Abstract
Increased left ventricular mass (LVM) is a strong independent predictor for adverse cardiovascular events, but conventional echocardiographic methods are limited by poor reproducibility and accuracy. We developed a novel method based on adding the mean wall thickness from the parasternal short axis view, to the left ventricular end-diastolic volume acquired using the biplane model of discs. The participants (n = 85) had various left ventricular geometries and were assessed using echocardiography followed immediately by cardiac magnetic resonance, as reference. We compared our novel two-dimensional (2D) method to various conventional one-dimensional (1D) and other 2D methods as well as the three-dimensional (3D) method. Our novel method had better reproducibility in intra-examiner [coefficients of variation (CV) 9% vs. 11–14%] and inter-examiner analysis (CV 9% vs. 10–20%). Accuracy was similar to the 3D method (mean difference ± 95% limits of agreement, CV): Novel: 2 ± 50 g, 15% vs. 3D: 2 ± 51 g, 16%; and better than the “linear” 1D method by Devereux (7 ± 76 g, 23%). Our novel method is simple, has considerable better reproducibility and accuracy than conventional “linear” 1D methods, and similar accuracy as the 3D-method. As the biplane model forms part of the standard echocardiographic protocol, it does not require specific training and provides a supplement to the modern echocardiographic report.
Similar content being viewed by others
Introduction
Increased left ventricular mass (LVM) is a strong independent predictor for adverse cardiovascular events1,2,3, and associated with impaired left ventricular (LV) myocardial function, coronary artery disease and arrhythmogenesis4. Unfavourable associations with increased LVM seem reversible through reduction of LVM5, but clinical responses to treatment and prognosis using echocardiography requires reliable LVM-quantification. As conventional methods for LVM-quantification lack reproducibility they are not suitable for serial comparisons, thereby not routinely deployed on individuals. This warrants a method with greater reproducibility to improve accuracy in detecting actual differences. Standard one-dimensional linear echocardiography (1DE) for LVM-quantification by the cube formula relies on a symmetrical shaped left ventricle (LV). Whilst technically simple, it is prone to inaccuracies6 and more suitable for comparison on a population level. Conversely, three-dimensional echocardiography (3DE) is independent of LV symmetry and has higher concordance with the reference method cardiac magnetic resonance (CMR)7,8. Since acquisition and analysis using 3DE are challenging and time-consuming, this presents disadvantages in busy echocardiography labs. We explored an alternative method to preserve the geometrical shape of the LV by applying the biplane model of discs by two-dimensional echocardiography (2DE) for LVM-quantification, without the need for troublesome epicardial boundary delineation in the apical views.
Our aims were to:
-
(1)
Develop a simpler, feasible and reproducible 2DE-based method for LVM-quantification that is less dependent on LV symmetry.
-
(2)
Compare various well-known echocardiographic methods for LVM-quantification as well as our novel method to CMR, amongst subjects with assorted LV geometries.
Methods
Study design
This is a single-centre prospective cohort feasibility study. We included patients scheduled for echocardiography > 18 years with sinus rhythm on the study days. Pregnant, breastfeeding, or claustrophobic patients were excluded. We aimed to include a wide variety of subjects with different LV geometries. All subjects were assessed at “baseline” using echocardiography immediately followed by CMR and re-assessed at “re-examination” using echocardiography after a median of 6 days (IQR 3–18). To limit effects of hydration status, all subjects were instructed to abstain from oral intake ≥ 5 h prior to both visits. The study was conducted in accordance with the second Helsinki declaration and approved by the Clinical Research Unit of the Heart Centre at the Rigshospitalet University Hospital of Copenhagen and by the local ethics committee of the Capital Region of Denmark, Protocol number H-16029778. All participants provided written informed consent.
Echocardiography: acquisition and analysis
One experienced sonographer performed all examinations at both baseline and re-examination, using a Vivid E95 ultrasound scanner (GE Healthcare, Norway), and M55c-D-matrix-array transducer (1.5–4.6 MHz) for 2DE and a 4 V-D volume-phased array transducer (1.5–4 MHz) for 3DE. Subjects were studied in the left lateral decubitus position with parasternal long-axis view (PLAX), short-axis view (SAX), apical four-chamber view (4CH), apical two-chamber view (2CH), and 3DE. We reduced depth to focus on the LV. Framerate for 2DE was 65 ± 7 frames/s, and for 3DE 26 ± 8 volumes/s. The 3DE full volume dataset was acquired from the apical window during breath-hold over four to six heart beats. The examinations were analysed using EchoPAC version 201.61 (GE Healthcare, Norway). End-diastole was defined as the first frame of mitral valve closure. We distinguished between the end-diastolic-volume (EDV) defined by the inner myocardial interface [endocardium (EDVENDO)] and by the outer myocardial interface [epicardium (EDVEPI)]. Conventional EDV (EDVENDO) was quantified by 3DE, 2DE using the biplane model9 and 1DE using the Teichholtz model10. LVM-quantifications were made at end-diastole. ECG-timing from PLAX was referenced to find the corresponding SAX-frame. All PLAX/SAX-measurements were made at the chordae level (Fig. 1). In PLAX the region between the mitral valve and papillary muscle, just beneath the attachment of the chordae to the papillary muscle. In SAX this corresponded to the visible attachments of the chordae to the papillary muscle. In this view, the mitral valve should not be visible, and chordae should be separated from LV wall. We delineated the boundaries in SAX by using the blood-endocardium interface, the inner boundary delineation and by the epicardium-blood/pericardium interface, outer boundary delineation. We didn’t use a leading-to-leading edge approach. Trabeculae or papillary muscles were considered part of the LV cavity, the pericardium was excluded from the delineation. 3DE LVM was quantified by the vendor-specific software package 4D Auto LVQ (EchoPAC, GE Healthcare, Norway). The full volume dataset was aligned for three apical views, which were manually adjusted, guided by the derived short-axis views.
The novel method
The novel method of LVM-quantification is based on adding the mean wall thickness (t) from a single SAX-recording to the EDVENDO acquired by the standard biplane model of discs in the apical 4CH- and 2CH-view. The t is calculated from SAX by conversion from the traced myocardial area, in the same manner as for the conventional 2DE methods (Step 1, Fig. 2). The EDVENDO is the sum of the sub-volumes of 30 unique discs, which are acquired during the conventional biplane delineation in the apical views. (Step 2, Fig. 2) A factor (k) adjusts the echocardiographic EDVENDO with CMR derived EDVENDO. EDVEPI is the summation of 30 larger sub-volumes quantified by adding t to each unique sub-volume from the EDVENDO-delineation. An apical cap is added for EDVEPI. (Step 3, Fig. 2) The difference between the quantified EDVEPI and the traced EDVENDO was multiplied with the myocardial gravity of 1.05 g/ml. (Step 4, Fig. 2). EDVENDO-delineations from each apical view including information regarding size of each unique disc from was exported from EchoPac and quantification of LVM was performed in Windows Excel 2010 (Microsoft Office Professional Plus). A more detailed description of the novel method, including formula for software implementation is available in the supplementary data.
Left ventricular mass quantification using echocardiography
We evaluated six different methods for LVM-quantification (Fig. 3)11,12,13,14. Four of these are widely recognized; A, D, E, F. All except B the novel method and C endo- and epicardial delineation in the 4CH- and 2CH-view by the biplane model of discs (BP), are recommended in current guidelines6.
Cardiac magnetic resonance: acquisition and analysis
Accuracy of the echocardiographic methods was defined by agreement according to CMR15,16,17. CMR images were obtained using a 1.5 Tesla system (GE Optima MR450W, GE Healthcare, Waukesha, WI) with a phased-array cardiac coil. Cine images were acquired during breath-hold using a steady-state free precession cine sequence with retrospective gating18. Slice thickness 8 mm, no gaps, field of view 300–360 mm, 25 phases/cycle. Analysis was performed using CVI42 (Circle Cardiovascular Imaging Inc., Version 5.6.5, Canada). End-diastole was defined as for echocardiography. Endocardium and epicardium were manually delineated in the short-axis-stack, papillary muscles were considered part of the LV volume. The subjects were classified in four groups according to age, gender and indexed values19 of the EDVENDO and LVM20; normal, dilatation, hypertrophy, dilatation and hypertrophy.
Reproducibility
Intra/inter-examiner examination were compared at baseline and test–retest variation between baseline and re-examination. All subjects were asked to walk around between baseline examinations and intra/inter-examiner exam. Intra-examiner exams (n = 13) were acquired and analysed by the same examiner who performed the baseline and re-examination exams, inter-examiner exams (n = 20) were acquired and analysed by another examiner.
Feasibility
Feasibility was estimated for the entire study cohort and for a small ‘all-comers’ cohort of twenty-six consecutive patients examined by a third sonographer during one week in our echo lab, no patients were excluded. Since 3DE is not routinely performed on all patients at our echo lab, we are unable to report reliable “all-comers” 3DE feasibility.
Classification of hypertrophy versus non-hypertrophy
Hypertrophy was defined by CMR according to age, gender and LVM-index20. The normal LVM-ranges for CMR were applied for the echocardiographic 3DE, novel method and BP-method20. Current echocardiographic guidelines were applied for normal LVM-ranges for the cube formula by Devereux (DEV), truncated ellipsoid (TE) and area-length (A-L)6. Sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) for detection of hypertrophy was evaluated.
Image quality
Image quality was graded as optimal, “analysis without effort” or suboptimal, “analysis with effort” or inadequate, “adequate analysis not possible”. We evaluated the image quality for each view separately; PLAX, SAX, apical views, 3D. For each method we evaluated the impact of image quality on the agreement to CMR and the test–retest variability.
Statistics
Continuous variables were expressed as mean and standard deviation (SD) and categorical values expressed as frequencies and percentage. The accuracy of echocardiographic methods was defined according to agreement with CMR; evaluated by the Bland–Altman-method (BA) using paired t-test presented as mean difference (bias), 95% limits of agreement (LOA)21, simple linear regression, Pearson’s correlation and intra class correlation coefficient (ICC). Reproducibility was assessed by the 95%LOA and by the coefficient of variation (CV) presented as percentages. P-values < 0.05 were considered statistically significant. Data analysis was performed in SPSS v25.0 (IBM Corp. IBM SPSS Statistics for Windows, Version 25.0. Armonk, NY).
Results
Study population
All 85 subjects had echocardiography and CMR at baseline. All were re-invited for re-examination; four subjects cancelled in advance, one subject was sent home because of technical problems, one subject did not show up for re-examination. Baseline characteristics of the population are presented in Table 1. Cardiac condition according to LV geometry in supplementary data, Table S5. Baseline EDVENDO and LVM by various methods in Table 2. We included all data for every methodology, although some subjects did not have feasible images for all methods. Data on the subjects with 100% feasibility (n = 59) are specified in supplementary data Table S7.
Feasibility and reproducibility
We report high feasibility for all methods, except the BP-method (74% vs. 95–100%) (Table 3). All-comers’ feasibility was lower; DEV 92%, TE/A-L/Novel 81%, BP 50%. We observe similar test–retest-variations of the 2D/3D-methods (14–18%) and larger test–retest-variation of the 1DE-method DEV (21%) (Table 3). The novel method has better reproducibility in intra- (CV 9% vs. 11–14%) and inter-examiner (CV 9% vs. 10–20%) analysis (Table 3).
Agreement of LVM quantification by echocardiography and CMR
Baseline agreements between echocardiography and CMR are visualized in BA-plots and linear regression-plots (Fig. 4). The novel method demonstrates equal distribution and limited proportional bias, based on the regression line (Fig. 4B). Table 4 presents the agreements between echocardiography and CMR at baseline. Figure 5 demonstrates the agreement of echocardiography and CMR among the defined LV geometries.
Impact of image quality
Table 5 presents the agreement between LVM from echocardiography compared to CMR and the test–retest variability with comparison between exams performed on different days. The novel method is equally affected by suboptimal image quality as the other methods.
Diagnostic performance on detecting hypertrophy using echocardiography
The sensitivity, specificity, PPV and NPV for all methods and divided by LV geometry are presented in Table 6.
Suggested normal reference interval for the novel method
Twenty-six subjects (31%) had no cardiac disease and no cardiovascular risk factors. The 95% CI for this subgroup was 46–96 g/m2; women 39–91 g/m2, men 63–94 g/m2.
Discussion
The main findings of our study are;
-
(1)
Our novel method has high feasibility and better intra/inter-examiner reproducibility than the other methods, especially the conventional “linear” 1DE-methods.
-
(2)
Accuracy of our novel method is similar to 3DE and greater across all four defined LV geometries than the other methods.
-
(3)
Our novel method is simple, does not require specific training and should not cause any considerable delay, since all images already form part of the standard echocardiographic protocol.
-
(4)
The formula can easily be integrated in any vendor specific echocardiographic analysis software for fast automatic quantification, and has the potential to provide a useful supplement to the modern echocardiographic report.
Requirements for a novel method
Any new method needs to be both reproducible and accurate compared to the reference method. For example; a method that always quantifies the LVM to 150 g is very reproducible but inaccurate and not able to detect differences. A method with solely high accuracy is not useful if reproducibility is poor when serial measurements are needed. Accuracy was evaluated according to agreement with CMR16,18. Reproducibility was evaluated according to test–retest variability and according to intra/inter-examiner variability. Feasibility was evaluated in the study population and among all-comers. The impact of image quality on accuracy and reproducibility was evaluated. Sensitivity and specificity for detecting hypertrophy was calculated. As envisaged, methods based on 2DE/3DE demonstrate better accuracy than 1DE. The novel method seems more accurate than other 1DE/2DE-methods, has high reproducibility, is equally affected by image quality compared to the other 2D/3D-methods recommended by guidelines6, demonstrates high sensitivity and specificity for hypertrophy (Table 6) and, moreover, performs best regardless of LV geometry (Fig. 5). BA plots (Fig. 4B) also reveal equal distribution and limited proportional bias, based on the regression line. An informative overview of the various methods are presented in Table 7.
The novel method versus conventional “linear” one-dimensional methods
It is unsurprising that the cube formula11 has been the most common method for LVM-quantification since the 1970’s, as this method is simple, feasible and useful in large population studies1,2,3. However, its simplicity makes it susceptible to measurement errors that make it less suitable for individual and serial measurements. For instance, recording a LVID of 45 mm with a small wall thickness measurement error of 11 mm instead of 10 mm yields 14% increase in LVM. Whereas wall thickness recordings using the novel method are derived from the whole circumferential area, and less sensitive to minor measurement errors. This vulnerability of methods deploying linear measurements to small differences that impact LVM measurement is reflected in the increased test–retest- and intra-/inter-examiner variations for DEV compared to the novel method (Table 3). High variations indicate decreased reproducibility and less ability to identify small yet significant real differences. Compared to the conventional method using DEV, our novel method presents lower variations and is thereby much more suited for monitoring serial measurements and comparing measurements by different examiners. Figure 6 demonstrates three patients with excellent image quality where methods deploying linear measurements fail to accurately quantify LVM: Example A has hypertrophic cardiomyopathy (HCM) and asymmetry, focal septal hypertrophy results in overestimation of t and consequently overestimation of LVM by 100 g. Whilst no echocardiographic method is ideal in focal/asymmetric HCM, 2DE/3DE correlate substantially better with CMR compared to 1DE. Example B has normal geometry with normal LVM and EDVENDO. However, the LV is short, only 78 mm, predisposing to overestimation of the LV length and consequently LVM by 44 g. Other methods overcome this pitfall and correlate better with CMR. Example C has severe aortic regurgitation, the LV is both dilated and hypertrophied. Small measurement errors are particularly magnified amongst patients with large LVs, resulting in both overestimation and large variations of 315–405 g despite very small, almost visually undetectable measured differences. This is also illustrated in Fig. 5 where this patient group “dilatation and hypertrophy” has large SDs, particularly amongst the method utilising linear measurements. Variations in LV geometry and size are common in cardiac disease, a cohort that particularly requires correct LVM-quantification, warranting exploration of improved methodologies.
The novel method versus other two-dimensional methods
Like A-L and TE, the novel method is based on adding the mean wall thickness from a single SAX-recording. What differs the novel method from these other two conventional 2D-methods is the biplane delineations of the endocardium in the apical 4CH- and 2CH-view, which preserves the LV geometric variations. Compared to the other 2D-methods, the novel method demonstrates similar feasibility of 95% in the study population (Table 3) and 81% among all-comers, slightly better reproducibility of 9% vs 10–11%, better agreement to CMR (Table 4) and overall better sensitivity, specificity, PPV and NPV (Table 6). Traditionally, the biplane model is applied for LV volumes and function and is better at correcting shape distortions compared to 1D-volume by Teichholtz or 2D-volume by A-L6. The biplane model has also been used for LVM-quantification; previous studies7,22,23,24,25,26,27 only report endo- and epicardial border delineation (Fig. 3C), not quantification (Fig. 3B). It is our experience that, the epicardium is more difficult to delineate than the endocardium. We envisaged advantages in measuring the myocardial thickness in another representative view and adding it to the EDVENDO to build up the EDVEPI. Compared to the BP-method, the novel method showed better feasibility, reproducibility and agreement to CMR. Several factors contribute to this observation: (1) reduced lateral resolution along the LV-walls impairs epicardial delineation in the apical views, (2) epicardial dropout, (3) echogenic pericardium, 4) small rotational errors causing the right ventricular wall to interfere with the epicardium of the inferior LV wall in the 2CH-view.
The novel method versus 3D echocardiography
The overall superiority of the novel method compared to 3DE is better reproducibility, better sensitivity and specificity for hypertrophy (Table 3) and better accuracy across all four defined LV geometries (Fig. 5). The novel methods require standard 2D images whereas it is our experience that 3D image acquisition is more troublesome. Improved 2D image quality with high spatial and temporal resolution may contribute to these findings. Furthermore, post-processing analysis of 3D images requires specific training and is more time-consuming; 3DE 142 s and Novel 79 s (Table S8, Supplementary data). On the contrary, the favourable characteristics of 3DE compared to the novel methods is the lack of geometrical assumption regarding LV shape and distribution of LV hypertrophy. Shape distortions beyond the 4CH/2CH-views are not accounted for in the novel method and, inherent to other 2DE-methods, inaccurate apical images and LV-foreshortening may underestimate the EDVENDO and LVM. The 3DE-agreement to CMR in our study is consistent with recent studies8, however, we observed underestimation of LVM by 30 g among subjects with hypertrophy (Fig. 5), mostly HCM. Only 62% of the patients with geometry profiles of “hypertrophy” were correctly classified as being hypertrophic by 3DE (Table S9 Supplementary data) compared to 92% for the novel method. This is similar to Chang et al.28, who also report underestimation of 20 g and similar LOA in HCM, probably secondary to interpolation of small segments of the epicardium in the apex. Because of their larger LVs, the group with both hypertrophy and dilatation should also be prone to potential errors caused by interpolation of the epicardium in the apex, but we didn’t observe the same pattern in this group. A plausible explanation for our findings may be slight overestimation of LVM by CMR in subjects with HCM and small/normal EDV. It may be hard to distinguish between trabeculae and LV cavity; delineation is easier among hypertrophied patients with increased EDV. Recognised difficulties in LVM-quantification amongst HCM are illustrated by relatively large LOAs’ when comparing 3DE to CMR29,30.
Estimation of mean wall thickness
We recommend acquiring t using 2DE delineations in SAX (Fig. 1, right panel), where it is easier to ensure centred/aligned measurements. Initially we compared both linear measurements in PLAX as 1DE and delineations in SAX as 2DE at three LV levels; mitral, chordae, mid-papillary (Fig. S1 Supplementary data). 2DE at chordae level performed slightly better (Table S3 Supplementary data). Current guidelines encourage 1DE-measurements at the mitral valve leaflet tip in PLAX and 2DE-delineations at the mid-papillary level in SAX6. However, results from Chetrit et al31, Guzzetti et al32 and our findings suggest that the measurement level corresponding to the mitral valve leaflet tip provides inaccurate quantification of LVM and that the preferred level is located more towards the mid-ventricular level. Similar to A-L, TE, and also DEV; the novel method is based on wall thickness estimation from a single imaging plane. Patients with distal wall thinning as in LV aneurysms, or basal septal hypertrophy are at risk of LVM-overestimation32, conversely patients with focal hypertrophy are at risk of LVM-underestimation. Measurements at several cross-sectional levels; at both base and apex may be considered with manifestly asymmetric geometry, although this may affect the feasibility and simplicity of the novel method. We did not include any patients with LV aneurysm. Because of deviating LV geometry in this patient group, the novel method and 3DE would hypothetically be more suited because of partial and full preservation of the LV shape. Furthermore, serial measurements with the novel method should not be particularly affected among patients with LV aneurysm, since the mass of fibrotic tissue does not have the ability to hypertrophy and remain constant over time.
Future aspects regarding implementation of the novel method
We aimed to improve and facilitate echocardiographic LVM-quantification by developing a method that is simple, reproducible, accurate and reliable for monitoring individuals using serial measurements, without impairing workflow. This novel method does not require specific training and has substantially less post-processing analysis time than 3DE (Table S8, Supplementary data). Time-efficacy may be even further improved by applying simultaneous bi-plane acquisition33, which most vendors provide already. Once integrated with the vendor specific echocardiographic analysis software, our novel method will provide an automated LVM, comparable to LVM by 3DE or CMR and with high reproducibility. Thus, it provides a useful supplement to the modern echocardiographic report.
Limitations
Our study population comprised mostly men (67%), young subjects (44 ± 14 years) with low BMI (25.5 ± 4.2 kg/m2), which may limit the applicability to other populations. We also recognize that developing and testing the model on the same population, may have biased the results and a validation cohort may have increased the strength of this study. Although all measurements were performed blinded, we were not blinded to the purpose of the study, which may have affected the results. We did not investigate potential vendor differences between 3DE and CMR. Further studies with various vendors or machines, contrast echocardiography, with/without contrast and with greater subject heterogeneity (including ages, obesity, LV shapes, hypertrophy and cardiac disorders) would provide corroboration, and widen interpretation and applicability of the findings. The novel method is not yet validated according to normal LVM-range or outcome. We recognise our report may pose challenges to interpreting established data that relied on less reproducible methodologies, although this is not unique to our observations/findings; future data is usually developed after adopting newer technologies and methods following a period of transition. For the subgroup without cardiac disease and without cardiovascular risk factors we found that LVM by the novel method ranged from 46 to 95 g/m2. We await validated normal LVM-ranges for both the 2DE/3DE-methods and hopefully, in time also for the novel method. We also recognize that there may be uncertainty regarding conventional geometrical classification according to relative wall thickness (RWT) and LVM index. Many clinical guidelines today are based upon linear measurements in PLAX. However, these measurements may also be ascertained by converting area to diameter using SAX-delineations. Theoretically, this may be a more accurate way of achieving RWT, since all segments of the LV are represented, not only the anterior–posterior segments.
Conclusion
Our novel method for LVM-quantification is simple, has considerable higher reproducibility and accuracy compared to conventional “linear” 1DE, and similar accuracy as 3DE. It can easily be integrated into any vendor specific echocardiographic software, and as the biplane model forms part of the standard echocardiographic protocol it does not require specific training and provides a supplement to the modern echocardiographic report.
Data availability
The data that supports the findings of the current study are available from the corresponding author upon reasonable request.
References
Levy, D., Garrison, R. J., Savage, D. D., Kannel, W. B. & Castelli, W. P. Prognostic implications of echocardiographically determined left ventricular mass in the Framingham heart study. N. Engl. J. Med. 322, 1561–1566 (1990).
Casale, P. N. et al. Value of echocardiographic measurement of left ventricular mass in predicting cardiovascular morbid events in hypertensive men. Ann. Intern. Med. 105, 173–178 (1986).
Barbieri, A. et al. Prognostic impact of left ventricular mass severity according to the classification proposed by the American Society of Echocardiography/European Association of Echocardiography. J. Am. Soc. Echocardiogr. 24, 1383–1391 (2011).
Vakili, B. A., Okin, P. M. & Devereux, R. B. Prognostic implications of left ventricular hypertrophy. Am. Heart J. 141, 334–341 (2001).
Koren, M. J., Ulin, R. J., Koren, A. T., Laragh, J. H. & Devereux, R. B. Left ventricular mass change during treatment and outcome in patients with essential hypertension. Am. J. Hypertens. 15, 1021–1028 (2002).
Lang, R. M. et al. Recommendations for cardiac chamber quantification by echocardiography in adults: An update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. Eur. Hear. J. Cardiovasc. Imaging 16, 233–271 (2015).
Takeuchi, M. et al. Measurement of left ventricular mass by real-time three-dimensional echocardiography: validation against magnetic resonance and comparison with two-dimensional and m-mode measurements. J. Am. Soc. Echocardiogr. 21, 1001–1005 (2008).
Mizukoshi, K. et al. Normal values of left ventricular mass index assessed by transthoracic three-dimensional echocardiography. J. Am. Soc. Echocardiogr. 29, 51–61 (2016).
Schiller, N. B. et al. Recommendations for quantitation of the left ventricle by two-dimensional echocardiography: American Society of Echocardiography Committee on Standards, Subcommittee on Quantitation of Two-Dimensional Echocardiograms. J. Am. Soc. Echocardiogr. 2, 358–367 (1989).
Teichholz, L. E., Kreulen, T., Herman, M. V. & Gorlin, R. Problems in echocardiographic volume determinations: Echocardiographic-angiographic correlations in the presence of absence of asynergy. Am. J. Cardiol. 37, 7–11 (1976).
Devereux, R. B. et al. Echocardiographic assessment of left ventricular hypertrophy: Comparison to necropsy findings. Am. J. Cardiol. 57, 450–458 (1986).
Geiser, E. A. & Bove, K. E. Calculation of left ventricular mass and relative wall thickness. Arch. Pathol. 97, 13–21 (1974).
Schiller, N. B. et al. Canine left ventricular mass estimation by two-dimensional echocardiography. Circulation 68, 210–216 (1983).
Wyatt, H. L. et al. Cross-sectional echocardiography. I. Analysis of mathematic models for quantifying mass of the left ventricle in dogs. Circulation 60, 1104–1113 (1979).
Pattynama, P. M. T., De Roos, A., Van der Wall, E. E. & Van Voorthuisen, A. E. Evaluation of cardiac function with magnetic resonance imaging. Am. Heart J. 128, 595–607 (1994).
Myerson, S. G., Bellenger, N. G. & Pennell, D. J. Assessment of left ventricular mass by cardiovascular magnetic resonance. Hypertension 39, 750–755 (2002).
Alfakih, K. et al. A comparison of left ventricular mass between two-dimensional echocardiography, using fundamental and tissue harmonic imaging, and cardiac MRI in patients with hypertension. Eur. J. Radiol. 52, 103–109 (2004).
Farber, N. J. et al. Ex vivo cardiovascular magnetic resonance measurements of right and left ventricular mass compared with direct mass measurement in excised hearts after transplantation: A first human SSFP comparison. J. Cardiovasc. Magn. Reson. 16, 1–10 (2014).
DuBois, D. & DuBois, E. A formula to estimate the approximate surface area if height and weight be known. Arch. Intern. Med. 17, 863–871 (1916).
Kawel-Boehm, N. et al. Normal values for cardiovascular magnetic resonance in adults and children. J. Cardiovasc. Magn. Reson. 17, 1–10 (2015).
Bland, J. M. & Altman, D. G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1, 307–310 (1986).
Losi, M. A. et al. Left ventricular mass in hypertrophic cardiomyopathy assessed by 2D-echocardiography: Validation with magnetic resonance imaging. J. Cardiovasc. Transl. Res. 13, 238–244 (2020).
Pacileo, G. et al. Assessment of left-ventricular mass and remodeling in obese adolescents: M-mode, 2D or 3D echocardiography?. J. Cardiovasc. Med. 14, 144–149 (2013).
Chuang, M. L. et al. Three-dimensional echocardiographic measurement of left ventricular mass: Comparison with magnetic resonance imaging and two-dimensional echocardiographic determinations in man. Int. J. Card. Imaging 16, 347–357 (2000).
Mor-Avi, V. et al. Fast measurement of left ventricular mass with real-time three-dimensional echocardiography: Comparison with magnetic resonance imaging. Circulation 110, 1814–1818 (2004).
Paczek, A. et al. Comparison of M-mode and two-dimensional echocardiographic algorithms used to estimate left ventricular mass: The coronary artery risk development in young adults study. J. Am. Soc. Echocardiogr. 8, 780–792 (1995).
Kusunose, K., Kwon, D. H., Motoki, H., Flamm, S. D. & Marwick, T. H. Comparison of three-dimensional echocardiographic findings to those of magnetic resonance imaging for determination of left ventricular mass in patients with ischemic and non-ischemic cardiomyopathy. Am. J. Cardiol. 112, 604–611 (2013).
Chang, S.-A. et al. Assessment of left ventricular mass in hypertrophic cardiomyopathy by real-time three-dimensional echocardiography using single-beat capture image. J. Am. Soc. Echocardiogr. 26, 436–442 (2013).
Bicudo, L. S. et al. Value of real time three-dimensional echocardiography in patients with hypertrophic cardiomyopathy: Comparison with two-dimensional echocardiography and magnetic resonance imaging. Echocardiography 25, 717–726 (2008).
Avegliano, G. P. et al. Utility of real time 3D echocardiography for the assessment of left ventricular mass in patients with hypertrophic cardiomyopathy: Comparison with cardiac magnetic resonance. Echocardiography 33, 431–436 (2016).
Chetrit, M. et al. Optimal technique for measurement of linear left ventricular dimensions. J. Am. Soc. Echocardiogr. 32, 476-483.e1 (2019).
Guzzetti, E. et al. Effect of regional upper septal hypertrophy on echocardiographic assessment of left ventricular mass and remodeling in aortic stenosis. J. Am. Soc. Echocardiogr. 34, 62–71 (2021).
Myhr, K. A. et al. Semi-automated estimation of left ventricular ejection fraction by two-dimensional and three-dimensional echocardiography is feasible, time-efficient, and reproducible. Echocardiography 35, 1795–1805 (2018).
Acknowledgements
We would like to thank radiograph and CMR-technologist Jesper Kromann (Department of Diagnostic Radiology, Rigshospitalet, Denmark) for technical support. This work was funded by the Augustinus Foundation (Application Number 16-3012) and by the research fund of The Heart Center at Rigshospitalet, Denmark.
Author information
Authors and Affiliations
Contributions
C.K., R.M.O., C.H. and N.V. conceived and planned the study. C.K. included the participants. C.K. and K.M. carried out the study. C.K. developed the novel method, performed the data analysis and wrote the manuscript. K.M. contributed to the inter-observer analysis. F.G. and R.M.A. provided critical feedback. All authors discussed the results and contributed to the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Kristensen, C.B., Myhr, K.A., Grund, F.F. et al. A new method to quantify left ventricular mass by 2D echocardiography. Sci Rep 12, 9980 (2022). https://doi.org/10.1038/s41598-022-13677-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-022-13677-1
- Springer Nature Limited