Abstract
A low-energy electronic recoil calibration of XENON1T, a dual-phase xenon time projection chamber, with an internal \({}^{37}\)Ar source was performed. This calibration source features a 35-day half-life and provides two mono-energetic lines at 2.82 keV and 0.27 keV. The photon yield and electron yield at 2.82 keV are measured to be (\(32.3\,\pm \,0.3\)) photons/keV and (\(40.6\,\pm \,0.5\)) electrons/keV, respectively, in agreement with other measurements and with NEST predictions. The electron yield at 0.27 keV is also measured and it is (\(68.0^{+6.3}_{-3.7}\)) electrons/keV. The \({}^{37}\)Ar calibration confirms that the detector is well-understood in the energy region close to the detection threshold, with the 2.82 keV line reconstructed at (\(2.83\,\pm \,0.02\)) keV, which further validates the model used to interpret the low-energy electronic recoil excess previously reported by XENON1T. The ability to efficiently remove argon with cryogenic distillation after the calibration proves that \({}^{37}\)Ar can be considered as a regular calibration source for multi-tonne xenon detectors.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
The XENON1T experiment, located underground at the INFN Laboratori Nazionali del Gran Sasso (LNGS) in Italy, was designed for direct detection of dark matter (DM) in the form of weakly interacting massive particles (WIMPs) [1, 2]. The detector was a liquid xenon dual-phase time projection chamber (LXe TPC), hosting a liquid volume with a small layer of gaseous xenon on top. It operated with a total of 3.2 t of ultra-pure xenon. It was surrounded by a stainless steel tank filled with water which operated as an active Cherenkov muon veto against cosmic radiation [3]. The cylindrical LXe TPC, with a height of 97 cm and a diameter of 96 cm, contained 2 t of LXe as a target. It was instrumented with 248 photomultiplier tubes (PMTs) arranged in two arrays at the top and bottom to efficiently detect the signals [4]. An electric drift field was applied between a cathode at the bottom of the LXe TPC and a grounded electrode at the top (named “gate”). With a third electrode, the anode located in the gaseous phase, an extraction field was applied.
In LXe, the total energy deposited by interacting particles is divided into excitation, ionization and heat [5,6,7,8,9]. The detectable energy is split between excitation and ionization, causing these channels to be anti-correlated, whereby the ratio of energy splitting is defined by the applied drift field. Some of the freed electrons from ionization recombine with Xe ions. Excited Xe atoms and Xe ions formed from recombination create Xe dimers which emit UV scintillation light (S1). The remaining free electrons are drifted to the gate by the applied field and extracted into the gas phase, generating a secondary scintillation signal (S2) that is proportional to the number of extracted electrons.
The light distribution pattern of the S2s on the top PMT array provides x-y coordinates (\(x,y=0\) at the center of the TPC), while the time difference between the S1 and S2 signals is used to measure the depth of the interaction z (\(z=0\) at the gate). This allows for a full 3D position reconstruction of the events. Furthermore, the ratio between the S2 and S1 signals makes it possible to discriminate between electronic recoils (ERs) and nuclear recoils (NRs). ERs are induced by \(\upbeta \)-particles, \(\upgamma \)-rays or neutrinos scattering off electrons, while NRs are induced by neutrons or neutrinos scattering off nuclei. WIMPs are expected to induce NRs, while other DM candidates can generate ERs. The difference in the S2/S1 ratio between ERs and NRs is caused by the different ionization densities at the interaction sites deposited by the two interaction types, thus resulting in different excitation-to-ionization and recombination ratios in the LXe. For ERs, the empirically-determined average energy required to produce a quantum (scintillation photon or ionization electron) is independent from the energy scale and includes the fraction of energy dissipated into heat, which is not detectable in LXe. For NRs, a quenching factor is considered to account for the larger fraction of energy lost into heat compared to the electronic channel [10].
In XENON1T, the unprecedentedly low background level of \((76 \pm 2)\,\text {events}\,/(\text {t}\,\cdot \,\,\text {y}\,\cdot \,\text {keV})\) for ERs between 1 keV and 30 keV opened up the possibility of investigating alternative DM candidates and other physics beyond the Standard Model. A significant excess with respect to background expectations was found in the low-energy ER spectrum between 1 keV and 7 keV [11]. The shape of the excess statistically allowed several possible explanations: solar axions, bosonic dark matter or neutrinos with enhanced magnetic moment. A trace amount of tritium is also able to explain the excess, the data however is also compatible with a null hypothesis. This is also valid for a possible contamination with \({}^{37}\)Ar, but its concentration could be tightly constrained and it was found to be negligible.
In XENONnT, the latest phase of the XENON project, no excess above background has been found, excluding interpretations of the XENON1T excess as due to physics beyond the Standard Model [12]. Despite this recent result, a further investigation of the XENON1T excess is still valuable, and it requires a robust description of the behaviour of the detector and validation of the energy reconstruction in the region of interest.
It was important to calibrate the low-energy ER region with high statistics in the whole TPC volume. Due to the size of the LXe TPC and the self-shielding properties of LXe, external calibration sources were unable to probe the internal regions of the detector. In XENON1T two internal calibration sources were primarily used. \({}^{83\text {m}}\)Kr, a source with a two-step decay producing lines at 32 keV and 9 keV, was used to study the spatial response of the detector [13]. The \(\upbeta \)-decay of \(^{212}\)Pb (via injection of \(^{220}\)Rn) was used to characterize the low-energy ER region, but produces a continuous spectrum, which complicates studies of spatial dependence [14]. Therefore, a low-energy, uniformly-distributed, and mono-energetic calibration source was desired to validate the detector response in the low energy region where the XENON1T excess was observed.
\({}^{37}\)Ar decays by electron capture into \(^{37}\)Cl (Q-value of 813.9 keV) with half-life of 35.01 days, emitting a mono-energetic neutrino [15]. The vacancy, which follows the electron capture from one of the shells (K-, L- or M-shell) is filled by an electron rearrangement accompanied by a cascade emission of x-rays and Auger-Meitner electrons. The overall energy deposits, corresponding to the K-, L- and M-shell electron binding energies, result in lines at 2.82 keV, 0.27 keV and 0.01 keV, with a branching ratio of 90.2%, 8.7% and 1.1%, respectively [16].
The \({}^{37}\)Ar K- and L-shell lines allow to study the low-energy detector response [17, 18], since they are well below the low-energy peak of \({}^{83\text {m}}\)Kr at 9 keV and close to the 1 keV energy threshold of XENON1T. In general, K-shell decays produce detectable S1s and S2s, while at L-shell energy only S2s are expected. For this reason, two different analyses are required and performed in the following.
This paper describes the calibration with \({}^{37}\)Ar performed for the first time in XENON1T at the end of the last science run and it is organized as follows: the \({}^{37}\)Ar source production, injection in the detector and removal after the end of the calibration period are described in Sect. 2. In Sect. 3 the standard analysis of \({}^{37}\)Ar K-shell signals, making use of both S1 and S2, is presented. The impact of these results on the XENON1T low-energy ER excess is discussed in Sect. 4. Section 5 focuses on the S2-only analysis which utilises only the ionization signal, lowering the energy threshold down to the energy of the L-shell decays at 0.27 keV. Conclusions are reported in Sect. 6.
2 Source production, injection and removal
2.1 Source production
The isotope \({}^{37}\)Ar is obtained by irradiation of argon enriched in \(^{36}\)Ar to 99.99 %, contained in a quartz glass container (ampule), with thermal neutrons:
The enrichment enhances the production of \({}^{37}\)Ar and reduces the production of undesired isotopes. The sample was irradiated at the TRIGA Mark II reactor located at the University of Mainz, which provides a flux of thermal neutrons of \(4.2 \times 10^{12}\, \text {n}/(\text {cm}^{2}\,\text {s})\).Footnote 1 The total activity produced during the irradiation was (100 ± 10) kBq. This value was calculated based on the neutron flux, the neutron capture cross-section of \(^{36}\)Ar and the irradiation time. The activity produced cannot be directly measured as the low-energy electrons and x-rays do not penetrate the walls of the quartz-ampule.
2.2 Dosing and insertion into XENON1T
A rate of O(10) Hz is needed in the detector to achieve a reasonable amount of data, much lower than the overall activity produced during irradiation. A dosing system is used to open the ampule with a guillotine mechanism. It is also used to reduce the activity to a suitable calibration dosage and inject the irradiated argon into the LXe TPC via the XENON1T purification system. Besides the device to open the ampule and store the \({}^{37}\)Ar, the dosing system consists of an arrangement of sections of well-defined volumes that are used to successively dilute a small portion of the activity with xenon. The amount of \({}^{37}\)Ar injected can be computed from the known volume sizes and the starting activity. With an initial activity of about 100 kBq, the dosing systems design enables doses of O(Bq) to be injected into the detector.
The evolution of the detector trigger rate is shown in Fig. 1, without any data selection applied. The flat rate in the beginning is due to the \({}^{83\text {m}}\)Kr source which was injected in parallel to allow simultaneous calibration with both sources. This simultaneous calibration is possible due to the difference in decay energy of both sources, which allows a clean separation of the respective events. The \({}^{37}\)Ar calibration started with a computed dose of 2.2 Bq injected on October 22nd, 2018. The injection was immediately visible as an increase in the event rate of the detector and is marked as day zero in Fig. 1. Two additional injections aimed at 8.0 Bq and 3.9 Bq, respectively, were performed in the following days to ensure sufficient statistics for the later analysis. The uncertainties on these values are estimated to be around 10%, dominated by the initial uncertainty in the activity produced during ampule irradiation. Inconsistencies between the expected doses and the rate increase shown in Fig. 1 can be caused by unknown systematics from the dosing system and the fact that different dosing procedures were tested. Injecting small activities at first avoids an accidental overdose and allows verification that the produced activity is in the expected order of magnitude. However a precise knowledge of the absolute activity is not necessary to perform the detector calibration.
After data processing, an energy cut in the S2 vs S1 parameter space was applied to select the 2.82 keV \({}^{37}\)Ar peak, together with a fiducial volume cut to reduce background events. More details on the selection can be found in Sect. 3.1. A detailed model [19] was fitted to the data and yielded injected activities of (2.6 ± 0.1) Bq, (4.3 ± 0.1) Bq and (2.1 ± 0.1) Bq. Note that this model estimates the activity injected into the purification system, while Fig. 1 only shows the rate increase caused by the injection of argon inside the sensitive, liquid filled volume of XENON1T.
2.3 \(^{37}\)Ar removal by cryogenic distillation
The \({}^{37}\)Ar half-life time of 35.01 days makes active removal of the isotope after a calibration run necessary, since a reduction of at least four orders of magnitude is required to return to normal background level. Due to the higher volatility of argon compared to xenon, it accumulates in the gaseous phase and makes removal via cryogenic distillation effective. Removal was performed using the distillation column initially designed to remove \(^{85}\)Kr, one of the major background sources of XENON1T [19, 20]. Distillation was conducted online during normal detector operation which allowed for continuous removal of \({}^{37}\)Ar from the gaseous phase.
The time evolution of the \({}^{37}\)Ar K-shell event rate during distillation is shown in Fig. 2. With the same selection criteria and cuts used in the injection phase, a simple exponential function, accounting for both \({}^{37}\)Ar decay and removal by distillation, was used to fit the process. The removal time constant found for the distillation is \(\tau _{Ar, dist} = (1.97\,\pm \,0.10)\) days. This means a reduction of the \({}^{37}\)Ar activity by an order of magnitude every (4.54\(\,\pm \,\)0.23) days. For this case, background level conditions were reached again after 24 days of distillation when the \({}^{37}\)Ar activity has been reduced by five orders of magnitude and can no longer be distinguished from the background level measured before the first injection. For comparison, a waiting time of approximately 570 days would have been necessary to reach the same low \({}^{37}\)Ar activity level via natural decay. The last data point in Fig. 2 exhibits a higher uncertainty because it refers to a day in which the livetime was lower. This last measurement is at 1.1\(\upsigma \) from the expected background level. However, there are no data available after this point to track the trend of the rate on longer time scales.
The difference for the time constant to the value of \(\tau _{Ar} = 1.7\) days obtained in Ref. [19] is caused by the extended dataset used, which is taking two more weeks of distillation into account. Additional testing of the detector was performed before the start of distillation. Due to this the detector was transitioning into a new equilibrium. The resulting instabilities, which cannot be modeled with this simple exponential function, caused the large residuals shown in Ref. [19].
Since it was possible to remove the source in a reasonable timescale, \({}^{37}\)Ar should be considered as a regular calibration source for multi-tonne xenon detectors equipped with on-site distillation capabilities.
3 S1–S2 analysis
In this analysis, \({}^{37}\)Ar K-shell events are considered where both signals, S1 and S2, are observed. The goal is to determine the absolute photon yield (PY, ph/keV) and electron yield (EY, e\(^-\)/keV) of the 2.82 keV energy deposition as well as the linearity of the detector response down to O(keV) energies.
As mentioned in Sect. 1 the energy deposited by an ER is shared between the scintillation and ionization channels as
where W is the average energy required to produce one quantum (photon or electron), while \(n_{\gamma }\) and \(n_{e}\) are the number of produced photons and electrons, respectively. For consistency with other XENON data analyses, \(W=13.7\pm 0.2~\mathrm eV/quanta\) [9] is used here, although other recent works suggest smaller values of \(W=11.5^{+0.2}_{-0.3}~\mathrm eV/quanta\) [22, 23]. The choice of a different W value impacts the determination of \(n_{\gamma }\) and \(n_e\) and the values of detector dependent constants \(g_1\) and \(g_2\), which parametrize the average size of the S1 and S2 signals per photon or electron, respectively, but it does not affect the energy reconstruction itself. The S1 (S2) signals in the XENON1T LXe TPC are proportional to the number of photons (electrons) produced at the interaction site. Therefore, Eq. (2) can be converted to observable quantities:
where \(\hat{E}\) denotes reconstructed energy. The S1 and S2 signals are corrected for position-dependent effects including light collection efficiency (for both S1 and S2). Additionally S2 is corrected for the loss of electrons due to attachment to electronegative impurities, which is a function of the drift time and follows an exponential law, with the electron lifetime as a decay parameter [13]. These corrections lead to the cS1 and cS2 variables seen in Eq. (3).
The \(g_1\) and \(g_2\) values, assuming a linear detector response, relate the detected light (S1) and charge (S2) yields to the intrinsic PY and EY for a given drift field:
where \(L_y\) (\(Q_y\)) is the light (charge) yield, i.e. the average cS1 (cS2) per unit energy E. The goal of this analysis is to first determine \(L_y\) and \(Q_y\) of the \({}^{37}\)Ar K-shell, and then extract \(g_1\) and \(g_2\) to calculate the PY and EY of LXe at the XENON1T drift field of 82 V/cm [24]. The value of \(g_2\) depends on how S2 is chosen. In the S1–S2 analysis in this paper the corrected S2 signal from the bottom PMT-array only is used.
3.1 Data selection
The data used in this analysis is from a period including roughly 3.2 days of the injection phase as shown in Fig. 1.
The calibration sources can be distinguished in the S1–S2 parameter space. For \({}^{37}\)Ar, all events with \(0\,< \text {S1} < 40~\text {photoelectrons (PE)}\) and \(500\,< \text {S2} < 2500~\text {PE}\) are selected. A fiducial volume cut of \(-93\,<\,z\,<\,-9\,\) cm and radius \(r < 42\) cm is also applied for the main results of this analysis, as described in Ref. [11].
The standard pre-selections of data are applied in order to exclude events happening when the data acquisition was in a busy condition, which may result in data loss, or when a single PMT detects a large fraction of the total signal. Further data quality cuts are applied. These include selections to ensure correct pairing of S1–S2 signals and to remove misclassified events caused by single electron emission, events dominated by one PMT, or events with an unreliable x–y position reconstruction. A more complete list of all applied selections with detailed explanations can be found in Ref. [13].
3.2 Data correction
Several corrections are applied to the data to account for non-uniform detector response throughout the LXe TPC [13]. The S1 signal is corrected to account for position-dependent light collection efficiency, particularly as a function of depth (z) but also of x and y. The S2 signal is corrected for both (a) charge loss as a function of z due to electronegative impurities and (b) light collection efficiency variation as a function of x and y, due to geometric effects on the S2 gain.
The corrections described above are determined from \({}^{83\text {m}}\)Kr calibration and applied in all XENON1T analyses. Using the dataset collected with both \({}^{37}\)Ar and \({}^{83\text {m}}\)Kr injected simultaneously, the same corrections are extracted using \({}^{37}\)Ar and compared to the standard results from \({}^{83\text {m}}\)Kr, based on earlier calibrations. The electron lifetime extracted from \({}^{37}\)Ar is approximately 10% higher than the one from \({}^{83\text {m}}\)Kr. This discrepancy, which was also observed in Ref. [13], is attributed to a slight non-uniformity of the drift field in the LXe TPC. Such an effect leads to position-dependent variations in the light and charge yields.
Indeed, the non-uniform drift field produces variations in the number of produced electrons and, consequently, the detected S2. This effect is not separable from the variation of S2 due to electrons absorption by impurities. Due to the different impact of drift field at different energies on charge yield variations, the estimation of electron lifetime for various sources may differ. Hence, in order to evaluate the true electron lifetime, energy-dependent correction should be applied for specific calibration sources.
For the S1–S2 analysis described in this section, the electron lifetime based on \({}^{83\text {m}}\)Kr is used in order to be consistent with other calibration points in this science run (see Fig. 5) as well as other XENON1T analyses. While a different choice of the electron lifetime impacts on the cS2 and \(g_2\) values, the energy reconstructed with Eq. (3) is not affected.
During the electron lifetime analysis, another effect was observed that motivated an additional correction for \({}^{37}\)Ar data: in the top \(\sim \)20 cm of the LXe TPC, the size of the observed S2 signals were on average a few percent smaller than expected, as shown in Fig. 3. The source of this bias was determined to be PMT afterpulsing. Afterpulses are due to residual gas molecules in the PMT vacuum that get ionized by a photoelectron and then drift to the photocathode generating a delayed signal. Afterpulses occur within a few microseconds after the primary S2 peak and thus are typically merged together with it [25]. As the afterpulse size is directly proportional to the primary S2 size, this results in a small but constant bias that is absorbed into the value of \(g_2\), which also applies to all comparable signals. Due to the low energy of the \({}^{37}\)Ar interaction the energy is deposited in a small volume, resulting in short S2. Additionally, at the top of the LXe TPC, the broadening by diffusion is reduced, so many \({}^{37}\)Ar S2 signals occuring there were reconstructed separately from their afterpulses, resulting in a relatively smaller S2 value. An additional correction is made on these events that have secondary S2s within 5 \(\upmu \)s of the primary and with total area larger than 60 PE (\(\sim \)2 extracted electrons). For events with this topology, the areas of the primary and secondary S2 signals are summed up to a total S2 including the afterpulse area. This afterpulse correction helps to ensure a linear energy scale across all energies and across the full fiducial volume. The effect of this correction is shown in Fig. 3.
3.3 Data analysis and results
After data selection and correction, the L\(_y\) and Q\(_y\) of \({}^{37}\)Ar are determined via fitting in the S1–S2 space. The event distribution for a mono-energetic source is generally expected to follow a two-dimensional Gaussian distribution. However, at low energies there are non-Gaussian effects to be considered, particularly for S1 where the central limit theorem does not apply, namely:
-
1.
with O(10) detected photons, the Gaussian approximation to the Poisson distribution does not fully hold;
-
2.
the PMT single-PE response is not a Gaussian [26];
-
3.
for xenon scintillation light wavelength, PMTs exhibit double-PE emission, a binomial process with \(p\sim 0.2\) [24, 27].
Empirically, a skew Gaussian distribution was used for low-energy S1 signals, as in Refs. [28,29,30].
In addition to the non-Gaussian effects described above, the sharply-dropping S1 detection efficiency for S1 \(\lesssim \) 5 PE [13] is taken into account, which has a position-dependent impact on the cS1 spectrum due to spatial variations in the light collection efficiency. To include this effect, cylindrical symmetry is assumed and the LXe TPC is divided into r-z bins of equal volumes. In each bin, a two-dimensional fit is performed in the space of uncorrected S1 and corrected S2 (cS2\(_\textrm{b,AP}\), b: S2 signal detected on the bottom PMT array; AP: afterpulse correction has been included). The full model, including the detection efficiency and the signal shape of the S1 distribution is given by:
where \(\xi \) and w are mean- and width-like parameters and \(\alpha \) is related to the skewness of the distribution, with \(\alpha =0\) corresponding to the standard Gaussian. The efficiency term \(\epsilon (S1)\) is fixed (see Fig. 4) from dedicated analyses [13], and N is a normalization constant.
The cS2 spectrum is modelled using a Gaussian, and the anti-correlation between the S2 (Gaussian) mean and the S1 (skew Gaussian) mean is accounted for in the full 2D fit. Two example fits are shown in Fig. 4, including the S1 and cS2\(_\textrm{b,AP}\) projections. It can be noticed that the S1 at the top of the TPC are much more affected by the detection efficiency, as they are generally smaller. This is because they are mainly detected with the bottom PMT array due to reflections at the liquid surface. A higher value of cS2\(_\textrm{b,AP}\) at the bottom of the TPC is also apparent. This is because the electron lifetime from \({}^{83\text {m}}\)Kr, which is smaller than the one derived from \({}^{37}\)Ar, is used to correct the signals. This leads to an over-correction of the \({}^{37}\)Ar signals at the bottom of the TPC.
The mean of the underlying skew Gaussian in Eq. (6) is given by
and is taken as the mean S1 value in each r-z bin. The S1 means are scaled according to the light collection map derived using \({}^{83\text {m}}\)Kr calibration data [13] and averaged to get an overall L\(_y\) to be compared with other calibration points. Q\(_y\) is determined by taking the sample mean of the fitted cS2\(_\textrm{b,AP}\) means. In the end, the light (charge) yields are found to be
The values of \(g_1\) and \(g_2\) can be obtained determining L\(_y\) and Q\(_y\) for other calibration points beyond \({}^{37}\)Ar. Equation (3) can be converted to
The additional calibration points used to determine \(g_1\) and \(g_2\) include the injected source \({}^{83\text {m}}\)Kr, as well as high-energy gamma lines from material radioactivity and the activated xenon lines, \(^{129\textrm{m}}\)Xe and \(^{131\textrm{m}}\)Xe, which are predominately produced during neutron calibration of the detector. Since no threshold effects are applicable for these sources, the L\(_y\) and Q\(_y\) are determined by simple 2D (standard) Gaussian fits.
The L\(_y\) and Q\(_y\) values for the various calibration sources are shown in Fig. 5, where it can be seen that a linear energy response is observed over a wide energy range, including the low-energy \({}^{37}\)Ar point.
Due to the high statistics for each calibration point, the uncertainty is dominated by systematics. In order to evaluate them, the value of \(\chi ^2\) is set equal to the number of degrees of freedom. Assuming that every calibration points is affected by the same systematics, they are estimated to be around 0.6%, as the values that satisfy this condition. Then a simple Monte-Carlo simulation is used to propagate these uncertainties on \(g_1\) and \(g_2\). \(L_y\) and \(Q_y\), sampled many times from Gaussian distributions which reflect such systematics, are fitted to evaluate the distribution of \(g_1\) and \(g_2\), whose standard deviation is taken as uncertainty. Potential systematic effects include small time dependencies not fully accounted for as well as the small drift field inhomogeneity hypothesized to explain the discrepancies in electron lifetime between different sources.
The linear fit to extract \(g_1\) and \(g_2\), taking into account these systematic uncertainties, yields
With \(g_1\) and \(g_2\) determined, the detector independent photon and electron yields at the \({}^{37}\)Ar K-shell energy are
These values are shown in Figs. 11 and 12 at the end of the paper, together with expected values from NEST (Noble Element Simulation Technique) [31] and the results from other measurements at different energies and drift fields. The photon yield is in disagreement (\(>5\upsigma \)) with the NEST prediction at the same drift field, while it is in agreement within 2\(\upsigma \) with measurements taken at same [18] or similar energies [32]. A \(>3\upsigma \) tension is also apparent with the values reported by PIXeY [17], which is not explained by the weak dependence on the drift field (Fig. 12, left panel). The measured electron yield is consistent with the predictions and other measurements at similar energies within 2\(\upsigma \).
Lastly, with the \(g_1\) and \(g_2\) values found above, the energy of any event can be estimated using cS1 and cS2 signals. The reconstructed energy spectrum of \({}^{37}\)Ar K-shell captures is shown in Fig. 6, along with 2 separate fits: a standard Gaussian and a skew Gaussian. As expected, the standard Gaussian does not describe the data as well as the skew Gaussian for this low-energy source. The mean of the skew Gaussian is (\(2.83 \pm 0.02\)) keV, consistent with the literature value 2.82 keV and with the uncertainty dominated by the systematics of \(g_1\) and \(g_2\). The full-width half-maximum of the K-shell peak is \((1.04 \pm 0.01)\) keV.
4 Implications for the XENON1T ER excess
The analysis reported here is relevant with respect to a previous result of XENON1T in which an excess of low-energy ER events was reported [11]. As mentioned in Sect. 1, XENONnT data has shown no ER excess [12]. However, it is still useful to reconsider the XENON1T ER data in the context of the improved knowledge obtained from the \({}^{37}\)Ar calibration. The running conditions of the detector at the time of \({}^{37}\)Ar calibration were mostly the same as for the scientific run in which the excess was observed. In particular drift field, temperature, pressure, charge yield and light yield were stable. This makes it possible to directly compare the data from the two periods.
The model used to fit the data is updated and compared to the one applied to the results reported in Ref. [11]. The improvements included in the model are (a) the effect of PMT afterpulsing in low-energy S2s and (b) the non-Gaussian response of the detector at low energies. A third effect is the impact of field inhomogeneity leading to different electron lifetime values for different sources, as mentioned in Sect. 3.2. However, the robust energy reconstruction described in Sect. 3.3 suggests that this effect does not have a large impact on mean reconstructed energy. Thus, its impact is implicitly included in the empirical resolution model.
4.1 PMT afterpulsing
The impact of the afterpulse correction (see Sect. 3.2) in the top 20 cm of the detector was investigated to cross-check the energy reconstruction as reported in Ref. [11]. The PMT afterpulsing rate was monitored throughout the operation of XENON1T. There were PMTs that displayed large increases in afterpulses consistent with xenon ions, suggesting these PMTs had small leaks that degraded their performance over time. The dataset analyzed in Ref. [11], called SR1, was recorded from February 2017 to February 2018, while the \({}^{37}\)Ar calibration was performed in October 2018. Therefore, any bias from PMT afterpulsing is expected to be smaller in SR1 than in the calibration reported in this work.
The afterpulse correction impacted 26% of events in the SR1 dataset with reconstructed energy below 15 keV, almost entirely in the top half of the LXe TPC, as expected. The S2 size for these events was increased on average by 3.3%, with a maximum change of 6.4%. This corresponds to the size of the deficit shown in the residuals of Fig. 3.
4.2 Non-Gaussian detector response
In the SR1 analysis [11] a Gaussian response was assumed over the full energy range for simplicity. However, as discussed in Ref. [28] and shown in this work, a skew Gaussian is a more realistic model at low energies. As this change impacts mono-energetic spectra most, an improved peak search is performed in the SR1 data with a new empirical resolution model.
The new empirical model is built using the calibration points of \({}^{37}\)Ar, \({}^{83\text {m}}\)Kr, \(^{131\mathrm m}\)Xe, and \(^{129\mathrm m}\)Xe. A skew Gaussian model (Eq. (6)) is used for all peaks instead of the typical Gaussian, and the skewness parameter (\(\alpha \)) and width parameter (w) are then fit as a function of the energy. The energy dependence of the skewness is modelled as a power law, while the width is modelled in the same way as the standard deviation of a Gaussian in similar contexts [11]. Those empirical fits are given by
with E expressed in keV.
4.3 Updated fit to SR1 ER data
A simplified version of the unbinned likelihood ratio analysis in Ref. [11] is performed, using the same exposure, background models, and search region, but without the partitioning of the dataset in time.
The result of the updated search for a mono-energetic peak is shown in Fig. 7. The best-fit mean is \((2.4 \pm 0.2)\) keV, to be compared with \((2.3 \pm 0.2)\) keV from the original analysis [11]. This indicates a small impact of the updated analysis on the energy reconstruction. The likelihood ratio curve for the peak mean is shown in Fig. 7 (right). The \({}^{37}\)Ar peak is disfavoured at \(\sim 2\sigma \) with the updated analysis, as it was already found with the previous analysis [11]. In conclusion, these changes have no impact on the interpretation of the ER data in XENON1T.
5 S2-only analysis
Any S1s from the \({}^{37}\)Ar L-shell decay (0.27 keV) are expected to fall well below the XENON1T energy threshold, which is determined by the 3-fold PMTs coincidence. However, the S2s are still detectable on account of the electron amplification in the gas phase. For this reason, the study of L-shell decay can only be performed dropping the requirement of having an S1, with an S2-only analysis. The XENON collaboration already made use of this technique to search for light dark matter signals in XENON10, XENON100 and XENON1T [33,34,35,36]. Figure 8 shows the impact of dropping the S1 requirement, which results in a L-shell peak with S2 below 1000 PE. The asymmetric shape of the peaks is due to electron absorption during drift which here is not corrected. It results that the fraction of K-shell events lost due to the S1 detection efficiency is about 8%. The events visible at the L-shell position with S1 and S2 are the results of a random S1 (e.g. dark counts) which is accidentally paired with a S2 from L-shell decay.
While this approach allows a substantial improvement of the energy threshold, it implies a worse energy resolution and no measurement of the depth of the interaction using the drift time information. This in turn results in poorer fiducialization and hence weaker background rejection.
The S2-only analysis is strictly necessary to study the L-shell, but it is also applied to the K-shell region as a cross check with the results from the standard S1–S2 analysis described in Sect. 3.
5.1 Data selection
The S2-only analysis is performed on a 6.9 days dataset acquired in the stable period after closing the krypton source and before distilling (day 7–14 after first injection).
The total S2 signal recorded by the top and bottom PMT-arrays is considered in this analysis, differently from the S1–S2 analysis. It is not possible to correct the S2 size depending on the interaction depth z. The S2s are only corrected depending on their x–y positions to account for position-dependent light collection efficiency. Moreover, to be consistent with the S1–S2 analysis, the afterpulse correction is applied, see Sect. 3.2. Nevertheless, the relationship between S2 width and depth for interactions at a given energy can be estimated with simulations. In this way, even for events without S1, a z coordinate can be statistically assigned depending on the signal width even in the L-shell energy region.
The pre-selections of data already listed in Sect. 3.1 are applied and events in which an S2 is recorded are selected. Events with S2 size lower than 90 PE (\(\sim \) 3 extracted electrons) are excluded. Events greater than 5000 PE, where \({}^{37}\)Ar signals are not expected, are also not used in this analysis.
The quality cuts applied are similar to those used in the search for light dark matter [35, 36], but they are adapted to the specific characteristics of \({}^{37}\)Ar signals, as described in the following:
-
Radial. Events reconstructed at a radius of \(r\,>\,26.5\,\,\)cm are removed. This cut is tighter than the corresponding cut in the S1–S2 analysis. Since z is not reconstructed, field distortion corrections cannot be applied and, hence, an inner region of the TPC where the field is more uniform must be selected.
-
S2 Width. Events with an S2 width outside [400, 3000] ns are removed. Since the width of S2 waveform is correlated with z, due to electrons diffusion during drift, excluded events correspond to events at the top and the bottom of the TPC.
-
S2 area fraction top. On average \(\sim \)64% of S2 light is expected to be seen by the top PMT array, given the detector geometry and the position where signals are produced. A selection was applied to remove events where this fraction is >69%, which are due to interactions in the gaseous xenon phase above the usual secondary scintillation region (‘gas events’).
-
Pile-up. Two dedicated cuts are implemented against pile-up of randomly emitted single-electron, which can be misidentified as S2s from real events. First, events whose S2 hit-pattern on the top array indicate two or more spatially separated small S2s which are reconstructed as a single event, as determined by a likelihood test, are removed. Second, events with two or more S2s or single electrons signals up to \(\sim \)1 ms before the largest S2 are not taken into account. This additionally removes cluttered waveforms, where reconstruction becomes difficult.
-
S2 Single scatter. A selection is applied to remove multiple scatter events, in which the secondary S2 has a hit-pattern compatible with that of a scatter following a primary interaction.
-
Cathode. For events where multiple S1s are present, it is checked if one of them paired with a main S2 indicates a reconstructed z at the cathode position. In this case the events are excluded.
The acceptances of the quality cuts and the trigger efficiency in a given z slice (\(-50\,<\,z\,<\,-30\,\) cm) are shown in Fig. 9, together with the signal model. S2 Width and Cathode cuts have a \(\sim \) 100% acceptance for the complete S2 range in this z slice, and they are not included in this plot. The Radial cut features the lowest acceptance, but it is almost flat with respect to S2 size. Due to the field distortion, events at the bottom of the TPC are reconstructed more inward compared to events at the top, and are less affected by Radial cut. This produce the step around 1000 PE in the transition region, where L-shell events at the top and K-shell events at the bottom of the TPC overlap.
The trigger efficiency, namely the ability to detect S2s depending on their size, has the largest impact in the lowest S2 region. The background contribution after the quality cuts are applied is negligible in the whole S2 size range.
5.2 Signal modeling
The expected signal from \({}^{37}\)Ar is evaluated following the same approach as in Ref. [35]. Since the S2 signal is uncorrected, it is not possible to use a simple Gaussian as it is done for the S1–S2 analysis. Instead, a parametric model is considered which describes the response of the detector to the \({}^{37}\)Ar energy deposition in the (S2, z) space. This model is built in three steps.
The first step is the calculation of the rate of events \(r_{p}(n_e)\) with \(n_e\) produced electrons:
where \(R_i\) is the rate of K-shell (\(i=K\)) or L-shell (\(i=L\)) decays, N the number of quanta produced, \(E_i\) the energy of the decay, W the average energy per quanta and \(p_i\) is the probability of producing an electron, with \(p_i\,=\,EY_{i}\,\cdot \,W\), where \(EY_i\) is the electron yield at the K-shell or L-shell energy.
The second step is the calculation of the rate \(r_d\) of events with k extracted electrons, which depends on z:
where \(\eta \) is the extraction efficiency, \(v_{d}\) the electron drift velocity and \(\tau \) is the electron lifetime.
The third step is the calculation of the final rate of events R(S2, z):
where \(\mu (k)\,=\,b(k)\cdot (\mu _{\textrm{SE}}\cdot k)\), \(\mu _{\textrm{SE}}\) is the single electron gain (i.e. PE detected per extracted electron) and b(k) is a reconstruction bias estimated from simulations, while
where \(\sqrt{k}\cdot \sigma _{\textrm{SE}}\) is the resolution for k electrons, \(s_{b}(k)\) is the gaussian smearing of b(k) and \(c \cdot \mu \) is the smearing accounting for x-y dependence of S2 (\(c\,=\,0.05\) as estimated from calibrations).
The parameters of the model have been set to the best-fit values from SR1 analysis in XENON1T [24] with the exception of the single electron gain \(\mu _{\textrm{SE}}\,=\,30.9\,~\mathrm {PE/e}^{-}\) which has been calculated from the value of \(g_2\) reported in Sect. 3.3. Five parameters, namely the two rates (\(R_K\), \(R_L\)), the two electron yields (\(EY_K\), \(EY_L\)) and the electron lifetime (for an independent estimate), are left free to vary and are determined from the fit to the data.
The cut acceptances and trigger efficiency are applied to R(S2, z) and then the predicted signal is projected onto uncorrected S2 to compare with the data.
5.3 Results
The model is fit to the data in the uncorrected S2 space with a binned maximum likelihood method. The fit is restricted to events whose depth, estimated on the basis of the model from S2 size and width, corresponds approximately to \(-85\, \text { cm} \,<\,z\,<\,-12\,\) cm, in order to exclude the top and bottom part of the LXe TPC. Moreover, it is performed on the central 95% region of the two peaks, since the model cannot perfectly describe the outer tails due to non accurate corrections of the field effects which also depend on z.
An extensive check of the events outside the central 95% fit region is performed. The temporal evolution of these events and their uniform spatial distribution inside the TPC are the same as for events in the central regions. It is found that these events originate from \({}^{37}\)Ar decays but are poorly reconstructed in x-y and thus end up in the signal tail region. The background contribution in the signal tails is also estimated to be negligible.
The comparison of the best fit with the data is shown in Fig. 10. The agreement is good for both L-shell and K-shell peaks (\(\chi ^2 / \text {ndf} = 135/111\)). There is a slight mismatch for the L-shell peak. This is a result of the electron lifetime, which is mainly driven by the K-shell distribution and, hence, non-optimal for the L-shell region, at whose energy less impact from field non-uniformity is expected. This value of the electron lifetime agrees with that based on \({}^{37}\)Ar obtained with the standard procedure (see Sect. 3.2), which only makes use of events from the K-shell.
Systematic uncertainties on the values obtained from the fit are studied by considering the variation of the parameters of the model inside their range of uncertainty [24]. The main contributor to the systematic uncertainties for the electron yields is the value of the single electron gain \(\mu _{\textrm{SE}}\), which has an uncertainty of 6% given by the sum in quadrature of the uncertainty on \(g_{2}\) and of the reconstruction bias. This occurs as a result of the anti-correlation between EY and the single electron gain, for a given S2. Varying other parameters in the model has a minor impact on the systematic uncertainties on the electron yield and on the branching ratio between electron captures of the L- and K-shell (L/K ratio). This results in a \(\sim \)7% systematic uncertainty on the electron yields and a \(\sim \)4% systematic uncertainty on the L/K ratio.
The electron yield best fit values for the \({}^{37}\)Ar peaks are:
The electron yield at 2.82 keV is in agreement with the result from S1–S2 analysis, see Eq. (14). As shown in Figs. 11 and 12, the electron yield at 0.27 keV is compatible with both NEST predictions [31] and with the extrapolations from the \(^{220}\)Rn calibration in XENON1T [24]. It is also in agreement within 1\(\upsigma \) with the measurement by PIXeY at a slightly higher drift field (99 V/cm) [17]. It adds information in the low-energy region (\(<\,1\) keV), where only few measurements are available [17, 37]. It has to be noted that no subtraction of the electrons produced directly by the decay and then drifted to the gas phase has been applied, differently from what performed in Ref. [38]. This allows a direct comparison with all the other measurements reported in this work, in which such a correction is also not applied.
The L/K ratio obtained from the best-fit model is
in agreement within 1\(\upsigma \) with the expected value of 9.67% [15, 16] . This suggests that the S2 detection efficiency and cut acceptances in the energy range between 0.27 keV and 2.82 keV are well understood. Furthermore, the good description of the \({}^{37}\)Ar signal for both peaks validates the modelling used for S2-only analysis in XENON1T [35].
6 Conclusions
A calibration was performed with the low-energy lines of \({}^{37}\)Ar at 2.82 keV (K-shell decay) and 0.27 keV (L-shell decay) in XENON1T. The mandatory removal of this source after calibration via cryogenic distillation was successfully demonstrated, reaching a reduction of activity by one order of magnitude every 4.5 days. This proves that \({}^{37}\)Ar can be used as a regular calibration source for multi-tonne xenon detectors.
Two analyses of the \({}^{37}\)Ar data are performed, one using both S1 and S2 signals, the other only requiring the detection of the S2. At the XENON1T drift field of 82 V/cm, the photon and electron yields at 2.82 keV are \((32.3 \pm 0.3)\,\)photons/keV and \((40.6 \pm 0.5)\,\)electrons/keV, respectively. The electron yield at 0.27 keV is \(68.0^{+6.3}_{-3.7}\,\)electrons/keV.
Since only a weak dependence from the drift field is expected at these energies, these values can be compared with measurements taken at different drift fields. As it is apparent from Figs. 11 and 12, there is in general a good agreement with other measurements at the same or similar energies. In particular agreement within 1\(\upsigma \) is observed among XENON1T and LUX, the two large detectors providing measurements at these energies. Some tension is observed with the measurements from PIXeY at 2.82 keV, which is much larger than the difference due to the higher drift field, whereas yields from Xurich II are closer to the results of this work. Disagreement is also apparent for the photon yield at 2.82 keV with respect to the NEST prediction at the same drift field. The electron yield value at 0.27 keV is in agreement with predictions and other measurements at the same or similar energies. This new measurement adds relevant information in an energy region where only two other experimental values are currently available.
The energy of the K-shell decay reconstructed with the combined energy scale is (2.83±0.02) keV, in agreement with the expected value. The results from the \({}^{37}\)Ar calibration suggests that XENON1T detector is well understood even in the low-energy ER region close to the detection threshold.
The outcomes of this work are used to inform a re-analysis of XENON1T science data in which a low-energy ER excess was previously reported. With the additional correction to the data and the updated resolution model described in this work, the main conclusions obtained in Ref. [11] remain unaltered.
References
E. Aprile et al. (XENON Collaboration), The XENON1T dark matter experiment. Eur. Phys. J. C 77, 881 (2017)
E. Aprile et al. (XENON Collaboration), Dark matter search results from a one ton-year exposure of XENON1T. Phys. Rev. Lett. 121, 111302 (2018)
E. Aprile et al. (XENON Collaboration), Conceptual design and simulation of a water Cherenkov muon veto for the XENON1T experiment. JINST 9, P11006 (2014)
E. Aprile et al. (XENON Collaboration), Lowering the radioactivity of the photomultiplier tubes for the XENON1T dark matter experiment. Eur. Phys. J. C 75(11), 546 (2015)
A. Hitachi, Properties of liquid xenon scintillation for dark matter searches. Astropart. Phys. 24, 247 (2005)
E. Aprile et al., Scintillation response of liquid xenon to low energy nuclear recoils. Phys. Rev. D 72, 072006 (2005)
A. Manzur et al., Scintillation efficiency and ionization yield of liquid xenon for monoenergetic nuclear recoils down to 4 keV. Phys. Rev. C 81, 025808 (2010)
P. Sorensen, C.E. Dahl, Nuclear recoil energy scale in liquid xenon with application to the direct detection of dark matter. Phys. Rev. D 83, 063501 (2011)
C.E. Dahl, The physics of background discrimination in liquid xenon, and first results from Xenon10 in the hunt for WIMP dark matter. Ph.D. thesis, Princeton University, AAT-3364532 (2009)
B. Lenardo et al., A global analysis of light and charge yields in liquid xenon. IEEE Trans. Nucl. Sci. 62, 6 (2015)
E. Aprile et al. (XENON Collaboration), X. Mougeot, Excess electronic recoil events in XENON1T. Phys. Rev. D 102, 072004 (2020)
E. Aprile et al. (XENON Collaboration), Search for new physics in electronic recoil data from XENONnT. Phys. Rev. Lett. 129, 161805 (2022)
E. Aprile et al. (XENON Collaboration), XENON1T dark matter data analysis: signal reconstruction, calibration and event selection. Phys. Rev. D 100, 052014 (2019)
R.F. Lang, A. Brown, E. Brown, M. Cervantes, S. Macmullin, A \(^{220}\)Rn source for the calibration of low-background experiments. JINST 11, P04004 (2016)
M.-M. Bé et al., Table of Radionuclides, Monographie BIPM-5, vol. 7. Bureau International des Poids et Mesures, Sèvres (2013)
V.P. Chechev, The evaluation of half-lives and other decay data used in nuclear astrophysics and cosmochronology. Phys. At. Nuclei 74, 1713–1717 (2011)
E.M. Boulton et al., Calibration of a two-phase xenon time projection chamber with a \(^{37}\)Ar source. JINST 12, P08004 (2017)
L. Baudis et al., The first dual-phase xenon TPC equipped with silicon photomultipliers and characterisation with \(^{37}\)Ar. Eur. Phys. J. C 80, 477 (2020)
E. Aprile et al. (XENON Collaboration), Application and modeling of an online distillation method to reduce krypton and argon in XENON1T. Prog. Theor. Exp. Phys. 5 (2022)
E. Aprile et al. (XENON Collaboration), Removing krypton from xenon by cryogenic distillation to the ppq level. Eur. Phys. J. C 77, 275 (2017)
G.J. Feldman, R.D. Cousins, A Unified approach to the classical statistical analysis of small signals. Phys. Rev. D. 57, 3873 (1998)
L. Baudis, P. Sanchez-Lucas, K. Thieme, A measurement of the mean electronic excitation energy of liquid xenon. Eur. Phys. J. C 81, 1060 (2021)
G. Anton et al. (EXO-200 Collaboration), Measurement of the scintillation and ionization response of liquid xenon at MeV energies in the EXO-200 experiment. Phys. Rev. C 101, 065501 (2020)
E. Aprile et al. (XENON Collaboration), XENON1T dark matter data analysis: signal & background models, and statistical inference. Phys. Rev. D 99, 112009 (2019)
P. Barrow et al., Qualification tests of the R11410–21 photomultiplier tubes for the XENON1T detector. JINST 12, P01024 (2017)
R. Saldanha et al., Model independent approach to the single photoelectron calibration of photomultiplier tubes. Nucl. Instrum. Methods Phys. Res. A 863 (2017)
C.H. Faham et al., Measurements of wavelength-dependent double photoelectron emission from single photons in VUV-sensitive photomultiplier tubes. JINST 10, P09010 (2015)
M. Szydagis et al., A review of basic energy reconstruction techniques in liquid xenon and argon detectors for dark matter and neutrino physics using NEST. Instruments 5(1), 13 (2021)
V.N. Lebedenko et al., Result from the first science run of the ZEPLIN-III dark matter search experiment. Phys. Rev. D 80, 052010 (2009)
D.S. Akerib et al. (LUX Collaboration), Discrimination of electronic recoils from nuclear recoils in two-phase xenon time projection chambers. Phys. Rev. D 102, 112002 (2020)
M. Szydagis et al., Noble Element Simulation Technique (v2.3.6) (2022). https://zenodo.org/record/6448408#.Y01Zo0xBw-U
D.S. Akerib et al. (LUX Collaboration), Tritium calibration of the LUX dark matter experiment. Phys. Rev. D 93, 072009 (2016)
J. Angle et al. (XENON10 Collaboration), Search for light dark matter in XENON10 Data. Phys. Rev. Lett. 107, 051301 (2011)
E. Aprile et al. (XENON Collaboration), Low-mass dark matter search using ionization signals in XENON100. Phys. Rev. D 94, 092001 (2016). (Erratum (fixing LUX points) Phys. Rev. D 95, 059901 (2017))
E. Aprile et al. (XENON Collaboration), Light dark matter search with ionization signals in XENON1T. Phys. Rev. Lett. 123, 251801 (2019)
E. Aprile et al. (XENON Collaboration), Emission of single and few electrons in XENON1T and limits on light dark matter. Phys. Rev. D 123, 251801 (2022)
D.S. Akerib et al. (LUX Collaboration), Ultra-low energy calibration of LUX detector using \(^{127}\)Xe electron capture. Phys. Rev. D 96, 112011 (2017)
P. Agnes et al. (DarkSide Collaboration), Calibration of the liquid argon ionization response to low energy electronic and nuclear recoils with DarkSide-50. Phys. Rev. D 104, 082005 (2021)
D.J. Temples et al., Measurement of charge and light yields for \(^{127}\)Xe L-shell electron captures in liquid xenon. Phys. Rev. D 104, 112001 (2021)
E. Aprile et al. (XENON Collaboration), Signal yields of keV electronic recoils and their discrimination from nuclear recoils in liquid xenon. Phys. Rev. D 97, 092007 (2018)
Acknowledgements
We gratefully acknowledge support from the National Science Foundation, Swiss National Science Foundation, German Ministry for Education and Research, Max Planck Gesellschaft, Deutsche Forschungsgemeinschaft, Helmholtz Association, Dutch Research Council (NWO), Weizmann Institute of Science, Israeli Science Foundation, Binational Science Foundation, Fundacao para a Ciencia e a Tecnologia, Région des Pays de la Loire, Knut and Alice Wallenberg Foundation, Kavli Foundation, JSPS Kakenhi and JST FOREST Program in Japan, Tsinghua University Initiative Scientific Research Program and Istituto Nazionale di Fisica Nucleare. This project has received funding/support from the European Union’s Horizon 2020 research and innovation programme under the Marie Skłodowska-Curie grant agreement No 860881-HIDDeN. Data processing is performed using infrastructures from the Open Science Grid, the European Grid Initiative and the Dutch national e-infrastructure with the support of SURF Cooperative. We are grateful to Laboratori Nazionali del Gran Sasso for hosting and supporting the XENON project.
Author information
Authors and Affiliations
Consortia
Corresponding authors
Additional information
C. Geppert and J. Riemer: Not member of the XENON Collaboration.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Funded by SCOAP3. SCOAP3 supports the goals of the International Year of Basic Sciences for Sustainable Development.
About this article
Cite this article
Aprile, E., Abe, K., Agostini, F. et al. Low-energy calibration of XENON1T with an internal \(^{{\textbf {37}}}\)Ar source. Eur. Phys. J. C 83, 542 (2023). https://doi.org/10.1140/epjc/s10052-023-11512-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1140/epjc/s10052-023-11512-z