Abstract
In this paper, a terahertz (THz) biosensor based on all-metal metamaterial is theoretically investigated and experimentally verified. This THz metamaterial biosensor uses stainless steel materials that are manufactured via laser-drilling technology. The simulation results show that the maximum refractive index sensitivity and the figure of merit of this metamaterial sensor are 294.95 GHz/RIU and 4.03, respectively. Then, bovine serum albumin was chosen as the detection substance to assess this biosensor’s effectiveness. The experiment results show that the detection sensitivity is 72.81 GHz/(ng/mm2) and the limit of detection is 0.035 mg/mL. This THz metamaterial biosensor is simple, cost-effective, easy to fabricate, and has great potential in various biosensing applications.
Avoid common mistakes on your manuscript.
Introduction
Nowadays, doctors usually collect serum from patients in the hospital for various examinations. Such as tumor detection [1,2,3] and virus detection [4,5,6] etc. The preliminary method for tumor detection is protein detection, because proteins are the components of many tumor markers, and these biomolecules with a lot of information exist in the serum. Serum albumins are the most abundant (52–62%) total water-soluble fraction proteins in the blood plasma [7, 8]. Furthermore, nutritional and physiological functions of serum albumins make them as essential bio-macromolecules. As one of the most popular serum albumins, bovine serum albumin (BSA) is used in a lot of research fields. Therefore, the research on the detection of BSA is quite important. There are many methods for detecting and determining the concentration of BSA, including electrochemical impedance spectroscopy [9], capillary electrophoresis [10], and light scattering techniques [11]. But they all have some disadvantages, such as complicated procedures, poor reproducibility or time-consuming. Therefore, developing new biosensors is of considerable significance and quite in demand.
The THz wave is between the microwave and infrared optical wave and in the transition region from electronics to photonics. Compared with light waves, the energy of photons in this band is very low. This means that terahertz waves will not cause radiation ionization damage to biological molecules. Many biological macromolecules also have unique fingerprints in the terahertz band [12,13,14,15]. Therefore, terahertz waves have received considerable attention in the field of biosensing [16, 17].
Metamaterials are artificial electromagnetic materials composed of subwavelength structures. Their unique electromagnetic resonance has many characteristics, such as a negative RI [18, 19], electromagnetically induced transparency [20, 21], and extreme environmental sensitivity [22, 23]. Metamaterials sensitive to the surrounding environment, especially those composed of subwavelength metal structures [24,25,26], have been widely used to detect various biomolecules. The combination of terahertz waves and metamaterials provides a new detection method for the biomedical molecules, which cannot only achieve label-free detection, but also refresh the resolution limit of existing sensors. In addition, detection can be completed simply and rapidly using a small amount of analyte with no chemical reagents.
At THz frequencies, metamaterial production usually relies on micron-level processing methods. Photolithography [27] or electron beam lithography [28] is mainly used to transfer micro-nano patterns from the photoresistor to the surface of the functional materials, and then wet [29, 30] or dry etching [31, 32] is required to complete the final processing of the metamaterials. Through the above steps, the construction of fine graphics can be achieved, but unfortunately, most of these methods demand expensive processing equipment, high standard operation environments, and cumbersome processing procedures. Laser drilling [33, 34] is the first practical laser processing technology, and it is also one of the main application fields of laser processing. The laser beam is highly concentrated in space and time. By focusing with a lens, the spot diameter can be reduced to micron level, and the laser power density of 105–1015 W/cm2 can be obtained. With such a high-power density, laser drilling can be carried out in almost any material. To the best of our knowledge, this is the first time to apply laser-drilling technology in the fabrication of metamaterial biosensors, which can significantly reduce the processing cost of metamaterial biosensors and promote their practical applications.
In this study, a highly sensitive terahertz biosensor for protein detections based on all-metal metamaterial was proposed, theoretically simulated and experimentally demonstrated. The device was simple to manufacture, cost-effective and quite stable. It was composed of stainless-steel material and manufactured using laser-drilling technology. At first, this metamaterial sensor was simulated and analyzed using the finite integral method, and the refractive index sensitivity was calculated. Then this THz metamaterial biosensor was fabricated and measured. The experiments confirmed the high sensitivity of this senor to external environment. BSA was chosen as the detection substance to assess the biosensor’s effectiveness. The Hill formula was used to fit the experimental data. A detection sensitivity of 72.81 GHz/(ng/mm2) and the limit of detection (LOD) of 0.035 mg/mL were obtained. The measurements were repeated three times to verify the biosensor’s reliability.
Design
Figure 1a shows the structure of the proposed all-metal metamaterial terahertz biosensor. A hollow dumbbell pattern with a periodic arrangement along the x and y directions was formed on a 50 μm thick stainless-steel plate (conductivity of 1.4 × 106 S/m). The period sizes Px and Py of the unit structure are 500 μm and 300 μm, respectively. The hollow dumbbell’s length L and gap H are 294 μm and 60 μm, respectively. The radius R of circles at both ends of the hollow dumbbell is 60 μm. The biosensor had an all-metal structural design and no traditional dielectric substrate. The terahertz wave is perpendicular incident to the surface of the metamaterial biosensor.
Methods and simulation
Then the three-dimensional full wave electromagnetic field simulation via the finite integral method (Commercial software CST) was used for the following simulations. Periodic boundary conditions were applied in the x and y directions, and the perfectly matched layer was used in the wave propagation direction of z. As shown in the upper right corner of Fig. 1a, the wave vector of the incident electromagnetic field kz was a plane wave propagating in the z-axis, and the electric and magnetic fields were polarized along the y-axis and x-axis, respectively. As shown in Fig. 1b, there was a transmission peak at 0.48 THz.
To study the physical mechanism of this resonance peak’s generation, the biosensor’s surface current and magnetic field at the resonance peak frequency were simulated. As shown on the left side of Fig. 2, the incident electromagnetic waves were polarized along the y axis, inducing charge oscillations at both ends of the opening, resulting in an electric dipole. The charge oscillations were accompanied by counter-rotating current oscillations along the rims of the two circular holes that comprised the apertures. This led to a pair of counter-oriented out-of-plane magnetic dipoles. As shown on the right side of Fig. 2, there was a pair of obvious opposite magnetic dipoles in the z axis connected end to end to form a toroidal dipole. Therefore, the metamaterial’s response was dominated by a combination of electric and toroidal dipoles.
Because the performances of the sensor are affected by the structure parameters, it is necessary to optimize the structural parameters during the design procedure. Figure 3 shows the effect of the structural size changes on the transmission spectra. As shown in Fig. 3a, when the length of the hollow dumbbell increased from 290 to 298 μm, the peak frequency of the transmission spectra red-shifted from 0.48 THz. As shown in Fig. 3b, when the gap of the hollow dumbbell increased from 56 to 64 μm, the peak frequency of the transmission spectra blue-shifted from 0.48 THz. As L and H increased, the resonance peak started to move toward the low frequency and high frequency, respectively. When the circle’s radius varied from 56 to 64 μm and the thickness of the stainless-steel varied from 40 to 60 μm, the position of the resonance peak slightly changed. Therefore, it is easier to adjust the resonant frequency of the metamaterial biosensor by adjusting the length of the hollow dumbbell L and the gap of the hollow dumbbell H.
It is also very important to study the influence of the incident angle and the polarization angle on the transmission spectra. The electromagnetic wave was vertically incident, and then the incident and polarization angles were changed. The definition of these angles is shown in Fig. 4a. The incident angle means θi in the y–z plane, and the polarization is θp in the x–y plane. As shown in Fig. 4b, when the incident angle changed from 0° to 15°, the difference of the resonance peak’s frequency was only 9 GHz. As shown in Fig. 4c, when the polarization angle increased from 0° to 15°, the difference in resonance peak frequency was almost 0 GHz, but the resonance peak amplitude decreased by about 0.1. This showed that the biosensor was almost insensitive to changes in the incident and polarization angles, which is beneficial for practical biosensor applications.
To explore the biosensor’s sensing performance, a 120-μm thin layer of analyte was added to this metamaterial biosensor as shown in Fig. 5a, then different transmission spectra of this metamaterial biosensor was simulated when the refractive index of the analyte changed as shown in Fig. 5b. The RI sensitivity S was defined as the ratio of the variations in the transmission peak position to the RI unit (S = Δf/Δn). As the analyte’s RI increased, the resonance peak frequency red-shifted. Then the resonance peak frequency shift corresponding to each RI was collected. Good linearity was observed. The fitting result in Fig. 5c shows that the sensitivity to the RI was 294.95 GHz/RIU.
The sensing performance was also quantified using the figure of merit (FOM), which was defined as:
where S is the sensitivity and FHWM is the full width at half maximum of the resonance peak. The FOM of this biosensor was 4.03.
For most metamaterial structures, they usually use dielectric materials as the substrates. However, this metamaterial biosensor proposed in this paper was based on an all-metal metamaterial with an all stainless-steel design and the air was used as the substrate. Compared with traditional dielectric materials, such as polyethylene terephthalate (PET), quartz, and silicon, the air has the lowest RI. To evaluate the role of the substrate, these metamaterial biosensors were simulated again using different substrates, and the refractive index sensitivities and FOM values were calculated subsequently. As shown in Fig. 6, as the RI of the substrate increased, the sensor’s RI sensitivity and FOM began to decrease. This result indicated that the biosensor with a lower substrate RI had a better sensing performance.
To further investigate the biosensor’s sensing principle, the electric field distribution diagrams were simulated, as shown in Fig. 7. The top and side views of the simulated electric field distribution demonstrated that the electric field energy was mainly concentrated in part of the stainless-steel holes. Therefore, it is crucial to make sure that the analyte was added into the holes.
Table 1 summarized the proposed sensor’s RI sensitivity and FOM, and compared them with other reported studies [35,36,37]. The other THz sensors are all based on traditional photolithography processes. It can be seen that the stainless-steel metamaterial biosensor we designed had excellent sensing performance using cheap laser-drilling technology.
Experiment
Materials and sample preparation
Then in order to demonstrate the biosensing ability of the proposed THz mmetamaterial sensor, protein detection was performed in the experiment. BSA and PBS buffer were both purchased from Sigma-Aldrich. The BSA solution was formulated in PBS buffer (pH = 7.4).
A microscope image of the fabricated metamaterial biosensor is shown in Fig. 8. The overall size of the metamaterial biosensor was 12 mm × 12 mm.
The concentrations of the formulated BSA solution were 0.2 mg/mL, 0.5 mg/mL, 2.0 mg/mL, and 4.0 mg/mL. The analyte was added to the biosensor surface by liquid deposition method. Each time, 150 μL BSA solution was transferred to the biosensor surface with a pipette gun, and the biosensor was dried on a 40 ℃ heating table. When the biosensor was heated at 40 ℃, the protein film formed faster and more uniformly. Each time before changing different concentrations of BSA solution, the stainless-steel sheet was put into deionized water and vibrated in an ultrasonic vibrator to ensure that the protein film from the previous process is cleared and the surface of this mmetamaterial biosensor is clean. Figure 9 showed the pictures and the microscopy pictures of the process of adding and drying the protein. As shown in Fig. 9a, the stainless steel sheet was clean, and then as shown in Fig. 9b, the BSA solution of one concentration was added to the stainless-steel sheet’s surface, and the solution stayed on this biosensor’s surface and did not pass through the holes due to the effect of water surface tension. After heating and drying, a thin layer of the BSA film was formed as shown in Fig. 9c.
In order to observe the situation inside the stainless-steel’s holes, one side of the stainless-steel was cut off, so that one side of the holes was revealed and observed with a microscope. As shown in Fig. 9d–g, when the BSA solution was dripped and dried, a thin layer of the BSA was added into the stainless-steel’s holes, which is mainly because the diameter of the holes is much larger than the size of the BSA proteins. This proves that the analyte to be detected can enter the detection sensitive region of our biosensor, which can greatly increase the sensitivity of this metamaterial biosensor.
Spectral measurements
All the spectral measurements were conducted using a continuous-wave THz spectroscopy system (TeraScan 1550, Toptica Photonics AG). The system consisted of dual-laser control (DLC) smart electronics, two distributed feedback (DFB) lasers, two fibre-coupled InGaAs photomixers, and four 90° off-axis parabolic mirrors as shown in Fig. 10. THz waves were collimated and focused on the sample through the 90° off-axis parabolic mirrors. All transmission spectra were obtained by scanning between 50 and 1220 GHz in step sizes of 40 MHz with an integration time of 10 ms operating in the fast scan mode to reduce the scanning time. The polarization of the terahertz wave was along the dumbbell ring’s opening direction.
Experiment results and discussion
The protein detection experiments of the metamaterial biosensor were carried out with four concentrations of the BSA solutions. Throughout the experiments, each group of BSA solutions was added in order from low to high. All the measurements were repeated three times.
As shown in Fig. 11a, as the concentration of the BSA solutions increased, the resonance peak frequency red-shifted. This trend is consistent with the simulation results. The decrease in the intensity of the resonance was due to the absorption of the terahertz wave by the BSA protein.
The relationship between the frequency shift and the concentration of the BSA solutions is not linear, which is common in biological experiments [38, 39]. The Hill model can characterize the binding ability between our metamaterial biosensor and biomolecules. Therefore, the Hill model [40] was used to fit the experimental data, as shown in Fig. 11b. The Hill equation is described as follows:
where the maximum peak frequency shift Δfmax is the saturation value, [BSA] is the concentration of the BSA solution, n is the Hill coefficient, and KD is the dissociation constant.
Using the fitting curve as shown in Fig. 11b, the Hill coefficient n was calculated to be 0.83, and the dissociation constant KD was calculated to be 2.87 mg/mL. Furthermore, Δfmax was approximately 166 GHz, which indicated the maximum peak frequency shift at the saturation concentration. The Hill fitting results of the resonant frequency peak confirmed the experiment’s reliability and accuracy.
The biosensor’s sensitivity S was deduced as follows [41]:
where Δfmax is 166 GHz and δmax is the BSA’s surface density. δmax was determined by [41]:
where MBSA = 66,430 g/mol is the estimated molecular mass of BSA [42], NA = 6.02 × 1023 mol−1 is Avogadro’s number, and PBSA = 6.96 nm [43] is the average length of one BSA molecule. δmax was 2.28 ng/mm2 and the biosensor’s BSA detection sensitivity was 72.81 GHz/(ng/mm2).
KD obtained using the Hill model showed that the dissociation constant was strongly related to the BSA, and the limit of detection (LOD) Clim of the BSA was calculated using the following equation [44]:
where Sf is the spectral resolution of 2 GHz. Equation (5) shows that smaller dissociation constant resulted in a lower detection limit. Thus, Clim was calculated to be 0.035 mg/mL.
Table 2 shows our biosensor’s BSA sensing performance compared with those reported studies. In the experiment, the lowest concentration of the BSA solution was 0.2 mg/mL and a frequency change of 10.8 GHz was obtained. Compared with Refs. [45,46,47], a relatively higher frequency change was achieved at the same BSA concentration. The Hill's formula was applied to analyze the metamaterial biosensor’s data. The calculated LOD of 0.035 mg/ml was significantly better than that in Ref. [45]. All these predict that our proposed all-metal THz metamaterial biosensor will have excellent performance in many biological and chemical applications.
Based on the excellent sensing performance of stainless steel biosensor, the stainless steel biosensor can be modified with specific antibody to achieve specific antigen detection in the future. And the thickness of stainless steel biosensor is only 50 μm. With the development of microfluidics technology and terahertz spectroscopy, it is hopeful to apply real-time measurement in vivo in the future.
Conclusion
In conclusion, a terahertz biosensor based on an all-metal metamaterial was used to measure protein concentrations. The biosensor was made of stainless-steel and prepared via laser-drilling technology. The maximum RI sensitivity and FOM calculated using CST electromagnetic simulation software are 294.95 GHz/RIU and 4.03, respectively. The sample was characterized using a continuous-wave THz spectrometer. The experimental results showed that, for the BSA analyte solution, the detection sensitivity and detection limit are 72.81 GHz/(ng/mm2) and 0.035 mg/mL, respectively. This biosensor has advantages of small shape, high detection sensitivity, low detection limits, reusability, easy to fabricate and cost-effective. These research results are of considerable significance for future applications in biomolecular detection and disease diagnosis.
Availability of data and materials
All data are fully available without restriction.
Abbreviations
- THz:
-
Terahertz
- RI:
-
Refractive index
- FOM:
-
Figure of merit
- BSA:
-
Bovine serum albumin
- PET:
-
Polyethylene terephthalate
- DLC:
-
Dual-laser control
- DFB:
-
Distributed feedback
- LOD:
-
Limit of detection
References
Fukutomi T, Sugano K, Kitoh T, Ohkura H, Ushiama M (2000) Combined measurement of the c-erbB-2 protein in breast carcinoma tissues and sera is useful as a sensitive tumor marker for monitoring tumor relapse. Int J Cancer 89(4):329–336
Escudero J, Auge J, Filella X, Torne A, Pahisa J, Molina R (2011) Comparison of serum human epididymis protein 4 with cancer antigen 125 as a tumor marker in patients with malignant and nonmalignant diseases. Clin Chem 57(11):1534–1544
Speeckaert M, Speeckaert R, Delanghe J (2013) Human epididymis protein 4 in cancer diagnostics: a promising and reliable tumor marker. Adv Clin Chem 59:1–21
Kauffman E, Jones S, Dupuis A, Ngo K, Bernard K, Kramer L (2003) Virus detection protocols for west nile virus in vertebrate and mosquito specimens. J Clin Microbiol 41(8):3661
Walter M, Menarguez-Palanca J, Peiper S (1992) Epstein-barr virus detection in neck metastases by polymerase chain reaction. Laryngoscope 102(5):481–485
Scotto J, Hadchouel M, Hery C, Yvart J, Brechot C (1983) Detection of hepatitis B virus DNA in serum by a simple spot hybridization technique: comparison with results for other viral markers. Hepatology 3(3):279–284
Jahanban-Esfahlan A, Panahi-Azar V (2016) Interaction of glutathione with bovine serum albumin: Spectroscopy and molecular docking. Food Chem 202:426–431
Roufegarinejad L, Jahanban-Esfahlan A, Sajed-Amin S, Panahi-Azar V, Tabibiazar M (2018) Molecular interactions of thymol with bovine serum albumin: Spectroscopic and molecular docking studies. J Mol Recognit 31(7):e2704
Pinto E, Soares D, Brett C (2008) Interaction of BSA protein with copper evaluated by electrochemical impedance spectroscopy and quartz crystal microbalance. Electrochim Acta 53(25):7460–7466
Liu C, Wei Y, Miao Y, Zhang Y (2012) Interactions between isoprenaline hydrochloride and bovine serum albumin (BSA) in capillary zone electrophoresis and affinity capillary electrophoresis. Pharmazie 67(10):839–843
Wu Z, Zhang J, Chen W, Hu J (2009) Studies on the Aggregation and interaction of Daidzin with Bovine serum albumin by fluorescence spectrum and laser light scattering. Acta Chim Sin 67(14):1609–1614
Xiong W, Shen J (2010) Fingerprint extraction from interference destruction terahertz spectrum. Opt Express 18(21):21798–21803
Pickwell E, Wallace V (2006) Biomedical application of terahertz technology. J Phys D 39(17):R301–R302
Bui T, Dao T, Dang L, Vu L, Ohi A, Nabatame T, Lee Y, Nagao T, Hoang C (2016) Metamaterial-enhanced vibrational absorption spectroscopy for the detection of protein molecules. Sci Rep 6:32123
Zhao G, Yu B, Zhang C (2009) Terahertz spectroscopic investigation of four kinds of vitamins. J Appl Phys 106(10):104702
Menikh A, Maccoll R, Mannella C, Zhang X (2015) Terahertz biosensing technology: frontiers and progress. ChemPhysChem 3(8):655–658
Hasebe T, Kawabe S, Matsui H, Tabata H (2012) Metallic mesh-based terahertz biosensing of single- and double-stranded DNA. J Appl Phys 112(9):094702
Dolling G, Wegener M, Soukoulis C, Linden S (2007) Negative-index metamaterial at 780 nm wavelength. Opt Lett 32(1):53–55
Wuestner S, Pusch A, Tsakmakidis K, Hamm J, Hess O (2010) Overcoming losses with gain in a negative refractive index metamaterial. Phys Rev Lett 105(12):127401
Liao Z, Liu S, Ma H, Li C, Jin B, Cui T (2016) Electromagnetically induced transparency metamaterial based on spoof localized surface plasmons at terahertz frequencies. Sci Rep 6:27596
Zhang F, Zhao Q, Zhou J, Wang S (2013) Polarization and incidence insensitive dielectric electromagnetically induced transparency metamaterial. Opt Express 21(17):19675–19680
Melik R, Unal E, Perkgoz N, Puttlitz C, Demir H (2010) Metamaterial based telemetric strain sensing in different materials. Opt Express 18(5):5000–5007
Al-Naib I, Jansen C, Koch M (2008) Thin-film sensing with planar asymmetric metamaterial resonators. Appl Phys Lett 93(8):083507
Danilov A, Tselikov G, Wu F, Kravets V, Ozerov I, Bedu F, Grigorenko A, Kabashin A (2018) Ultra-narrow surface lattice resonances in plasmonic metamaterial arrays for biosensing applications. Biosens Bioelectron 104:102–112
Lu Y, Yang M, Zhang Z, Liang L, Li J, Yao J (2019) Dual-wavelength terahertz sensing based on anisotropic Fano resonance metamaterials. Appl Opt 58(7):1667–1674
Zhou H, Yang C, Hu D, Li D, Hui X, Zhang F, Chen M, Mu X (2019) Terahertz biosensing based on bi-layer metamaterial absorbers toward ultra-high sensitivity and simple fabrication. Appl Phys Lett 115(14):143507
Yahiaoui R, Takano K, Miyamaru F, Hangyo M, Mounaix P (2014) Terahertz metamolecules deposited on thin flexible polymer: design, fabrication and experimental characterization. J Opt 16(9):094194
Pease R, Ballantyne J, Henderson R, Voshchenkoy A, Yau L (1974) Applications of the electron beam exposure system. IEEE Trans Electron Devices 22(7):393–399
Kovacs G, Maluf N, Peterser K (1998) Bulk micromachining of silicon. Proc IEEE 86(8):1536–1551
Youtsey C, Adesida I, Bulman G (1997) Highly anisotropic photoenhanced wet etching of n-type GaN. Appl Phys Lett 71(15):2151–2153
Gloersen P (1975) Ion-beam etching. J Vac Sci Technol 12(1):28–35
Matsuo S, Adachi Y (1982) Reactive ion beam etching using a broad beam ECR ion source. Jpn J Appl Phys 21:L4
Ancona A, Roeser F, Rademaker K, Limpert J, Nolte S, Tunnermann A (2008) High speed laser drilling of metals using a high repetition rate, high average power ultrafast fiber CPA system. Opt Express 16(12):8958–8968
Ancona A, Doering S, Jauregui C, Roeser F, Limpert J, Nolte S, Tuennermann A (2009) Femtosecond and picosecond laser drilling of metals at high repetition rates and average powers. Opt Lett 34(21):3304–3306
Zhang Z, Ding H, Yan X, Liang L, Wei D, Wang M, Yang Q, Yao J (2018) Sensitive detection of cancer cell apoptosis based on the non-bianisotropic metamaterials biosensors in terahertz frequency. Opt Mater Express 8(3):659–667
Lin S, Xu X, Hu F, Chen Z, Wang Y, Zhang L, Peng Z, Li D, Zeng L, Chen Y (2021) Using antibody modified terahertz metamaterial biosensor to detect concentration of carcinoembryonic antigen. IEEE J Sel Top Quantum Electron 27(4):6900207
Li D, Hu F, Zhang H, Chen Z, Zhou Y (2021) Identification of early-stage cervical cancer tissue using metamaterial terahertz biosensor with two resonant absorption frequencies. IEEE J Sel Top Quantum Electron 27(99):1–1
Cui N, Guan M, Xu M, Fang W, Zeng Y (2020) Design and application of terahertz metamaterial sensor based on DSRRs in clinical quantitative detection of carcinoembryonic antigen. Opt Express 28(11):16834–16844
Cai J, Zhang C, Liang C, Min S, Li W (2019) Solution-processed large-area gold nanocheckerboard metasurfaces on flexible plastics for plasmonic biomolecular sensing. Adv Opt Mater 7(19):1900516
Kim K, Song J, Kee J, Liu Q, Lo G, Park M (2013) Label-free biosensor based on an electrical tracing-assisted silicon microring resonator with a low-cost broadband source. Biosens Bioelectron 46:15–21
Chen L, Chan C, Ni K, Hu P, Li T, Wong W, Balamurali P, Menon R, Shaillender M, Neu B, Poh C, Dong X, Ang X, Zu P, Tou Z, Leong K (2013) Label-free fiber-optic interferometric immunosensors based on waist-enlarged fusion taper. Sense Actuators B Chem 178:176–184
Hirayama K, Akashi S, Furuya M, Fukuhara K (1990) Rapid confirmation and revision of the primary structure of bovine serum albumin by ESIMS and Frit-FAB LC/MS. Biochem Biophy Res Commun 173(2):639–646
Axelsson I (1978) Characterization of proteins and other macromolecules by agarose gel chromatography. J Chromatogr 152(1):21–32
Maguis S, Laffont G, Ferdinand P, Carbonnier B, Kham K, Mekhalif T, Millot M (2008) Biofunctionalized titled fiber bragg gratings for label-free immunosensing. Opt Express 16(23):19049–19062
Wang S, Xia L, Mao H, Jiang X, Yan S, Wang H, Wei D, Cui H, Du C (2016) Terahertz biosensing based on a polarization-insensitive metamaterial. IEEE Photonics Technol Lett 28(9):986–989
Li Y, Chen X, Hu F, Li D, Teng H, Rong Q, Zhang W, Han J, Liang H (2019) Four resonators based high sensitive terahertz metamaterial biosensor used for measuring concentration of protein. J Phys D Appl Phys 52(9):095195
Lan F, Luo F, Mazumder P, Yang Z, Meng L, Bao Z, Zhou J, Zhang Y, Liang S, Shi Z, Khan A, Zhang Z, Wang L, Yin J, Zeng H (2019) Dual-band refractometric terahertz biosensing with intense wave-matter-overlap microfluidic channel. Biomed Opt Express 10(8):3789–4379
Acknowledgements
We thank the Centre for THz Research, China Jiliang University for the use of their equipment.
Funding
This work was supported by the National Natural Science Foundation of China (61875251, 61875179), the Fundamental Research Funds for the Provincial Universities of Zhejiang (2020YW08) and the Outstanding Student Achievement Cultivation Program of China Jiliang University (2019YW27).
Author information
Authors and Affiliations
Contributions
FZ conceived the research and supervised the whole work, GW conducted simulations, experiment and analyses, and wrote the draft of the manuscript. TL and ZH assisted in the analysis and revised the manuscript. JL and JQ assisted in fabrication of metamaterials and experimental measurement. All the authors reviewed the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Wang, G., Zhu, F., Lang, T. et al. All-metal terahertz metamaterial biosensor for protein detection. Nanoscale Res Lett 16, 109 (2021). https://doi.org/10.1186/s11671-021-03566-3
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s11671-021-03566-3