Background
Variable flip angle 3D TSE (SPACE) is a promising technique for vessel wall imaging that provides high resolution and the freedom of 3D retrospective formatting compared with 2D TSE. However it is difficult to completely suppress residual luminal blood which may mimic plaque leading to difficulties in diagnosis [Fan. JMRI. 2010:31], especially after contrast enhancement (CE). Additionally, a more efficient imaging protocol is needed that covers both carotid and intracranial arteries as atherosclerosis is a systematic disease that usually affects multiple vascular beds. The purpose of this work is to develop a 3D combined carotid and intracranial vessel wall imaging protocol with high resolution (0.7mm3 isotropic) and excellent blood suppression both pre- and post-CE.
Methods
A DANTE module [Li. MRM. 2012:68] was added before each SPACE readout train in order to improve the suppression of slow flow. The DANTE pulse train parameters were optimized based on simulations and phantom study to minimize SNR loss and banding artifacts while maintaining sufficient flow dephasing effects. A 16-channel head/neck coil was used together with a 4-channel special purpose surface coil on a 3T scanner (Siemens Verio). A coronal imaging slab slightly rotated transversely was oriented to cover both carotid and vertibrobasilar arterial systems with one scan. Healthy subjects (n=14) and symptomatic patients (n=6) were scanned with DANTE-SPACE, SPACE-only and T1w 2D TSE in a randomized order pre- and post-CE (0.1mmol/kg). DANTE-SPACE parameters include: FA = 10; train length = 100; Gxyz = 25 mT/m; TR = 780ms; spatial resolution = 0.7mm3 isotropic; 80 slices; scan time = 5'57".
Results
DANTE-SPACE improved both arterial and venous blood suppression compared with SPACE-only (Figure 1). Five (pre-CE) and 9 (post-CE) out of 14 healthy subjects had apparent residual blood with SPACE-only whereas 2 (post-CE) subject had apparent residual blood with DANTE-SPACE. Blinded region-of-interest analysis showed 33% (pre-CE) and 98% (post-CE) improvement in wall CNR with DANTE-SPACE over SPACE-only when residual blood was present (both p<0.001) (Figure 2A). Also vessel wall area measured from SPACE-only images was significantly larger (5.2%, p<0.001) than that from DANTE-SPACE images, possibly caused by incomplete blood suppression. Example images from the patient study were shown in Figure 2B where DANTE-SPACE helped diagnose a patient with internal carotid artery dissection and in Figure 2C where DANTE-SPACE helped identify a vulnerable plaque with calcification, intra-plaque haemorrhage and thin fibrous cap.
Conclusions
DANTE-SPACE significantly improved blood suppression and vessel wall CNR both pre- and post-CE when SPACE-only resulted in residual blood. Combined high-resolution carotid and intracranial imaging using DANTE-SPACE was feasible and time-efficient (<6 min), with the potential ability to identify different plaque components including calcification, intra-plaque haemorrhage and fibrous cap.
Funding
NIH/NHLBI: R01HL096119.
Author information
Authors and Affiliations
Rights and permissions
This article is published under an open access license. Please check the 'Copyright Information' section either on this page or in the PDF for details of this license and what re-use is permitted. If your intended use exceeds what is permitted by the license or if you are unable to locate the licence and re-use information, please contact the Rights and Permissions team.
About this article
Cite this article
Xie, Y., Yang, Q., Fan, Z. et al. Improved black-blood imaging using DANTE-SPACE for combined carotid and intracranial vessel wall evaluation. J Cardiovasc Magn Reson 17 (Suppl 1), O17 (2015). https://doi.org/10.1186/1532-429X-17-S1-O17
Published:
DOI: https://doi.org/10.1186/1532-429X-17-S1-O17