Background

Intravascular catheters are widely used and are indispensable for proper patient management. However, its use implies several risks, which are mainly infectious. Catheter-related bloodstream infection (C-RBSI) is the more important entity which is associated to high rates of morbidity, mortality, and sanitary costs [1].

In general, in those catheters that are inserted for more than 8 days, infection is mainly acquired by an intra-luminal route (66%) because of hub contamination. Hub colonization after manipulation is responsible for 29% to 60% of catheter-related infections (CRI) [2].

Recently, it has been shown that the use of closed needleless connectors is effective against microorganism penetration by hub contamination. Some studies tested the in vitro efficacy of different needleless access devices to prevent the ingress of microorganisms [3]-[8]. However, these studies have heterogeneous designs, and they are all performed by instilling high concentrations of microorganisms through the catheter.

Our study purpose was to create an in vitro model lasting long enough be used to compare various connectors simulating the real daily handling of these devices based on blood culture bottles.

Methods

Setting

This in vitro study has being carried out in the Laboratory of the Clinical Microbiology and Infectious Disease Department of the Hospital General Universitario Gregorio Marañón.

Study design

The model consisted of 40 blood culture bottles with a peripheral venous cannula inserted under sterile conditions and left in place (Figure 1). In each bottle, a disinfectable needle-free closed connector (CLAVE™, CareFusion, Spain) was used to close the cannulas. Twice a day, each cannula was manipulated while instilling 1 mL of either sterile saline or propofol. Manipulation of the bottles was divided into four models: ten bottles (five saline, five propofol) were manipulated with clean gloves and disinfecting the connector with alcohol (controls), ten bottles (five saline, five propofol) were manipulated without gloves (hands), ten bottles (five saline, five propofol) were manipulated with gloves impregnated with a 0.5 McFarland ATCC 29213 Staphylococcus aureus solution, and ten bottles (five saline, five propofol) were manipulated with gloves impregnated with a 0.05 McFarland ATCC 29213 S. aureus solution. Additional photograph files show this in more detail (see Additional files 1 and 2).

Figure 1
figure 1

Experimental model of the blood culture with the cannula and the CLAVE™.

Before handling the connectors in the hand model, a base count and phenotypic identification of the colonizing microorganisms on the manipulators' hands were performed until the bottles became positive.

Laboratory procedure

The bottles were incubated in a BACTEC 9120 System (Becton Dickinson Microbiology Systems, Maryland, DE, USA) up to 10 days at 37°C under continuous agitation. Each day, it was observed whether there was positivity in the bottle fluid (alert from the BACTEC 9120), and in case it occurred, 100 μL of the fluid was cultured into the blood, MacConkey, and Brucella agar and incubated under aerobic and anaerobic conditions for 48 h at 37°C. Microorganism identification was performed using standard procedures [9].

At the end of the incubation time (10 days, 20 mL), the negative bottles were also tested by culturing 100 μL of the fluid into the blood, MacConkey, and Brucella agar and incubated under aerobic and anaerobic conditions for 48 h at 37°C.

The different study variables were annotated in a data collection form.

Statistical analysis

The qualitative variables appear with their frequency distribution. The quantitative variables are summarized as the median with interquartile range (IQR). Continuous variables were compared using the median test for non-normally distributed variables. The chi-square or Fisher exact test was used to compare categorical variables.

The Kaplan-Meier survival curve was used to compare the time to positivity with the degree of contamination.

All statistical tests were two-tailed. Statistical significance was set at p < 0.05 for all the tests. The statistical analysis was performed with SPSS 18.0.

Ethics

This experimental design does not include human subjects and does not use human tissue or samples, so it was exempt from approval of the local ethics committee.

Results and discussion

Overall, all bottles in the control group were negative at the end of the incubation time. In contrast, almost all bottles (38/40) in the three contamination experiments were positive during the incubation time. In the hand model with saline, we recovered the same Staphylococcus epidermidis strain in four out of the five bottles, which was phenotypically coincident with that isolated from the manipulators' hands. In the remaining bottle, we recovered a Staphylococcus warneri strain which was also the same strain we isolated from the manipulators' hands. In the hand model with propofol, all bottles recovered a Micrococcus sp., which was phenotypically coincident with that isolated from the manipulators' hands. In the remaining models with S. aureus with both saline and propofol, the ATCC 29213 S. aureus was recovered in 18 out of the 20 bottles (Table 1).

Table 1 Characteristics of the experiment regarding contamination model and type of instillation fluid

Regarding the median (IQR) time to positivity, there were no differences in either the hand model (saline 76 h (65.92 to 125.01) vs. propofol 75 h (51.45 to 99.00), p > 0.05) or in the 0.5 McFarland S. aureus model (saline 105 h (45.97 to 239.94) vs. propofol 148 h (66.04 to 236.58), p > 0.05). In contrast, in the 0.05 McFarland S. aureus model, the median (IQR) time to positivity was significantly higher when instilling saline than when instilling propofol: 240 h (154.82 to 360.00) vs. 66 h (58.01 to 69.11), p = 0.008. Moreover, when comparing the median (IQR) time to positivity between the three models with saline, the 0.05 McFarland S. aureus model was also significantly higher than the hands model: 240 h (154.82 to 360.00) vs. 76 h (65.95 to 125.00), p = 0.016 (Table 2, Figure 2).

Table 2 Median time to positivity of the bottles comparing different contamination models and different fluids
Figure 2
figure 2

Main results. MF, McFarland; SA, Staphylococcus aureus; MTP, median time to positivity.

We have created an in vitro model of connector handling with gloves impregnated with a 0.05 McFarland S. aureus solution while instilling saline, lasting long enough to be used as a useful model to test the efficacy of closed needleless connectors.

Some recent in vitro studies used standard models of contamination to compare different needleless connectors against the capacity of microorganisms to ingress through the catheter lumen. However, some of them were performed using a two-phase method, in which, first, the connectors were contaminated with a single known microorganism (such as S. aureus or S. epidermidis, which is commonly associated to catheter infection) at different concentrations and, second, they were manipulated to instill contaminated infusion fluids [3],[4],[6],[7],[10],[11]. But this methodology did not simulate the real practice of connector handling.

We compared two solutions of S. aureus of different concentrations by colonizing the surface of a blood culture bottle (while instilling sterile fluids), which simulates better the real daily practice, instead of instilling a contaminated fluid directly through the catheter. Besides, we tested not only a known concentration of a single known microorganism, but also the microorganisms from the flora of the manipulators' hands, as it has been demonstrated that these are highly colonized and could be a potential source of contamination [12]. With this model based on a simple daily handling, we have demonstrated that manipulation of connectors using gloves impregnated with a 0.05 McFarland solution of S. aureus while instilling sterile saline had longer times of positivity, allowing us to prove this model in the comparison of other types of connectors with time enough to detect differences between them regarding their efficacy to prevent the entry of microorganisms through the catheter.

Our study also supports the importance of disinfection before handling a connector, as in our model, all bottles manipulated without gloves turned positive up to 3 to 4 days after catheter insertion.

Conclusions

A daily connector handling with a 0.05 McFarland S. aureus solution while instilling saline proved to be a useful model to test the efficacy of closed needleless connectors. Future studies must be performed with a larger sample and comparing other types of connectors.

Authors' information

MG is a young researcher at the Microbiology Department, whose main project is based on the prevention, diagnosis, and treatment of C-RBSI. MP is a research nurse from the Cardiac Surgery Postoperative Care Unit, who daily manipulates central venous catheters. LA is responsible for the BACTEC area and is a specialist in statistics. PM is responsible for the diagnosis area of catheter colonization and C-RBSI. EB is the head of the Clinical Microbiology and Infectious Diseases Department and an expert on C-RBSI.