Abstract
A rapid microfluidic based capillary electrophoresis immunoassay (CEIA) was developed for on-line monitoring of glucagon secretion from pancreatic islets of Langerhans. In the device, a cell chamber containing living islets was perfused with buffers containing either high or low glucose concentration. Perfusate was continuously sampled by electroosmosis through a separate channel on the chip. The perfusate was mixed on-line with fluorescein isothiocyanate-labeled glucagon (FITC-glucagon) and monoclonal anti-glucagon antibody. To minimize sample dilution, the on-chip mixing ratio of sampled perfusate to reagents was maximized by allowing reagents to only be added by diffusion. Every 6 s, the reaction mixture was injected onto a 1.5-cm separation channel where free FITC-glucagon and the FITC-glucagon–antibody complex were separated under an electric field of 700 V cm−1. The immunoassay had a detection limit of 1 nM. Groups of islets were quantitatively monitored for changes in glucagon secretion as the glucose concentration was decreased from 15 to 1 mM in the perfusate revealing a pulse of glucagon secretion during a step change. The highly automated system should be enable studies of the regulation of glucagon and its potential role in diabetes and obesity. The method also further demonstrates the potential of rapid CEIA on microfluidic systems for monitoring cellular function.
Similar content being viewed by others
References
El-Ali J, Sorger PK, Jensen KF (2006) Nature 442:403–411
Wlodkowic D, Cooper JM (2010) Curr Opin Chem Biol 14:556–567
Salieb-Beugelaar GB, Simone G, Arora A, Philippi A, Manz A (2010) Anal Chem 82:4848–4864
Cheng W, Klauke N, Smith G, Cooper JM (2010) Electrophoresis 31:1405–1413
Clark AM, Sousa KM, Chisolm CN, MacDougald OA, Kennedy RT (2010) Anal Bioanal Chem 397:2939–2947
Easley CJ, Rocheleau JV, Head WS, Piston DW (2009) Anal Chem 81:9086–9095
Huang WH, Cheng W, Zhang Z, Pang DW, Wang ZL, Cheng JK, Cui DF (2004) Anal Chem 76:483–488
Goto M, Sato K, Murakami A, Tokeshi M, Kitamori T (2005) Anal Chem 77:2125–2131
Roper MG, Shackman JG, Dahlgren GM, Kennedy RT (2003) Anal Chem 75:4711–4717
Li MW, Spence DM, Martin RS (2005) Electroanal 17:1171–1180
Tokuyama T, Fujii S, Sato K, Abo M, Okubo A (2005) Anal Chem 77:3309–3314
Shackman JG, Dahlgren GM, Peters JL, Kennedy RT (2005) Lab Chip 5:56–63
Reid KR, Kennedy RT (2009) Anal Chem 81:6837–6842
Zhang XY, Grimley A, Bertram R, Roper MG (2010) Anal Chem 82:6704–6711
Schultz NM, Kennedy RT (1993) Anal Chem 65:3161–3165
Dishinger JF, Reid KR, Kennedy RT (2009) Anal Chem 81:3119–3127
Norman AW, Litwack G (1997) Hormones, 2nd edn. Academic, San Diego, CA
D'Alessio D (2011) Diabetes Obes Metab 13:126–132
Gromada J, Franklin I, Wollheim CB (2007) Endocr Rev 28:84–116
Bagger JI, Knop FK, Holst JJ, Vilsboll T (2011) Diabetes Obes Metab 13:965–971
Salehi A, de la Cour CD, Hakanson R, Lundquist I (2004) Regul Pept 118:143–150
Göpel S, Zhang Q, Eliasson L, Ma X-S, Galvanovskis J, Kanno T, Salehi A, Rorsman P (2004) J Physiol 556:711–726
Hentz NG, Richardson JM, Sportsman JR, Daijo J, Sittampalam GS (1997) Anal Chem 69:4994–5000
Jacobson SC, Ermakov SV, Ramsey JM (1999) Anal Chem 71:3273–3276
Shackman JG, Watson CJ, Kennedy RT (2004) J Chromatogr A 1040:273–282
Edwards JL, Kennedy RT (2005) Anal Chem 77:2201–2209
German I, Kennedy RT (2000) J Chromatogr B 742:353–362
Guillo C, Roper MG (2011) J Chromatogr A 1218:4059–4064
Franklin I, Gromada J, Gjinovci A, Theander S, Wollheim CB (2005) Diabetes 54:1808–1815
Walker JN, Ramracheya R, Zhang Q, Johnson PRV, Braun M, Rorsman P (2011) Diabetes Obes Metab 13:95–105
Ortega FJ, Moreno-Navarrete JM, Sbataer M, Ricart W, Fruhbeck G, Fernandez-Real JM (2011) Euro J Endocrin 165:639–645
Acknowledgments
This work was supported by National Institutes of Health grant R37 DK46960. J.G.S. was supported by an Eli Lilly & Co. fellowship in association with the University of Michigan. K.R.R. was partially supported by a U.S. Dept. of Education Graduate Assistance in Areas of National Need fellowship.
Author information
Authors and Affiliations
Corresponding author
Electronic supplementary material
Below is the link to the electronic supplementary material.
Fig. S1
(PDF 19 kb)
Rights and permissions
About this article
Cite this article
Shackman, J.G., Reid, K.R., Dugan, C.E. et al. Dynamic monitoring of glucagon secretion from living cells on a microfluidic chip. Anal Bioanal Chem 402, 2797–2803 (2012). https://doi.org/10.1007/s00216-012-5755-7
Received:
Revised:
Accepted:
Published:
Issue Date: