Abstract
A circuit diagram of the current integrator intended for the measurements of a positive particle-beam charge arriving at a target is described. The device converts the charge into a number of pulses (target current into the pulse repetition rate). The conversion factor is 109 pulses/C (109 Hz/A), the error is no more than ≤±1%, the dead time is absent, the input impedance is close to zero, and the temperature drift is no more than ≤1.2 × 10–12 А/°С.
Similar content being viewed by others
REFERENCES
Brovchenko, V.G., Vorotnikov, P.E., and Molchanov, Yu.D., Elektronnye ustroistva na elektrostaticheskikh uskoritelyakh (Electron Equipment on Electrostatic Accelerators), Moscow: Atomizdat, 1968, p. 99.
Kolombet, E.A., Mikroelektronnye sredstva obrabotki analogovykh signalov (Microelectron Means for Processing Analog Signals), Moscow: Radio i Svyaz', 1991, p. 299.
Antonov, O.V., Buzumurga, B.Z., Burdeinii, V.A., et al., Prib. Tekh. Eksp., 1989, no. 1, p. 120.
Shchagin, A.V. and Lysenko, V.F., Prib. Tekh. Eksp., 1985, no. 5, p. 120.
Batvinov, L.P., Vergunov, A.D., Glazunov, L.S., et al., Vopr. At. Nauki Tekh., Ser.: Tekh. Fiz. Eksp., 1983, no. 1(22), p. 26.