Introduction

Positron emission tomography (PET) scanners dedicated to small animals emerged in the mid-1990s [1], as animal model-based research of human disease proved to be an essential and extensively used research tool. Molecular imaging enables to measure in vivo molecular pathways and interactions, not only to early identify and describe the biological nature of a disease but also to follow its evolution. Furthermore, it provides direct biological data for the development of potential therapies, as well as for the assessment of their effects [2].

Technical, physical, and physiological factors influence the experimental PET with [18F]-fluorodeoxyglucose (FDG) outcomes [3]. For instance, PET scanner design, acquisition time, and image analysis may differ between studies and centres—if the scanner has a high-count rate capability, the activity may be reduced, and PET scan duration increased, thereby keeping ALARA (as low as reasonably achievable) principle in mind. Additionally, many factors such as diet, room temperature, and anaesthesia impact markedly 18F-FDG uptake by normal tissues of small animals [4, 5]. In 2019, Mannheim et al. [6] detected differences among four preclinical imaging facilities due to different standard PET imaging protocols for 18F-FDG, including animal preparation and handling, experimental equipment used, personal performing the PET studies, and image analysis. In the same year, McDougald et al. performed a phantom multicentre study recommending acquisition and reconstruction PET/CT protocols, which would improve in vivo measurements independent of scanner design [7]. Besides the incorrect or inappropriate statistical analysis of results or the insufficient sample sizes pointed out by Prinz et al. [8] as one of the reasons for the lack of reproducibility, the absence of a global standard protocol limits repeatability, reproducibility, and reliability of data.

Stout et al. [9] proposed a complete report of materials and methods to assist a manuscript submission, while Mannheim et al. [10] published a review addressing the current status and prospects of small animal imaging standards, including PET. Furthermore, a checklist of recommendations to improve the reporting of research involving animals, the ARRIVE (Animal Research: Reporting of In Vivo Experiments) guidelines 2.0. were recently launched [11]. Although these works provide valuable information to plan and conduct animal studies, ensuring reproducibility through a detailed report, the authors still mentioned the need to develop specific animal preparation and care protocols for a particular biomarker. This issue had already been emphasized by Chalmers and Glasziou [12] while examining the causes and degree of waste in producing and reporting research evidence. The authors concluded that a loss of about 50% in the quality of research design and methods, adequacy of publication practices, and quality of reports of research would lead to a loss greater than 85% [12]. Tackling this problem has become an international effort to ensure that data from in vivo studies provide reliable results to the scientific knowledge [13,14,15].

Therefore, the present work aims to propose a guideline to provide a minimum standard for the preparation and care of mice for PET imaging with 18F-FDG. The global standard protocol focuses on the practical steps that can be controlled and suggested for broad adoption. For this purpose, recommendations and good practices described in relevant literature were assessed.

Principle

PET is a non-invasive imaging modality that quantifies the 3-dimensional distribution of a retained positron-emitting compound as a function of time. Thus, allowing to measure physiological, biochemical, and pharmacokinetic parameters such as: blood flow, glucose and oxygen metabolism, pre/postsynaptic receptor density and affinity, neurotransmitter release, enzymatic activity, drug distribution and uptake, gene expression, among others [16]. Hence, PET technology is a standard indication in oncology, neurology, and cardiology [17, 18]. Furthermore, PET imaging enables to conduct longitudinal studies, thus, each animal can be its own control, eliminating inter-individual variability. Consequently, there is a considerable reduction of the number of animals (complying with the 3Rs principles), and cost associated with the experiments, enabling a quite fast availability of results.

The most used tracer is the glucose analogue, FDG. The uptake of 18F-FDG reflects the regional glucose consumption, allowing to study altered metabolic states associated with a wide variety of brain and heart disorders and diseases, as well as in oncology [2]. 18F-FDG gained relevance quite fast regarding the detection and biological characterization of tumour tissue [19], monitoring of tumour progression and transformation [20], treatment planning, and follow-up [21, 22]. Additionally, 18F-FDG is commonly used as a gold standard when assessing other imaging probes [23,24,25].

Before starting the experiment, perform cross-calibration of the clocks used and between PET scanner and dose calibrator. Osborne et al. [26] outlined a set of quality control guidelines for small animal imaging, encompassing expert-driven suggestions and alternatives for laboratories that do not have access to specialized phantoms for testing. Consult the Directive 2010/63/EU on the protection of animals used for scientific purposes. This Directive sets out legal requirements to implement the 3Rs principles of replacement, reduction, and refinement, harmonizing animal research legislation throughout the European Union, thus ensuring high standards of animal welfare and scientific research.

It is of utmost importance to select the appropriate procedures for each study and to ensure that the parameters set during the experiment remain constant. In 18F-FDG PET imaging, animal preparation and care can compromise the research results by possibly masking glucose metabolic activity of targets due to a high glucose consumption in different background tissues. Hence, the main purpose is to maintain and optimize the 18F-FDG in the target structures. In the following, a generally applicable protocol is outlined.

Mouse preparation and care prior to 18F-FDG PET imaging

Animal handling

During a PET study, where the interaction between the animals and the person performing the PET imaging is often of short duration and low frequency, mice should be handled indirectly in a home cage tunnel. Gouveia and Hurst [27] investigated the duration and frequency of handling required for effective familiarization with non-aversive handling methods, namely tunnel and cup handling, compared to tail handling. The findings demonstrated that 2 s of handling during cage cleaning were sufficient to familiarize mice with tunnel handling, while brief but more frequent handling was needed for cup handling. Even after experiencing complete immobilization by scruff restraint, mice familiarized with tunnel handling continue to interact with the handler, while maintaining the level of anxiety reduced. As for mice picked up by the tail, even when handled briefly and infrequently, they showed strong aversion and anxiety.

Housing

The minimum cage height and floor size should be 12 cm and 330 cm2, respectively, for one individual housed adult, according to the Directive 2010/63/EU. Isolate the mouse in a clean cage prior to 18F-FDG administration, removing food but always keeping water available ad libitum. In the work of Cao et al. [28], fasting promoted the general activity of the animal due to food search and, consequently, ingestion of litter or faeces (coprophagy), stress, and aggression. As prevention, remove bedding material during the fasting period, as Prior et al. [29] advocate, but maintain environmental enrichment to enhance the physical and psychological well-being of the mouse, fulfilling its need to find shelter. If relevant, consult the work of Baumans and Loo [30], who studied the possibilities of environmental refinement in terms of benefit to the animal.

Temperature

The metabolic rate of rodents is minimal and theoretically equal to the basal metabolism rate in the so-called thermoneutral zone (26–34 °C) [31]. In this range, body temperature is essentially regulated by conduction and heat convection [32]. Due to the body surface area to body volume ratios and metabolic rate (practically tenfold higher than larger mammalian species), the thermal mass to keep the body warm is also reduced, being associated with the high rodent metabolism which makes them more susceptible to hypothermia [33]. Therefore, set the room temperature between 18 and 26 °C [32, 34] and keep the body temperature of the mouse between 36.5 and 38 ºC employing heating devices (electrical or microwavable heating pads, warm air devices, hot water bags or bottles, circulation of hot water or hair blankets, lamps).

While planning your PET studies, consider that the body temperature peak occurs at the beginning of the scoto-phase, in contrast with the lowest temperature value that is observed at the middle of the photo-phase [35, 36]. Hence, fasting and anaesthesia effects on murine temperature may result in hypothermia. After 7–8 h of fasting, the metabolic rate may be suppressed up to 30% regarding the basal metabolism rate, accompanied by a temperature drop to 15 °C [37], and 15–20 min of general anaesthesia induces a decrease in temperature in the centre of the body of approximately 4–10 °C [38]. The possible consequences of hypothermia are cardiac arrhythmias, decreased protection against infections, long post-anaesthetic recovery time, a decrease in the minimum alveolar concentration values (increased toxicity potential) [39], and death within minutes (5–6 min at 20 ºC) [40]. If measurements or injections are performed on/into the tail vein, temperature monitoring is even more relevant since the blood flow is sensitive to temperature due to thermoregulation through the tail.

Blood glucose level and diet conditions

The circadian rhythm controls several physiological factors such as glucose metabolism: glucose concentration increases at the beginning of the photo-phase (presenting a peak in its middle) and decreases with the beginning of the scoto-phase, remaining at a base level [41]—Froy [42] summarized findings linking the circadian rhythm and several physiological parameters, including glucose metabolism. Whenever possible, schedule your PET experiments to the same time of day to minimize this variable.

Fasting is required to decrease blood glucose level, avoiding competition between glucose analogue (FDG) and endogenous glucose for its transporters to ensure uptake by the target. For more detail, Dolat and Sazgarnia [43] presented a review to compare the results of some clinical studies regarding the effects of fasting on the function of important organs; and Jensen et al. [41] presented a review about the effects of fasting in mice, providing evidence for fasting-induced changes in hormone balance, body weight, metabolism, hepatic enzymes, cardiovascular parameters, body temperature, and toxicology responses.

Concerning the optimal duration of fasting, Fuger et al. [4] kept mice fasting for 8 to 12 h and Woo et al. [44] for 20 h. Lee et al. [45] and Deleye et al. [46], both noticed that metabolic effects were attenuated after 20 h of fasting. According to the review of Rowland [47] on the important physiological effects of deprivation or restricted access to either food or fluids in common laboratory animals, a 24 h fasting represents an acceptable weight loss (< 10%), since they are physiologically prepared to tolerate acute suppression of food for periods in multiples of 24 h of duration and periodicity. However, fasting duration should be as short as possible when mice need to be repeatedly scanned within a short period, as it results in substantial weight loss [46]. A reasonable fasting period was found by Siikanen et al. [48]: 4 h was considered sufficient to achieve a stable blood glucose level, being near the 5–6 h advised for a better comparison to humans by Jensen et al. [41].

Prior to administering 18F-FDG, measure the blood glucose level (106–278 mg/dL [49])—a Glucometer or a similar device can be used for this purpose. Collect a 3 µl blood sample [50] using a syringe (0.6–1.0 mm needle) or a puncture in the lateral or ventral vein of the tail [32, 49]. If cannulation for 18F-FDG administration is unsuccessful, blood may be used from this puncture, and 18F-FDG should be injected on the opposite side [50]. Moreover, heat the site with hot water, for example, to induce vasodilation.

Anaesthesia

Inhalational anaesthesia is the method of choice and the commonly used inhalable anaesthetic is isoflurane. For isoflurane, the carrier gas is usually 100% oxygen, mitigating the effects of respiratory depression and hypoxia [39]. For more detail, the work of Gargiulo et al. [51] reviews the existing literature on anaesthetic protocols adopted in mice for molecular imaging studies and a review article that discusses the science of mouse anaesthesia was recently published by Navarro et al. [39].

In general, the procedure for inducing anaesthesia in mice involves the following steps:

  • Apply sterile unmedicated eye drops or ointment to prevent corneal damage and desiccation.

  • Place the mouse in the induction chamber where it is initially anaesthetized and latch the lid.

  • Set the vaporizer concentration between 4–5% + 0.8–1 L/min [51] until the mouse becomes immobile.

  • Reduce the vaporizer setting to the appropriate level (1–3% + 0.8–1 L/min [51]) for a stable heart rate and blood pressure.

  • The mouse should remain in the induction chamber for 3 to 5 min after initial anaesthetic induction before moving it to the PET instrument.

  • Once the mouse is ready, transfer it from the induction chamber onto the bed of the PET camera with a breathing device and mask that supplies a constant flow of the anaesthetic gas, keeping the mouse safely anaesthetized during PET scanning.

  • Note: if the animal has subclinical respiratory disease, resulting in hypoventilation and mild respiratory compromise, in the book entitled Biology and Medicine of Rabbits and Rodents, Harkness et al. [31] advised pre-oxygenation with 100% oxygen for 3 to 5 min prior to anaesthesia induction.

18F-FDG administration

First, make sure that clocks of the dose calibrator and scanner are synchronized (official local time within 1 min) [52]. Then, maintain the mouse warm for a normal metabolism, reducing interscapular brown fat uptake. For a fast systemic product biodistribution, apply intravenous bolus as administration route using indwelling cannulation to avoid partial paravenous injection (with local 18F-FDG retention in surrounding tissues), which is a common error due to the small size and fragility of murine tail veins [32, 34].

The volume of 18F-FDG administered should be less than 10% of the total blood volume [17]. Table 1 shows the recommended intravenous administration volume (V, mL/kg), maximum volume (VMax, mL), and needle gauge (G) [32]. Due to the small total blood volume of adult mice (1.6–3.2 mL [31]), the volume of 18F-FDG that can be administered is limited. The total activity contained in that volume must be concentrated to obtain an acceptable signal [34]. The activity value depends mainly on the performance of the PET system (sensitivity or enhanced technology), i.e. the minimum recommended 18F-FDG activity and PET acquisition duration must be adjusted to obtain a quality image within acceptable limits. One may decide to increase the activity or, preferably, the duration of the acquisition. In all cases, the administered activity, within the field-of-view, should not exceed the peak count rate capability of the scanner in use. Here are some examples of 18F-FDG activities applied in mice: Inveon (Siemens, 2009)—9 MBq [53], PETbox4 (UCLA, 2013)—1.5 MBq [54], ClairvivoPET (Shimadzu, 2016)—5.8 MBq [55], IRIS (2017)—7.5 MBq [56], MADPET4 (Bruker, 2017)—11.5 MBq [57], β-CUBE (Molecubes, 2018)—6.5 MBq [58], G8 (Sofie Biosciences, 2018)—1.96 MBq [59], PKU-PET-II (Peking University, 2018)—48 MBq [60], Xtrim-PET (Parto Negar Persia, 2019)—8 MBq [61], nanoScan (Mediso, 2021)—14 MBq [62], and easyPET.3D (RI-TE, 2021) 7.5 MBq [63].

Table 1 Good practice volume (V, mL/kg), possible maximum volume (VMax, mL), and recommended needle size (G) for intravenous injection in mice

After preparing the syringe, measure and record its activity, as well as its volume and time of measurement. Insert a short catheter (25–28 G needle) filled with heparinized saline right through the skin and into the lateral tail vein—excellent lighting would be helpful, and a light localized heating (warm water, lamp, or heating pad) simplifies the procedure. Advance the tip a couple of millimetres. At that point, there should be no resistance felt while administering 18F-FDG, if the placement has been successful. In the end, record the time of 18F-FDG administration, route, and site. Record also the residual activity measured in the syringe and catheter, including the time of measurement, to correct the injected dose, especially if the material is left in place during scanning, as it may lead to errors in quantification given the small volume [3].

During 18F-FDG PET imaging

Start positioning the mouse with a slight inclination and the head above the tail, enabling maximal costal movement and avoiding thoracic compression, as suggested by Balaban and Hampshire [33] in their review of the challenges of small animal non-invasive imaging. Then, check if the target is included in the field-of-view. Prevent the animal's eyes from being directly in contact with heat, inhalational anaesthetic agent, or any surface, and prevent as well short- or long-term adverse effects by adjusting the depth of anaesthesia or supportive care, maintaining the physiological state of the mouse as close as possible to normal, to reduce pain and stress and predict possible complications:

  • Visual monitoring includes watching the respiration (classifies the character of breathing), the colour of mucous membrane and skin (oxygenation state: blue—poor oxygenation; pale—poor blood perfusion), and the general behaviour of the mouse [3].

  • Electronic equipment available for small animal monitoring include rectal probe (body temperature, 36.5–38 °C), electrocardiogram (heart function and rate, 350–700 beats/min), small pneumatic pillow (respiration rate, 80–220 breaths/min), clip-on-sensors (pulse oximeter, 1.63–2.17 mL/g/h), cuff sensor (blood pressure, systolic: 133–160 mmHg; diastolic: 102–110 mmHg) [31, 32, 49].

Mouse care after 18F-FDG PET imaging

Set previously the room temperature between 18 and 26 ºC and prepare a cage stocked with paper and nesting material. During recovery, isolate the animal—if it is not fully awake, it may injure other animals in the cage—and maintain fluid supply ad libitum, as well as monitoring of the vital signs until the mouse is fully recovered. Lastly, disinfect the bed of the PET device with a suitable solution or dispose of the protective material, between groups of animals in the same health condition or individual animals, to prevent disease transmission and possible radioactive contamination due to urine loss during the exam.

Procedures for specific indications

Oncology

The time of day the PET experiments are performed at is especially important in oncological treatments, as 18F-FDG uptake may vary due to chronotolerance and radiotolerance. Several studies demonstrate this dependence of response to medication and radiation treatments on the time of the day: Sephton et al. [64] reported an earlier mortality for metastatic breast cancer patients with an abnormal cortisol rhythm, proving the importance of diurnal rhythm changes to survival; Lévi et al. [65, 66] discussed the importance of timing in cancer therapy, considering the circadian rhythm responsible for predictable changes in the tolerance and efficacy of anticancer agents, as well as in tumour induction or development; the increase in chemotherapy tolerance through chronomodulated chemotherapy in recurrent and metastatic head and neck squamous cell carcinoma was also demonstrated by Chen et al. [67]. Therefore, try to perform the PET studies at the same time of the day or reverse the day/night cycle by changing the light sequence in the vivarium.

During 18F-FDG administration and uptake period, keep mice under isoflurane anaesthesia to markedly decrease the background (brown fat and skeletal muscle), while the heart and blood clearance (liver and kidneys) signals are increased. This recommendation agrees with the protocol for imaging tumour xenografts with 18F-FDG by Fueger et al. [4] and the recent data from Pattison et al. [68], suggesting that isoflurane may improve biodistribution of 18F-FDG for tumour imaging.

According to the guidelines for the welfare and use of animals in cancer research, imaging sessions should be limited to 2–3 h in a 24 h period and do not exceed five sessions in a 1–2-week period with no more than one session/day [65].

Neurology

As previously mentioned, fasting is required in 18F-FDG PET imaging in mice. For neurological purposes, it is particularly important given that the cerebral 18F-FDG uptake varies inversely with blood glucose level (i.e. higher in fasting mice), as Wong et al. [5] concluded while studying the effects of dietary condition (fasting versus non-fasting) and blood glucose level on the kinetics and uptake of 18F-FDG in mice using intraperitoneal and intravenous routes. Simultaneously, they verified that the cerebral glucose metabolic rate does not differ considerably under different dietary states and administration routes. In agreement, Deleye et al. [46] observed a high brain uptake even after 20 h of fasting, recommending corrections for blood glucose levels to attenuate the effect of different fasting durations in brain 18F-FDG PET imaging.

Once more, schedule your PET experiments in order to favour target uptake. For instance, Krueger et al. [69] studied the effect of the circadian rhythm on 18F-FDG uptake in human tumour xenografts, every 4 h over a period from 8 AM to 8 PM. Their results revealed that 18F-FDG uptake in the tumour models analysed, and other organs (muscle, heart, and liver) remained stable throughout the day, while the circadian rhythm changed the uptake in the brain from noon to 4 PM.

Toyama et al. [70] proved that anaesthesia reduces neural activity and metabolism inducing a significant high standard deviation of binding potential. However, Bascuñana et al. [71] showed that conscious mice during 18F-FDG uptake had similar values of whole brain compared to mice anaesthetized with isoflurane, whereas the cortex and cerebellum uptakes were higher for conscious and unconscious mice, respectively. Moreover, Matsumura et al. [72] demonstrated that 18F-FDG uptake was almost the same as that in conscious state (no anaesthesia) when anaesthesia was induced 40 min after 18F-FDG administration. Hence, the 18F-FDG administration and uptake should be conscious over 40 min followed by unconscious PET imaging.

Concerning the administration route, intravenous infusion over approximately 5 min instead of a bolus injection showed an optimal accuracy of the initial portion of the blood curve for mouse brain kinetic modelling and allowed for longer blood sampling intervals as stated by Alf et al. and Vanhove et al. [34, 73]. The work of Mizuma et al. [74] established an in vivo brain PET imaging method in conscious mice to assess physiological neural functions, reporting a significant brain 18F-FDG uptake increase in mice receiving automatic injection via an indwelling catheter in the tail vein. Besides, the incidence of adverse effects (cardiovascular failure, for example) is decreased through a slow rate of injection [34].

Cardiology

In studies concerning thoracic malignancies, with or without myocardial involvement, fasting followed by high-fat diet can be advantageous for a greater suppression of physiological myocardial 18F-FDG uptake, as Langah et al. [75] showed by comparing the effects of four preparation protocols in rodents. Thus, the dietary conditions are again relevant to consider, as example, it was useful to detect inflammatory lesions in the early stage of cardiac sarcoidosis, in the work of Okumura et al. [76]. This prerequisite also induces a decrease in insulin levels. Consequently, glucose uptake in the background tissues is reduced enhancing 18F-FDG uptake in the target. The previously mentioned work of Wong et al. [5] showed that myocardium and skeletal muscle 18F-FDG uptake constants are highly affected by dietary conditions (lower in fasting mice) but not by blood glucose level. Similarly, Kreissl et al. [77] found a low glucose metabolic rate in the myocardium and skeletal muscle of fasted mice but not in the brain, while evaluating the effect of insulin stimulation and dietary changes on brain, myocardium, and skeletal muscle 18F-FDG kinetics and uptake in mice. In the same work, insulin injection increased the 18F-FDG uptake rate constant for myocardium (in non-fasted mice) and skeletal muscle (independently of the dietary state). However, in the brain, the parameter measured was not impacted by fasting or administration of insulin.

Like in brain, use conscious 18F-FDG administration and uptake over 40 min followed by unconscious PET scanning, since isoflurane alters myocardial glucose metabolism [4, 70] masking any effect of heparin or fasting. Thackeray et al. [78] tested suitable approaches for suppression of cardiomyocyte uptake in mouse models, confirming that the effect of isoflurane can be minimized by conscious 18F-FDG administration and uptake over 40 min prior to anaesthesia induction.

Dynamic studies

Unconscious 18F-FDG administration and uptake is a prerequisite for performing dynamic PET studies, i.e. the mouse is injected when placed inside the imaging instrument and scanned continuously during a certain time to identify changes in 18F-FDG biodistribution. These scans are required to measure the so-called arterial input function along with tissue time activity curve for kinetic modelling of PET data. In 2005, Laforest et al. reviewed several techniques for measurement of the blood input function in rodents and Meyer et al. evaluated a standardized arterial input function method in mice [79, 80]. Kinetic modelling with PET aims to estimate physiological parameters by extracting transfer rate constants, volumes of distribution, or binding potentials from the radiopharmaceutical to the target. Consult the publication of Amirrashedi et al. [81] for current trends in preclinical PET imaging, including the key factors to consider for kinetic modelling.

Final considerations

Technical, physical, and physiological parameters can affect the outcomes obtained from preclinical imaging techniques. Therefore, standardization of experimental procedures is essential to achieve repeatable, reproducible, and reliable data. Animal preparation and care particularly impact the recorded 18F-FDG PET data. Current publications concerning the research report improvement and the quality and reliability of published research amplification. Still, none defines a global standard protocol for preclinical 18F-FDG PET imaging procedures. The present overview represents a key milestone in the effort to acquire 18F-FDG PET more robust data with less variability, enabling the use of fewer mice and ensuring high standards of animal welfare, thus complying with the 3Rs policy. This guideline focuses on the practical steps for mice preparation and care, which can be controlled and suggested for broad adoption prior, during and after 18F-FDG PET imaging. Overall, researchers are encouraged to follow the recommendations outlined herein during the planning and performing of their studies, assuring savings in cost and time, as better results.