Background

Video-assisted thoracoscopic surgery (VATS) has been widely used in the diagnosis and treatment of a variety of thoracic diseases. To facilitate endoscopic visualization of intrathoracic structures, collapse of the ipsilateral lung is required. One-lung ventilation though double-lumen endotracheal tube (DLET) intubation, provides excellent intrathoracic exposure. However, DLET intubation can lead to a higher incidence of tracheobronchial injury, a longer time of DLET placement and an increased cost for patients [1, 2]. Besides, DLET intubation is not suitable for pediatrics or patients with difficult airways. Two-lung ventilation with CO2 insufflated into unilateral thoracic cavity can also afforded an excellent working space to perform many thoracoscopic surgeries including sympathectomies, wedge resections and esophagectomies [1, 3, 4]. However, insufflation of CO2 creates physiologic effects similar to a tension pneumothorax. Acute changes of intrathoracic pressure during the initial insufflation of CO2 can compromise cardiopulmonary function and patient safety [5,6,7]. Currently, few reports have observed the characteristics of intrathoracic pressure after artificial CO2 pneumothorax has been established and its potential impacts on cardiorespiratory function. During laparoscopy, Jacobs found that the real intraabdominal pressure peaks temporarily reached up to 1.8 times of the pre-set value, namely pressure overshoot [8].

Therefore, we conducted the present study to observe the real intrathoracic pressure during artificial pneumothorax and compared the hemodynamic and respiratory effects of intrathoracic pressure overshoot in patients who underwent thoracoscopic esophagectomy with two different insufflators.

Methods

Study design and ethics

This study was approved by the ethics committee of Daping Hospital, Army Medical University (IRB 2014-9) and written informed consent was obtained from all subjects participating in the trial. The trial was registered prior to patient enrollment at clinicaltrials.gov (NCT02330536, Principal investigator: Qingxiang Mao, Date of registration: December 24, 2014). All patients were randomly allocated to two groups, Stryker insufflator (STR) group and Storz insufflator (STO) group, by a computerized randomization table. This article adheres to the applicable Consolidated Standards of Reporting Trials (CONSORT) guidelines.

Patient population

From August 2016 to July 2020, a total of 228 patients who were scheduled for elective thoracoscopic esophagectomies with artificial pneumothorax were enrolled (Figure S1). Inclusion criteria included a definite diagnosis of esophageal cancer and a willingness to be treated by thoracoscopy combined with carbon dioxide artificial pneumothorax. Exclusion criteria included: age > 80 years, severe cardiovascular or respiratory diseases, American Society of Anesthesiologist (ASA) Grade > 3 and being unable or unwilling to give written consent to this clinical study. Patients who were converted to thoracotomy (including 3 cases with unexpected massive hemorrhage and 5 cases with severe thoracic adhesion), had significant intraoperative blood loss (> 400ml) or sustained severe arrhythmia during the operation were also excluded from the study. This study finally included 172 males and 28 females with a mean age of 66.75 years (Table 1). All of the surgeries were performed by the same team in the Department of Thoracic Surgery, Daping Hospital, Army Medical University (Chongqing, China).

Table 1 Baseline characteristics of participating patients, values are mean (SD) or number (proportion)

Anesthesia management

At admission to the operating room, the patients were monitored with electrocardiogram (ECG), oxygen saturation (SpO2), non-invasive blood pressure (NBP) and bispectral index (BIS). Anesthesia was induced with midazolam (30 to 40 µg.kg−1), sufentanil (0.5 µg.kg−1), propofol (1.5 to 2.0 mg.kg−1), and cisatracurium (0.2 mg.kg−1). Trachea was intubated with a single-lumen endotracheal tube. Anesthesia was maintained with inhaled sevoflurane and continuous infusion of remifentanil and propofol to maintain the BIS value within 40-60. Catherization of the right internal jugular vein and the left radial artery were performed to monitor central venous pressure (CVP) and arterial blood pressure (ABP), respectively.

Patients were infused with 6 ml.kg−1 of Ringer’s lactate solution before anesthesia induction and infused with 6-8 ml.kg−1.hour−1 fluid including crystalloid and colloid (2:1 ratio of crystalloid to colloid) during the operation. If intraoperative systolic blood pressure (SBP) was lower than 20% of the baseline or less than 90 mmHg (hypotension) and lasted more than 3 min, norepinephrine (0.01-0.1 µg.kg−1.min−1) would be administrated to maintain the blood pressure over the hypotension threshold.

Mechanical ventilation with 100% oxygen was initiated with a tidal volume of 7 ml.kg−1 and respiratory rate of 13 breaths.min−1 (non-artificial pneumothorax period). During artificial pneumothorax, the tidal volume was 5ml.kg−1 and the respiratory rate was 20 breaths.min−1, and the inspiratory: expiratory ratio was maintained at 1:2.

Induction and maintenance of artificial pneumothorax

Patients were placed in a 30° left-lateral tilt position. After a disposable trocar was inserted into the pleural cavity, the CO2 was insufflated into the thoracic cavity via an insufflator. The insufflation pressure and the flow rate were set as 8 mmHg and 20 l.min−1. An invasive blood pressure transducer filled with saline was connected to another trocar and used to monitor the real-time intrathoracic pressure. The patients were randomly divided into two groups: the STR group, in which artificial pneumothorax was performed by Stryker insufflator F102 (Stryker corporation, Kalamazoo, MI, USA); the STO group, in which artificial pneumothorax was performed by Storz insufflator 0U32540 (Karl Storz GmbH & Co, Tuttlingen, Germany).

Surgical procedure of VATS

Briefly, patients were intubated with a single-lumen endotracheal tube and placed in the semi-prone position. Four trocars were placed at four sites, including the camera trocar (12 mm) in the sixth intercostal space on the anterior axillary line, one 5-mm trocar in the sixth intercostal space on the anterior axillary line, one 5-mm trocar in the sixth intercostal space on the posterior axillary line, and one 12-mm trocar in the eighth intercostal space on the posterior axillary line (Fig. 1a). After artificial pneumothorax was performed, the esophagus was mobilized caudally to the esophageal hiatus and cranially to the thoracic entrance, accompanied by lymphadenectomy of supraphrenic, paraesophageal, peripulmonary vein, subcarinal, right and left recurrent laryngeal nerve, and paratracheal stations (Fig. 1b and d).

Fig. 1
figure 1

Thoracoscopic procedures of thoracoscopic-laparoscopic esophagectomy. a Patients were placed in the semi-prone position and four trocars were placed at four sites. b-d Intraoperative images during artificial pneumothorax

Data collection

A video camera was used to record the changes in intrathoracic pressure, ABP, ECG, SpO2, CVP, and airway pressure of the patients in two groups during artificial pneumothorax. The arterial blood gas was measured at 5 min before artificial pneumothorax started (T1) and 5 min after artificial pneumothorax stopped (T2). The usage of vasopressors and the postoperative recovery time were also recorded.

Statistical analysis

Numerical data were reported as mean ± standard deviation and was statistically analyzed using 2-sided t test or two-way ANOVA followed by the post hoc Turkey method. Count data were analyzed with the Chi-square test. Multiple linear regression analysis was conducted for the correlation analysis between pressure overshoot and cardiopulmonary function changes. A value of P < 0.05 was considered statistically significant. SPSS 19.0 software (SPSS Inc., Chicago, IL, USA) was used for statistical analysis. Almost 200 patients per group were required to detect a 15% difference in the pressure overshoot value between two group (alpha = 0.01, power = 0.9). The level of significance was set at P < 0.01.

Results

Patient characteristics

No significant difference was found in the height, weight, age, and gender between the STR group and the STO group (P > 0.05). In addition, the clinical parameters including operation time, pneumothorax time, and anesthesia recovery time were similar between the two groups (P > 0.05). The vasopressor usage rate was significant higher in STR group than in STO group (STR: 68%, STO: 43%, P = 0.001). However, there was no significant difference between the two groups in intraoperative blood loss (STR: 122 ± 45.9 ml, STO: 118 ± 55.8 ml, P = 0.790), fluid replacement volume (STR: 2182 ± 503.7 ml, STO: 2290 ± 455.4 ml, P = 0.489) and urine volume (STR: 454 ± 241.1 ml, STO: 453 ± 206.2 ml, P = 0.991) (Table 1).

The profiles of thoracic pressure overshoot and its impact on the circulatory and respiratory functions

Constant fluctuations in intrathoracic pressure of the two groups were observed during the entire artificial pneumothorax (Fig. 2a and d). One possible reason was that the surgical suction and gas leakage from the trocar puncture would decrease the intrathoracic pressure and re-initiate the CO2 insufflation. In addition to CO2 insufflation, mechanical ventilation can also cause periodic fluctuations in intrathoracic pressure of the pneumothorax side. However, this increased intrathoracic pressure induced by the mechanical ventilation was generally less than 3 mmHg (Fig. 2b and d). Thus, in this study, the threshold of intrathoracic pressure overshoot events caused by CO2 insufflation was defined as an increased pressure that was greater than 3 mmHg. We found that pressure overshoot occurred in both of the two groups during the artificial pneumothorax, and there was no significant difference in the frequency (STR: 38 ± 7.4 times.hour−1; STO: 32 ± 9.9 times.hour−1, P = 0.238) and the magnitude (STR: 8.9 ± 1.60 mmHg; STO: 9.8 ± 3.13 mmHg, P = 0.551) of pressure overshoot (defined as the maximum thoracic pressure minus eight mmHg) between the two groups. However, a longer pressure overshoot duration (5.3 ± 0.86 s vs. 1.2 ± 0.31s, P < 0.01) as well as a larger increase in peak airway pressure (4.8 cmH2O vs. 0.9 cmH2O, P < 0.01), a greater rise in CVP (8.2 cmH2O vs. 4.9 cmH2O, P < 0.01), a greater decrease in SBP (5.6 mmHg vs. 1.1 mmHg, P < 0.01), and a higher percentage of patients with significant SBP drop (drop > 10 mmHg) were found in the STR group than those in the STO group. No significant difference was observed in the magnitude of heart rate reduction (0 ± 1.4 bpm vs. 0 ± 1.3 bpm, P = 0.534) and the percentage of patients with SpO2 reduction (0% vs. 0%, P = 1.000) between the two groups (Table 2).

Fig. 2
figure 2

Examples for intrathoracic pressure changes in the STR and the STO group. a Typical graph of intrathoracic pressure changes in the STR group during artificial pneumothorax within 60 min. b A detailed view of the changes in intrathoracic pressure during the 500 – 560 s in a. c Typical graph of intrathoracic pressure changes in the STO group during artificial pneumothorax within 60 min. d A detailed view of the changes in intrathoracic pressure during the 350 – 410 s in c

Table 2 The profiles of intrathoracic pressure overshoot and its impact on the circulatory and respiratory functions

Cardiovascular inhibition caused by the thoracic pressure overshoot was highly correlated with the duration of overshoot

Since the higher incidence and magnitude of SBP drop and a longer overshoot duration were found in the STR group, we assumed that whether the cardiovascular inhibition caused by the thoracic pressure overshoot depended on the overshoot duration. To address this question, we pooled the data from the STR and STO group together and performed the multiple linear regression analysis for the correlation between the magnitude of thoracic pressure overshoot, the duration of pressure overshoot, the increase of peak airway pressure, the increase of CVP and the SBP drop, respectively. The analysis indicated that there was no significant correlation between the SBP reduction and the intrathoracic pressure overshoot (R = 0.00, P = 0.996) (Fig. 3a) while a strong correlation was found between the decrease in blood pressure and the duration of intrathoracic pressure overshoot (R = 0.76, P < 0.01) (Fig. 3b). Additionally, the decrease in blood pressure and the change in peak airway pressure were moderately correlated (R = 0.47, P < 0.01) (Fig. 3c). A weak correlation was also detected between the decrease in blood pressure and the increase in central venous pressure (R = 0.22, P < 0.01) (Fig. 3d).

Fig. 3
figure 3

The correlation between the SBP reduction and the magnitude of pressure overshoot, duration of overshoot, increase of peak airway pressure and increase of CVP, respectively. a Scatter plot of the magnitude of pressure overshoot and the SBP reduction. b Scatter plot of the duration of overshoot and the SBP reduction. c Scatter plot of the increase in peak airway pressure and the SBP reduction. d Scatter plot of the increase in CVP and the SBP reduction. e Trend graph showed the relationship between the duration of pressure overshoot and the incidence of SBP drop (defined as SBP drop was more than 5 mmHg during pressure overshoot)

To investigate the effect of duration of pressure overshoot on the incidence of SBP drop with different magnitudes (> 3 mmHg, > 10 mmHg), the duration of intrathoracic pressure overshoot was divided by one second interval and into 10 levels as below: ≤ 1 s, 1-2 s, 2-3 s, 3-4 s, 4-5 s, 5-6 s, 6-7 s, 7-8 s, 8-9 s and > 9 s. The incidences of SBP drop with different magnitudes all increased with the overshooting time. When the overshoot time exceeded 2 s, the incidence of more than 3 mmHg SBP drop rapidly increased to 57.14%, and when the overshoot time exceeded 6 s, the incidence of more than 10 mmHg SBP drop rapidly increased to 45.45%, which indicated that there was a threshold effect of overshoot time on the SBP drop (Fig. 3e).

Changes of blood gas analysis after artificial pneumothorax

To investigate the effect of artificial pneumothorax on the respiratory function, we compared the blood gas analysis results between the STR group and the STO group at 5 min before the artificial pneumothorax (T1) and 5 min after the artificial pneumothorax stopped (T2). The results showed that arterial partial pressure of CO2 (PaCO2) significantly increased (P < 0.01) while the pH value (P < 0.01) and base excess (BE) significantly decreased (P < 0.01) at T2 than those at T1 in both groups. No significant difference was found in blood lactic acid (Lac) and arterial partial pressure of oxygen (PaO2) between T1 and T2 in both groups. There were no statistically differences of blood gas analysis between two groups at either T1 or T2 (P > 0.05) (Table 3).

Table 3 Comparison of blood gas analysis before and after artificial pneumothorax between the STR and the STO groups

Discussion

CO2 artificial pneumothorax has been widely used in cardiothoracic surgery, which not only helps surgeons obtain the optimal visualization of surgical field and operation space, but also makes it possible to perform thoracoscopic surgery for patients who are not suitable for intubation with DLET(such as children or patients with difficult airway) [9, 10]. However, CO2 artificial pneumothorax may also have several limitations, including the occurrence of intrathoracic pressure overshoot. Although the insufflation pressure of CO2 is usually fixed, the real thoracic pressure is not presumably constant due to the need for surgical suction and gas leakage from the trocar puncture. The insufflator will insufflate CO2 into the chest cavity again to reach the pre-set pressure. The insufflation volume may exceed the actual required volume because of the delayed response between the pressure monitoring system and the flow control regulator [11]. Then the intrathoracic pressure will be higher than the pre-set value, that is, the pressure overshoot that occurs. Modern insufflators have been adopted by several ways to reduce the magnitude and frequency of pressure overshoot, including constant pressure and variable flow mode (namely, the closer the insufflation pressure is to the pre-set value, the lower the flow rate is), active pressure relief, etc. However, intrathoracic pressure overshoot still cannot be completely eliminated. Jacobs et al. reported that the intra-abdominal pressure is not constant during CO2 pneumoperitoneum, and the pressure overshoot reached up to 1.8 times of the pre-set value [8]. In this study, we also found that intrathoracic pressure is not constant during CO2 artificial pneumothorax, which was consistent with previous reports [8, 11, 12]. We also found that the intrathoracic pressure overshot value can reach up to 7 times of the pre-set value.

Intrathoracic pressure overshoot can impair the cardiovascular function, which is mainly due to the compresses on the heart and large blood vessels, leading to the obstructions in venous return and the inhibition of cardiac systolic and diastolic functions [13, 14]. The severity of cardiovascular instability depends on the pneumothorax pressure, the insufflation flowrate, and the individual compensatory to pneumothorax [15]. Currently, CO2 pneumothorax with insufflation pressure of 8 mmHg is widely accepted as the following cardiovascular compromise is less obvious [16, 17]. In previous reported cardiovascular collapse events during CO2 insufflation, the insufflation pressure was higher than 8 mmHg which implied a potential higher intrathoracic pressure overshoot and a more severe cardiovascular comprise [5, 6]. In present study, the CO2 insufflation pressure was set at 8 mmHg. However, we still found that pressure overshoot during artificial pneumothorax could compromise cardiovascular function as evidenced by the significant SBP drop (>10 mmHg) events affected 51% patients in the STR group and 9% patients in the STO group.

Interestingly, we found that the frequency and magnitude of pressure overshoot in the STO group were not significantly different from those in the STR group except for the shorter duration of pressure overshoot (1.2 s vs. 5.3 s). The pressure overshoot in the STO group appeared to have less effect on cardiopulmonary function. The percentage of vasopressors usage, the increase in peak airway pressure, the increase in central venous pressure and the decrease in blood pressure during artificial pneumothorax in the STO group were all less than those in the STR group. Then a correlation regression analysis was performed and found that SBP drop was highly correlated with the pressure overshoot duration (R = 0.76) while not the magnitude of pressure overshoot. The longer the pressure overshoot lasts, the higher the risk of blood pressure reduction goes. We also found that there was a threshold effect of overshoot duration time on the SBP drop which is that overshoot duration longer than 2 or 6 s will lead to a significant increased incidence of SBP drop > 3 mmHg or incidence of SBP drop > 10 mmHg, respectively. These results imply that different brands of insufflators may have different characteristics of pressure overshoot due to discrepancies in electrical mechanism, which may exert distinctive effects on the cardiovascular function. Besides, a significant correlation was also detected between SBP reduction and peak airway pressure increase. One possible explanation may be since the peak airway pressure always occurred in the inspiratory phase, and the inspiration of lung would reduce the intrathoracic space and further increase the pressure overshoot. Then the further increased pressure overshoot may lead to SAP reduction. Whether the increase in peak airway pressure is the direct cause of the decrease in SBP still needs to be confirmed by further experiments.

CO2 pneumothorax can lead to the increased airway pressure, the decreased oxygenation, and the hypercapnia [18]. In addition to the impact of artificial pneumothorax, intrathoracic pressure overshoot can further increase the peak inspiratory pressure which may cause more stress damage to the pulmonary alveoli [19, 20]. However, in our study there was no significant decrease in the oxygenation index of the STR group when compared with the STO group. This could be because that the duration and magnitude of the increased peak inspiratory pressure of the STR group are not long and high enough to exert significant alveolar damage. Both the STR group and the STO group had obvious hypercapnia and acidosis after artificial pneumothorax, but there was no significant difference between the two groups, indicating that the difference of pressure overshoot between those two groups have little effect on the absorption and discharge of CO2.

This study has several limitations. First, this is a single-center randomized controlled study and we excluded patients who were converted to thoracotomy, had massive intraoperative blood loss or sustained severe arrhythmia during the operation, the potential selection bias may limit generalization of our results to other populations. Second, we only analyzed the intraoperative circulatory and respiratory functions during intrathoracic pressure overshoot, the potential damage of pressure overshoot to other organs and the long-term outcomes of patients were not studied. Another potential effect of capnothorax is that intraoperative venous bleeding may be covered when intrathoracic pressure exceeds venous pressure and the negative intrathoracic pressure during spontaneous inspiration could siphon blood into the pleural cavity [21]. This potential risk for patient’s outcome needs further evaluation. Third, we only compared the overshoot characteristics of two types of insufflators. A further work is needed to clarify the potential effects of other types of insufflators on the intrathoracic pressure. Last, all the patients in this study are adults, whether the pressure overshoot has a different cardiopulmonary impact on children patients needs to be further explored.

In conclusion, during CO2 artificial pneumothorax, intrathoracic pressure overshoot can occur and lead to cardiovascular compromise. The longer the pressure overshoot lasts, the higher the risk of blood pressure reduction goes.