Abstract
Series dead space (VdS) is assumed to be represented by that volume exhaled until alveolar gas is observed. Phase II of the single breath CO2-diagram contains the (flow, concentration and sequence weighted) distribution off all stationary interfaces (SI) expired before phase III. We describe a new method to estimate the mean value of VdS based on the differentiation of phase II. This approximation of VdS is called the ‘Pre Interface Expirate’ (PIE) and is compared in this study with the integrative approach of Langley. Tidal volume (Vt) and PEEP were varied from 71 to 123% and from 0 to 6 cmH2O respectively.
The estimation of VdS by differentiation of phase II (PIE) shows excellent reproducibility and depends only on phase II — not on phase III and IV as does VdS-Langley. PIE does not depend on Vt and PEEP per se but reflects the distension of convective airways due to elevated end-inspiratory airway pressure.
Our results confirm the predictions of Paiva's model calculations in that the size of VdS is determined by the distension of airways rather than by the altered position of the SI.
Similar content being viewed by others
References
Cumming G, Horsfield K, Preston SB: Diffusion equilibrium of the lungs examined by nodal analysis. Respir Physiol 12: 329–345, 1971.
Paiva M: Computation of the boundary conditions for diffusion in the human lung. Comput Biomed Res 5: 585–595, 1972.
Horsfield K, Cumming G: Functional consequences of airway morphology. J Appl Physiol 24 (3): 384–390, 1968.
Olsson SG, Fletcher R, Jonson B, Nordstroem L, Prakash O: Clinical studies of gas exchange during ventilatory support — a method using the Siemens-Elema CO2 analyzer. Br J Anaesth 52: 491–498, 1980.
Langley F, Even P, Duroux P, Nocolas RL, Cumming G: Ventilatory consequences of unilateral pulmonary artery occlusion. Les Colloques de l'Institut National de la Sante et de la Recherche Medicale 51: 209–214, 1975.
Fletcher R, Jonson B, Cumming G, Brew J: The concept of deadspace with special reference to the single breath test for carbon dioxide. Br J Anaesth 53: 77–94, 1981.
Fletcher R, Jonson B: Deadspace and the single breath test for carbon dioxide during anaesthesia and artificial ventilation. Br J Anaesth 56: 109–119, 1984.
Brunner J, Wolff G: Reliable estimation of series dead space in ventilated patients. Clin Physiol: in press, 1984.
Brunner J, Langenstein H, Wolff G: Direct accurate gas flow measurement in the patient: Compensation for unavoidable error. Med Prog Technol 9: 233–238, 1983.
Brunner J, Wolff G, Langenstein H, Cumming G: Reliable detection of inspiration and expiration by computer. J. Appl Physiol: in press, 1984.
Paiva M: Model analysis of inert gas mixing in the lung. Bull Europ Physiopath Resp 18: 189–201, 1982.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Wolff, G., Brunner, J.X. Series dead space volume assessed as the mean value of a distribution function. International Journal of Clinical Monitoring and Computing 1, 177–181 (1984). https://doi.org/10.1007/BF01872769
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01872769