Determination of N-Methylpyrrolidine in Cefepime for Injection by Capillary Electrophoresis with Indirect UV Detection

  • Published:
Chromatographia Aims and scope Submit manuscript

Abstract

N-Methylpyrrolidine in cefepime for injection was determined by capillary electrophoresis with indirect UV detection. Best results were achieved with background electrolyte consisting of 10 mM creatinine adjusted to pH 3.8 with formic acid and an applied voltage of 30 kV in a bare fused-silica capillary. Indirect UV detection was performed at a wavelength of 225 nm. The application of a small amount of inlet pressure during the separation assisted the attainment of a stable baseline. The optimized method was validated regarding selectivity, linearity, accuracy, precision, ruggedness, repeatability and detection limits. Careful control of capillary conditioning enabled migration time precision values of <0.2% RSD. The use of an internal standard enabled precision values of <1% RSD to be obtained for peak area ratios.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to V. B. Sunderland.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Liu, H., Sunderland, V. Determination of N-Methylpyrrolidine in Cefepime for Injection by Capillary Electrophoresis with Indirect UV Detection. Chromatographia 59, 653–657 (2004). https://doi.org/10.1365/s10337-004-0191-2

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1365/s10337-004-0191-2

Keywords

Navigation