Abstract
An ionization chamber with an operating supply voltage of 10 V has been designed for use at the ITER. The sensitivity of the chamber filled with air at atmospheric pressure is 100–200 times higher than the sensitivity of the chamber pumped down to a pressure of 10−3–10−4 Torr. It is shown that application of the supply voltage modulation technique allows the pickup noise to be substantially reduced; as a result, the amplifier can be located at a large (70–100 m) distance from the chamber. Results of the experiments aimed at testing this technique on the T-10 facility are described.
Similar content being viewed by others
References
Gott, Yu.V. and Stepanenko, M.M., Rev. Sci. Instrum., 2005, vol. 7, p. 073506.
Barykov, I.A., Gott, Yu.V., and Stepanenko, M.M., Prib. Tekh. Eksp., 2006, no. 2, p. 133 [Instrum. Exp. Tech. (Engl. Transl.), no. 2, p. 270].
Veksler, V., Groshev, L., and Isaev, B., Ionizatsionnye metody issledovaniya izluchenii (Ionization Methods for Studying Radiations), Moscow: Gostekhteorizdat, 1949.
Chuklaev, S.V., Grudskii, M.Ya., and Artem’ev, V.A., Vtorichno-emissionnye detektory ioniziruyushchikh izluchenii (Secondary-Emission Ionizing-Radiation Detectors), Moscow: Energoatomizdat, 1995.
Author information
Authors and Affiliations
Corresponding author
Additional information
Original Russian Text © Yu.V. Gott, M.M. Stepanenko, 2009, published in Pribory i Tekhnika Eksperimenta, 2009, No. 2, pp. 117–121.
Rights and permissions
About this article
Cite this article
Gott, Y.V., Stepanenko, M.M. A low-voltage ionization chamber for the ITER. Instrum Exp Tech 52, 260–264 (2009). https://doi.org/10.1134/S0020441209020250
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1134/S0020441209020250