A new, validated, and selective RP-HPLC technique was established for quantitative estimation of Dicyclomine HCl and Simethicone in the pharmaceutical oral liquid drop formulations. The chromatographic conditions were adjusted using Luna C18 (25 cm × 0.46 cm, 5 μm) column, water-methanol mixture (80:20, v/v) as solvent system and pH adjustment up to 7.4 using orthophosphoric acid. The absorbance was observed at 289 nm λmax injecting 20 μl sample solution and maintain the ambient temperature (30 °C) with 1 ml per min flow rate.
Results
The retention times of Dicyclomine HCl and Simethicone were determined 2.962 and 4.091 min, respectively. The developed analytical technique was validated for accuracy, precision, linearity, specificity, sensitivity, ruggedness, and robustness as per ICH guidelines. The LOD was found to be 0.72 and 0.57 μg per ml for Dicyclomine HCl and Simethicone, respectively. But, LOQ was 2.19 and 1.73 μg per ml, respectively.
Conclusions
This validated method showed good precision (RSD% < 1) with acceptable linearity (≥ 0.999). Intra- and inter-day relative RSD of retention times and AUC (area under curve) were found to be less than 2.0%. The method was also significantly useful as economic technique for simultaneous quantitative determination of Dicyclomine HCl and Simethicone in the bulk and pharmaceutical formulation.
Similar content being viewed by others
Background
Dicyclomine hydrochloride (DIC) is 2-(diethylamino) ethyl-1-cyclohexylcyclohexane-1-carboxylate hydrochloride chemically is used as a parasympatholytic and antispasmodic agent. This drug blocks effect of acetylcholine on parasympathetic receptors in the smooth muscle along with direct relaxing effect on smooth muscle [1,2,3].
Simethicone (SIM), α-(trimethylsilyl)-ω-methylpoly[oxy(dimethylsilylene)], is present with silicon dioxide to enhance defoaming properties of silicon. It is used as an anti-foaming agent to minimize bloating, distress, or pain due to excessive gas [4,5,6,7,8].
The structures of these drugs are given in Figs. 1 and 2.
According to the literature, several RP-HPLC methods were availed for quantitative determination of DIC and SIM individually in pharmaceutical dosages forms [9,10,11,12,13,14,15,16,17,18,19,20,21]. But, there is no RP-HPLC technique was identified until now for simultaneous estimation of both drug combinations in the bulk and marketed formulation. Our research work focused on developing an economical, validated, and sensitive method for the simultaneous determinations of DIC and SIM in oral liquid drop formulation.
Methods
The analytical samples of DIC and SIM were provided as gift samples by Blue Cross Laboratories Pvt Ltd., Mumbai. They also provided that the gift sample of dicyclomine hydrochloride and simethicone contain 99.87% and 97.38% equivalent quantity of pure drug, respectively. An oral liquid drop formulation (Meftal-Spas Drops manufactured by Blue Cross Laboratories Ltd.) used for the analysis was obtained from market labeled with 10 mg/ml dicyclomine hydrochloride and 40 mg/ml simethicone in the each unit (Fig. 3). Methanol and water of HPLC grades, orthophosphoric acid of analytical grade were purchased from Sigma and Himedia Chemicals supplied by S. K. Traders Indore (M.P.). The samples were observed at 289 nm using the Shimadzu VP Series HPLC system associated with photo diode array detector, LC-10AT pump, and manual injector of 20 μl volumes with rheodyne valve. The samples were analyzed under optimized chromatographic conditions on a Luna C18 (5 μm, 25 cm × 0.46 cm i.d.) phenomenex column. An ultrasonic vibrations degasser was used for mobile phase prior to use.
Chromatographic conditions
A methanol and water mixture (20:80) was used as mobile phase. The optimized solvent system was passed through Luna C18 column with 1 ml/min flow rate for the interval of 8 min as total analysis time. After injection of 20 μl sample, elution of sample was observed by photo diode array detector at 289 nm.
Preparation of solvent system
A mixture of water 800 ml (80%) and 200 ml HPLC grade methanol (20%) was prepared along with degassing on ultrasonic water bath. The pH of 7.4 for mixture was adjusted using orthophosphoric acid. The final resultant mixture was filtered through 0.45-μm membrane filter.
Preparation of working stock solution
The accurately weighed 10 mg of drug was dissolved using methanol-water mixture in 100-ml volumetric flask. Then, stock solution of 100 μg/ml was obtained through adjust the volume up to the mark using same solvent.
Preparation of sample solution
A volume of 10 ml of oral drop containing amount equivalent to 100 mg Dicyclomine HCl and 400 mg Simethicone was dissolved using about 80 ml of solvent in 100-ml volumetric flask then sonicated for 10 min to obtain clear solution. A volume was adjusted with the same solvent and filtered using 0.45-μ membrane filter. Dilutions prepared to reduce the strength of resultant solution by serial dilution technique and then 20 μl of above solutions were injected into the column under the set chromatographic conditions.
System suitability study
The system suitability was studied with the help of theoretical plates, tailing factor, and resolution. All the results of above parameters demonstrate the suitability of system for the analysis of this drug combination reported in Table 1.
Purity assessment (percentage purity w/w)
The content of each drug in the marketed formulation was checked and calculated the percentage purity. This study was also confirmed the purity of procured drug gift sample.
Method validation
The developed analytical method was validated according to ICH guidelines (Q2B) [22, 23].
Linearity study
The linearity was observed as linear regression study of absorbance for both drugs with respect to concentration in range 5 to 40 μg/ml. Several aliquots of known strengths were prepared with dilution of working stock solution using mobile phase. The AUC and concentration data for all the aliquots were determined and reported in Table 2. The linearity curves were obtained by plotting a graph between relative AUC and concentration shown in Figs. 4 and 5. Linearity curves were constructed using relative peak area to avoid very high values for intercept.
Accuracy or recovery study
This study was accomplished with standard addition method (n = 3) at 50%, 100%, and 150% analytical concentration. In this method, known quantity of DIC and SIM was mixed to pre-estimated sample solution and further proceeds to the developed HPLC method. Then, experimental and true values were compared. The percentage recovery of Dicyclomine HCl and Simethicone is reported in Table 3.
Precision study
The study was analyzed by taking six independent quantitative estimation of standard (10 μg/ml Dicyclomine HCl + 40 μg/ml Simethicone) as well as sample solution (oral drops formulation) on same instrument reported in Tables 4 and 5.
Intra-day and inter-day repeatability study
The drug estimation of standard drug solution was performed for thrice on the same day and three consecutive days over duration of a week. The comparison of results was used for the repeatability study of developed analytical method. All the results are shown in Tables 6 and 7.
Sensitivity
The sensitivity of measurement for Dicyclomine HCl and Simethicone by using developed RP-HPLC method was estimated as LOD and LOQ. These indicate the quantitation limit of individual method which is lowest measurable quantity of analyte in the sample. These parameters were determined with the help of equation given as follows:
LOD = 3.3 × N/B and LOQ = 10 × N/B
where N is the standard deviation of the peak areas for drug (n = 5) that consider as measure of noise, and B is the slope of the corresponding calibration curve.
The calculation was completed using the above formulas and found the results given as follows:
Dicyclomine HCl LOQ = 10 × 69.37/316.34 = 2.19 μg/ml
Simethicone LOQ = 10 × 1231.39/7092.7 = 1.73 μg/ml
Dicyclomine HCl LOD = 3.3 × 69.37/316.34 = 0.72 μg/ml
Simethicone LOD = 3.3 × 1231.39/7092.7 = 0.57 μg/ml
Robustness
A solution containing 10 μg/ml Dicyclomine HCl and 40 μg/ml Simethicone was prepared and estimated by two different analysts under the same operational and environmental conditions. A result of reproducibility is compared for robustness study of analytical method reported in Table 8.
Result and discussion
This research work focused on the adequate development of RP-HPLC method with subsequent validations for quantitative analysis of DIC and SIM in oral drop formulation using C18 column and PDA detector. The water-methanol mixture (80:20, v/v) was used as mobile phase with adjustment of pH to 7.4 using orthophosphoric acid. The isosbestic point was used as maximum wavelength for the identification of drug samples elution from HPLC column which was found out through scanning of UV spectrum. The RP-HPLC method was performed at various chromatographic conditions with different changes in solvent composition and flow rate, etc. The system suitability parameters like peak resolution, tailing factor, and theoretical plates were used for the optimization of RP-HPLC method which was reported in Table 1.
The resolution was found to be more than 2 that indicate good separation of DIC and SIM with developed RP-HPLC method. The values of asymmetric/tailing factors and theoretical plates for DIC and SIM were also favorable for efficient separation (asymmetric/tailing factors and theoretical plates should be ≤ 2.0 and > 2000, respectively). When a mixture of Dicyclomine HCl (10 μg/ml) and Simethicone (40 μg/ml) was run into HPLC under optimized chromatographic conditions, we were found two symmetric, well-resolved peaks for the retention of DIC and SIM along with retention time at 2.962 and 4.091 min, respectively. A HPLC chromatogram is given in Fig. 3.
The assay was also performed for each drug in marked formulation to check its purity and calculated the percentage purity. The percentage purity of DIC and SIM was found to be 101.06 and 96.98%, respectively.
The validation of developed analytical method was completed with respect to parameters like linearity, specificity, accuracy, precision, repeatability, sensitivity, and ruggedness study.
The different concentrations of working standard solutions of DIC and SIM were given linearity between the ranges of 5 and 40 μg/ml with significant R square value reported in Table 2. The linear line equations for Dicyclomine HCl and Simethicone were y = 0.1494x + 0.0007 and y = 0.7418x − 0.0039, respectively.
The accuracy of the developed RP-HPLC method was determined by calculating recoveries of Dicyclomine HCl and Simethicone by method of standard addition. Recovery studies were performed in triplicate, and results are reported in Table 3. All the drugs were found to be stable with a recovery of more than 98%. The observed percentage recovery of DIC and SIM was comparable with the corresponding labeled amounts. System and method precision was observed by quantitative analysis of six independent standard as well as oral drop samples. For the repeatability study, the intra-day and inter-day quantitative estimation were carried out for three selected samples concentrations of DIC and SIM. The LOQ for DIC and SIM was found to be 2.19 and 1.73 μg/ml, respectively. The LOD for DIC and SIM was found to be 0.72 and 0.033 μg/ml, respectively (Table 3). Robustness of developed method was determined by deliberately changing the analyst at same experimental conditions. All the parameters are summarized in Table 9 were confirmed about the good precision (RSD% < 1) with acceptable linearity (≥ 0.999) for validated method. The relative RSD with different studies for this method was also found to be less than 2.0%. The proposed method was successfully applied to the determination of DIC and SIM in their combined dosage form.
Conclusion
We were found on the basis of reported literature that there was not a significant method for the simultaneous estimation of dicyclomine hydrochloride and simethicone in pharmaceutical formulation. All the methods were reported either only for another combinations or individual drugs. Therefore, it needs to develop new method for their quantitative analysis in the bulk or pharmaceutical formulation. We had developed an analytical RP-HPLC technique successfully for simultaneously determination of these drugs in oral drops formulation. All the validation results for developed method were found to be under the limits as per ICH guidelines. All the results showed that the developed method is new, simple, rapid, and validate. Furthermore, it can be employed for the analysis of Dicyclomine HCl and Simethicone in bulk and oral liquid dosage form. The simplicity, reproducibility, and cost-effectiveness of method complete the objective of research work.
Availability of data and materials
The research work has been carried out by us, and we assure you that it can be provided to you whenever required.
Abbreviations
- RP-HPLC:
-
Reverse phase high performance liquid chromatography
- ICH:
-
International Conference on Harmonization
- AUC:
-
Area under curve
- LOD:
-
Limit of detection
- LOQ:
-
Limit of quantification
- RSD:
-
Relative standard deviation
References
Prajapati D, Raj H (2012) Simultaneous estimation of mefenamic acid and dicyclomine hydrochloride by RP-HPLC method. Int J Pharm Biosci 3:611–625
Shah D, Rana JP, Chhalotiya UK, Baldania S, Bhatt K (2014) Development and validation of a liquid chromatographic method for estimation of dicyclomine hydrochloride, mefenamic acid and paracetamol in tablets. Indian J Pharm Sci 76:529
British Pharmacopoeia (2015) Medicines and Healthcare products Regulatory Agency, London, Vol. 2, Dicycloverine Hydrochloride, I-736.
da Silva Acunha CF, dos Santos JHZ (2011) An analytical method for quantifying dimethicone in a 30% simethicone emulsion using gas chromatography. Braz J Anal Chem 6:278–285
Meier R, Steuerwald M (2007) Review of the therapeutic uses of simethicone in gastroenterology. Schweiz Z Ganzheitsmed 19:380
Moore DE, Liu TX, Miao WG, Edwards A, Elliss R (2002) A RP-LC method with evaporative light scattering detection for the assay of simethicone in pharmaceutical formulations. J Pharmaceut Biomed 30:273–278
Satyanarayana P, Pavankumar KLNNSVK, Srinivasa RA, Subrahmanya SG (2015) Method development and validation for simultaneous estimation of ranitidine,domperidone and simethicone in bulk and pharmaceutical dosage form by using RP-HPLC method. WJPPS 4(11):1146–1158
Madhavi LB, Noorjahan MD, Madhukar A (2018) A new RP-HPLC method for the simultaneous estimation of magaldrate and simethicone in bulk and tablet dosage form as per ICH guidelines. J Syn Natu Chem 3(1):31–34
Bachani MH, Acharya DS, Shah KV (2013) Development and validation of HPLC method for simultaneous estimation of dicyclomine hydrochloride, acetaminophen and clidinium bromide in solid dosage form. Int J Pharm Pharm Sci 5:462–466
Donda ST, Baviskar VB, Deshmukh PK, Bari SB, Patil PO (2014) Development and validation of a reversed-phase HPLC method for the simultaneous estimation of dicyclomine hydrochloride and famotidine in bulk and tablets. J Chil Chem Soc 59:2662–2665
Gaikwad H, Borde S, Suralkar M, Choudhari V, Kuchekar B (2014) Development and validation of RP-HPLC method for simultaneous determination of tramadol hydrochloride, paracetamol and dicyclomine hydrochloride by using design of experiment software (DOE). Int J Pharm Sci 4:792–801
Hedpara RJ, Chhatarala V, Agola A (2013) A validated RP-HPLC method for the simultaneous estimation of paracetamol and dicyclomine hydrochloride in pharmaceutical formulations. IJPSR 4:2400–2405
Khadge E, Yadav S, Rao J (2017) Reversed phase high performance liquid chromatography method development and validation for simultaneous estimation of dicyclomine hydrochloride, paracetamol and mefenamic acid in bulk and tablet dosage form. Asian J Pharm Clin Res 10:393–397
Kumar A, Chawla P, Porwal P, Rawal R, Anghore D (2018) Development and validation of mefenamic acid, dicyclomine HCl and pamabrom in marketed formulation by HPLC. Pharm Anal Acta 9:594
Lokhande S, Mhetre S, Pekamwar S, Kalyankar T (2012) Development and validation of reverse phase HPLC method for simultaneous estimation of dicyclomine hydrochloride, mefenamic acid and paracetamol in tablet dosage form. World J Pharm Pharm Sci 1:968–980
Neelima K, Prasad YR (2013) Analytical method development and validation for simultaneous estimation of dextropropoxyphene Hcl, dicyclomine and paracetamol in bulk and capsule formulation by RPHPLC. Indo Am J Pharm Res 3:1225–1232
Sharma H, Vishakha K, Kumar K, Bhatta HP (2016) Validated RP-HPLC method for simultaneous estimation of paracetamol, pamabrom and dicyclomine, hydrochloride in bulk and pharmaceutical dosage form. IJPSR 7:316–324
Shrikrishna B, Mulgund S, Ranpise N (2014) Development and validation of RP-HPLC method for simultaneous determination of dicyclomine and mefanamic acid. J Pharm Res 13:16–19
Sireesha T, Kumari K (2014) Durga sai RV, Naik KS Development and validation of new analytical methods for the simultaneous estimation of paracetamol and dicyclomine hydrochloride in bulk and pharmaceutical dosage forms by using RP-HPLC and UV methods. World J Pharm Res 3:1584–1602
Vijjigiri C, Daravath B, Kamarapu S (2013) Method development and validation of RP-HPLC method for simultaneous estimation of dicyclomine hydrochloride and diclofenac potassium in tablet dosage forms. Int J Pharm Bio Sci 3:255–264
Wadher S, Kalyankar T, Kshirsagar J, Anitha K (2017) Simultaneous determination of famotidine and dicyclomine HCl in combined tablet dosage form by UV-spectrophotometer. RJPT 10:408–413
Shah DA, Rana Jainika P, Chhalotiya Usmangani K, Baldania SL, Bhatt KK (2014) Development and validation of a liquid chromatographic method for estimation of dicyclomine hydrochloride, mefenamic acid and paracetamol in tablets. Indian J Pharm Sci 76(6):529–534
Kanubhai TH, Nayan K, Patel Mukesh C (2013) A rapid, validated RP-HPLC method for the simultaneous determination of cleaning validation and cross-contamination of 12 beta-lactam compounds. Sci Pharm 81:151–165
Acknowledgements
The authors are grateful to the Department of Pharmaceutical Chemistry, Mittal Institute of Pharmacy for providing the facilities for research work.
Funding
Not applicable
Author information
Authors and Affiliations
Contributions
We have assured that “all authors have read and approved the manuscript.” All the authors have equal contribution and participation in this research work. PK has analyzed all samples on HPLC instrument and completed the experimental work. He had completed his work under the supervision of SK and MP who help him to elaborate the methodology as well as theoretical approach. AM also helped him in their experimental work and guides to resolve the complications. PW helped him in the data analysis, statistical study, and manuscript editing.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Not applicable
Consent for publication
Not applicable
Competing interests
Not applicable
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Pandey, P.K., Patel, M., Manigauha, A. et al. Simultaneous estimation for Dicyclomine HCl and Simethicone in bulk and oral liquid drop formulation: an RP-HPLC method development and validation. Futur J Pharm Sci 6, 12 (2020). https://doi.org/10.1186/s43094-020-00029-x
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s43094-020-00029-x