Background

Diagnostic x-ray examinations as a basic tool for preserving and recovery of human health play an important role in helping the radiologist to discover a defect, monitor the improvement of diseases and assess the cure reaction. In the United States, medical radiation exposure is approximately the middle of whole radiation exposure from man-made and natural sources [1]. This is serious as the medical application of ionizing radiation over 95% of all artificial radiation exposures and is the major radiation source after natural radiation [2]. Ionization radiation can both induce damaging consequences and supply a strong instrument for diagnostic goals. The relation has been demonstrated among radiation exposures and the occurrence of cancer. It is so important to evaluate the dose received at the x-ray examination. The dose assessment is used either as a representation for radiation risk or as a stage in the real approximation of the risk. There are three basic concepts for radiation protection that are justification, optimization, and dose limit. The optimization needs the usage of particular protocols to convince the remaining doses at a reasonably achievable low level (ALARA) [3]. One of the fundamental provisions for optimization is the perception of patient doses orderly; dosimetry is advised to appraise the dose of patients. Now measurement dose of the patient is considered as a perfect section of a quality assurance program. For optimization, the dose reference level (DRL) is one of the most useful means. DRL is described by two quantities of dose area product and entrance skin dose (ESD) [4]. In the United States, the dosimetry and quality control of x-ray instruments regularly have shown and the establishment of DRLs has played substantial roles in decreasing patient doses [5]. ESD for the 1964 to 2004 decrees 50–70%. A similar x-ray examination in various countries may have different values. As a result, DRLs can be just defined for a city or country as local diagnostic reference levels (LDRLs), while nationwide surveys make national dose reference levels (NDRLs) [6]. However, by comparison another diagnostic imaging patient dose in radiography is low, its quota to the collective dose is considerably caused to the repeated utilization of it. Dose-Area-Product (DAP) is a result of the area of a patient that is exposed multiplied by the radiation dose [7]. The unit of DAP is Gy*m2. Measurement of DAP is appropriate to achieve the optimum level of protection in radiological tests of patients. In utilization, the ionization chamber is located vertically to the beam central and in a position to totally cut off the total region of the x-ray beam. Compounding data of DAP with data on x-ray field size; can be applied to characterize the mean dose created by the x-ray. Then it must be located accurate reading. The reading from a DAP-meter can be varied by either changing the x-ray factors (kVp, mAs), or altering the field or both. Digital radiology over the last decade may demonstrate the major technological development in medical imaging. Although digital radiology has the potential to decreases patient doses, they also have the potential to notably enhance them. This study evaluated the patient dose with a DAP meter and LDRLs for common radiography examination by conventional and digital radiography in South Khorasan, Iran.

Methods

This study was performed in South Khorasan, Iran in the departments of radiology at the educational hospitals affiliated with Birjand University of Medical Sciences department during the period September 2019 to October 2020. There were four active radiology rooms in three educational hospitals of Birjand city (Emam reza, Valias and Razi). The technical specifications of the devices are shown in Table 1. They were chosen from patients randomly who had been referred to the radiographic department of hospitals in South Khorasan. Data were accumulated from physical parameters such as kVp, mAs, and patient data (age, sex, and weight), and type of projection which are as follows: neck (AP), lumbar spine (AP and Lat), skull (AP and Lat), abdomen (AP), pelvis (AP) and CXR. All unite used were conventional and digital radiography (Table 1). All projections and measurements were performed on four different x-ray machine models. The first step for the examined x-ray machine performed a quality control test. Total filtration for all units was 3.5 mm aluminum at 80 kV. Also, the DAP-meter for each type of x-ray projection registers by itself.

Table 1 The technical specifications of the devices

DAP was measured by a DAP-meter (KERMAX-plus SPD, model 120–131 HS). It is capable of measuring output X-ray tubes, with an energy range of 40–150 kVp (less than 3 mm Al). The condition recommended is as followed: temperature between + 10 °C and + 50 °C, humidity between 30 and 75% without condensation, and pressure between 700 and 1060 hPa. It transfers data to a connected computer every 5 ms.

DAP-meter must be calibrated before being used. Calibration process was done according to the method suggested by NRPB protocol, to achieve DAP values [8]. The DAP had a monitor and a detector. The detector was fixed under the beam collimator. A calibrated X-ray test device piranha model was used to measure the performance of the examined X-ray machine and for quality control. The X-ray machines were calibrated. Digital images were acquired by the complete automatic system. An essential quality control test for all x-ray machines including accuracy of the timer, kVp, mA linearity, mAs reciprocity, half value layer (HVL) check, and output check) was done to ensure the high accuracy. The absorbed dose (in unit of Gy) multiplied by the area of tissue irradiated is defined as DAP which reflects dose in the radiation field [9] The DAP values were measured in cGy*cm2 and then were converted to Gy*m2. Average DAP values were calculated from the measurements for each device for the 8 conventional examinations considered in this study: Neck(AP), Skull(lat), Skull(AP), Abdomen, CXR, Lumbar(Lat), Lumbar(AP),Pelvic. The third quartile DAP values were then calculated from the results for each radiographic examination type and view and adopted as the LDRL in Birjand.

Results

In this study radiographic and demographic information of 477 patients was collected. Table 2 shows the radiological and patient parameters (weigh patient, high patient, kVp, mAs). Table 3 displays the average, minimum, maximum, mean, median, first quartile, third quartile of DAP for X-ray examination. Observed DAP for CXR was Minimum of average and quartile3th DAP is CXR of conventional radiography. DAP for Digital radiography for CXR is 0.11. The observed DAP for Chest PA examination was gained to be in the range from 0.12 Gy.cm2 to 0.42 Gy.cm2 with an average value of 0.23 Gy.cm2 which is above the value of the reference level (0.12 Gy.cm2) for Chest PA given in Table 3. For Chest lateral the observed DAP was found to be in the range from 0.18 Gy.cm2 to 1.48 Gy.cm2 with an average value of 0.65 Gy.cm2 and are therefore twice the reference value (0.3 Gy.cm2) given in Table 3.

Table 2 The radiological and patient parameters (weigh of patient, height of patient, kVp, mAs)
Table 3 Shows the average, minimum, maximum, mean, median, first quartile, third quartile of DAP for X-ray examination

Table 4 displays DRLs for different projection radiographies in this study, and our results were checked with DRLs reported by NRPB, Japan, and NDRL. According to Table 4, the 3rd quartile of the measured by digital radiography is upper than convolution radiography. The dose of all projection is lower than in other studies. Figure 1 shows the different means of dose obtain by digital and convolution radiography. DRL for digital radiography is upper than convolution radiography.

Table 4 Displays DRLs for different projection radiographies in this study, and our results were check with DRLs of other study
Fig. 1
figure 1

DRL values for different radiology examinations

Discussion

The linear no-threshold theory says that low dose of ionizing radiation is also not safe. Therefore, decrease of the doses from medical exposures is a serious stage into radiation protection. Measurement of radiation dose in patients is crucial to optimize radiological protection in x-ray centers. This study presents useful information about the radiation dose received by patients undergoing common radiology procedures at hospitals affiliated to Birjad university of medical sciences. It can be used for comparing patients' dose with the established DRL values in order to refuse unnecessary radiation exposure than needed. Table 3 shows a wide range of DAP values for radiographic procedures. These variations were also observed in a same device, for a specific procedure for different patients and in different devices for similar procedures. For instance, the maximum-to-minimum ratio of DAP for individual patients varied from 0.1 for chest LAT to 42 for skull AP/PA. Other studies also have reported such wide variations in diagnostic radiography practice [13, 14]. According to our results, the average DAP value for patients are consistent with reported by many studies have been done in Iran [10, 15,16,17,18,19,20,21]. The DAP value measured in this study for the CXR is consistent with the values reported by Faghihi et al. [22] and Smans et al. [12] for neonatal chest x-rays, whereas, it was lower than values reported by Bahreyni Toossi et al. [23]. Both agreement and discordance were present between our study and other study performed by chest x-rays in Iran. Similarly, our results are inconsistent with DAP value reported by Zewdu et al. for patients during chest x-rays [22]. Variation in the applied film to source distance (FSD) and exposure parameters (kVp and mAs), patient habitus, use or not use of anti-scatter grid and x-ray unites used in these studies are major reasons for the variation of data. We found no significant correlation between patients' sex and weight with DAP values, whereas, a significant correlation was seen for the applied kVp and mAs. These results are consistent with the literature [6]. Variation in typical doses delivered by different x-ray rooms and departments is still substantial, indicating that there is further scope for patient dose reduction in those departments at the top end of the dose range. This verifies the continuing usefulness of reference doses for identifying them. Digital devices have revolutionized medical imaging. Digital radiography can help reduce the patient's dose as long as the operator is trained. There are solutions to reduce the dose and increase the quality of the image, which can be improved by training technologists. As mentioned, improper use of digital radiography can lead to an increase in the dose. What was also shown in this study was that the dose from radiographic convolution imaging was lower than from digital radiography. This is a sign of poor training for technologists. On the other hand, the dose range and the difference between the lowest and highest doses are very high, and this shows that it is possible to provide a suitable image in a specific projection by reducing the dose. Therefore, we need to work hard to reduce the dose.

Conclusions

Variation in typical doses delivered by different x-ray rooms and departments is still substantial, indicating that there is further scope for patient dose reduction in those departments at the top end of the dose range. Training radiation workers and implementing a QA program for devices are necessary for lower patient doses and lower costs for medical health services.