008) Radiation dose in CT colonography—trends in time and differences between daily practice and screening protocols. Eur Radiol 18:2222–2230
Geleijns J, Broerse JJ, Shaw MP, et al. (1997) Patient dose due to colon examination: dose assessment and results from a survey in the Netherlands. Radiology 204:553–559
Yanch JC, Behrman RH, Hendricks MJ, McCall JH. Increased radiation dose to overweight and obese patients from radiographic examinations. Radiology 2009. doi:10.1148/radiol.2521080141
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Neri, E., Faggioni, L., Cerri, F. et al. CT colonography versus double-contrast barium enema for screening of colorectal cancer: comparison of radiation burden. Abdom Imaging 35, 596–601 (2010). https://doi.org/10.1007/s00261-009-9568-x
Received:
Accepted:
Published:
Issue Date: