Conclusion
Both DC and AC measurement bridges should be calibrated for dynamic use. This paper has described methods of calibration for both types and has shown how the equipment requirement may be kept to a minimum while maintaining the accuracy of the calibration. Methods of measurement of the input and output voltage amplitudes are also suggested.
Similar content being viewed by others
References
Dixon, M., “Errors in Dynamic Force Measurement,” Strain, 24 (4), 139–142 (Nov. 1988).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Dixon, M.J. Dynamic calibration methods for transducer instrumentation. Exp Tech 14, 51–53 (1990). https://doi.org/10.1111/j.1747-1567.1990.tb01130.x