Abstract
Accurate determination of resection margins in breast specimens is important as complete removal of malignancy is a prerequisite for patients’ outcome. Mammography (DM) as 2D-technique provides only limited value in margin assessment. Therefore, we investigated whether cone-beam computed tomography (CBCT) or digital breast tomosynthesis (DBT) has incremental value in assessing margins to microcalcifications. Three independent readers investigated breast specimens for presence of microcalcifications and the smallest distance to margins. Histopathology served as gold standard. Microcalcifications were detected in 15 out of 21 included specimens (71%). Pooled sensitivity for DM, DBT and CBCT for microcalcifications compared to preoperative DM was 0.98 (CI 0.94–0.99), 0.83 (CI 0.73–0.94) and 0.94 (CI 0.87–0.99), pooled specificity was 0.99 (CI 0.99–0.99), 0.73 (CI 0.51–0.96) and 0.60 (CI 0.35–0.85). Mean measurement error for margin determination for DM, DBT and CBCT was 10 mm, 14 mm and 6 mm (p = 0.002) with significant difference between CBCT and the other devices (p < 0.03). Mean reading time required by the readers to analyze DM, DBT and CBCT, was 36, 43 and 54 s (p < 0.001). Although DM allows reliable detection of microcalcifications, measurement of resection margin was significantly more accurate with CBCT. Thus, a combination of methods or improved CBCT might provide a more accurate determination of disease-free margins in breast specimens.
Similar content being viewed by others
Introduction
Microcalcifications in the breast can be an indicator of malignancy1,2,3. In about 32% of cases, microcalcifications are the only imaging feature of ductal carcinoma in situ (DCIS) or invasive cancer of the breast4. Currently, the most important standard method for detecting breast microcalcifications is mammography, which is recommended as a cancer screening measure5 and for women with a suspected breast tumor. Suspicious microcalcifications in the breast can be evaluated by vacuum biopsies with histopathological examination6. If malignancy is histologically confirmed, additional surgery is performed. The completeness of removal of all suspicious microcalcifications can be controlled by specimen mammography during surgery.
However, mammography is a two-dimensional examination technique, which is only suitable to a limited extent for tumor size measurement7 and the examination of mostly amorphous resected tissue especially in dense breast tissue8. Due to superimposition phenomena, it is only possible to make limited statements about the tumor extension or the distances between microcalcifications and resection margin. Nevertheless, the complete removal of tumor tissue is one of the most important predictors for patients’ outcome9,10. From these points of view, it seems very reasonable to utilize three-dimensional examination techniques for the evaluation of resected specimens.
Therefore, the presented study applies the three-dimensional imaging methods digital breast tomosynthesis (DBT) and cone-beam computed tomography (CBCT) in addition to digital mammography (DM). DBT represents a mammographic technique, involving mammograms from different angles to create computed tomosynthesis layered reconstructions of the breast11,12. Especially in the case of inhomogeneously dense breast parenchyma, DBT shows diagnostic advantages over conventional DM8,13,14,15,16,17. In detection of microcalcifications it shows similar results as DM18,19,20,21,22. However, due to a limited scan angle DBT cannot produce isotropic imaging data. In contrast, CBCT is a fully three-dimensional imaging technique without breast compression, which allows reconstructions with isotropic voxels and a high spatial resolution that even enables the visualization of microcalcifications to different extents, unlike conventional computed tomography23,24,25. In addition, in other studies dedicated breast CBCT dose proved to be similar to or even less than DM dose26.
In this study, we aimed to investigate whether CBCT has additional value over DM and DBT in assessing resection margins of breast specimens with microcalcifications. Our hypothesis was that the CBCT had a lower error in determining the resection margin compared to the other two modalities.
Results
Cohort characteristics
For our study we selected the first 50 patients who received breast surgery in our institution over a period of 6 months. After applying all exclusion criteria 21 patients could finally be included in the presented study (more details for patient inclusion are given in Fig. 1). Patients were between 24 and 69 years old. Specimens were withdrawn from the right (n = 7) and left (n = 14) breast.
Histopathological findings
Patients’ pre-operative proven histopathologic diagnoses included DCIS, invasive carcinoma, a combination of both and fibrocystic lesions and are provided in Table 1. Microcalcifications were detected in 15 resected specimens according to mammography as the reference standard.
Measures of diagnostic performance
The pooled sensitivity for DM, DBT and CBCT for detection of microcalcifications in specimens in comparison to the preoperative DM (Table 2) were 0.98 (CI 0.94–0.99), 0.83 (CI 0.73–0.94) and 0.94 (CI 0.87–0.99). The pooled specificity for DM, DBT and CBCT for microcalcifications (Table 2) were 0.99 (CI 0.99–0.99), 0.73 (CI 0.51–0.96) and 0.60 (CI 0.35–0.85).
The mean maximal size of microcalcifications measured in the DM of the specimens was 0.37 mm (SD 0.08). Two specimens showed single clusters of microcalcifications that measured up to 0.79 and 0.86 mm.
The mean measurement errors for the smallest distance of delineated microcalcifications to the resection margin for DM, DBT and CBCT (Table 2) were 10 mm, 14 mm and 6 mm (p = 0.002). There was a significant difference between CBCT and the other devices (CBCT vs. DM with p = 0.03 and CBCT vs. DBT with p = 0.006). No significant difference was observed between DM and DBT (p = 0.81). Imaging examples of two representative specimens are provided in Figs. 2 and 3. Both show clear differences in the projection of microcalcifications in relation to the resection margins between the three modalities.
The mean absorbed dose for DM, DBT and CBCT was 109 µSv, 132 µSv and 2167 µSv (SD 5, 18, 112).
Inter-rater reliability, reading time and confidence
The inter-rater reliability for resection margin measurements was 0.88 (CI 0.73–0.95), 0.93 (CI 0.82–0.98) and 0.78 (CI 0.57–0.92) for DM, DBT and CBCT.
The mean reading time for DM, DBT and CBCT was 36, 43 and 54 s. There was a significant difference in reading time between CBCT and DM (p < 0.001), also between CBCT and DBT (p = 0.03). However, no significant difference in reading time was detected between DM and DBT (p = 0.12).
The mean confidence in detection of microcalcifications (Likert scale 1–4 with 1 being very safe and 4 being very uncertain) was 1.1, 1.7 and 1.5 for DM, DBT and CBCT (p < 0.001). There was a significant difference between DM and DBT (p = 0.006), but no significant difference between the others.
The mean confidence for measurement of the smallest distance of delineated microcalcifications to the resection margin (Likert scale 1–4 with 1 being very safe and 4 being very uncertain) was 2.0, 2.0 and 1.8 for DM, DBT and CBCT (p = 0.15) without any significant differences between the modalities.
Discussion
In our study, after breast surgery DM showed the highest sensitivity and specificity for the detection of microcalcifications in specimens in comparison to the reference standard of preoperative DM. One reason to explain this advantage of DM over DBT with the lowest sensitivity and over CBCT with the lowest specificity might be explained due to the better spatial resolution of DM especially in comparison to CBCT. Whereas CBCT data of the specimen were reconstructed and analyzed with an isometric spatial resolution of 100 µm, the pixel size in mammography and tomosynthesis was 85 µm and therefore only slightly less than the resolution of CBCT. Although spatial resolution is generally lower at DBT, some studies indicate, that the detection and characterization of microcalcifications and associated carcinomas do not lead to significant differences between DBT and DM18,19,20,21,22. Previous studies showed reduced detectability of microcalcifications of dedicated breast CT systems in comparison to DM27, which could be improved over time with improvement of methods, so that most of even small microcalcifications are reproducible by CBCT systems28,29.
The possible appearance of several artifacts in DBT like blurring-ripple artifacts with shadowing and zipper artifacts or bright-edge artifacts12 is known. In CBCT imaging beam hardening artifacts, exponential edge-gradient effects, aliasing or ring artifacts can appear30. With both methods some artifacts might reduce the visibility of microcalcifications, whilst others might even imitate the appearance of microcalcifications. We also noted artifacts with our CBCT system. In addition to ring artifacts (Fig. 4) or metal-related artifacts around the marking wire (Figs. 2C,D and 3B,C), they especially appeared in the middle of specimens. These central axis artifacts in CBCT images were represented as central hyperdense spots in some slices, that might have been misinterpreted as microcalcifications. But as they were always accompanied by a central hypodense spot in adjacent slices, this typical representation and the central location made it easy to be identified as an artifact (Fig. 4). Metal-related artifacts, for example due to beam hardening, were seen around the wires in CBCT and less in tomosynthesis and made it difficult to evaluate the adjacent tissue (Figs. 2C,D and 3B,C). This might impair the evaluation of directly adjacent microcalcifications. But as none of the mentioned artifacts appeared in DM, which is performed in two projections after wire marking, a location of microcalcifications close to wires can be preoperatively identified. This information might be taken into account, when assessing the resection margin of a specimen with CBCT and in case of unchanged wire location the according distance might be evaluated.
In addition, the visualization of microcalcifications in CBCT can be affected by scan parameters such as applied dose, tube voltage, detector quality or size of investigated specimens. Whereas the detectability of microcalcifications is significantly increased following a higher dose, it is decreased with more volume of the investigated object31. Therefore, the visibility of microcalcifications might be improved with carefully adapted scan protocols, reduced artifacts, improved detector and CT technology and better post imaging processing and reconstructions including metal artifact reduction.
Rößler et al. demonstrated that another advanced dedicated breast CT with photon-counting technology even showed a comparable or superior performance for the detection of microcalcifications and lesions in specimens compared to DM and DBT32.
Only recently, Wetzl et al.33 confirmed that the sensitivity and specificity for the detection of microcalcifications in breast specimens is at least equal or better for photon-counting breast CT compared to DBT with DM as standard reference. Therefore, with a dedicated breast CT system the visibility of microcalcifications in specimens might be further improved.
In addition, we found that CBCT was the most accurate method with the least measurement errors for determining the distance of microcalcifications, if present, to the resection margin with significant better results compared to DM and DBT. In our clinical practice a specimen is investigated with DM in two projections and the minimal detectable resection margin is measured on those images. In our study we equally investigated the specimens with DM in two projections. Still, the specimen was regularly simply placed only in the two main orientations (craniocaudal and mediolateral) directly on the investigating table. Accordingly, the distribution of the frequently irregular specimen on the investigating table follows the principles of gravitas with the possibility of unwanted tilts. Therefore, the specimen is object of not always influenceable and not anatomical correct superimpositions which altogether influence the result of the measurements in DM imaging. For DBT the specimen was equally placed on the investigation table, i.e. DBT was performed directly after the first DM projection with the same device and without repositioning of the specimen with 25 projections over a 50° angle. In contrast, in CBCT the specimen was carefully placed in a custom made and adaptive storage aid for investigation (compare Supplementary Fig. S1). Therefore, only with CBCT true volumetric positioning of the specimen could be performed. In addition, CBCT allows high contrast imaging with no tissue overlap and multiplanar reconstructions. This might explain the better distance measurements with CBCT than with DM and even with DBT in comparison to histopathological measurements as gold standard. It must however be conceded that histopathology itself is susceptible to a certain degree of error regarding the detection and analysis of microcalcifications in specimens. Microcalcifications might not be detected due to their small size or their location. However, in the case that microcalcifications known by mammography are not detected histopathologically, further radiographic examinations of the histology sections and the specimen and appropriate histopathological analyses were performed. Still, this might potentially be a source of error and further analysis in studies with more samples should be conducted.
The exposure time for all three methods varied in the range of seconds with 0.3 s for DM and up to 10 s for CBCT. For the investigation of specimens regarding microcalcifications and the measurement of the resection margin to microcalcifications there was no significant difference in reading time between DM and DBT in our study. Only CBCT imaging resulted in a significant longer reading time in comparison to DM or DBT. This is most likely owed to the greater amount of imaging data in CBCT due to its 3D acquisition and presentation of the specimens. Nevertheless, the difference is in the range of seconds and therefore does not relevantly delay the operation time until the result can be communicated to the breast surgeon. Therefore, the difference in examination and reading time might be tolerable considering the benefits of a more exact resection margin assessment and radiological report for the surgeon following CBCT.
Still, a breast CT system or another suitable CBCT as in our case has to be available and it should be located within easy reach of the operating theatre and the pathology department for immediate imaging during surgery, as the processing and transport of the specimen should not be significantly delayed due to a change in imaging modality. For even faster direct intraoperative imaging a dedicated specimen CT might be a conceivable future development. In contrast to other studies that proved dedicated breast CBCT dose to be similar to or even less than DM dose24, in our study, dose measurements showed a much higher dose for CBCT than for DM in two projections or DBT. As we only investigated specimens and as our CT system was not used for in vivo breast imaging this has no further relevance.
Regarding the latest promising studies with further improved breast CT imaging technology for specimens32,33 as well as for breast imaging, which might even be additionally improved by contrast media application25, more focus should be laid on breast CT. Especially in preoperative imaging dedicated breast CT might be a promising alternative to the more expensive breast MRI. In imaging of specimens, CBCT might soon become a realistic alternative for DM, especially as a cancer free margin of specimens is extremely important for later outcome of patients10. One step further, in combination with preoperative breast CT even more advantages might be expected by specimen CBCT. One may even speculate that the assessment of non-calcified tumor parts could possibly be easier in the postoperative CBCT examinations of the specimens, if they are compared with preoperative contrast-enhanced breast CT examinations. It can be assumed that this would improve the ability to assess resection margins in postoperative CBCT and help with the guidance of surgical decisions.
In addition, several other methods like micro-CT, specimen MRI, high frequency ultrasound, even PET or electromagnetic imaging and optical depiction are theoretically available for specimen analysis regarding the cancer detection and correct depiction of resection margins34. Certainly, a volumetric imaging with high contrast and spatial resolution and better reconstruction modes would be of advantage, but there is still room for quality improvement and more studies should underline these alternative ways.
This study is limited by its monocentric design. Due to the exclusion of some patients for several reasons, a selection bias might have occurred, although in our view this should not have had a relevant influence on the results of this study.
In conclusion, in this study we were able to show that measurement of the resection margin to microcalcifications was significantly more accurate with CBCT compared to DM and DBT, most likely due to better 3D-placement and analysis of specimens. A combination of methods or improving CT technology might be more promising for improved determination of disease-free margins of breast specimens with microcalcifications than DM alone. However, further studies have to be performed, especially regarding additional criteria for margin assessment of specimens like soft tissue changes associated with breast cancer.
Methods
Study population
In our study we enrolled consecutive patients who received breast surgery (ablation as well as breast conserving surgery) in our institution and who presented preoperatively to the Radiology Department over a period of six months. Subjects were excluded from participation if they were minor, declined participation, required frozen sectioning, were enrolled in different studies, had incomplete preoperative imaging or in case of interference with clinical workflow.
The Ethics Committee of the University of Freiburg approved this prospective study (No. 507/12), which was in accordance with the Helsinki Declaration and all subjects provided written informed consent.
Radiologic examination of specimens
After breast surgery, resected specimens were immediately and without any further procession examined with DM, digital breast tomosynthesis, and cone-beam computed tomography. DM was performed with 24 kV and 9 mAs (Mammomat Inspiration, Siemens Healthineers, Erlangen, Germany) and included two projections after positioning the specimen with the help of suture markings in craniocaudal and mediolateral orientation. DBT was performed with 25 projections over a scan angle of 50 degrees with 24 kV and 71mAs in the same device directly after DM in craniocaudal orientation, therefore no repositioning of the specimen was necessary for DBT examination (Mammomat Inspiration, Siemens Healthineers, Erlangen, Germany). CBCT was performed with 96 kV and 36 mAs (Verity, Planmed Oy, Helsinki, Finland, originally designed for musculoskeletal imaging). Our CBCT scanner features an un-binned pixel-size of 127 µm at the detector with a magnification of 1.5, resulting in a geometrical resolution of 85 µm. Un-binned CBCT data of the specimen were acquired and reconstructed on an 100 µm isometric grid due to the given reconstruction algorithm. The pixel size of DM and DBT was 85 µm.
The exposure time for mammography was 0.3 s, for tomosynthesis 2.4 s and for CBCT 10 s.
Dose measurement was performed for DM, DBT and CBCT using thermoluminescent dosimeter (TLD), that were centrally placed in an imaging phantom of breast specimen like amorphous tissue consisting of 100 g Gallus gallus domesticus fatty and muscular tissue. Subsequently TLDs were analyzed with a TLD reader and the absorbed dose was calculated.
Positioning in the CBCT was done with a custom made styrofoam storage aid. Oriented to the suture markings, this allowed an axially correct examination of the specimens.
Directly after the radiologic investigation the specimens were stored in formaldehyde and analyzed at the pathological institute.
After screening the first 50 patients we performed an interim analysis. In this analysis the primary endpoint of the study was met, and the trial was closed.
Radiologic analysis of preoperative mammography and specimens
The result of the histopathological examination served as the reference standard for resection margin. A consensus reading of mammography by two radiologists with over 25 and 5 years of experience in breast imaging served as the reference standard for the presence of microcalcifications.
The images of the examinations were pseudonymized and submitted to three blinded raters in a randomized fashion. All raters were breast radiologists with an experience in breast imaging of 7, 8 and 26 years. The raters assessed the examinations of the different imaging modalities at least 4 weeks apart to avoid recall bias. In each evaluation cycle, the studies were randomized in a different order. Image evaluation was performed on approved diagnostic monitors under constant light conditions. The assessment criteria included the presence of microcalcifications and their diagnostic reliably (judged by Likert scale 1–4 with 1 being very safe and 4 being very uncertain). Only if microcalcifications could be delineated, the smallest distance of those microcalcifications to the resection margin was measured (in mm, compare Fig. 2) and the accessibility of this measurement was evaluated (Likert scale 1–4 with 1 being very good accessibility and 4 being very poor accessibility). The reading time was recorded using a stopwatch.
In addition, one radiologist measured the maximal size of microcalcifications in the mammography of all specimens.
Standard pathological assessment
Gross section was conducted according to standardized routine gross section protocols. After formalin fixation (minimum 24 h) all tissue specimens were macroscopically examined and documented. Resection margins (medial, lateral, cranial, caudal, ventral and dorsal) were ink-dyed. Next, sections of 0.4 cm thickness were cut and observed for macroscopic conspicuities by an experienced pathologist. Tissue specimens with a maximal diameter of 3.0 cm or less were completely embedded. From tissue specimens with a diameter above 3.0 cm macroscopic conspicuous areas as well as clinical labeled areas were embedded with reference to the nearest resection margins. Of note, macroscopic conspicuous areas were individually described, measured in three dimensions and documented in relation to all nearest resection margins.
After formalin fixation and paraffin embedding, tissue slices of 3 μm thickness were automatically Hematoxilin and Eosin stained using the Dako Cover Stainer.
All slides were histologically examined by an experienced pathologists for histological abnormalities. Histopathological reports included histological findings (e.g. usual ductal hyperplasia, fibrocystic change and columnar cell changes). Microcalcifications, defined as deposits below 0.5 mm, were identified using HE staining and documented as far as histologically detectable. If, despite clinical "indication of microcalcification", no microcalcifications were histologically detectable, supplementary sectional stages were prepared. If the result was not sufficient, an X-ray of all blocks was requested and corresponding blocks were completely serially sectioned and HE stained.
In case of a malignant tumor, histological reporting comprised WHO classification, UICC stage, tumor grading according to Silverstein (DCIS) and Elston & Ellis (tumor grading), the presence or absence of lymphatic, vascular and perineural invasion and the histological assessment of the medial, lateral, cranial, caudal, ventral and dorsal resection margin in cm using an ocular micrometer.
In the case of a differential diagnosis of UDH/DCIS/invasive carcinoma, additional immunohistochemical tests were performed for Cytokeratin 5/6, p63, oestrogen receptor protein, progesterone receptor protein, HER2/neu and MIB1.
Statistical analysis
Statistical analysis and presentation of the collected data were performed using the statistical program R version 4.0.2 (The R Foundation for Statistical Computing).
Descriptive data were presented using absolute frequency, mean value (MV) and confidence intervals (CI).
Inter-rater reliability for the detection of microcalcifications were calculated with Fleiss kappa. Inter-rater reliability for the measurements of resection margin were calculated with intraclass correlation (twoway, agreement). The error in the measurement of the resection margin was calculated by subtracting the evaluators' measurements from the reference standard. The error of measurement, the reading time, the confidence for detection of microcalcifications and the confidence for measurement of microcalcifications were compared between the three modalities with Friedman rank sum test and post hoc pairwise comparison using Nemenyi multiple comparison test.
Sensitivity and specificity were calculated for the detection of microcalcifications for each modality separately pooled for all raters. P-values were corrected with Bonferroni adjustment. A p-value < 0.05 was considered significant.
Data availability
The datasets generated and analyzed during the current study are available from the corresponding author on reasonable request.
References
Bent, C. K., Bassett, L. W., D’Orsi, C. J. & Sayre, J. W. The positive predictive value of bi-rads microcalcification descriptors and final assessment categories. Am. J. Roentgenol. 194, 1378–1383 (2010).
Stomper, P. C., Geradts, J., Edge, S. B. & Levine, E. G. Mammographic predictors of the presence and size of invasive carcinomas associated with malignant microcalcification lesions without a mass. Am. J. Roentgenol. 181, 1679–1684 (2003).
Tot, T., Gere, M., Hofmeyer, S., Bauer, A. & Pellas, U. The clinical value of detecting microcalcifications on a mammogram. Semin. Cancer Biol. 14, 1044–1579 (2019).
Weigel, S. et al. Calcifications in digital mammographic screening: improvement of early detection of invasive breast cancers?. Radiology 255, 738–745 (2010).
Duffy, S. W. et al. Mammography screening reduces rates of advanced and fatal breast cancers: Results in 549,091 women. Cancer 126, 2971–2979 (2020).
Kim, S. Y. et al. Evaluation of malignancy risk stratification of microcalcifications detected on mammography: A study based on the 5th edition of BI-RADS. Ann. Surg. Oncol. 22, 2895–2901 (2015).
Thomas, J. et al. Radiological and pathological size estimations of pure ductal carcinoma in situ of the breast, specimen handling and the influence on the success of breast conservation surgery: A review of 2564 cases from the Sloane Project. Br. J. Cancer. 102, 285–293 (2010).
Lowry, K. P. et al. Screening performance of digital breast tomosynthesis vs. digital mammography in community practice by patient age, screening round, and breast density. JAMA Netw. Open 3, e2011792. https://doi.org/10.1001/jamanetworkopen.2020.11792 (2020).
Veronesi, U. et al. Twenty-year follow-up of a randomized study comparing breast-conserving surgery with radical mastectomy for early breast cancer. N. Engl. J. Med. 347, 1227–1232 (2002).
Houssami, N., Macaskill, P., Marinovich, M. L. & Morrow, M. The association of surgical margins and local recurrence in women with early-stage invasive breast cancer treated with breast-conserving therapy: a meta-analysis. Ann. Surg. Oncol. 21, 717–730 (2014).
Chong, A., Weinstein, S. P., McDonald, E. S. & Conant, E. F. Digital breast tomosynthesis: Concepts and clinical practice. Radiology 292, 1–14 (2019).
Tirada, N. et al. Digital breast tomosynthesis: Physics, artifacts, and quality control considerations. Radiographics 39, 413–426 (2019).
Förnvik, D. et al. Breast tomosynthesis: Accuracy of tumor measurement compared with digital mammography and ultrasonography. Acta Radiol. 51, 240–247 (2010).
Ciatto, S. et al. Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): A prospective com-parison study. Lancet Oncol. 14, 583–589 (2013).
Gilbert, F. J., Tucker, L. & Young, K. C. Digital breast tomosynthesis (DBT): A review of the evidence for use as a screening tool. Clin. Radiol. 71, 141–150 (2016).
Heywang-Köbrunner, S. H., Jänsch, A., Hacker, A., Weinand, S. & Vogelmann, T. Digital breast tomosynthesis (DBT) plus synthesised two-dimensional mammography (s2D) in breast cancer screening is associated with higher cancer detection and lower recalls compared to digital mammography (DM) alone: Results of a systematic review and meta-analysis. Eur. Radiol. https://doi.org/10.1007/s00330-021-08308-8 (2021).
Canelo-Aybar, C. et al. Digital breast tomosynthesis compared to diagnostic mammographic projections (including magnification) among women recalled at screening mammography: A systematic review for the european commission initiative on breast cancer (ECIBC). Cancer Med. 10, 2191–2204 (2021).
Spangler, M. L. et al. Detection and classification of calcifications on digital breast tomosynthesis and 2D digital mammography: A comparison. AJR Am. J. Roentgenol. 196, 320–324 (2011).
Gilbert, F. J. et al. Accuracy of digital breast tomosynthesis for depicting breast cancer subgroups in a UK retrospective reading study (TOMMY trial). Radiology 277, 697–706 (2015).
Clauser, P. et al. Diagnostic performance of digital breast tomosynthesis with a wide scan angle compared to full-field digital mammography for the detection and characterization of microcalcifications. Eur. J. Radiol. 85, 2161–2168 (2016).
Choi, J. S. et al. Comparison of synthetic and digital mammography with digital breast tomosynthesis or alone for the detection and classification of microcalcifications. Eur. Radiol. 29, 319–329 (2019).
Kilic, P. et al. Comparison of diagnostic performances in the evaluation of breast microcalcifications: Synthetic mammography versus full-field digital mammography. Ir. J. Med. Sci. https://doi.org/10.1007/s11845-021-02744-7 (2021).
Lindfors, K. K., Boone, J. M., Newell, M. S. & D’Orsi, C. J. Dedicated breast computed tomography: The optimal cross-sectional imaging solution?. Radiol. Clin. North Am. 48, 1043–1054 (2010).
Wienbeck, S., Lotz, J. & Fischer, U. Review of clinical studies and first clinical experiences with a commercially available cone-beam breast CT in Europe. Clin. Imaging. 42, 50–59 (2017).
O’Connell, A. M., Marini, T. J. & Kawakyu-O’Connor, D. T. Cone-beam breast computed tomography: Time for a new paradigm in breast imaging. J. Clin. Med. 10, 5135. https://doi.org/10.3390/jcm10215135 (2021).
O’Connell, A. M. & Kawakyu-O’Connor, D. Dedicated cone-beam breast computed tomography and diagnostic mammography: Comparison of radiation dose, patient comfort, and qualitative review of imaging findings in BI-RADS 4 and 5 lesions. J. Clin. Imaging Sci. 2, 7. https://doi.org/10.4103/2156-7514.93274 (2012).
Lindfors, K. K. et al. Dedicated breast CT: Initial clinical experience. Radiology 246, 725–733 (2008).
O’Connell, A. et al. Cone-beam CT for breast imaging: Radiation dose, breast coverage, and image quality. AJR Am. J. Roentgenol. 195, 496–509 (2010).
He, N. et al. The utility of breast cone-beam computed tomography, ultrasound, and digital mammography for detecting malignant breast tumors: a prospective study with 212 patients. Eur. J. Radiol. 85, 392–403 (2016).
Nagarajappa, A. K., Dwivedi, N. & Tiwari, R. Artifacts: The downturn of CBCT image. J. Int. Soc. Prev. Commun. Dent. 5, 440–445 (2015).
Lai, C. J. et al. Visibility of microcalcification in cone beam breast CT: Effects of X-ray tube voltage and radiation dose. Med. Phys. 34, 2995–3004 (2007).
Rößler, A. C. et al. Performance of photon-counting breast computed tomography, digital mammography, and digital breast tomosynthesis in evaluating breast specimens. Acad. Radiol. 24, 184–190 (2017).
Wetzl, M. et al. Detection of microcalcifications in spiral breast computed tomography with photon-counting detector is feasible: A specimen study. Diagn. (Basel) 11, 848. https://doi.org/10.3390/diagnostics11050848 (2021).
Maloney, B. W. et al. Review of methods for intraoperative margin detection for breast conserving surgery. J. Biomed. Opt. 23, 1–19 (2018).
Acknowledgements
This study was supported by a fellow seeding Grant CCC/Deutsche Krebshilfe 2015 to J.N..
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Contributions
J.N. conceived the idea. J.N. and T.E. were responsible for patient selection, T.E. for surgery and providing specimens. P.B. was responsible for the histopathological analysis. J.S.Y. prepared the datasets. M.P. and J.N. performed and M.P. analyzed the dose measurements. J.N., C.N. and M.W.B. analyzed the data. The statistical analysis was mainly performed by J.N. C.N. and J.N. wrote the paper. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Neubauer, C., Yilmaz, J.S., Bronsert, P. et al. Accuracy of cone-beam computed tomography, digital mammography and digital breast tomosynthesis for microcalcifications and margins to microcalcifications in breast specimens. Sci Rep 12, 17639 (2022). https://doi.org/10.1038/s41598-022-21616-3
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-022-21616-3
- Springer Nature Limited