1 Introduction

Liquid scintillators have always been an important part of many experiments in neutrino physics. Today, great success has been achieved in carrying out precision measurements of the properties of neutrinos (antineutrinos) in experiments such as KamLAND [1], Borexino [2], Daya Bay [3], RENO [4], Double Chooz [5]. These experiments have successfully deployed large volumes of organic liquid scintillator to detect neutrinos with high efficiency. Presently the next-generation neutrino experiment JUNO with 20 kt liquid scintillator is being under construction in China [6].

A liquid organic scintillator is usually composed of a solvent and scintillation dopants (fluors): a primary fluor with a concentration of a few grams per litre of solvent and a secondary wavelength shifting fluor with substantially lower concentration. Recently, linear alkylbenzene (LAB) is gaining popularity as a base solvent. Its advantages are low toxicity, high transparency, high flash point and relatively low cost. 2,5-Diphenyloxazole, also known as PPO, has been widely used as the primary fluor. 1,4-Bis(2-methylstyryl)benzene (bis-MSB) or 1,4-bis(5-phenyloxazol-2-yl)benzene (POPOP) are used as the second wavelength shifting fluor. The standard LAB-based liquid scintillator with the addition of PPO and bis-MSB or POPOP has been successfully applied in large-scale experiments such as Daya Bay [7] and RENO [8].

A new organosilicon fluor for a LAB-based liquid scintillator, so called NOL 37 (Nanostructured Organosilicon Luminophore 37) [9], is proposed and developed as an alternative to the “standard” composition of fluors (PPO + bis MSB or POPOP). A picture of samples of the LAB-based liquid scintillators with NOL 37 and PPO fluors is shown in Fig. 1. One of advantages of this fluor is that there is no need to use an additional wavelength shifting fluor anymore. In this paper, a liquid scintillator, consisting of LAB as a solvent and NOL 37 as a fluor has been extensively studied.

Fig. 1
figure 1

The sample of the LAB-based scintillator with NOL 37 (vial on the right) and “standard” scintillator with PPO (vial on the left)

2 Absorbtion and emission spectra of the NOL 37

Molecules of the organosilicon fluor have donor and acceptor sites that can effectively produce scintillation light through the mechanism of Förster resonance energy transfer (FRET). The excitation energy is absorbed by a donor site of the molecule and transferred via silicon to an acceptor site, which emits scintillation photons in response. This approach provides a high conversion efficiency of ionization energy into scintillation light with a high Stokes shift of about 100 nm compared to the standard approach when two fluors are used. Typical absorption and emission spectra of NOL 37 are shown in Fig. 2. The small overlap area of the absorption and emission spectra indicates that reabsorption is very small. Absorption and emission spectra measurements were performed using the scientific equipment of Research Sharing Center “Center for Polymer Research”, ISPM RAS.

The maximum of the emission spectrum of NOL 37 lies in the region of about 490–500 nm. Therefore, there is a mismatch of emission spectrum of NOL 37 and sensitivity curves of contemporary bialkali photocathodes usually used in photomultiplier tubes (PMTs). This problem has to be solved by adjusting the sensitivity of photomultiplier tubes to match the emission spectra of new scintillation fluors. One option is to use PMTs with a multialkali photocathode with their sensitivity maximum corresponding to the maximum of the emission spectrum of the new fluors.

The new fluor will also help to solve the problem of solvent purification, which is so critical for large-scale scintillator detectors. It is possible to use unpurified solvents because the emission spectrum with a maximum at about 490–500 nm lies in the region where the transparency of even “raw” unpurified solvents is high. This property is a great advantage, since there will be no need to create huge multistage purification systems.

3 Light yield and emission kinetics measurements of LAB-based scintillator with NOL 37 fluor

The main properties of the LAB-based scintillator with NOL 37 fluor such as light yield and emission kinetics have been extensively characterized and studied. The measurements of the light yield and the emission kinetics were carried out using a specialized setup. The liquid scintillator mixtures were poured into a cylindrical quartz cuvette. The diameter and height of the cuvette are 20 mm and 10 mm respectively. The measurement of the light yield was performed by a Photonis XP5301B PMT with high quantum efficiency bialkali photocathode. The PMT signals are digitized by LeCroy 2249W charge-to-digital converter. The cuvette was irradiated by 662 keV gamma-quanta from a \(\mathrm {^{137}Cs}\) radioactive source. Figure 3 shows the measured energy spectra of LAB-based scintillators with different concentrations of the NOL 37 fluor.

Fig. 2
figure 2

Typical absorption (black colour) and emission spectra (red colour) of the NOL 37 fluor

Fig. 3
figure 3

Results of light yield measurements of liquid scintillators with NOL 37 fluor under irradiation by 662 keV gamma-quanta from \(\mathrm {^{137}Cs}\) radioactive source

The light yield of the scintillators with NOL 37 fluor was evaluated by normalizing to the LAB-based scintillator with the addition of 1.5 g/L PPO. Due to the small size of the cuvette, there is no need to add a wavelength shifter to scintillator. The energy spectrum of LAB + 1.5 g/L PPO is also shown in Fig. 3. The position of the Compton edge and the corresponding charge were determined. The light yield values were obtained in arbitrary units.

The dependence of the PMT quantum efficiency on wavelength was measured carefully in our lab. The mean value of the PMT quantum efficiency averaged over emission spectra of PPO and NOL 37 in the wavelength range between 300 and 700 nm was calculated using the weighted arithmetic mean formula:

$$\begin{aligned} \left\langle \eta \right\rangle = \frac{\sum _{\lambda = 300}^{700} S_{\lambda }\cdot \eta _\lambda }{\sum _{\lambda = 300}^{700} S_{\lambda }} \end{aligned}$$
(1)

where \(S_{\lambda }\) is the intensity of the emission spectrum of the scintillation fluor for a given wavelength \(\lambda \), and \(\eta _\lambda \) is the quantum efficiency of the PMT corresponding to this wavelength.

The PMT quantum efficiency vs wavelength curve and two emission spectra of PPO and NOL 37 are shown in Fig. 4.

Fig. 4
figure 4

XP5301B quantum efficiency curve (black colour) and emission spectra of PPO (blue colour) and NOL 37 (red colour)

The average quantum efficiency of the XP5301B is \(\left\langle \eta \right\rangle _{\textrm{PPO}}=42.1\%\) and \(\left\langle \eta \right\rangle _{\textrm{NOL 37}}=19.1\%\) over the emission spectra of PPO and NOL 37 respectively.

In Fig. 5 the dependences of the light yield of LAB-based scintillator on the concentration of NOL 37 and PPO fluors are shown, taking into account the quantum efficiency of the XP5301B. The light yield of the scintillators with NOL 37 is almost two times higher than the light yield of the scintillator with PPO. The ratio of light yields \( LY_{\textrm{NOL 37}}/LY_{\textrm{PPO}} \) was determined to be \(1.87 \pm 0.08\) at fluors concentrations of 4 g/L, where \(LY_{\textrm{NOL 37}}\) and \(LY_{\textrm{PPO}}\) are light yields of liquid scintillators with NOL 37 and PPO respectively. Light yields of both liquid scintillators are already saturated at fluor concentration of 4 g/l.

Fig. 5
figure 5

Dependence of the light yield on the concentration of the NOL 37 (black colour) and PPO (red colour) fluors

The emission kinetics spectra were measured by a time correlated single-photon counting technique [10]. In this case, two fast PMTs (Hamamatsu R6427 and Photonis XP3112PA) and wide-range time-to-digital converter with a channel width of 75 ps were used to measure the emission kinetics. In Fig. 6 the spectra of the emission kinetics of LAB-based scintillators with NOL 37 and PPO fluors under irradiation by gamma-quanta from \(\mathrm {^{137}Cs}\) radioactive source are shown.

Fig. 6
figure 6

Light emission kinetics of liquid scintillators with NOL 37 fluors

Table 1 Decay time constants of scintillation pulse of LAB-based scintillators with new fluors

The kinetics of the scintillation process in the LAB-based scintillators with new fluor has a complex multicomponent character. The scintillation pulse is characterized by three decay time constants, \(\tau _1\), \(\tau _2\) and \(\tau _3\). The obtained decay time constants for scintillator mixtures with the different concentrations of NOL 37 are represented in Table 1. The values in the brackets are the contributions of the components to the total light yield. It should be noted that the contribution of the fastest component is more than 80% of the total light yield. The emission kinetics of LAB + 1.5 g/L PPO is characterized by decay time constants \(\tau _1=6.8 \pm 0.5\) ns (59%), \(\tau _2=16.3 \pm 0.8\) ns (37%) and \(\tau _3=64 \pm 4\) ns (4%).

Fig. 7
figure 7

Light yield of the developed LAB-based scintillator with NOL 37 kept stable between June 2018 and June 2022

4 Light yield and emission kinetics stability of LAB-based scintillators with NOL 37 fluor

Modern large-scale experiments in astroparticle and neutrino physics are designed to run for many years of operation, typically 10 years or more. Therefore, the long-term stability of the scintillator is one of the most important parameters. To perform long-term measurements for new developments is a difficult task. However, the stability of the light yield and the emission kinetics for four full years from June 2018 to June 2022 has been measured. EJ204 plastic scintillator of 2 cm in diameter and 1 cm thick was chosen as a reference scintillator, relative to which the light yield has been measured. The light yield stability of the EJ204 and the PMT’s sensitivity were monitored by the average number of photoelectrons in the PMT when the scintillator was irradiated with alpha-particles from the \(\mathrm {^{241}Am}\) radioactive source. The PMT’s sensitivity has been controlled throughout the measurements using a calibrated radio-luminescent source (RLS) [11] coupled to the PMT’s window by counting single-photoelectron pulses.

The measurement results of the light yield stability of the developed LAB-based scintillator with NOL 37 are shown in Fig. 7. The emission kinetics stability of the developed liquid scintillator was also studied. Figure 8 shows the emission kinetics spectra measured in June 2019 and June 2021.

Fig. 8
figure 8

Light emission kinetics of LAB-based scintillator with NOL 37 measured in June 2019 ( red colour) and June 2021 (blue colour)

As can be seen from Figs. 7 and  8, no changes in the light yield and the emission kinetics of the developed liquid scintillator were found over three years of measurements.

5 Conclusion

We developed new organosilicon fluor and LAB-based liquid scintillator with such fluor. The advantage of the liquid scintillator with new fluor is its high light yield, almost two times higher than the light yield of the scintillator with PPO. It should be noted that the liquid scintillator with the new fluor has a slightly slower light emission kinetics in comparison with scintillators with PPO. The measurement results of newly developed scintillator represent an intermediate step of an ongoing development programme. It is expected that further optimization of the composition of new organosilicon fluors will improve further the scintillation properties of liquid scintillators. Studies of the new organosilicon fluors show very promising results and we believe that the new fluors will be successfully applied in the next-generation large-scale experiments in neutrino and astroparticle physics.