Abstract
A method for detecting thorium by its decay products has been tested on PRIPYAT six-crystal 4π γ-ray coincidence spectrometers. This method makes it possible to measure large-volume samples of any shape with a higher sensitivity and a shorter acquisition time than when an HPGe spectrometer with a comparable cost is used.
Similar content being viewed by others
References
ENSDF Web Search. Isotope Explorer. Version 3.0, http://ie.lbl.gov/ensdf/
Andjelic, T., Vukotic, P., Zizic, R., et al., Abstracts of Papers, Proc. of XXIII Symp. of Yugoslav Soc. of Radiation Protection, Donji Milanovac, Serbia, 2005, p. 317.
Adamovich, A.A., Korneev, S.V., Martsynkevich, B.A., et al., Vestsi NAN Belarusi, Ser. Fiz.-Tekh. Nauk, 2004, no. 3, p. 110.
Cooper, J.A. and Perkins, R.W., Nucl. Instrum. Methods Phys. Res., 1972, vol. 99, p. 125.
Berestov, A.V., Gutko, V.I., and Khilmanovich, A.M., Prib. Tekh. Eksp., 1990, no. 2, p. 67.
Andrukhovich, S.K., Antovich, N.M., Berestov, A.V., et al., Prib. Tekh. Eksp., 1999, no. 1, p. 123 [Instrum. Exp. Tech. (Engl. Transl.), no. 1, p. 111].
Berestov, A.V. and Zyazyulya, F.E., Prib. Tekh. Eksp., 1990, no. 2, p. 67.
GammaVision-32. Gamma-Ray HPGe Spectral Analysis. A66-B32. ORTEC, 2003.
Author information
Authors and Affiliations
Additional information
Original Russian Text © N.M. Antovic, N. Svrkota, P. Vukotic, S.K. Andrukhovich, A.V. Berestov, S.N. Harkusha, F.E. Ziaziulia, 2008, published in Pribory i Tekhnika Eksperimenta, 2008, No. 4, pp. 126–131.
Rights and permissions
About this article
Cite this article
Antovic, N.M., Svrkota, N., Vukotic, P. et al. Measuring the activity of thorium by its decay products using a multidetector γ-ray coincidence spectrometer. Instrum Exp Tech 51, 606–611 (2008). https://doi.org/10.1134/S0020441208040180
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1134/S0020441208040180