Abstract
We have evaluated the potential of flow injection chemical reaction interface isotope-ratio mass spectrometry to replace radioactive labeling techniques in material balance studies. A sample is flow injected and transmitted through a desolvation system followed by combustion to form 13CO2 with a microwave-powered chemical reaction interface. We can detect trace amounts of a 13C-labeled drug (3′-azido-3′-deoxythymidine, AZT) in urine or feces. Our ability to quantify less than 100 ng/mL of excess 13C (∼1 µg/mL of 13C-labeled AZT) from a sample equivalent to 10 µL of urine is superior to previous detection limits for 13C in urine that use off-line combustion methods. Parallel studies using 14C-labeled AZT showed that our stable isotope method provides comparable percent excretion data for urine and feces. These results support previous findings that mass balance studies could be carried out with isotope-ratio mass spectrometer, here using doses as low as 1–2 mg/kg.
Article PDF
Similar content being viewed by others
References
von Unruh, G. E.; Hauber, D. H.; Schoeller, D. A.; Hayes, J. M. Biomed. Mass Spectrom. 1974, 1, 345–349.
Nakagawa, A.; Kitagawa, A.; Asami, M.; Nakamura, K.; Schoeller, D. A.; Slater, R.; Minagawa, M.; Kaplan, I. R. Biomed. Mass Spectrom. 1985, 12, 502–506.
Browne, T. R.; Szabo, G. K.; Ajami, A.; Wagner, D. J. Clin. Pharmacol. 1993, 33, 246–252.
Teffera, Y.; Kusmierz, J. J.; Abramson, F. P. Anal. Chem. 1996, 68, 1888–1894.
Chen, P.; Teffera, Y.; Abramson, F. P. Proceedings of the 45th ASMS Conference on Mass Spectrometry Allied Topics, Palm Springs, CA, 1–5 June 1997; p 707.
McLean, M.; Vestal, M. L.; Teffera, Y.; Abramson, F. P. J. Chromatogr. A 1996, 732, 189–199.
Abramson, F. P. Mass Spectrom. Rev. 1994, 13, 341–356.
Horning, M. G.; Boucher, E. A.; Stafford, M.; Horning, E. C. Clin. Chim. Acta 1972, 37, 381–386.
Ahmed, A. E.; Jacob, S.; Loh, J.-P.; Samra, S. K.; Nokta, M.; Pollard, R. B. J. Pharmacol. Exp. Therap. 1991, 257, 479–486.
Miller, J. C.; Miller, J. N. Statistics for Analytical Chemistry, 3rd ed.; Ellis Horwood Limited: West Sussex, 1993; p 117.
De Miranda, P.; Burnette, T. C.; Good, S. S. Drug Metab. Dispos 1990, 18, 315–320.
Abramson, F. P.; Teffera, Y.; Kusmierz, J.; Steenwyk, R. C.; Pearson, P. G. Drug Metab. Dispos 1996, 24, 697–701.
Goldthwaite, C. A., Jr.; Hsieh, F.-Y.; Womble, S. W.; Nobes, B. J.; Blair, I. A.; Klunk, L. J.; Mayol, R. F. Anal. Chem. 1996, 68, 2996–3001.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Chen, P., Teffera, Y., Black, G.E. et al. Flow injection with chemical reaction interface-isotope ratio mass spectrometry: an alternative to off-line combustion for detecting low levels of enriched 13C in mass balance studies. J Am Soc Mass Spectrom 10, 153–158 (1999). https://doi.org/10.1016/S1044-0305(98)00133-0
Received:
Revised:
Accepted:
Issue Date: