Baseline errors in cuff electromagnetic flowmeters

  • Published:
Medical and biological engineering Aims and scope Submit manuscript

Abstract

An account is given of sources of baseline error which can occur in electromagnetic flowmeters used for measurement of blood flow in uncut blood vessels. A criterion is suggested for the baseline stability of a given system. The following factors are considered: phase drift, harmonics, coil leakage, common mode signals, multiple earths, balanced earths, eddy currents in metallic components, defective magnet switch, electrode effects, tissue phase angle, and external interference. Attention is drawn to the presence of axial potential gradient in electromagnetic flowmeters generally, and the way in which this can vary both with experimental conditions and with different types of flowmeter heads. Ways in which such an axial potential gradient can affect baseline are described. Electrode effects are considered in some detail, and it is shown how they can be reduced by recessing the electrodes in the wall of the flow tube.

Sommaire

Dans cet article il est question des sources d'erreurs de la ligne de base pouvant survenir dans les résultats fournis par les débitmètres électromagnétiques employés à mesurer le flux sanguin dans les vaisseaux intacts. Un critère de stabilité de la ligne de base est proposé pour chaque système. Les facteurs suivants sont étudiés: la dérive, les harmoniques, les pertes électromagnétiques de la bobine, les terres multiples, les terres équilibrées, les courants de Foucault dans les composantes métalliques, les effets d'électrodes, l'angle de phase des tissus et l'interférence externe. La présence du gradient de potentiel axial dans tous les débitmètres électromagnétiques est soulignée ainsi que les variations pouvant survenir dans ce phénomène en fonction des conditions expérimentales et les différents types de capteurs. Une description des effets possibles du gradient de potentiel axial est donnée. Les effets de l'électrode sont étudiés en détail et la possibilité de les réduire en plaçant les électrodes dans les parois du tube de débit est indiquée.

Zusammenfassung

Mögliche Ursachen von Fehlern bezüglich der Basislinie (Nullpunkt) bei elektromagnetischen Geräten zur unblutigen Blutstromstärkemessung werden zusammengestellt. Ein Kriterium der Nullpunktstabilität im Messsystem wird vorgeschlagen. Folgende Faktoren werden betrachtet: Phasenverschiebung, Harmonizität, Spulenverluste, gleichphasige Signale, mehrfache Erdung, abgeglichene Erden, Oberwellen in metallischen Komponenten, fehlerhafte Magnetschalter, Elektrodeneffekte, Phasenwinkel des Gewebes und externe Interferenz. Es wird darauf hingewiesen, dass im allgemeinen in elektromagnetischen Strömungsmessern axiale Potentialgradienten existieren. Einflüsse der Versuchsbedingungen und der Verwendung verschiedener Typen von Ableitelektroden auf diese Gradienten werden diskutiert. Es wird beschrieben, wie axiale Potentialgradienten die Basislinie beeinflussen. Elektrodeneffekte werden in Einzelheiten erörtert, und es wird gezeigt, dass sie vermindert werden können, wenn die Elektroden in die Wand der Strömungsröhre einbezogen werden.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Gessner, U. (1961)Biophys. J. 1, 627.

    Google Scholar 

  • Goldman, S. C., Marple, N. B. andScolnik, W. I. (1963)J. appl. Physiol. 18, 652.

    Google Scholar 

  • Hognestad, M. (1963) inProceedings of the 5th Int. Conference Med. Electronics. Ed.F. H. Bostem Desoer, Liège.

    Google Scholar 

  • Hutcheon, I. C. andHarrison, D. N. (1960)Proc. I.E.E. 107B, 73.

    Google Scholar 

  • Kolin, A. (1960)Med. Phys. 3, 151.

    Google Scholar 

  • Kolin, A., Herrold, G. andAssali, N. (1958)Proc. Soc. exp. Biol. Med. 98, 550.

    Google Scholar 

  • Kolin, A. andWisshaupt, R. (1963)I.E.E.E. Trans. Bio-Med. Electron. BME-10, 60.

    Article  Google Scholar 

  • Kurlbaum, F. (1899)Ann. Phys. Lpz. 67, 846.

    Google Scholar 

  • Richardson, A. W., Denison, A. B. andGreen, H. D. (1952)Circulation,5, 430.

    Google Scholar 

  • Schwan, H. P. (1957)Ann. N.Y. Acad. Sci. 65, 1007.

    Google Scholar 

  • Shercliff, J. A. (1962)The Theory of Electromagnetic Flow Measurement. Cambridge University Press, Cambridge.

    Google Scholar 

  • Wyatt, D. G. (1961a)Physics Med. Biol. 5, 289.

    Article  Google Scholar 

  • Wyatt, D. G. (1961b)Electronic Engng. 33, 650.

    Google Scholar 

  • Wyatt, D. G. (1964a)J. Physiol. Lond. 173, 8P.

  • Wyatt, D. G. (1964b)Nature, Lond. 204, 1294.

    Article  Google Scholar 

  • Wyatt, D. G. (1965)Electronic Engng 37, 16.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Wyatt, D.G. Baseline errors in cuff electromagnetic flowmeters. Med. & biol. Engng. 4, 17–45 (1966). https://doi.org/10.1007/BF02474784

Download citation

  • Received:

  • Revised:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02474784

Keywords

Navigation