Introduction

Metal on polyethylene (MOP) has traditionally been the most frequently used bearing system in total hip arthroplasty (THA) [1]. However, wear-induced osteolysis continues to be a common cause of failure for this bearing surface [1, 2]. Metal-on-metal (MOM) THA has emerged as an attractive alternative bearing surface in THA because of its potential for decreased wear and reduced dislocation rates [3, 4].

Although the cobalt-chromium alloys used for MOM THA have decreased volumetric wear [5], they have been shown to release up to 500 times more particles as compared to the MOP implants [6]. These released metal particles can undergo oxidation, reabsorption and eventual release in the blood [7]. The metal ion levels in patients with MOM implants have been shown to be higher as compared to patients with other bearing surfaces [8, 9]. The increasing number of joint replacements especially in younger patients who are exposed to orthopaedic metal alloys is raising questions as to the long-term effects of exposure. Concerns include possible local tissue toxicity, hypersensitivity, altered lymphocyte concentrations, osteolysis, irreversible chromosomal damage and carcinogenicity [1015]. Arthroprosthetic cobaltism is now a recognized serious complication characterized by neurological, endocrine and cardiac symptoms [1618].

Several clinical and in vitro hip simulator studies have shown a number of factors that influence the amount of metal ions produced in MOM bearing surfaces. These have primarily included modularity of the prostheses [1923], overall position of the implanted components [2428] and the size of the femoral head [23, 2836]. Most of the clinical studies related to metal ion blood levels in MOM bearing surfaces have evaluated only MOM hip resurfacing (to see the effect of component position) [2528] or compared large-head MOM hip resurfacing with small-head MOM THA (to see the effect of femoral head size) [23, 2830, 32, 33, 36].

Although the bearing surface is similar in both MOM hip resurfacing and MOM THA, the corrosion at the head-neck junction (trunnion) in MOM THA can be an additional source of metal ion release and this could be dependent on the femoral head size [1923]. A few studies have actually looked at the effect of femoral head size [31, 34, 35] on the metal ion levels in MOM THA. However, these studies have mainly compared metal ion levels in small-head (<36 mm) and large-head (≥36 mm) MOM THA. Corrosion at the head-neck junction can be particularly problematic in large diameter MOM THA because of higher friction, which increases rotational movement at the head and neck junction [20]. This in turn can produce high metal ion levels. Whether there could be any differences in metal ion levels due to component size or specific implant type in large-head MOM THA has not yet been thoroughly studied. To our knowledge, there are only two studies [20, 37] comparing metal ions and clinical outcomes of different large-head MOM total hip prostheses. However, these studies did not evaluate the effect of femoral head size on metal ion levels in their series of large-head MOM THA.

The primary objective of this study was to evaluate the distribution of metal ion levels among the three different large-head MOM total hip systems: two monoblock large-head systems (the Durom Socket, Zimmer, Warsaw, IN, USA and the Birmingham socket, Smith and Nephew, Memphis, TN, USA) and one modular metal-on-metal total hip system (Pinnacle, Depuy Orthopedics, Warsaw, IN, USA). The secondary objectives were to assess whether position of the implanted prostheses, patient demographics or factors such as activity levels influence overall blood metal ion levels and whether the short-term functional outcome differs with specific large-head MOM implant type.

Materials and methods

This was a cross-sectional cohort study recruiting 54 patients who had received three different large-head MOM prostheses. The three hip systems included two monoblock large heads (Durom or Birmingham socket) and one modular system (Pinnacle). All the stems used for these implants had a 12/14 taper for head-neck junction. The exclusion criteria included diagnosis other than osteoarthritis, infection, metal components elsewhere in the body and renal insufficiency. The minimum time for follow-up was selected as 1 year to allow for the fact that metal ion levels are known to be elevated in the initial run-in period lasting for approximately 1 year postoperatively [3, 35].

The mean age of the patients was 59.7 years with the mean follow-up time of 41 months (range, 12 to 60 months). There were 19 patients in each of the monoblock large head groups (Durom or Birmingham socket) operated by a single surgeon (JP). The third modular Pinnacle group had 16 patients operated by another surgeon (BB). Both the surgeons used a posterior approach to expose the hip and cementless fixation of the components. The median head size used in the Birmingham group was 52 mm (range 44 mm to 56 mm) and in the Durom group, 48 mm (range 42 to 54 mm). The median head size used in the modular Pinnacle group was 40 mm (range 36 to 44 mm).

Following institutional review board approval, all the patients were contacted by telephone and asked if they would be interested in participating in the study. Patients agreeing to participate were invited to attend dedicated clinics for clinical, radiological and biochemical evaluation.

Clinical evaluation

This included assessment of Harris hip score, Western Ontario and McMaster Universities (WOMAC) index, short form 36 (SF 36), and University of California Los Angeles (UCLA) activity score (1, ‘no physical activity, dependent on others’ to 10, ‘regular participation in impact sports’). Range of motion (ROM) was calculated as part of the Harris hip score.

Biochemical evaluation

Venous whole blood samples were obtained for cobalt (Co) and chromium (Cr) levels. Blood in the first 40 patients assessed was initially sent to two separate laboratories, Trace elements Laboratory, London, Ontario, Canada and Alberta Centre for Toxicology, Calgary, Alberta. Both the laboratories utilized an octopole reaction system (ORS) inductively coupled plasma mass spectrometer (ICPMS) for measuring the whole blood metal ion levels. The ORS-ICPMS method has been previously described by Pei et al. [38]. Due to a high correlation in metal ion results between the laboratories, the final group of patients was assessed in just one location (Calgary). All blood ion levels from the Calgary laboratory were used in our data analysis.

Radiological evaluation

Standardized anteroposterior (AP) and cross table lateral radiographs of the pelvis and hip were assessed. All AP radiographs were taken with the legs in 15° of internal rotation. The best pre- and postoperative films were selected for each patient and assessed using Dicom software (United Kingdom) for calculation of inclination and anteversion of the acetabular components.

Statistical analysis

This was performed on all the collected data to assess for any differences between the three groups in terms of demographics, overall blood metal ion levels, acetabular component position and clinical outcomes. As this was an exploratory study, we did not conduct an a priori sample size calculation. Analysis of variance or t tests were used for the comparisons of means (for continuous variables with near symmetrical distribution) across device type. For continuous variables that did not have approximately symmetrical and normal distributions, the medians (with the 25th percentile and 75th percentile) were calculated. Mann–Whitney non-parametric tests were used for comparisons of these variables across device types (large modular vs Pinnacle devices). Binary (yes/no) and categorical variables were evaluated as percentages. Corrected χ2 tests were used to compare binary and categorical variables (calculated as percentages) across device types. We considered two-tailed probability values <0.05 to indicate statistical significance for all statistical tests.

Results

The blood metal ion levels in the larger monoblock large heads (with a Durom or Birmingham socket) were significantly raised compared to the Pinnacle group (Figure 1). Median Co levels were 2.8 and 3.3 μg/l in the Durom and Birmingham groups, respectively, compared to only 0.52 μg/l in the Pinnacle group (p < 0.001). Median Cr levels were 2 and 2.2 μg/l in the Durom and Birmingham groups, respectively, compared to only 1.2 μg/l in the Pinnacle group (p < 0.001). In all the groups, however, the whole blood metal ion levels were within an acceptable safe range when compared to the upper safe limit of metal ion levels as set forth by various studies [3942].

Figure 1
figure 1

Box and whisker plots showing the distributions of chromium and cobalt levels for all patients by implant type. For each device type, the top and bottom of the ‘box’ represents the 75th and 25th percentiles of the data. The heavy horizontal line through the box is the median. Circles outside of the ‘whiskers’ represent outliers. The top and bottom of the whiskers are the maximum and minimum values that are not outliers.

Although there were observed differences in abduction and anteversion angles for the different devices (Figures 2 and 3), these differences were not shown to be statistically significant. Further, the abduction and anteversion angles were found to be within acceptable limits for most of the patients and did not seem to correlate with blood metal ion levels (Figures 2 and 3).

Figure 2
figure 2

Relationship between acetabular anteversion and cobalt levels by implant type.

Figure 3
figure 3

Relationship between acetabular abduction angle and cobalt levels by implant type.

As summarized in Table 1, there were no statistically significant differences between the three groups in terms of age, sex, functional outcome, UCLA activity score or follow-up time. The ROM scores (calculated as part of the Harris Hip Scores) were also not significantly different between the three groups. None of these factors was found to be related to whole blood metal ion levels in the patients. There was no dislocation in any of the patients. The only complication reported was infection following a dental abscess in one of the patients in the Durom group.

Table 1 Patient characteristics by implant type

Discussion

Measurement of metal ion levels in the blood following MOM arthroplasty is important as higher levels have been found to be associated with prosthetic failure [2123, 37]. There have also been reports of metal ion toxicity in a few patients [1618]. Unfortunately with the large numbers of different combinations of implants [29, 30, 34, 39] as well as different methods of analysis to determine minute traces of metal ion levels with the blood, it is difficult to compare published studies on this subject. A safe level of metal ions in blood is also hard to define [40]. A serum Cr level of >17 μg/L and a Co of >19 μg/l is more likely to be associated with metallosis leading to tissue damage [4042]. The United Kingdom’s Medicines and Healthcare Products Regulatory Agency [42] has advised that patients with MOM implants who have pain, prosthetic malposition or implants that are known to have excessive failure rates should be evaluated with the measurement of serum cobalt. Whilst normal serum cobalt and chromium levels are known and it is recognized that they are elevated following metal-on-metal THR, toxic levels are yet to be established [40]. Every effort should be made to try and reduce these when considering MOM arthroplasty.

The current study was primarily aimed at evaluating blood metal ions in three commonly used larger head MOM THA systems. These hip systems included sockets from Zimmer (Durom), Smith and Nephew (Birmingham) and Depuy (Pinnacle). The Durom is a one-piece metal socket device with plasma coating. The Birmingham cup is also a one-piece metal acetabular socket with beads on the outer surface coated with hydroxyapatite. Both these monoblock sockets articulate with a large metal head. The Pinnacle, on the other hand, is a modular system with a metal liner (Ultamet) sitting inside the metal shell. The size of the metal head that articulates with the metal liner in the Pinnacle system, although larger than traditional metal on poly bearings, is smaller than the head size in the monoblock systems (Durom or Birmingham).

The most important finding of the study was a clear demonstration of statistically significant differences in the metal ion levels in patients following a monoblock large-head MOM arthroplasty system compared to a smaller modular MOM hip arthroplasty. In our series, the smaller head size (36 to 44 mm) appeared to produce less metal ions whilst at the same time was large enough to increase hip stability and provide a range of motion and function comparable to the larger head size (>44 mm).

Size of the femoral head has been shown to be an important predictor of blood metal ion levels in MOM arthroplasty [23, 2836]. Although a larger femoral head size theoretically produces less volumetric wear [3] (and hence lower metal ions in blood), this has not been consistently shown to be the case in various studies on the subject. Most of the studies evaluating the effect of femoral head size on blood metal ion levels have specifically looked for the same in hip resurfacing procedures or compared ion levels in MOM resurfacing and MOM total hip arthroplasty [2830, 33, 36, 43]. These studies have been summarized in Table 2. However, a MOM THA may behave differently as compared to the MOM resurfacing as far as the amount of metal ions released is concerned. This could be due to a higher modularity of MOM THA systems and the presence of additional MOM surfaces like the head-neck junction, which could be prone to corrosion and metal ion release [1923]. As such, the effect of femoral head size in a MOM THA may be different from MOM resurfacing. Only four studies, to our knowledge, have compared the metal ion levels in small and large-head MOM THA [31, 32, 34, 35]. However, none of these found metal ion levels to be consistently dependent on the size of the femoral head. These studies are summarized in Table 3.

Table 2 Effect of femoral head size on metal ion concentrations
Table 3 Effect of femoral head size on metal ion concentrations

There have been several reports over the last few years suggesting a higher than anticipated failure rate of large-head monoblock MOM prostheses. This has raised concern and has led not only to the withdrawal of some devices from the market but also advice from the British Orthopaedic Association to consider carefully and possibly avoid the use of large diameter MOM bearings [44]. These failures are related to higher circulating levels of metal ions in the blood [37].

There are a number of implant design factors that could potentially influence the production of metal ions including the fluid film to surface roughness ratio (Lambda ratio), the surface roughness of a material as well as hardness and sphericity [45]. However, it has not yet been clearly established if specific implant types are related to the higher failure rate of large-head MOM THA and if metal ion levels (which correlate with failure) differ with different large-head MOM THA implant systems. Only two studies [20, 37] prior to the current study have actually looked at that. These studies have been summarized in Table 4.

Table 4 Comparison of metal ion levels and clinical outcomes in various large-head MOM THA systems

Several authors have looked at the effect of increasing head diameter on the wear rates of a MOM THA [4649]. Unlike MOP THA, the wear rate for MOM THA has been shown to be lower with increasing head diameter. Liu et al. explained that this is based on the fact that a larger diameter head increases the entrainment velocity, thereby improving the lubrication and decreasing wear with a MOM bearing [50]. Bowsher et al. [51] in their study showed a reduction in wear with a size 40 mm bearing when compared to size 28 mm. However, the authors actually found an increase in wear with 56 mm bearings. This may be related to the clearance of the couple. In a study by Leslie et al. [52], there was actually a reduction in wear rates of larger head MOM couples (55 mm head) as compared to smaller heads (39 mm) with similar radial clearances. The authors explained their results based on the reduction in inlet pressure gradients with increasing head size. Thus, multiple factors and not just the head size may affect the wear rate (and hence the metal ion levels) in a MOM THA.

Increasing the femoral head size in THA is beneficial in many respects, especially with regard to joint stability and avoidance of component impingement [20]. In a comprehensive review of the literature published by Cross et al. [53], the authors concluded that increasing femoral head size decreases the risk of postoperative dislocation and improves impingement-free ROM. However, volumetric wear increases with large femoral heads on polyethylene and increases corrosion of the stem in large metal-on-metal modular THA. Also, the authors found that the range of motion and impingement is not much different once the head size goes beyond 36 mm. This is explained by the fact that it is the component impingement that restricts the ROM in small diameter heads, while in the larger diameter heads, it is actually the bony impingement (and not component impingement) that restricts ROM beyond one stage [53]. Thus, there may not be any mechanical advantage (as far as ROM/impingement is concerned) once the head size goes beyond 36 or 40 mm.

With increasing concerns regarding the higher failure rate of large-head MOM THAs, it is important to identify the safe upper limit for femoral head size in MOM THA. Few reports have compared the outcomes of different sized bearings in MOM THR. However, none to our knowledge has actually compared large head (36 to 44 mm) sizes to even larger (>48 mm) sizes as in our study. Pattyn et al. [36] have published a prospective study on comparing three metal-on-metal bearing surfaces, namely the Durom socket, the Birmingham socket and the Metasul total hip system, which is a modular total hip system. This study identified ion levels to be lower in one of the larger head resurfacing groups which appears to conflict with the data of the current study. Bernstein et al. [34] compared two groups of patients, one with head sizes of 28 and 36 mm and another group comparing size 40 and 44 mm heads and found no significant relationship between metal ion levels in whole blood and the size of the femoral heads. They used a similar design to the current study, but the maximum head size examined was only 44 mm, which may be the reason for their conclusion.

Although the short functional outcome scores (WOMAC, SF-36 and Harris Hip Score) were similar in all three MOM THA systems in the current study, metal ions were found to be higher in the larger monoblock heads as compared to the smaller head modular Pinnacle system. One possible explanation for the difference in the blood metal ions could be a mismatch between the head and neck size of the monoblock head implants. A higher mismatch in the head and neck size can lead to increased rotational movement and torque at the head-neck junction [20]. This can, in turn, lead to higher wear at the head neck-junction and higher release of metal ion levels. In fact, some studies have shown increased corrosion and fretting as a result of wear at the trunnion-head interface (thereby increasing metal ion release in the body) as a result of large diameter MOM THA [54, 55]. However the concept of increased torque at the head-neck junction with a larger head is just a theoretical explanation and our study can in no way prove this concept.

In the current study, age, BMI, medical comorbidities (smoking, diabetes, etc.) and postsurgery activity levels were not found to have an effect on blood metal ion levels. These results were very much similar to those reported previously [2, 2022, 2437]. Further, the blood metal ion levels were not found to have a statistically significant relationship to anteversion or abduction angle of the acetabular component as measured on the radiographs. This was clearly in contrast to the findings of previous studies that have found higher blood metal ion levels with decreased anteversion and increased abduction angles of the acetabular component [2528]. One of the potential reasons for this was the relatively smaller number of patients and lack of sufficient power in our study to evaluate this effect.

There were a few limitations of the current study. The cross-sectional cohort design makes it far less ideal as compared to a randomized control trial. Also, the number of patients enrolled in the study was small. This was an exploratory study. A larger study population would have increased the power of our study and enabled us to better evaluate the effect of other factors (like component position) on the level of metal ions in the blood. Our strict exclusion criteria and patients inability to participate in this call back study further reduced the numbers. Another potential limitation is the inability to rule out the effect of different manufacturing techniques utilized by the three companies as partly responsible for the differences in the metal ions. Although we strongly believe that a larger head was the major reason behind higher metal ion levels in the Durom/Birmingham systems, we cannot really conclude that there are no other factors contributing to this difference in the metal ion levels in the three MOM hip systems included in our study. Also, although there was no dislocation in any of the patients in all the three groups, the study is underpowered to make any analysis on this given the small risk of dislocation following total hip arthroplasty now.

In the current series of patients, only one patient underwent a revision procedure. This was following haematogenous infection after a dental abscess which happened a couple of months after our review for the current study. The lead author (JP) actually has a series of 41 Durom MOM THR. Eight of these patients have now undergone revision procedures for implant failure related to prosthetic design. These patients presented with persistent pain and the absence of bony ingrowth in the acetabular socket. None of them had evidence of metallosis at the time of revision. Therefore, whilst in our study the short-term functional results are very similar between the monoblock systems, this may not represent the true overall long-term outcomes. Despite these limitations, this study indicates that there are significant differences in metal ion levels among large head devices, and it provides valuable information that can be used for designing more comprehensive evaluations of factors related to metal ion levels.

Conclusion

In summary, the current study is one of a few of its kind to compare blood metal ions and short-term clinical outcomes in large-head (36 to 44 mm) modular MOM THA systems with even larger (>48 mm) monoblock MOM THA systems. We have shown higher blood metal ion levels in the larger monoblock MOM THA systems. The Pinnacle type modular systems (with head size in range of 36 to 44 mm) seem to be large enough (similar to ≥48 mm heads) to provide advantages of joint stability, better function and lesser impingement (as against small diameter heads ≤32 mm heads). At the same time, the modular systems have a femoral head that is small enough so as not to cause an excessive increase in metal ion levels in the blood as seen with monoblock head designs with larger metal heads (>48 mm). We believe that when considering a metal-on-metal THR, the use of a monoblock large-head system offers no apparent advantage to the smaller head modular system. In our practice, we are no longer using this design and the safest strategy, when considering MOM bearings, is to use a modular system with a smaller head.