Abstract
A pulse train programmer for a spin–echo NMR relaxometer–diffusiometer is based on an AT89C51-24PC microcontroller. The programmer ensures the arbitrary generation of pulse trains with a minimum train step of 30 μs and recording of up to 30 000 10-bit amplitudes of spin–echo NMR signals with a minimum quantization step of 3 μs. The all-in-one one-board programmer is connected to a computer serial asynchronous data port, which is compatible with NMR relaxometer–diffusiometers of almost any design.
Similar content being viewed by others
REFERENCES
Odnokristal'nye mikro-EVM (Single-Chip Microcomputer), Moscow: MIKAP, 1994.
Fedorkov, B.G. and Telets, V.A., Mikroskhemy TsAP i ATsP: funktsionirovanie, parametry, primenenie (DAC and ADC Microcircuits: Functioning, Parameters, and Application), Moscow: Energoatomizdat, 1990.
Idiyatullin, D.Sh., Cand. Sci. (Phys.-Math.) Dissertation, Kazan, Kazan Inst. of Biochemistry and Biophysics, Kazan Scientific Center, RAS, 1996.
Teplov, V.Yu. and Bochkarev, V.V., Abstracts of Papers, Konferentsiya “Struktura i dinamika molekulyarnykh sistem” (Conf. on the Structure and Dynamics of Molecular Systems), Ioshkar-Ola-Kazan-Moscow, 1998, part 2, p. 28.
Chizhik, V.I., Yadernaya magnitnaya relaksatsiya (Nuclear Magnetic Relaxation), St. Petersburg: St.-Peterb. Gos. Univ., 2000.
Idiatullin, D.Sh., Skirda, V.D., and Smirnov, V.S., USSR Inventor' Certificate no. 1578608, Byull. Izobret., 1990, no. 26, p. 156.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Teplov, V.Y., Romanov, V.I., Anisimov, A.V. et al. Pulse Train Microprocessor Programmer for NMR Relaxometers-Diffusiometers. Instruments and Experimental Techniques 45, 770–775 (2002). https://doi.org/10.1023/A:1021422920448
Issue Date:
DOI: https://doi.org/10.1023/A:1021422920448