Introduction

Extracorporeal carbon dioxide removal (ECCO2R), or “partial ECMO”, is a form of extracorporeal life support therapy which has received renewed interest as an adjunct or alternative to invasive mechanical ventilatory support for patients suffering from reversible acute hypercapnic respiratory lung failure [1]. A clinical need for alternative respiratory support therapies providing carbon dioxide (CO2) removal exists for both patients experiencing acute exacerbation of chronic obstructive pulmonary disorder (COPD) as well as those with acute respiratory distress syndrome. In acute exacerbation of COPD, patients who do not respond effectively to drug therapy exhibit increasing respiratory acidosis and hypercapnia, while remaining normoxic due to supplemental oxygen [2]. Despite the advent and effectiveness of non-invasive ventilation, invasive mechanical ventilation is still required by 26 % to 54 % of patients [3, 4]. The prognosis for COPD patients requiring invasive mechanical ventilation for ventilatory failure is poor, with hospital survival ranging from 76 % to as low as 31 % [59]. In respiratory distress syndrome, lung protective ventilation using lower tidal volumes may be complicated by significant CO2 retention, which can lead to arrhythmias and severe dyspnea [10].

Advances in gas exchanger, pump and catheter technologies, in conjunction with the recognized utility of lower flow rate venovenous support modes, have reduced the risks associated with extracorporeal life support, thus opening the door for consideration of its use in earlier stages of acute lung failure [1114]. The Hemolung respiratory assist system (ALung Technologies, Pittsburgh, PA) is a novel artificial lung device which provides extracorporeal CO2 removal with a cannula and at blood flow rates similar to those used in continuous renal replacement therapy. This device was designed to provide partial respiratory support through a single 15.5-Fr venovenous catheter at blood flow rates of 350–450 ml/min. The system is unique in that it uses active mixing of the blood adjacent to the hollow fiber membranes (HFM) to increase its gas exchange efficiency, allowing a smaller fiber surface area and reduced flow rates. The levels of oxygenation provided are secondary and recognized to be subtherapeutic.

The purpose of the in vivo study presented in this paper was to demonstrate the safety and performance of the Hemolung for extracorporeal CO2 removal in standing, awake sheep over an 8-day period. The design and methodology of this study was developed to support regulatory approval for clinical testing of the Hemolung. The objective in using healthy sheep was both to maximize the consistency in the length of the study, and to allow evaluation of safety independent of risks associated with disease states or mechanical ventilation. Based on its design and performance in vitro, it was hypothesized that the Hemolung could provide consistent CO2 removal exceeding 50 ml/min at an arterial CO2 partial pressure (PCO2) of 45 mmHg and at blood flow rates in the range 350–450 ml/min. It was further hypothesized that this level of CO2 exchange could be provided without failure or clinically significant adverse reaction in healthy awake sheep for an extended period of 8 days.

Materials and methods

Animal selection, preparation and baseline assessment

This study was approved by the University of Pittsburgh Institutional Animal Care and Use Committee, and the care and handling of the animals were conducted in accordance with the standards set forth in the USDA Animal Welfare Act and in the current guide for the care and use of laboratory animals prepared by the Institute of Laboratory Animal Resources, National Academy of Sciences. Eight Dorset crossbred sheep weighing 49–115 kg were used for the study. Prior to surgery, the animals were premedicated with atropine (0.05 mg/kg, intramuscular) followed by initial anesthetic induction with ketamine hydrochloric acid (30 mg/kg, intramuscular). Endotracheal intubation was achieved under direct visualization. A surgical plane of anesthesia was achieved using isoflurane (1–3 % to effect) followed by the placement of an oral–gastric tube for abdominal decompression. Monitoring lines were placed in the left external jugular vein and carotid artery. Baseline assessment included arterial pressure, heart rate, blood gases, blood chemistry, complete blood count, activated partial thromboplastin time (aPTT), and plasma-free hemoglobin (PfHb).

Description of Hemolung respiratory assist system

The primary component of the Hemolung system is a cartridge which houses a cylindrical bundle of siloxane and heparin-coated HFM positioned around a spinning core which simultaneously drives blood flow (centrifugally) through the cartridge and CO2 removal from the blood into the oxygen sweep gas flowing under negative pressure through the fiber lumens (Fig. 1b). The total fiber surface area of the gas exchange cartridge is 0.59 m2. The cartridge is magnetically mated to and driven by an integrated control console (Fig. 1a). Device performance is controlled from the console monitor using two parameters: rotational speed of the cartridge core (rpm) and sweep gas flow (l/min) through the fibers. The system was interfaced with the sheep through a custom dual lumen 15.5-Fr catheter (Fig. 1c) similar to a dialysis catheter, but with lower flow resistance and greater kink resistance. The volume of the circuit is 280 ml and is primed with sterile 0.9 % sodium chloride containing one unit of heparin per milliliter immediately prior to the procedure.

Fig. 1
figure 1

Hemolung respiratory assist system: a controller with cartridge magnetically mounted; b schematic of integral gas exchange/pump unit; c dual lumen catheter

Surgical procedure

Following placement of monitoring lines and baseline assessments, the animals were rotated to a left lateral decubitus position. The Hemolung catheter was inserted using a percutaneous Seldinger technique [15]. Following guide wire placement and vessel dilation, a heparin bolus (80 U/kg) was administered. The Hemolung catheter was then advanced into the vasculature under the guidance of fluoroscopy. The tip of the catheter was advanced to the junction of the right atrium and the superior vena cava. After confirming position, the blood tubing was connected, and extracorporeal flow through the Hemolung circuit immediately initiated. The catheter was secured to the skin with sutures. Hemolung gas flow was set to 8 l/min and the rotational speed was adjusted to obtain a blood flow of approximately 400 ml/min. A systemic intravenous (IV) heparin drip (100 U heparin per milliliter normal saline) was used to maintain aPTT levels in the range 46–70 s (1.5–2.3 times baseline) during the chronic phase of the trial.

Animal recovery, maintenance, monitoring and measurements

After securing the catheter and initiating Hemolung circulation, anesthesia was discontinued and the animal was extubated. The animal was then transported to the ICU where it was cared for in individual stanchions. Vital signs were monitored hourly over the course of the study, including heart rate, arterial blood pressure, output, and intake. In addition, all animals received prophylactic antibiotic (cephalexin, 3 g/day IV) and analgesic (flunixin meglumine, 1 mg/kg IV twice per day for the first 3 days, or longer as needed).

A series of pathology parameters were measured at regular intervals to assess the health of the animal, the performance of the Hemolung device, and the effect of the device on the sheep. These measurements included arterial and venous blood gases, complete blood count, serum chemistry, coagulation parameters, PfHb, urine samples and blood smears.

Hemolung measurements

The Hemolung cartridge and controller performance was assessed every hour throughout the duration of each trial. Critical measurements included CO2 removal, blood flow, pump speed in revolutions per minute (rpm), and sweep gas flow rate. CO2 removal from the blood into the fiber lumens of the device occurs by diffusion across the wall of the hollow fibers, driven by the difference in PCO2 between the blood and the gas flowing through the fibers. Since variations in blood PCO2 values would result in variable CO2 removal rates, critical evaluation of the CO2 exchange performance required a linear normalization of CO2 removal. The normalized CO2 exchange rate was calculated based on a moving average of venous PCO2, using the following equation:

$$ {\text{VCO}}_{ 2}^{\text{norm}} {\text{ = VCO}}_{ 2}^{\text{actual}} \times \frac{{ 4 5 {\text{ mmHg}}}}{{{\text{PvCO}}_{ 2} }} $$

Study termination, device removal, and necropsy

After 8 days of data collection the study was terminated. Blood samples were obtained for a complete set of terminal blood studies including blood culture. Before the animals were euthanized they were heparinized (100 IU/kg, IV) for anticoagulation. After flushing with heparinized saline, the device was examined for evidence of thrombus deposition. A complete necropsy examination of each animal was performed with the Hemolung catheter in situ. Following explant, the catheter was rinsed with saline and inspected for thrombus and mechanical integrity. The luminal vascular surfaces of the animal were evaluated for potential gross damage related to catheter placement and operation. The organs were evaluated for pathology and evidence of embolic injury. Particular attention was paid to the lungs and the pulmonary arterial vasculature. Representative samples from each of the major organs (lungs, heart, kidneys, liver, and spleen) and the adrenal glands were taken for histopathology.

Results

Eight healthy sheep were successfully placed on Hemolung therapy under anesthesia, resuscitated, and monitored for up to 8 days. The Hemolung catheter was percutaneously inserted without incident. Recovery from anesthesia was uneventful in all animals. Mechanically, the Hemolung performed as intended with no bearing failures. The Hemolung controller also performed as expected with no loss of sweep gas flow or other failures. Six of the eight animals remained on the device for the entire 8 days. In the second and sixth animals, the Hemolung catheter became dislodged causing premature cessation of treatment. No animals died unexpectedly during the study. The following is a summary of the major findings from the measurements and observations collected prior to, during, and immediately following each animal trial.

Average daily blood flow rates through the Hemolung device are shown in Fig. 2 with corresponding average pump speeds. The normalized CO2 removal rates for each animal are shown in Fig. 3, which includes a table listing the daily average venous PCO2 values. The target blood flow rate of 350–450 ml/min was obtained at nominal pump speeds. This flow rate successfully supported the target minimum CO2 removal rate of 50 ml/min at a venous PCO2 of 45 mmHg without degradation in all but one animal. In the fourth animal, which remained on Hemolung therapy for 8 days, a bleeding complication occurred on the third day from a damaged venous monitoring catheter. Blood counts in this animal demonstrated moderate to marked anemia due to bleeding at the site of the venous monitoring line (left neck, opposite side to Hemolung catheter). A blood transfusion was administered on the third day. No other animals experienced bleeding complications.

Fig. 2
figure 2

Average flow rate of blood through the Hemolung circuit (left axis) plotted with average pump motor speed (right axis)

Fig. 3
figure 3

Normalized CO2 exchange rate achieved by the Hemolung device in each animal over time. The minimum target level is shown as 50 ml/min at a PCO2 of 45 mmHg. Average daily venous CO2 tensions corresponding to average daily normalized CO2 removal rates in each animal are tabulated below the graph

The results associated with hemocompatibility and organ function are listed in Table 1. The statistical significance of changes over time were evaluated using a two-tailed heteroscedastic Student’s t test. Hemocompatibility measures included hematocrit, platelet count, white blood cell (WBC) count, fibrinogen, PfHb and total bilirubin. Figure 4 shows individual PfHb measurements for each animal over the course of treatment. Kidney and liver function were assessed using aspartate transaminase (AST), alanine transaminase (ALT), creatinine, and urea nitrogen (BUN). No animal showed any signs of infection and all remained normothermic throughout treatment.

Table 1 Hematology and organ function
Fig. 4
figure 4

PfHb over time in each animal. Time zero is immediately following initiation of flow through the Hemolung device while the animal was still under sedation

All necropsy and histopathology results were unremarkable, with the exception of the animal which experienced the bleeding complication. There were no gross anatomic lesions found in any of the sheep. All visceral organs were considered to be within normal limits. No intimal vessel damage or vascular clots were seen. Thrombus at the catheter insertion site was seen in some animals. This was considered a normal finding consistent with central venous catheterization. There was no definitive evidence of thromboembolic injury in any of the organs examined, including the lungs. The pulmonary vasculature in all sheep was within normal limits with no evidence of reactive endothelium or thrombus. At necropsy in the fourth animal which experienced the bleeding complication at the site of the venous monitoring line, a kink in the line at the point where it entered the vein was observed. In this animal, there was evidence of minimal subacute to chronic inflammation in the lung, seen as alveolar proteinosis, type II cell hyperplasia, eosinophilic infiltration, and pigmented macrophages. These findings were considered to be a result of the blood transfusion.

Following explant, thrombi of various sizes were found entrained in the bottom gap below the rotating core of three of the eight devices. These thrombi were stable and unable to pass through the fiber bundle, which effectively acted as a filter, and did not affect device performance. Varying amounts of adherent thrombus were found on the fiber bundles, primarily at the warp threads of the fiber mat and the potting junctions, and did not appear to affect gas exchange. The spinning core bearings were all free of any intrusion of blood.

Discussion

The primary aim of this study was to demonstrate the safety and effective performance of the Hemolung over an 8-day period. The in vivo model of healthy awake and standing sheep for an 8-day period was used to maximize the ability to evaluate long-term device performance and reliability, and to detect adverse reactions. The primary measurements made to evaluate device performance were the rate of CO2 removal, the rate of blood flow through the device, and the speed of the rotating core of the fiber bundle (other device performance measures were also monitored and recorded). The safety of the therapy provided by the Hemolung was assessed by measurements of animal health, including physiologic monitoring (blood pressure, pulse, and temperature), hematology (complete blood count), serum chemistry, coagulation studies, PfHb, and urinalysis.

The CO2 exchange levels normalized to a venous PCO2 of 45 mmHg were not only consistent over the 8-day period with little degradation (on average, 91 % of initial on day 8), but exceeded the target level by as much as 40 % in all but one animal. In animal four in which CO2 exchange levels were in the range 42–53 ml/min (normalized), the bleeding complication associated with a damaged venous monitoring catheter resulted in a hematocrit drop of 15–17 % during the first 4 days of the trial. CO2 transfer in the lung has been shown to decline as a result of reduced hematocrit [16]. In that study, a reduction in hematocrit from 45 % to 15 % was associated with a 50 % reduction in CO2 exchange. This same effect would be expected in the Hemolung and was therefore the likely reason for the reduced CO2 removal rates in this animal.

The target blood flow rate for the Hemolung was 350–450 ml/min, which was consistently observed in all animals. The target blood flow rate was based on the optimal range for achieving clinically significant levels of CO2 removal and on the goal of minimizing exposure of the blood to the extracorporeal system. A fundamental objective in the design process for the Hemolung device was to eliminate the invasiveness, complexity and complications associated with high blood flow ECMO and ECCO2R systems used in earlier respiratory assist studies. This aim was achieved through integration of a centrifugal pump within the gas exchanger. The shear and secondary flow patterns set up by the spinning core in the blood adjacent to the surface of the HFMs increases the gas exchange efficiency allowing very low blood flow through the device, as well as simplification of the circuit. A reduced blood flow rate also allowed a reduction in the catheter size needed for percutaneous access to the circulatory system, and therefore the ability to utilize a single venous insertion site with a smaller dual lumen catheter, as opposed to the need for separate cannulations with larger bore catheters required in higher flow applications, which increases the risk of cannulation-associated complications. Furthermore, the intended indications for low flow partial respiratory support excludes patients requiring venoarterial cannulation, which has been shown to be associated with significantly greater mortality [17].

The hematology and organ function measurements made to assess device safety showed changes, but there were no trends or events which indicated significant adverse safety concerns (Table 1). In this study, only one of the eight animals (animal five) exceeded the 40 mg/dl PfHb threshold for clinically significant hemolysis during the last 48 h of the study (Fig. 4). At the time of elevated PfHb levels for this animal, total bilirubin remained within the normal range at 0.3 mg/dl or below, but both hemoglobin and hematocrit fell. Necropsy and explant analysis revealed nothing suggesting that mechanical, device-related hemolysis had occurred in this animal, and pump speed remained stable, indicating no increase in flow resistance through the circuit. Initial increases in the liver function markers AST and ALT were considered to not be indicative of liver damage, but rather a result of muscle damage caused by the cannulation procedure since both enzymes returned to normal over time, and liver necropsy and histopathology results were normal.

The changes seen in hematocrit, in the absence of hemolysis, were likely due to the frequent blood collections for analytic purposes (approximately 40–50 ml of blood per day was drawn for testing) and were not believed to have been caused by the device. Additionally, the hematocrit showed a marked drop perioperatively, even prior to initiation of extracorporeal circulation. It is hypothesized that the perioperative hemodilution due to intravenous fluids, as well as sequestration of red blood cells in the spleen, shifts in body fluids and vasodilation induced by anesthesia, all contributed to this change in hematocrit [18].

Conclusion

The results of 8-day trials in six awake and standing sheep supported by the Hemolung demonstrate that the Hemolung can consistently achieve clinically relevant levels of CO2 removal without failure and without significant risk of adverse reactions. Normocapnic CO2 exchange rates of 50–70 ml/min were achieved through a venovenous 15.5-Fr dual lumen catheter inserted percutaneously in the right jugular vein at blood flow rates in the range 350–450 ml/min. There were no complications or difficulties experienced with catheter insertion or positioning. The system controller also performed successfully without failure, and was easily monitored and operated by veterinary technicians without substantial respiratory support training. These results support the conclusion that the Hemolung respiratory assist system may be a safe and effective method of ECCO2R for partial respiratory support in patients suffering from acute hypercapnic respiratory failure, and that progression to pilot clinical trials in humans can be considered.