Abstract
It is significant to develop an efficient early detection and prediction method for ovarian cancer via a facile and low-cost approach. To address such issues, herein, we develop a novel circulating tumor cell (CTC) detection method to sensitively detect ovarian cancer by using a flexible graphene-based biosensor on polyethylene terephthalate (PET) substrate. The results show that the graphene-based flexible biosensor demonstrates sensitive and rapid detection for ovarian cancer cells: it delivers obvious different responses for cell culture medium and cancer solution, different cancer cells and cancer cell solution with different concentrations; it demonstrates high sensitivity for detecting several tens of ovarian cancer cells per ml; moreover, the flexible graphene biosensor is very suitable for rapid and sensitive detection of ovarian cancer cells within 5 s. This work provides a low-cost and facile graphene biosensor fabrication strategy to sensitively and rapidly detect / identify CTC ovarian cancer cells.
Graphical Abstract

Introduction
Ovarian cancer is the second most common gynecological cancer and has the highest mortality among the gynecological cancers [1, 2]. So far, the ovarian cancer patients are generally diagnosed very late due to the nonspecific symptoms of ovarian cancer and the lack of effective early screening methods. Imaging combined with carbohydrate antigen CA125 can be used in detection, diagnosing recurrences after surgery or chemotherapy. CA125 is not a single accurate marker for ovarian cancer because it is affected by numerous factors and has high false positive predictive value. The sensitivity of elevated CA125 (> 35 U/mL) which we used in the diagnosis of the recurrence of ovarian cancer is less than 70% [3]. Ultrasound examination and radiological examination also have no adequate sensitivity nor specificity in early detection and recurrence diagnosis. The 5-year survival rates of the ovarian cancer at stage I and II are 90% and 70% separately [4]. Despite of the advance in the surgical treatment and adjuvant therapy, the 5-year survival rate of the ovarian cancer at advanced stage is less than 30% [4]. Early detection of ovarian cancer is related to the obvious higher 5-year survival rate, and early diagnosis of recurrence is also important. Several new approaches such as the TP53 autoantibody, DNA methylation assays, microRNA algorithms, Pap-like cytologic analysis have been reported to improve sensitivity of the early detection of ovarian cancer [5]. However, it is urgent but still challengeable to develop new detection method with higher sensitivity for all stage of ovarian cancer.
Recently researchers found that the early-stage tumors can shed into cancer cells into bloodstream and cause metastasis [6]. Cells enter into the peripheral bloodstream through intravasation from primary tumors, recurrences, or metastases called circulating tumor cells which can be used as diagnostic or prognostic biomarkers for solid tumors [7]. The CTCs are rare in peripheral blood, and the detection methods need high sensitivity and specificity. In recent years, immunomagnetic separation, microfluidic separation, filter-based methods, and ligand-targeted PCR were reported in the detection of CTC [8,9,10,11]. To date, cell search system of Janssen Diagnostics is the only US Food and Drug Administration (FDA)-approved CTC detection method, which can be used to monitor patients with metastatic breast, colorectal, and prostate cancers [12,13,14]. The detection of CTC in ovarian cancer provides a noninvasive diagnostic method and has an advantage when it is difficult for biopsy. However, the detection rate of CTC in early stage of ovarian cancer is still low. The new methods to detect CTC with higher sensitivity are still needed. If we can easily detect CTC in ovarian cancer patients, it can be useful both in the early tumor detection, monitoring recurrence, and treatment effect.
Graphene, a two-dimensional semiconductor was isolated by Andre Geim and Kostia Novoselov in 2004 [15]. Recently, graphene-like 2D materials have been widely applied in sensing and bio-sensing, energy conversion and storage, catalysis, composites and coatings, electronics and biomedical field [15]. Graphene sensor is a promising candidate to detect cancer biomarkers due to its unique structure and excellent electrical performance, which had been developed to detect carcinoembryonic antigen, prostate specific antigen, carbohydrate antigen 19–9 and 15–3 [16,17,18,19]. Compared to conventional graphene-based biosensors fabricated on rigid SiO2/Si substrates by conventional photolithography, electrode evaporation, lift-off and sensor package process using precious facilities, it is significant to develop a low-cost and facile approach to fabricate flexible graphene-based biosensors with high sensitivity and rapid detect speed.
To address such issues, herein, we develop a novel and facile approach to fabricate graphene-based flexible biosensor on PET substrate. Two electrodes were directly fabricated on graphene/PET by using silver paste, and the cell pool was directly constructed by using silicone gel; this flexible biosensor can be made by hand in any laboratories without needing photolithography process and precious facilities. Surprisingly, our graphene-based flexible biosensors demonstrate highly sensitive and can rapidly detect for ovarian cancer cells. As far as we know, there are not any reports about flexible graphene-based biosensors for detection of ovarian cancer cells yet.
Materials and Methods
Growth and Transfer of Graphene Film
In this work, the graphene film was grown on the surface of Cu foil (Alfa Aesar, No. 13382) by chemical vapor deposition (CVD) [20]. Firstly, the surface oxide of Cu foil was removed by 20% hydrochloric acid solution for 5 min; then, the Cu foil was cleaned by de-ionized water for several times and then, dried with nitrogen flow. The cleaned Cu foil was put on a quartz boat and put into the quartz tube of CVD furnace. The furnace chamber was pumped down to 1 × 10–2 Pa. The furnace temperature was increased up to 1000 °C for 20 min with50 sccm of 99.999% H2, and then, 50 sccm of 99.999% methane was introduced into the tube for growth of large-area graphene film for 20 min. Finally, the CVD furnace was cooled down room temperature with CH4/H2 gas flow.
The large-area graphene/Cu foil was cut into many desired pieces. Then, the PMMA was spin-coated on the surface of graphene/Cu foil, forming the PMMA/graphene/Cu sandwich-like structure. Subsequently, the underlying Cu foil was etched with 1 M FeCl3 solution. The PMMA/graphene was cleaned in DI water for 30 min and then transferred onto the PET substrate. Finally, the PMMA was removed by acetone, and graphene/PET sample was obtained.
Fabrication of Graphene-Based Biosensors
The fabrication procedure of graphene-based biosensors is described as follows. Firstly, about 1 cm × 2 cm graphene film on Cu foil was transferred onto 1 cm × 2 cm PET substrate by PMMA-assisted wet transfer method. Then, two electrodes were fabricated near around the center of graphene/PET film by using silver paste. Finally, in order to test the electrical response of the cancer cell solution, a cell pool with several millimeter length and width, and about 1 mm-height was constructed by silicone gel at the edge of the electrode. After the silicone gel of the cell pool is completely solidified, one can use Agilent 4155B semiconductor analyzer to check whether the graphene biosensor can work normally.
Culture of SKOV3 Ovarian Cancer Cells
SKOV3 ovarian cancer cell series (provided by the public laboratory of the Second Affiliated Hospital of West China) were cultured in RPMI-1640 (Transgene, France) complete medium containing 10% calf serum (MRC, USA) under the condition of 5% CO2 and 37 °C.
Preparation of Cell Solution and Electrical Measurement
The cancer cells were diluted to a specific concentration with cell culture medium. Take 50μL cell solution with pipette to the groove for measurement. The electrical signal was recorded by Agilent 4155B semiconductor analyzer.
Results and Discussion
The photograph of 10 × 10 cm2 large-area CVD-grown graphene film on Cu foil is shown in Fig. 1 a. From Fig. 1a, one can observe that compared to bare Cu foil with bright metallic color, the color of graphene/Cu is a little bit darker. Corresponding Raman spectrum of graphene/Cu is shown in Fig. 1b. As shown in Fig. 1b, the Raman peaks at 1580 cm−1and 2680 cm−1 correspond to G and 2D peaks of graphene film. In order to further check the quality of graphene film, we have measured the Raman spectrum of monolayer graphene film transferred onto SiO2/Si substrate, as shown in Fig. 1c. One can observe that the ratio between IG and I2D is lower than 0.5, which confirms that the thickness of the graphene is monolayer; one can also observe that D peak is very low and nearly cannot be observed, which suggests that the quality of the graphene film is very high and the defects are very few.
The photographs of flexible graphene biosensors on PET substrates are shown in Fig. 2. The cancer cell solution can be added into the cell pool, and the electrical signal of graphene biosensor can be obtained from two silver paste electrodes. The electrical response for cell culture medium, and CTC solution were measured.
The time dependence of current response for such liquids is recorded at a fixed voltage of 0.01 V before and after putting them into the cell pool. As shown in Fig. 3, one can observe that before dipping such liquid, the current keeps constant; when such liquids were put into the cell pool, the current quickly decreases and then slowly keep a new balance. The response is defined as η = (I0 − I)/I0*100%, where I0 is the current just before dipping the liquid, and I is maximal (or minimal) value after dipping the liquid at some time. One can see that after one put such liquids, the resistance of grapheme increases. The electrical response for bare cell culture medium, and CTC solution before and after dipping solution 200 s are 2.96%, and 37.04%, respectively. Obviously, compared to bare cell culture medium, the electrical response for CTC solution even with 30 cells/ml is very significant, which suggests that the flexible graphene-based biosensor is very sensitive for cancer cell detection.
The time dependence of electrical signal for bare cell culture medium and CTC solution with 30 cancer cells/ml is further analyzed from Fig. 3. As shown in Fig. 4, one can observe that compared to that of bare cell culture medium, the electrical response for CTC solution (even as 30 cells/ml) is very sensitive and fast. After dipping the cell solution, it needs only 2.1, 2.0, 4.5, 7.5, 10.5, 28.5 s to reach response of 5%, 10%, 15%, 20%, 25%, 30%, while the response of corresponding cell culture medium only increases from 0.15 to 1.3%. That means, the flexible graphene biosensor is very suitable for rapid and sensitive detection within 5 s.
We further investigated the electrical response for two kinds of CTC cancer cells (SUDHL8 cells and OCILYS cells) with same concentrations of 10,000 (10 K)/ml. As shown in Fig. 5, the time dependence of current for two different cancer cells with a little bit different trend and a large difference in electrical response. That means, the graphene biosensor is promising to be used to identify different cancer cells.
The time dependence of electrical current and response for SUDHL 8 cancer cell solution with various cell concentrations of 10,000 (10 K)/ml and 100 K/ml has also been investigated. As shown in Fig. 6, one can observe that solution with lower cancer cell concentration solution shows higher current, which suggests that the cancer cell tends to be insulating and many more cells are not beneficial to conductive. The time dependence of response for two concentration solutions shows similar change trends and the response for lower concentration solution is a little bit higher than that of higher concentration solution. These results show that biosensor can be used to identify cancer solution with different concentrations.
As mentioned above, the results show that cheap and flexible graphene-based biosensor demonstrates different response for cell culture medium and cancer solution, different cancer cells and cancer cell solution with different concentrations, which suggests that such flexible graphene-based biosensor is promising to be used to detect and identify CTC ovarian cancer cells.
Conclusion
In order to develop an efficient early detection method particularly for ovarian cancer, we develop a very simple graphene-based flexible biosensor on PET substrate. This flexible biosensor consists of a cell pool and two electrodes and compares the electric signal before and after adding cell solution, which shows high sensitivity and fast detection speed. It shows obvious different responses for cell culture medium and cancer solution, different cancer cells and cancer cell solution with different concentrations. Our work indicates that flexible graphene-based biosensor is promising to be used to sensitively and rapidly detect/identify CTC ovarian cancer cells.
Availability of Data and Materials
Authors can confirm that all relevant data are included in the article and its supplementary information files.
Abbreviations
- CA:
-
Carbohydrate antigen
- CTC:
-
Circulating tumor cell
- PET:
-
Polyethylene terephthalate
- FDA:
-
Food and Drug Administration
- CVD:
-
Chemical vapor deposition
References
Torre LA, Trabert B, DeSantis CE, Miller KD, Samimi G, Runowicz CD, Gaudet MM, Jemal A, Siegel RL (2018) Ovarian cancer statistics, 2018. CA Cancer J Clin 68:284–296
Cannistra SA (2004) Cancer of the ovary. N Engl J Med 351:2519–2529
Guo N, Peng Z (2017) Does serum CA125 have clinical value for follow-up monitoring of postoperative patients with epithelial ovarian cancer? Results of a 12-year study. J Ovarian Res 10:14
Gupta KK, Gupta VK, Naumann RW (2019) Ovarian cancer: screening and future directions. Int J Gynecol Cancer 29:195–200
Nebgen DR, Lu KH, Bast RC Jr (2019) Novel approaches to ovarian cancer screening. Curr Oncol Rep 21:75
Hosseini H, Obradović MMS, Hoffmann M, Harper KL, Sosa MS, Werner-Klein M, Nanduri LK, Werno C, Ehrl C, Maneck M, Patwary N, Haunschild G, Gužvić M, Reimelt C, Grauvogl M, Eichner N, Weber F, Hartkopf AD, Taran FA, Brucker SY, Fehm T, Rack B, Buchholz S, Spang R, Meister G, Aguirre-Ghiso JA, Klein CA (2016) Early dissemination seeds metastasis in breast cancer. Nature 540:552–558
Alix-Panabieres C, Pantel K (2013) Circulating tumor cells: liquid biopsy of cancer. Clin Chem 59:110–118
Dementeva N, Kokova D, Mayboroda OA (2017) Current methods of the Circulating Tumor Cells (CTC) analysis: a brief overview. Curr Pharm Des 23:4726–4728
Burinaru TA, Avram M, Avram A, Mărculescu C, Ţîncu B, Ţucureanu V, Matei A, Militaru M (2018) Detection of circulating tumor cells using microfluidics. ACS Comb Sci 20:107–126
Rawal S, Ao Z, Datar RH, Agarwal A (2017) Microfilter-based capture and release of viable circulating tumor cells. Methods Mol Biol 1634:93–105
Chen X, Zhou F, Li X, Yang G, Zhang L, Ren S, Zhao C, Deng Q, Li W, Gao G, Li A, Zhou C (2015) Folate receptor-positive circulating tumor cell detected by LT-PCR-based method as a diagnostic biomarker for non-small-cell lung cancer. J Thorac Oncol 10:1163–1171
De Bono JS, Scher HI, Montgomery RB, Parker C, Miller MC, Tissing H, Doyle GV, Terstappen LW, Pienta KJ, Raghavan D (2008) Circulating tumor cells predict survival benefit from treatment in metastatic castration-resistant prostate cancer. Clin Cancer Res 14:6302–6309
Cristofanilli M, Budd GT, Ellis MJ, Stopeck A, Matera J, Miller MC, Reuben JM, Doyle GV, Allard WJ, Terstappen LW, Hayes DF (2004) Circulating tumor cells, disease progression, and survival in metastatic breast cancer. N Engl J Med 351:781–791
Cohen SJ, Punt CJ, Iannotti N, Saidman BH, Sabbath KD, Gabrail NY, Picus J, Morse M, Mitchell E, Miller MC, Doyle GV, Tissing H, Terstappen LW, Meropol NJ (2008) Relationship of circulating tumor cells to tumor response, progression-free survival, and overall survival in patients with metastatic colorectal cancer. J Clin Oncol 26:3213–3221
Bollella P, Fusco G, Tortolini C, Sanzò G, Favero G, Gorton L, Antiochia R (2017) Beyond graphene: electrochemical sensors and biosensors for biomarkers detection. Biosens Bioelectron 89(Pt 1):152–166
Samanman S, Numnuam A, Limbut W, Kanatharana P, Thavarungkul P (2015) Highly-sensitive label-free electrochemical carcinoembryonic antigen immunosensor based on a novel Au nanoparticles-graphene-chitosan nanocomposite cryogel electrode. Anal Chim Acta 853:521–532
Jang HD, Kim SK, Chang H, Choi JW (2015) 3D label-free prostate specific antigen (PSA) immune-sensor based on graphene-gold composites. Biosens Bioelectron 63:546–551
Yang F, Yang Z, Zhuo Y, Chai YQ, Yuan R (2015) Ultrasensitive electrochemical immunosensor for carbohydrate antigen 19–9 using Au/porous graphene nanocomposites as platform and Au@Pd core/shell bimetallic functionalized graphene nanocomposites as signal enhancers. Biosens Bioelectron 66:356–362
Lipatov A, Varezhnikov A, Wilson P, Sysoev V, Kolmakov A, Sinitskii A (2013) Highly selective gas sensor arrays based on thermally reduced graphene oxide. Nanoscale 5:5426–5434
Liu JB, LiPJ ChenYF (2014) Large-area synthesis of high-quality and uniform monolayer graphene without unexpected bilayer regions. J Alloys Compd 615:415–418
Acknowledgements
Not applicable.
Funding
The research was supported by Science and Technology Plan Projects from Sichuan (Grant Nos. 2021YFG0142 and 19ZDYF1743).
Author information
Authors and Affiliations
Contributions
Conceptualization was contributed by LH and YC; Methodology was contributed by LH and QW; Data curation was contributed by LH and YC; Writing—Original Draft Preparation, was contributed by LH; Writing—Review & Editing, was contributed by YC, AZ, YG; Funding acquisition was contributed by AZ and YG. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The author declares no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Han, L., Wan, Q., Zheng, A. et al. Demonstration of a Flexible Graphene-Based Biosensor for Sensitive and Rapid Detection of Ovarian Cancer Cells. Nanoscale Res Lett 16, 181 (2021). https://doi.org/10.1186/s11671-021-03633-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s11671-021-03633-9
Keywords
- Flexible graphene-based biosensor
- Circulating tumor cell (CTC)
- Ovarian cancer