Noninvasive measurement of blood glucose concentrations by analysing fourier transform infra-red absorbance spectra through oral mucosa

  • Molecular Sensing
  • Published:
Medical & Biological Engineering & Computing Aims and scope Submit manuscript

Abstract

Whether Fourier transform infra-red spectroscopy with an attenuated total reflection prism could be applied for noninvasive glucose measurement through oral mucosa was evaluated. As a result, the same absorbance peak at 1033 cm−1 as in glucose aqueous solution was found in the absorbance spectra through mucous membrane. However, these glucose specific peaks were interfered with by the baseline drifts owing to prism attachment and the background spectra from body constituents other than glucose. Therefore, to eliminate these interferences, the calibration curve between the second derivatives of the absorbance peak at 1033 cm−1 and those at 2920 cm−1 was calculated (r=0·910). By using this calibration curve, the spectral changes due to prism attachment were first eliminated. Secondly, by obtaining the second derivative of the difference between the postprandial absorbance peak and the fasting sample as a characteristic of an individual, high correlations between the corrected second derivatives of absorbance spectra through the mucous membrane of the lip at 1033 cm−1 and the increases in blood glucose concentrations above fasting levels were observed (r=0·910). In conclusion, it was suggested that Fourier transform infra-red spectroscopy could be useful for noninvasive monitoring of glucose through oral mucosa.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • BACK, D.M., MICHALSKA, D.F. and POLAVARAPU, P.L. (1984) Fourier transform infrared spectroscopy as a powerful tool for the study of carbohydrates in aqueous solutions. Appl. Spectrosc., 38, 173–180.

  • BAUER, B. and FLOYD, T.A. (1987) Monitoring of glucose in biological fluid by Fourier transformation infrared spectroscopy with a cylindrical internal reflectance cell. Anal, Chim. Acta, 197, 295–301.

  • CONWAY, J.M., NORRIS, K.H. and BODWELL, C.E. (1984) A new approach for the estimation of body composition: infrared interactance. Am. J. Clin. Nutr., 40, 1123–1130.

  • HOPK1NSON, J.H., Movsxou, C., REYNOLDS, N. and NEWBERY, J.E. (1987) Application of attenuated total reflection of carbohydrates and biological whole cell samples in aqueous solution. Analyst, 112, 501–505.

  • KAISER, N. (1978) Laser absorption spectroscopy with an ATR prism. Appl. Phys., 17, 1–4.

  • KAISER, N. (1979) Methods for determining the contents of metabolic products in the blood. IEEE Trans., BME-26, 597–600.

  • KAJIWARA, K., FUKUSHIMA, H., SHICHIRI, M., KAWAMORI, R. and YAMASAKI, Y. (1989a) Non-invasive measurement of glucose concentrations. J. Artif. Organs, 18, 1328–1331.

  • KAJIWARA, K., FUKUSHIMA, H., SAKAKIDA, M. and SHICHIRI, M. (1989b) Development of artificial endocrine pancreas-possibility of non-invasive measurement of blood glucose using infra-red spectroscopy with an ATR prism. In Recent advances in insulin therapy. LEE, T.H., HuH, K.B. and BABA, S. (Eds.), Excerpta Medica, Elsevier Science Publishers B.V., Amsterdam, 357–360.

  • KAJIWARA, K., FUKUSHIMA, H., SAKAKIDA, M. and SHICHIRI, M. (1990) Non-invasive measurement of glucose concentrations. Interferences of biosubstances with the spectrum of glucose. Jap. J. Artif Organs, 12, 881–884.

  • KAJIWARA, K., FUKUSHIMA, H., SAKAKIDA, M., HASHIGUCHI, Y. and

  • SHICHIRI, M. (1991a) Development of non-invasive measurement of glucose concentrations - Application of subtraction method. Ibid., 20, 183–187.

  • KAJIWARA, K., FUKUSHIMA, H., KISHIKAWA, H., NISHIDA, K., HASHIGUCHI, Y., SAKAKIDA, M., UEHARA, M. and SHICHIRI, M. (1991b) Spectroscopic quantitative analysis of blood glucose by Fourier transform infrared spectroscopy with an attenuated total reflection prism. Med. Prog. Through Technol., 17, (in press).

  • LANZA, E. (1983) Determination of moisture, protein, fat and calories in raw pork and beef by near infrared spectroscopy. J. Food Sci., 48, 471–474.

  • MIYAZAWA, T., SH1MANOUCH1, T. and MIZUSHIMA, S. (1956) Characteristic infrared bands of monosubstituted amides. J. Chem. Phys., 24, 408–418.

  • SHICHIRI, M., KAWAMORI, R., YAMASAKI, Y., HAKUI, N. and ABE, H. (1982) Wearable artificial endocrine pancreas with needle-type glucose sensor. The Lancet, il, 1129–1131.

  • SHICHIRI, M., KAWAMORI, R., YAMASAKI, Y., HAKUI, N. and ABE, H. (1984) Closed-loop glycemic control with a wearable artificial endocrine pancreas. Variations in daily insulin requirements to glycemic response. Diabetes, 33, 1200–1202.

  • SHICH1RI, M., YAMASAKI, Y., NAO, K., SEKIYA, M. and UH)A, N. (1988) In viva characteristics of needle-type glucose sensor. Measurement of subcutaneous glucose concentrations in human volunteers. Harm. Metab. Res., 20, Suppl., 17–20.

  • SHICHIRI, M. (1989) The future of insulin therapy. Towards a closed-loop system. In Proceedings of 13th IDF Congress.

  • LARKINS, R.G., ZIMMET, P.Z. and CHISHOLM, D.J. (Eds.), Excerpta Medica, Amsterdam, 23–27.

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kajiwara, K., Uemura, T., Kishikawa, H. et al. Noninvasive measurement of blood glucose concentrations by analysing fourier transform infra-red absorbance spectra through oral mucosa. Med. Biol. Eng. Comput. 31 (Suppl 1), S17–S22 (1993). https://doi.org/10.1007/BF02446645

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02446645

Keywords

Navigation