Abstract
In high precision Fourier transform spectrometry, proper handling of the phase correction is essential if the full potential wave number accuracy of the data is to be preserved. One-sided interferograms are shown to be especially sensitive to phase error, and this sensitivity is quantitatively related to the signal-to-noise ratio. The role of the truncation/apodization function used to obtain the phase correction is also discussed, as is the special problem of emission spectra.
Similar content being viewed by others
References
A. Savitsky, M. J. E. Golay,Anal. Chem. 1964,36, 1627.
R. N. Bracewell,The Fourier Transform and Its Applications, McGraw-Hill, New York, 1965.
G. A. Vanasse, H. Sakai,Progress in Optics 1967,6, 259.
J. W. Brault, in:High Resolution Astronomy (A. Benz, M. Huber, M. Mayer, eds.), Geneva Observatory, Sauverny, Switzerland, 1985, p. 1.
J. Connes, in:Aspen International Conference on Fourier Spectroscopy, AFCRL Rep. No. 114 71-0019, 1971, p. 83.
L. Mertz,Transformations in Optics, Wiley, New York, 1965.
I. Wynne-Jones,Infrared Phys. 1984,24, 273.
J. Kauppinen,Infrared Phys. 1976,16, 359.
B. Carli, F. Forni, F. Mencaraglia,Int. J. Infrared Millimeter Waves 1982,3, 529.
Author information
Authors and Affiliations
Additional information
National Solar Observatory, NOAO, operated by the Association of Universities for Research in Astronomy, Inc., under contract with the National Science Foundation
Rights and permissions
About this article
Cite this article
Brault, J.W. High precision fourier transform spectrometry: The critical role of phase corrections. Mikrochim Acta 93, 215–227 (1987). https://doi.org/10.1007/BF01201691
Received: