Background

Diacerein is chemically 1,8-diacetoxy-3-carboxyanthraquinone and is also known as diacetylrhein. This drug is used in the treatment of osteoarthritis. After absorption, the drug is metabolized to its active metabolite rhein [1]. Diacerein and rhein are anthraquinone compounds that ameliorate the course of osteoarthritis [24]. Aceclofenac is chemically (2-[(2,6-dichlorophenyl)amino] phenyl acetoxyacetic acid) [57]. It has analgesic properties and a good tolerability profile in a variety of painful conditions.

A combined fixed dose of 50 mg diacerein and 100 mg aceclofenac is available commercially as tablets and are widely used for the treatment of osteoarthritis. The two main aspects of drug product analysis that play an important role in shelf life determinations are assay of active drug and degradants generated during the stability study. However, most of reference HPLC methods are less sensitive and not stability indicating [812]. Comparison of the proposed HPLC method with the reference HPLC methods is shown in Table 1.

Table 1 Comparison of the proposed HPLC method with the reference HPLC methods

Stability indicating RP-HPLC method for the simultaneous determination of aceclofenac and diacerein in tablet dosage form is not available in any pharmacopeia. Hence, it was felt essential to develop and validate a sensitive, accurate and stability indicating RP-HPLC method for the simultaneous determination of aceclofenac and diacerein in tablet dosage form. The chemical structures of diacerein and aceclofenac used in this work are shown in Figure 1. Overlain UV spectra of diacerein and aceclofenac are shown in Figure 2.

Figure 1
figure 1

The chemical structures of diacerein and aceclofenac.

Figure 2
figure 2

Overlain UV spectra of diacerein and aceclofenac.

Methods

Chemicals

Diacerein and aceclofenac working standards were obtained from Sir Sayyed College Aurangabad, India. Glacial acetic acid, HPLC grade acetonitrile, sodium hydroxide, hydrochloric acid and hydrogen peroxide were obtained from Merck Ltd., Mumbai, India. The 0.45-μm nylon filter was obtained from Advanced Micro Devices Pvt Ltd. (Ambala Cantt, India). The combination product of diacerein and aceclofenac label claim (diacerein 50 mg and aceclofenac 100 mg) branded tablets were purchased from the Indian market. Double distilled water was used throughout the experiment.

Equipment

The chromatographic system used was an Agilent-1100 series (Bangalore, India) comprised of degasser, quaternary pump, auto injector, column compartment and photodiode array detector. The system was controlled through Empower Software (Orlando, Florida).

Chromatographic conditions

Kromasil C-18, 150 × 4.6 mm, 3.5-μm column was used. The mobile phase consists of double distilled water (pH 2.7 with glacial acetic acid)-acetonitrile (45:55 v/v). Isocratic elution was at a flow rate of 1 mL/min and column temperature at 25°C. Detector wavelength was kept at 256 nm using a photodiode array detector.

Preparation of standard solution

The standard stock solution was prepared by dissolving 25 mg diacerein and 50 mg of aceclofenac. Working standard with 10 mL of dimethyl acetamide and dilution with the mobile phase up to 100 mL. Standard solution containing 25 μg/mL of diacerein and 50 μg/mL of aceclofenac was prepared by transferring 5 mL of standard stock solution to 50-mL volumetric flask and diluting with the mobile phase.

Sample preparation

The average fill weight of 20 capsules was determined. Accurately weighed sample powder (100 mg of aceclofenac and 50 mg of diacerein) was transferred into a 200-mL volumetric flask. Dimethyl acetamide (20 mL) was added and sonicated for about 20 min to dissolve; then, mobile phase was added to obtain a volume of 200 mL. The centrifuged solution was filtered through 0.45-μm filter. From the filtered solution, 5 mL of solution was transferred into a 50-mL volumetric flask and diluted to volume with mobile phase.

System suitability studies

The resolution, number of theoretical plates and peak asymmetry were calculated for the standard solutions and are shown in Table 2. The values obtained demonstrated the suitability of the system for the analysis of these drugs in combination. The typical chromatogram of the standard solution is shown in Figure 3.

Table 2 System suitability studies
Figure 3
figure 3

A typical chromatogram of the tablet: diacerein and aceclofenac.

Forced degradation study of drug product

Transferred sample powder equivalent to 100 mg of aceclofenac and 50 mg of diacerein into a 200-mL volumetric flask was added with 20 mL of dimethyl acetamide and sonicated for about 20 min to dissolve and was treated with acid, alkali and peroxide as described in Table 3. The stress studies were performed as per International Conference on Harmonisation (ICH) recommendation. Results of stressed studies are summarized in Table 3. Figures 4, 5 and 6 show degradation products formed during the degradation study.

Table 3 Summary of forced degradation results
Figure 4
figure 4

Chromatogram of the acid hydrolysis-degraded tablet solution.

Figure 5
figure 5

Chromatogram of the base hydrolysis-degraded tablet solution.

Figure 6
figure 6

Chromatogram of the peroxide-degraded tablet solution.

Results and discussion

Optimization of the chromatographic conditions

The main criteria of a successful HPLC method development for determination of diacerein and aceclofenac in tablet dosage form was the method should be able to determine the assay of both drugs in single run. The method should be robust, stability indicating, free of interference from degradation products and useful for routine use in the quality control laboratory.

Our objective of the chromatographic method development was to achieve a peak tailing factor that is <2, run time up to 10 min, along with a resolution that is >5 between diacerein and aceclofenac. In order to optimize the LC separation of aceclofenac and diacerein, initially, the retention behavior of both the components was studied in the pH range of 2.5-6.8 using mobile phases of buffer (pH 2.5-6.8) and acetonitrile, methanol as organic modifier.

To ensure the resolution between diacerein and aceclofenac not to less than 5, the method was as fast as possible, an isocratic run was optimized using the mobile phase consisted of double distilled water (pH adjusted 2.7 with glacial acetic acid)-acetonitrile (45:55 v/v), isocratic elution at flow rate of 1 mL/min and column temperature at 25°C. Detector wavelength was kept at 256 nm using a photodiode array detector. This was selected as an appropriate chromatographic condition, which gave good resolution and acceptable peak parameters for both diacerein and aceclofenac. The analytes of this combination had adequate retentions, peak shape, less tailing and more resolution. The chromatographic analysis time was 10 min. In optimized conditions, diacerein, aceclofenac and their degradants were well separated.

Experimental

Method validation

As per ICH guidelines [13] the method validation parameters were checked for linearity, precision, accuracy, limit of detection, limit of quantitation and robustness.

Specificity

Photodiode array detection was used as an evidence of the specificity of the method. Complete separation of diacerein and aceclofenac was noticed in the presence of tablet excipients. In addition, there was no interference at the retention time of diacerein and aceclofenac in the chromatogram of placebo solution. In peak purity analysis with photo diode detector, purity angle was less than purity threshold for both the analytes. This shows that the peak of analytes was pure, and excipients in the formulation did not interfere with the peak of analytes.

Linearity

Linearity of the method was tested from 10% to 300% of the targeted level of the assay concentration (diacerein 25 μg/mL, aceclofenac 50 μg/mL) for both analytes. Linearity solutions were injected. The calibration graphs were obtained by plotting the peak area ratio against the concentration of the drugs. The equations of the calibration curves for diacerein and aceclofenac obtained were y = 11 , 194 x + 2 , 477 and y = 11 , 926 x - 2 , 248 , respectively. In the simultaneous determination, the calibration graphs were found to be linear in the aforementioned concentrations with correlation coefficient of 0.9999 for diacerein and 0.9997 for aceclofenac.

Precision

The repeatability of the analytical method was evaluated by assaying six sample solutions of diacerein 25 μg/mL and aceclofenac 50 μg/mL during the same day, under the same experimental conditions. Intermediate precision was evaluated by assaying solutions on different days. Peak areas were determined and compared. Precision was expressed as percentage relative standard deviation (RSD% < 2). From the data obtained, the developed RP-HPLC method was found to be precise, and results are reported in Table 4. The results obtained by the proposed method for repeatability were compared with those of the literature HPLC method (7) and found that %difference between %mean assay (n = 6) is 0.2% as shown in Table 6.

Accuracy (recovery test)

Accuracy of the method was studied by recovery experiments. The recovery experiments were performed by spiking solution of known amounts of the drugs in the placebo. The recovery was performed at three levels: 50%, 100% and 150% of the label claim of the tablet (100 mg of aceclofenac and 50 mg of diacerein). Placebo equivalent to one tablet was transferred into a 200-mL volumetric flask, and the amounts of aceclofenac and diacerein at 50%, 100% and 150% of the label claim of the tablet were added. Three samples were prepared for each recovery level. The solutions were then analyzed, and the percentage recoveries were calculated. The recovery values for diacerein and aceclofenac are shown in Table 4.

Table 4 Results from validation studies

LOD and LOQ

The LOD and LOQ for diacerein and aceclofenac was determined at a signal-to-noise ratio of 3:1 and 10:1, respectively, by injecting a series of dilute solutions with known concentrations. The LOD and LOQ are as shown in Table 4.

Robustness

The robustness of a method is the ability to remain unaffected by small changes in parameters. To determine the robustness of the method, experimental conditions were purposely altered, and system suitability parameters were evaluated. The flow rate of the mobile phase was 1.0 mL/min. To study the effect of flow rate on system suitability parameters, it was changed to 0.2 units from 1.0 mL/min to1.2 mL/min and 0.8 mL/min. The effect of column temperature on system suitability parameters was studied at 20°C and 30°C instead of 25°C. The effect of pH variation of buffer on system suitability parameters was studied at pH 2.5 and pH 2.9 instead of pH 2.7. The effect of organic variation (acetonitrile) in mobile phase on system suitability parameters was studied at 90% and 110% instead of 100%. Robustness results are shown in Table 5.

Table 5 System suitability parameters and robustness
Table 6 Determination of aceclofenac in tablets and comparison with the reference method

Conclusions

The developed simple LC method for assay determination of diacerein and aceclofenac is linear, precise, accurate and specific. The method was validated to the requirements of ICH, and the results were satisfactory. The developed stability-indicating analytical method can be used for the routine analysis of production samples, where sample load is higher and high throughput is essential for faster delivery of results. Overall, the method provides a high throughput solution for determination of diacerein and aceclofenac.