The advent of improved digital imaging modalities in diagnostic and therapy is fast making conventional films a nonexistent entity. However, several radiotherapy centers still persist with film for performing quality assurance (QA) tests. This paper investigates the feasibility of using a diagnostic computed radiography (CR) system as a QA tool in radiotherapy. QA tests such as light field congruence, field size verification, determination of radiation isocentre size, multileaf collimator (MLC) check and determination of isocentric shift for stereotactic radiosurgery (SRS) were performed and compared with film. The maximum variation observed between CR and film was 0.4 mm for field size verification, −0.13 mm for the radiation isocentre size check, 0.77 for MLC check and −0.1 mm for isocentric shift using the Winston Lutz test tool for SRS QA. From these results obtained with the CR it is concluded that a diagnostic CR system can be an excellent cost-effective digital alternative to therapy film as a tool for QA in radiotherapy.
Similar content being viewed by others
References
Langmack, K.A.,Portal imaging. Br J Radiol., 74(885): p. 789–804, 2001.
Sonoda, M., Takano, M., Miyahara, J. and Kato, H.,Computed radiography utilizing scanning laser stimulated luminescence. Radiology, 148(3): p. 833–8, 1983.
Wilenzick, R.M., C.R. Merritt, and S. Balter,Megavoltage portal films using computed radiographic imaging with photostimulable phosphors. Med Phys, 14(3): p. 389–92, 1987.
Gur, D., Deutsch, M., Fuhrman, C.R., Clayton, P.A., Weiser, J.C., Rosenthal, M.S. and Bukovitz, A.G.,The use of storage phosphors for portal imaging in radiation therapy: therapists’perception of image quality. Med Phys, 16(1): p. 132–6, 1989.
Whittington, R., Bloch, P., Hutchinson, D. and Bjarngard, B.E.,Verification of prostate treatment setup using computed radiography for portal imaging. J Appl Clin Med Phys, 3(2):p. 88–96, 2002.
Curtin-Savard, A.a.P., E.B.,An electronic portal imaging device as a physics tool, Med Dosim, 22(2): p. 101–5, 1997.
Dunscombe, P., Humphreys, S. and Leszczynski, K.,A test tool for the visual verification of light and radiation fields using film or an electronic portal imaging device. Med Phys, 26(2): p. 239–43, 1999.
Liu, G., van Doorn, T., and E. Bezak,The linear accelerator mechanical and radiation isocentre assessment with an electronic portal imaging device (EPID). Australas Phys Eng Sci Med., 27(3): p. 111–117, 2004.
Olch, A.J.,Evaluation of a computed radiography system for megavoltage photon beam dosimetry. Med Phys, 32(9): p.2987–99, 2005.
Seggern, H.V.,Photostimulable x-ray storage phosphors: a review of present understanding. Braz. J. Phys., 29(2): p. 254–268, 1999.
Rasband, W.S.,ImageJ, U.S.N.I.o. Health, Editor, http://rsb.info.nih.gov/ij/: Bethesda, Maryland, USA, 2005.
Gonzalez, A., Castro, I. and Martinez, J.A.,A procedure to determine the radiation isocenter size in a linear accelerator. Med Phys, 31(6): p. 1489–93, 2004.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Peace, T., Subramanian, B. & Ravindran, P. An experimental study on using a diagnostic computed radiography system as a quality assurance tool in radiotherapy. Australas. Phys. Eng. Sci. Med. 31, 226–234 (2008). https://doi.org/10.1007/BF03179349
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF03179349