Background

Early onset scoliosis (EOS), which first presents in children under the age of 5 years, has high morbidity and mortality rates due to chest wall deformities that restrict pulmonary development. Current treatment methods include casting, bracing, rib expansion and spine distraction. Conservative treatments such as bracing or casting are not always effective. Surgical treatments such as rib expansion and growing rod instrumentation typically require multiple surgeries and involve many complications. Most growing rods are lengthened at 6 month intervals [1]. Complications include infection, rod breakage, screw pull-out, joint fusion, and junctional kyphosis. Constructs that lengthen magnetically are under investigation. Although these would reduce repeated surgeries, they are relatively stiff and bulky, the elongating section cannot be contoured, and MRI is contraindicated. Physicians have suggested that a more flexible growing rod construct might result in a more flexible spine with fewer surgical complications, provided construct strength is sufficient for curve correction.

Polymer rods have been previously investigated for use in adult, short segment, lumbar spine surgery [2, 3]. The material, polyetheretherketone (PEEK), has a lower modulus than traditional rod materials, which might allow for greater range of motion (ROM) than the current metal rods of cobalt-chrome (CoCr) or titanium (Ti) alloys. However, it is not known whether PEEK rods of the required length and diameter have sufficient stability to withstand physiological loads and provide distraction and curve correction, even in very young children with low physiological demands. The purpose of this study was to determine changes to the biomechanical properties of skeletally immature spines after implantation of simulated growing rod constructs with a range of clinically relevant structural properties. The hypothesis was that ROM of spines instrumented with PEEK rods would be both much greater than metal rods and significantly lower than noninstrumented controls. It was expected that ROM with PEEK rods would remain closer to controls than to metal rod constructs, and so be unlikely to provide sufficient stability. Further, adjacent segment motion was expected to be lower with polymer rods compared to conventional systems.

Methods

In vitro biomechanical tests were conducted on six skeletally immature porcine thoracic spines (domestic pigs, 10-14 weeks of age, body mass 35-40 kg). Spines were harvested after death of swine that had been utilized for other studies (approved by IACUC, University of Cincinnati) which had not involved the spine.

Spines were sectioned to include vertebrae T1-T13. Specimens were tested before and after instrumentation. Paired pedicle screws were inserted into T3 and T4 for the proximal anchor, and into T10 and T11 for the distal anchor (Figure 1). An open intervertebral joint remained above and below the surgical construct. Specimens were tested 1) before dual rod insertion, followed by 2) PEEK rods (6.25 mm dia, n=6) (Figure 2), 3) Ti rods (4 mm dia, n=6), or 4) CoCr rods (5 mm dia, n=4). Tests were conducted in lateral bending (LB) and flexion-extension (FE) by applying continuous loads to peak moments exceeding ±5 Nm using a materials test system with cable-pulley attachments. Five cycles were applied, with the fourth analyzed. Vertebral positions at each level were measured from arrays of markers (LEDs) using high definition video. Rotations were calculated using a customized program (MATLAB). Range of motion was defined as the maximum side-to-side rotation for each level. ROM for the treated region was determined by adding ROMs at each instrumented level (T3-T11). Differences in ROM by treatment were determined by two-tailed paired t-tests and Bonferroni correction based on four primary comparisons: PEEK vs control and PEEK vs CoCr, in LB and FE (α = 0.05/4 = 0.0125).

Figure 1
figure 1

Radiograph of spine test specimen Spine test specimen with dual titanium rod construct, coronal view. The rods are anchored using two pairs of pedicle screws at the proximal end, and another two pairs at the distal end. One noninstrumented adjacent disc is above and below the instrumented region.

Figure 2
figure 2

Photograph of spine test specimen Spine test specimen with dual PEEK rods, with one rod visible, mounted for flexion-extension test. At each vertebra, a marker array with 3 white LEDs was inserted into the anterior aspect for video motion analysis.

Results

Lateral bending

In lateral bending, ROM after each treatment was lower than noninstrumented control, including treatment with PEEK rods, at each of the instrumented levels. PEEK rods allowed greater ROM than both Ti or CoCr rods at every instrumented level. ROM was greater at the proximal and distal noninstrumented segments of instrumented specimens compared to control. Mean ROM at proximal and distal noninstrumented levels was lower for PEEK than for Ti and CoCr. Combining instrumented levels, ROM for spines with PEEK rods was 35% of noninstrumented controls (p<0.0001), and 2.7 x greater than spines with CoCr rods (p<0.01).

Flexion-extension

In flexion-extension, ROM after each treatment was lower than noninstrumented control, including treatment with PEEK rods, at each of the instrumented levels. PEEK rods usually allowed greater ROM than Ti or CoCr rods at individual levels, but variability was greater in FE than in LB. Mean ROM at proximal and distal noninstrumented levels was lower for PEEK than for Ti and CoCr. Combining instrumented levels (Figure 3), ROM for spines with PEEK rods was 27% of noninstrumented control (p<0.001), and 1.8x greater than spines with CoCr rods (p<0.01).

Figure 3
figure 3

Range of motion (ROM) over the instrumented region of the spine for four test conditions Range of motion for control, non-instrumented, spine segments, and for spines instrumented with PEEK, titanium, and cobalt-chrome (CoCr) rods. *** p< 0.0001, ** p< 0.001, * p< 0.01.

Discussion

The large reduction in ROM after instrumentation with PEEK rods indicated significantly increased spine stability due to the construct. Flexible polymeric growing rod constructs significantly decreased range of motion compared to noninstrumented controls. ROM with PEEK rods remained significantly greater than ROM with cobalt-chrome alloy rods. The hypothesis that PEEK rod constructs would clearly not provide sufficient stability for growing rod constructs was not supported. The hypothesis that PEEK rods would provide increased flexibility compared with metal rods was supported. Further, smaller increases in ROM at proximal and distal adjacent discs occurred with PEEK compared to the metal rods, which may decrease propensity for junctional kyphosis [4].

The concept of flexible growing rods has been explored with computational modeling [5]. To the investigators’ knowledge, this is the first biomechanical study of flexible growing rods. Many other studies are necessary to determine the clinical potential of this concept prior to translation. Limitations include in vitro tests on physiologically normal quadruped spines, and intact rods that did not contain any distraction mechanism. Further tests are needed in buckling and torsion, and strength and fatigue properties are essential. Physiological loads of body weight, activity, and curve correction are not yet well defined. The next stages of this preliminary work will likely investigate the effect of a realistic distraction mechanism. Molding of the initial rod configuration into curves to better approximate a desirable sagittal profile may be incorporated. However, results of this early feasibility study indicate that continued investigation into the concept and potential application of flexible growing rods for early onset scoliosis remains warranted.

Conclusions

In this pilot study, flexible polymeric growing rod constructs decreased range of motion compared to noninstrumented controls, and significantly increased ROM compared to CoCr rod constructs. Growing rods of more flexible materials may eventually help improve treatment options for young patients with severe spinal deformity. Retention of more spine flexibility would likely allow for fewer complications, and higher satisfaction for patients, parents, and caregivers.

This is the extended abstract of IRSSD 2014 program book [6].