Abstract

The schematic and results of studying an interferometer for a wavelength of 8 mm with homodyne frequency conversion for measuring the electron density of a steady-state plasma are described. The design of its waveguide system allows the drift of the indicated initial phase to be reduced to a level of ±0.1°/h. The maximum phase shift measured is 360°, the tolerable signal-power decay in plasma is 16 dB, and the maximum phase measurement error is ∼5%.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Dushin, L.A., SVCh interferometry dlya izmereniya plotnosti plazmy v impul’snom gazovom razryade (Microwave Interferometers for Measurements of Plasma Density in Pulsed Gas Discharge), Moscow: Atomizdat, 1973.

    Google Scholar 

  2. Akhtar, K., Scharer, J.E., Tysk, S.M., et al., Rev. Sci. Instrum., 2003, vol. 74, no. 2, p. 996.

    Article  ADS  Google Scholar 

  3. Nagornyi, D.A., Nagornyi, A.G., and Voznyi, V.I., Prib. Tekh. Eksp., 2005, no. 2, p. 98 [Instrum. Exp. Tech. (Engl. Transl.), no. 2, p. 225].

  4. Malykh, N.I., Yampol’skii, E.S., Garanzha, I.N., et al., Prib. Tekh. Eksp., 1983, no. 5, p. 233.

  5. Malykh, L.Ya., Malykh, N.I., Prepelkin, N.F., and Yampol’skii, E.S., in Diagnostika plazmy (Plasma Diagnostics), Moscow: Atomizdat, 1968, issue 2, p. 247.

    Google Scholar 

  6. Plasma Diagnostic Techniques, Huddlestone, R.H. and Leonard, S.L., Eds., New York: Academic, 1965. Translated under the title Diagnostika plazmy, Moscow: Mir, 1976.

    Google Scholar 

  7. Voznyi, V.I., Miroshnichenko, V.I., Mordik, S.N., et al., Vopr. At. Nauki Tekh., Ser. 3: Plazm. Electron. Novye Met. Uskoren., 2003, no. 4, p. 284.

Download references

Author information

Authors and Affiliations

Authors

Additional information

Original Russian Text © D.A. Nagornyi, A.G. Nagornyi, C.C. Sidorov, 2006, published in Pribory i Tekhnika Eksperimenta, 2006, No. 6, pp. 96–99.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Nagornyi, D.A., Nagornyi, A.G. A microwave interferometer with increased stability for diagnostics of steady-state plasma. Instrum Exp Tech 49, 831–833 (2006). https://doi.org/10.1134/S002044120606011X

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1134/S002044120606011X

PACS numbers

Navigation