Abstract
It was shown in a theoretical study and confirmed by experiment that a spiral multicapillary column had maximum efficiency if the bunch of capillaries was additionally coiled around its longitudinal axis to produce an integral number of coils. This technique made it possible to manufacture gas-chromatographic columns with performance as high as 12 to 16 thousand theoretical plates. These columns can find various applications, especially if quick separation is required.
Similar content being viewed by others
References
V. N. Sidel’nikov and Yu. V. Patrushev, Ross. Khim. Zh. 47(1), 23 (2003).
V. M. Gruznov, A. T. Shishmarev, V. G. Filonenko, et al., Zh. Anal. Khim. 54(9), 957 (1999) [J. Anal. Chem. 54 (9), 850 (1999)].
A. P. Efimenko and I. I. Naumenko, RF Patent No. 2114427, Byull. Izobret., No. 10 (1998).
W. S. Cooke, Today Chemist at Work 5(1), 16–20 (1996).
V. V. Malakhov, V. N. Sidel’nikov, and V. A. Utkin, Dokl. Akad. Nauk 329(6), 739 (1993).
V. P. Soldatov, A. P. Efimenko, and I. I. Naumenko, USSR Inventor’s Certificate No. 1485126, Byull. Izobret., No. 21 (1989).
A. P. Efimenko, I. I. Naumenko, and V. K. Soboleva, Zh. Fiz. Khim. 81(3), 4882 (2007) [Russ. J. Phys. Chem. A 81 (3), 410 (2007)].
Author information
Authors and Affiliations
Corresponding author
Additional information
Original Russian Text © A.P. Efimenko, I.I. Naumenko, V.K. Soboleva, 2008, published in Zhurnal Fizicheskoi Khimii, 2008, Vol. 82, No. 8, pp. 1560–1565.
Rights and permissions
About this article
Cite this article
Efimenko, A.P., Naumenko, I.I. & Soboleva, V.K. Spiral multicapillary columns. Russ. J. Phys. Chem. 82, 1388–1392 (2008). https://doi.org/10.1134/S0036024408080268
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1134/S0036024408080268