Background

Immunological test systems for diagnostics of type I hypersensitivity currently involve whole allergen extracts, highly purified allergens from the extracts, and their recombinant analogues obtained by gene engineering techniques. The majority of test systems use extracts and allergen mixtures.

Extracts are heterogeneous protein mixtures isolated from natural materials containing allergenic and non-allergenic material. They allow estimating the reactivity of a patient’s serum toward all potentially allergenic components. Furthermore, allergen extracts contain both species-specific proteins and components with epitopes demonstrating high cross-reactivity with proteins from other origins. This complicates the identification of the primary allergy source [1]. Moreover, the standardization of extracts composition is yet another application problem. It has been shown in a number of studies that allergen extracts produced by number of manufacturers differ considerably in composition and activity [2, 3], leading to discrepancy between analysis results in various test systems [4].

In vitro allergy diagnostics with standardized recombinant allergens and highly purified components of allergen extracts enables getting more reproducible results. It should be noted that the current number of recombinant allergens does not cover the entire spectrum of potentially allergenic proteins present in extracts. This is why the use of only individual protein components is not recommended to identify the allergy source, because it may give a false-negative result in the case of immune response to a protein that is not included in that range. Therefore allergy diagnostics is mostly carried out taking into account clinical history, results of skin tests and/or measurement of sIgE to allergen extracts, whereas molecular-based allergy diagnostics employing individual allergen components is applied for polysensitized patients [5, 6] in order to identify the main sensitizing component, for example, for subsequent allergen specific immunotherapy (ASIT) [7, 8] and for prediction and monitoring of treatment efficiency [9].

Some cases showed that analysis with recombinant allergens increased the diagnostics accuracy [10, 11]. Currently, the most widely used test system based only on recombinant allergens is ImmunoCAP ISAC®, manufactured by Phadia (at present trademark of Thermo Fischer Scientific Inc.).

Though individual allergen proteins have been used in studies on the efficacy of diagnostics with recombinant allergens since the 1990s [12], there are a few examples of simultaneous multiplex analysis of sIgE both to whole allergen extracts and their recombinant proteins. It seems that the biological microchip is the most convenient instrument for simultaneous analyzing of sIgE to both individual molecules whole allergen extracts in one serum sample during one assay. The purpose of this work was to compare the diagnostic accuracy of the method for detection of type I hypersensitivity using whole allergen extracts of birch pollen, cat dander and timothy grass pollen (diagnostically significant in central Russia) and their corresponding recombinant proteins on the biological microchips. Comparison of the diagnostic accuracy was made using receiver operating curve (ROC) analysis.

Methods

Patients

The study was carried out using patients’ blood serum samples provided by the Federal State Budgetary Institution Polyclinic #1 of the Business Administration for the President of the Russian Federation. Primary selection of the patients was based on the presence of allergy symptoms in the anamnesis: seasonal allergic rhinitis during early spring or summer period as well as year-round rhinitis remissive in outdoor conditions.

Sera from patients demonstrating positive skin prick test (SPT) (reaction >2 mm wheal diameter at least to one of the extracts: birch pollen, timothy grass pollen or cat dander) were chosen for the study. Sera from patients exhibiting positive SPT to the allergen were approved as positive to this allergen. Sera from patients demonstrating negative SPT to the allergen (reaction <2 mm wheal diameter) were approved as negative to this allergen. SPT was performed using salt aqueous extracts, histamine dihydrochloride 0.1 % as a positive control and NaCl 0.9 % as a negative control.

Totally, sera from 139 patients were chosen: 56, 73, and 33 of them were sensitized to birch pollen, cat dander and timothy grass pollen, respectively. The data of the patients’ sera were analyzed anonymously.

Analysis of blood sera on biochips

Patients’ blood sera were analyzed by the multiplex fluorescent immunoassay method on biological microchips/biochips (EIMB RAS).The biochip manufacturing and procedure of biochip-based immunoassay of allergen-specific IgE have been described previously [13, 14]. Biochip was the set of semispherical hydrogel elements of 0.1 nL containing the following immobilized antigens: birch pollen, recombinant allergens of birch pollen—Bet v 1 and Bet v 2; cat dander, recombinant allergen of cat dander—Fel d 1; timothy grass pollen, recombinant allergens of timothy grass pollen—Phl p 1, Phl p 5, Phl p 7 and Phl p 12 (Fig. 1).

Biochip-based assay allows analyzing of sIgE in the range from 0.15 to 100 IU/ml [12]. We used the concentration 0.35 IU/ml as the cut-off for distinguishing between positive and negative results according to the WHO recommendation.

Statistical analysis

ROC analysis was performed to estimate the diagnostic accuracy of using allergen extracts and their recombinant proteins. The ROC curves were built and the areas under the curve (AUC) were calculated using MedCalc program, version 15.2.2 [15, 16].When constructing ROC curves for a mixture of recombinant proteins, the result was considered as positive if at least one of the proteins showed positive signal for a chosen cut-off.

Results and discussion

In this study we analyzed 139 blood serum samples from patients with allergy to birch pollen (56 cases), cat dander (73 cases) or timothy grass pollen (33 cases). These allergens are most widespread and diagnostically significant in central Russia. For each allergen we determined the level of sIgE to the natural extracts and to their recombinant proteins: Bet v 1, Bet v 2, Phl p 1, Phl p 5, Phl p 7, Phl p 12, Fel d 1. Fig. 1 shows the biochip structure (a) and the example of fluorescent image after the immunoassay (b) Data were analyzed with ROC analysis. The results are presented in Table 1 and Fig. 2. Statistical analysis of the results for diagnostic accuracy of using the natural birch pollen extract and its recombinant components has shown that the AUC was 0.95 both for the extract and for its major recombinant protein Bet v 1 or 0.83 for the minor component profilin Bet v 2. For the combination Bet v 1 + Bet v 2 the AUC was the same and equaled to 0.95.

Fig. 1
figure 1

a Schematic arrangement of immobilized allergen extracts, their recombinant proteins, and marker spots (М) on biological microchips. Each antigen was immobilized in quadruplicate. Non-specific binding was monitored using spots that contain no proteins (empty gel). b An example of fluorescent image of a biochip after immunoassay of a blood serum sample from a patient with allergy to birch pollen and cat dander

Table 1 Comparison of ROC analysis results for methods of identification of type I hypersensitivity using allergen extracts, recombinant components, and recombinant components groups
Fig. 2
figure 2

Comparison of ROC curves for methods of identification of type I hypersensitivity using allergen extracts, recombinant components, and combinations of recombinant components. a 1-Bet v 1, 2-Bet v 2, 3-Bet v 1 and Bet v 2, 4—birch pollen extract. b 1-Fel d 1, 2—cat dander extract and Fel d 1, 3-cat dander extract. c 1-Phl p 1, 2-Phl p 5, 3-Phl p 7, 4-Phl p 12, 5-Phl p 1, Phl p 5, Phl p 7, Phl p 12, 6-timothy grass pollen extract

Analysis of diagnostic accuracy of using the cat dander allergen has shown that the AUC both for the extract and for the recombinant Fel d 1 was the same and equaled to 0.76.

The results obtained for the birch and cat allergens are in accordance with available literature data concerning the frequency of occurrence of immune response to various allergenic components. Sensitization to Bet v 1 as major allergen occurs in a considerable fraction of patients who have allergy to birch pollen (60–90 %) [17], while as little as 12 % of patients have sensitization to other recombinant components (Bet v 2, Bet v 4, Bet v 6, Bet v 8, etc.) [18]. According to various estimates, sensitization to Fel d 1 occurs in 80–95 % of patients with hypersensitivity to cat dander, which results in a high efficiency of the diagnostic test employing this recombinant protein. From the data obtained it can be assumed that for diagnostics purposes allergen extracts can be replaced by a combination of the corresponding recombinant components in case of birch pollen or by the major recombinant protein Fel d 1 for cat dander. However it should be noted that both false positive and false negative results may be obtained in this case. For example, interaction of Fel d 1 allergen with IgE specific to dog dander [19] is observed in 25 % of patients with allergy to dog dander. On the other hand, a patient may show sensitization to minor allergen components that only occur in whole allergen extracts. In these cases, the usage of only recombinant allergens could give false negative results.

To estimate the diagnostic accuracy of analysis in case of allergy to timothy grass, we used natural extract and four different recombinant proteins: Phl p 1, Phl p 5 and Phl p 7, Phl p 12. The obtained data show that the AUC for the extract was 0.83, for Phl p 1 and combination of Phl p 1 and Phl p 5 was 0.71, whereas for Phl p 5, Phl p 7 and Phl p 12 AUC ranged within 0.65–0.66. The AUC was 0.74 for the combination of 4 recombinant components.

Phl p 1 and Phl p 5 are considered to be the major timothy grass pollen components. According to literature data, Phl p 1 is involved in sensitization process for more than 90 % of the patients with timothy grass allergy  [20, 21], so sIgE to Phl p 1 is sufficient for diagnostics of timothy grass pollen allergy for Central European population [22]. However, according to our data, the accuracy of using this recombinant component is lower than using the extract. The addition of the major component Phl p 5, which occurs in up to 80 % of timothy grass pollen sensitized patients, does not increase the AUC value.

The efficiency of the diagnostics with using solely minor cross-reactive components calcium-binding protein Phl p 7 and profilin Phl p 12 is rather poor, but their inclusion in the analysis along with the major proteins improves AUC from 0.71 to 0.74. Thus, it can be concluded that, unlike the situations with birch and cat allergens, the use of available recombinant proteins to timothy grass pollen does not allow reaching the diagnostic accuracy for the natural extract.

In the current work biochips were used as a research instrument for the comparison of diagnostic approaches based on natural extracts and their recombinant components. Furthermore, after modification and validation the biochips could appear as practically significant method combining second- and third-line approaches of allergy diagnostics.

Conclusions

Multiplex analysis of serum samples obtained from patients with allergy to birch and cat dander carried out on biological microchips has shown that the use of recombinant allergens give accuracy comparable to the natural extracts. In the case of timothy grass, the use of the recombinant components may be insufficient for the identification of allergy source. In some cases, simultaneous analysis using both extracts and their individual protein components is preferable. This approach can be efficiently implemented by means of the protein biochip technology.