Effect of Elevated Acetabular Cup on Contact and Failure Analysis in Hip Implants for Different Microseparations and Cup Inclinations Under Routine Gait Activities Using In Silico Approach

  • Original Article
  • Published:
Indian Journal of Orthopaedics Aims and scope Submit manuscript

Abstract

Objectives

The acetabular cup design plays a critical role in reducing contact stress between femur head acetabular cup. Many studies used ellipsoidal and spheroidal geometry in acetabular cup design to effectively reduce contact stress. The present study focuses on elevated acetabular cup rim with round corner design to reduce contact stress with round corner geometry.

Methods

The cobalt chromium femur head and cup are considered for finite element (FE) model of hip resurfacing. The gait loads of routine activities of humans like normal walking, stair ascending and descending and sitting down and getting up gait activities are applied to the developed 3D FE model. Five microseparations of 0.5, 1, 1.5, 2 and 2.5 mm are considered in the present study. The acetabular cup inclination angle considered for this study are 35°, 45°, 55°, 65° and 75°. The contact stress and von Mises stress plot for each gait activities under these microseparations are analyzed for betterment of longevity of implants.

Results

Overall elevated cup rim design helped in reducing contact stress to a greater extent than conventional cup with different geometries. Also, the predicted von Mises stress for all the parameters considered in the current study are well within the yield strength of CoCr material. Therefore, elevated cup rim could be used as a better alternative to spline and, ellipsoidal and circular geometries of cup.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig.1
Fig.2
Fig.3
Fig.4
Fig.5
Fig.6
Fig.7

Similar content being viewed by others

Data availability

Data will be made available on request.

References

  1. Uddin, M. S., & Chan, G. W. C. (2019). Reducing stress concentration on the cup rim of hip implants under edge loading. International Journal for Numerical Methods in Biomedical Engineering., 35(1), e3149.

    Article  PubMed  Google Scholar 

  2. Wang, L., Liu, X., Li, D., Liu, F., & Jin, Z. (2014). Contact mechanics studies of an ellipsoidal contact bearing surface of metal-on-metal hip prostheses under micro-lateralization. Medical Engineering & Physics., 36(4), 419–424.

    Article  Google Scholar 

  3. Shankar S, Nithyaprakash R, Selvamani K, Ponappa K, Uddin M, Santhosh B. Effect of Radial Clearance, Corner Radius and Micro-Lateralization on Contact Stress of Metallic and Ceramic Hip Prosthesis-A Finite Element Analysis. Defence Science Journal. 2022;72(3).

  4. Cerveri, P., Manzotti, A., & Baroni, G. (2014). Patient-specific acetabular shape modelling: Comparison among sphere, ellipsoid and conchoid parameterisations. Computer Methods in Biomechanics and Biomedical Engineering., 17(5), 560–567.

    Article  PubMed  Google Scholar 

  5. de Villiers, D., & Collins, S. (2020). Wear of large diameter ceramic-on-ceramic hip bearings under standard and microseparation conditions. Biotribology., 21, 100117.

    Article  Google Scholar 

  6. Hart, A., Buddhdev, P., Winship, P., Faria, N., Powell, J., & Skinner, J. (2008). Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing. Hip International., 18(3), 212–219.

    Article  CAS  PubMed  Google Scholar 

  7. Al-Hajjar, M., Leslie, I. J., Tipper, J., Williams, S., Fisher, J., & Jennings, L. M. (2010). Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX® delta ceramic-on-ceramic total hip replacement. Journal of Biomedical Materials Research Part B, 95(2), 263–268.

    Article  Google Scholar 

  8. Shankar, S., Nithyaprakash, R., Santhosh, B., Gur, A. K., & Pramanik, A. (2020). Experimental and submodeling technique to investigate the wear of silicon nitride against Ti6Al4V alloy with bio-lubricants for various gait activities. Tribology International., 151, 106529.

    Article  CAS  Google Scholar 

  9. Shankar, S., Nithyaprakash, R., Abbas, G., Pramanik, A., Basak, A. K., & Prakash, C. (2021). In-vitro tribological study and submodeling finite element technique in analyzing wear of zirconia toughened alumina against alumina with bio-lubricants for hip implants. Medical Engineering & Physics., 98, 83–90.

    Article  CAS  Google Scholar 

  10. Feyzi, M., Fallahnezhad, K., Taylor, M., & Hashemi, R. (2021). A review on the finite element simulation of fretting wear and corrosion in the taper junction of hip replacement implants. Computers in Biology and Medicine., 130, 104196.

    Article  CAS  PubMed  Google Scholar 

  11. Freitas, A., Barin, F. R., Battaglion, L. R., da Costa, H. I., Santos, E. D., Rosado, H. A., et al. (2021). Proposal for a new fixation method for pauwels type III femoral neck fracture-metaphyseal stem: A finite-element analysis. Indian Journal of Orthopaedics., 55(Suppl 2), 378–384.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Kaku, N., Tanaka, A., Tagomori, H., & Tsumura, H. (2020). Finite element analysis of stress distribution in flat and elevated-rim polyethylene acetabular liners. Clinics in Orthopedic Surgery., 12(3), 291.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Daniel, M., Iglič, A., & Kralj-Iglič, V. (2008). Hip contact stress during normal and staircase walking: The influence of acetabular anteversion angle and lateral coverage of the acetabulum. Journal of applied biomechanics., 24(1), 88–93.

    Article  PubMed  Google Scholar 

  14. Liu, F., & Fisher, J. (2017). Effect of an edge at cup rim on contact stress during micro-separation in ceramic-on-ceramic hip joints. Tribology international., 113, 323–329.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Hua, X., Wroblewski, B. M., Jin, Z., & Wang, L. (2012). The effect of cup inclination and wear on the contact mechanics and cement fixation for ultra high molecular weight polyethylene total hip replacements. Medical Engineering & Physics., 34(3), 318–325.

    Article  Google Scholar 

  16. Chegini, S., Beck, M., & Ferguson, S. J. (2009). The effects of impingement and dysplasia on stress distributions in the hip joint during sitting and walking: A finite element analysis. Journal of Orthopaedic Research., 27(2), 195–201.

    Article  PubMed  Google Scholar 

  17. Liu, F., Williams, S., & Fisher, J. (2015). Effect of microseparation on contact mechanics in metal-on-metal hip replacements—A finite element analysis. Journal of Biomedical Materials Research Part B, 103(6), 1312–1319.

    Article  CAS  Google Scholar 

  18. Gkiatas, I., Karasavvidis, T., Sharma, A. K., Xiang, W., Malahias, M.-A., Chalmers, B. P., et al. (2022). Use of elevated liners in primary total hip arthroplasty: A systematic review of the literature. European Journal of Orthopaedic Surgery & Traumatology, 32(4), 587–594.

    Article  Google Scholar 

  19. Krull, P., Steinbrück, A., Grimberg, A. W., Melsheimer, O., Morlock, M., & Perka, C. (2022). Modified acetabular component liner designs are not superior to standard liners at reducing the risk of revision: An analysis of 151,096 cementless total hip arthroplasties from the German Arthroplasty Registry. The Bone & Joint Journal, 104(7), 801–810.

    Article  Google Scholar 

  20. Geng, X., Li, Y., Li, F., Wang, X., Zhang, K., Liu, Z., et al. (2020). A new 3D printing porous trabecular titanium metal acetabular cup for primary total hip arthroplasty: A minimum 2-year follow-up of 92 consecutive patients. Journal of Orthopaedic Surgery and Research, 15, 1–8.

    Google Scholar 

  21. Uddin, M. (2015). Contact of dual mobility implants: Effects of cup wear and inclination. Computer Methods in Biomechanics and Biomedical Engineering, 18(15), 1611–1621.

    Article  CAS  PubMed  Google Scholar 

  22. Shankar, S., & Nithyaprakash, R. (2014). Effect of radial clearance on wear and contact pressure of hard-on-hard hip prostheses using finite element concepts. Tribology Transactions, 57(5), 814–820.

    Article  CAS  Google Scholar 

  23. Yu, Z., & Zhang, L. (2023). Design of low wear artificial hip joint considering 3D physiological loading and motion. Wear, 523, 204744.

    Article  CAS  Google Scholar 

  24. Sariali, E., Stewart, T., Jin, Z., & Fisher, J. (2012). Effect of cup abduction angle and head lateral microseparation on contact stresses in ceramic-on-ceramic total hip arthroplasty. Journal of Biomechanics., 45(2), 390–393.

    Article  PubMed  Google Scholar 

  25. Shankar, S., Nithyaprakash, R., Santhosh, B., Uddin, M., & Pramanik, A. (2020). Finite element submodeling technique to analyze the contact pressure and wear of hard bearing couples in hip prosthesis. Computer Methods in Biomechanics and Biomedical Engineering., 23(8), 422–431.

    Article  CAS  PubMed  Google Scholar 

  26. Shankar, S., Nithyaprakash, R., Sugunesh, P., Uddin, M., & Pramanik, A. (2020). Contact stress and wear analysis of zirconia against alumina for normal and physically demanding loads in hip prosthesis. Journal of Bionic Engineering., 17, 1045–1058.

    Article  Google Scholar 

  27. Ruhr M, Baetz J, Pueschel K, Morlock MM. Influence of acetabular cup thickness on seating and primary stability in total hip arthroplasty. Journal of Orthopaedic Research®. 2022;40(9):2139–46.

  28. Alpkaya AT, Mihcin S. Dynamic computational wear model of PEEK-on-XLPE bearing couple in total hip replacements. Medical Engineering & Physics. 2023:104006.

  29. Varady, P. A., Glitsch, U., & Augat, P. (2015). Loads in the hip joint during physically demanding occupational tasks: A motion analysis study. Journal of Biomechanics, 48(12), 3227–3233.

    Article  PubMed  Google Scholar 

Download references

Funding

No funding was received.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to R. Nithyaprakash.

Ethics declarations

Conflict of Interest

R. Nithyaprakash, S. Shankar, R. Naveenkumar, A.J. Kirthik Raja, R.Gokul Anand, G.Abbas, Chander Prakash, Alokesh Pramanik, Animesh Kumar Basak declare that they have no conflict of interest.

Ethical Standard Statement

This article does not contain any studies with human or animal subjects performed by the any of the authors.

Informed Consent

For this type of study informed consent is not required.

Research Involving Human Participants

No human participants involved in this study.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nithyaprakash, R., Shankar, S., Naveenkumar, R. et al. Effect of Elevated Acetabular Cup on Contact and Failure Analysis in Hip Implants for Different Microseparations and Cup Inclinations Under Routine Gait Activities Using In Silico Approach. JOIO (2024). https://doi.org/10.1007/s43465-024-01146-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1007/s43465-024-01146-y

Keywords

Navigation