Abstract
Purpose
Growing rods are the gold-standard for treatment of early onset scoliosis (EOS). However, these implanted rods experience frequent fractures, requiring additional surgery. A recent study by the U.S. Food and Drug Administration (FDA) identified four common rod fracture locations. Leveraging this data, Agarwal et al. were able to correlate these fractures to high-stress regions using a novel finite element analysis (FEA) framework for one patient. The current study aims to further validate this framework through FEA modeling extended to multiple patients.
Methods
Three patient-specific FEA models were developed to match the pre-operative patient data taken from both registry and biplanar radiographs. The surgical procedure was then simulated to match the post-operative deformity. Body weight and flexion bending (1 Nm) loads were then applied and the output stress data on the rods were analyzed.
Results
Radiographic data showed fracture locations at the mid-construct, adjacent to the distal and tandem connector across the patients. Stress analysis from the FEA showed these failure locations matched local high-stress regions for all fractures observed. These results qualitatively validate the efficacy of the FEA framework by showing a decent correlation between localized high-stress regions and the actual fracture sites in the patients.
Conclusions
This patient-specific, in-silico framework has huge potential to be used as a surgical tool to predict sites prone to fracture in growing rod implants. This prospective information would therefore be vital for surgical planning, besides helping optimize implant design for reducing rod failures.
Similar content being viewed by others
Introduction
Treatment of excessive deformity (coronal Cobb angle > 40°) in early onset scoliosis (EOS) has always been challenging due to concurrent management of progressive adolescent growth along with correction of the spinal deformity in these young patients [1, 2]. These growth-friendly surgical techniques can be broadly classified as distraction-based, guided growth, or compression-based techniques [3]. Common examples of these include traditional growing rods, magnetically controlled growing rods, SHILLA growth guidance technique, and, more recently, anterior vertebral body tethers [2, 3]. Barring the latter, a common complication across these techniques is rod fractures. Multiple studies have reported a high incidence (> 25%) of rod fractures endemic to the most used traditional and magnetic growing rod techniques [4,5,6,7,8,9]. Given the high usage and significant failure incidence of these rods in EOS, research focused on prospective identification of potential failure-prone regions in rods might facilitate improved implant design as well as surgical planning.
Finite element analysis (FEA) is a commonly used tool for performing such evaluations in medical devices, along with traditional benchtop testing options [10, 11]. FEA helps overcome problems inherent to physical experiments for evaluating EOS therapies, i.e., limited availability of relevant scoliotic cadavers, variability of deformity manifestation, and the limited accuracy of stress measurements on implanted constructs [12]. Furthermore, with the advent of patient-specific FEA, the research can be highly specialized/focused and generate meaningful data when compared to a more generalized approach [13]. For example, a study by Henao et al. showed how a patient-specific FEA successfully reproduced the biomechanics of neurological injury during scoliosis correction maneuvers when compared to clinical cases with and without intra-operative neurological complications [14]. In the future, modeling frameworks like these could be used as a tool to aid pre-operative surgical planning. Specifically, the biomechanical data obtained from patient-specific FEA can be applied to simulate different spinal disorders and thus lead to optimization of surgical planning.
In the field of scoliosis, current literature using patient-specific FEA is focused on two specific areas. In the first area, studies focus on validating the modeling framework by quantifying the variation in modeling predictions when compared to clinical outcomes (i.e., Cobb angle correction, kyphosis correction, and spinal height change) [15, 16]. For example, a study conducted by Little and Adam showed low variation in the anatomical measurements of spinal parameters (such as Cobb angle, kyphosis, lordosis, thoracic spinal heights, etc.) across FEA model predictions compared with the mean intra-observer variability for three patients [15]. The second area focuses on the same clinical outcomes but with different correction maneuvers and different instrumentation strategies [17,18,19,20,21]. For example, a recent study by Wang et al. explored the effect of various configurations of rod contouring on three-dimensional spinal correction. The key outcome to evaluate different configurations was to understand their effect on the bone-screw forces stemming from screw pull-out; a known complication in adolescent idiopathic scoliosis surgeries [18].
Building on this work, Agarwal et al. developed a novel patient-specific computational modeling framework and correlated high-stress regions on traditional growing rods with clinical fracture-prone locations for a single patient [22]. This FEA model was developed using clinical registry data obtained as part of a study conducted by the U.S. Food and Drug Administration (FDA) on retrieved failed traditional growing rods across 36 patients [23]. The results of this study verified the proof-of-concept modeling framework, with two of the three rod fracture-prone regions (high-stress locations) matching the retrieval data. The study, however, was limited in that data from a single-patient FEA model. The results of the previous study were compared against broad conclusions from the previous retrieval analysis data [22, 23]. The aim of the current study is to extend this verified framework to multiple EOS patients with patient-specific rod-failure location data that aids in understanding the growing rod failures from a biomechanical perspective. Therefore, the objective of this study is to develop three patient-specific finite element models, simulate the traditional growing rod surgery, and validate the high-stress regions on growing rods with respect to the clinical rod fracture locations. This information about the failure-prone regions on implants would be valuable to the surgeons/end users to supplement surgical planning. Furthermore, this biomechanical analysis may aid manufacturers in device design and development, to optimize implant design for reducing rod failures.
Methods
This study leverages the existing computational modeling framework to identify and validate high-stress regions on traditional growing rods against prior retrieval analysis fracture data [22]. In the current study, the FEA modeling framework outlined in the sections below is applied to three patients, simulating traditional growing rod surgery, to compare rod-stress data to their respective rod-failure data obtained from the clinical registry. The hypothesis is that the clinical fracture locations will match the high-stress locations from the FEA models.
Pre-operative scoliotic FE model development
Three patient-specific FEA models of the thoracolumbar spine (T1-S1) were developed to match the pre-operative (pre-op) scoliosis curve to corresponding patient registry (Growing Spine Study Group, San Diego, CA) and biplanar radiographs, as described previously [22]. In brief, a healthy pediatric spine FEA model was modified using a custom MATLAB script (MATLAB Inc., Natick, MA) to induce patient-specific biplanar deformity [8, 9, 22]. Input data for necessary parameters like Cobb angle, thoracic kyphosis, lumbar lordosis, and the spinal height (T1-S1) were measured using Surgimap software (Surgimap, Nemaris Inc., New York, NY) from the patient radiographs (Table 1). The final pre-op parameters for the FEA models were within ± 5° of the radiographic measurements [24, 25].
Post-operative scoliotic FE model development
Following the creation of the pre-operative FEA models, patient-specific instrumentation was created using SolidWorks (Dassault Systèmes, SolidWorks Corporation, Waltham, MA, USA) for the three FEA models. Specific data regarding location, material, and geometry of the implants were obtained from post-operative patient radiographs and registry data (Table 2). All designed constructs comprised dual traditional growing rod constructs; multiple tandem connectors and pedicle screws customized for each patient. The instrumentation was then implanted into the pre-operative model. An element size of 0.5 mm was chosen based on the prior mesh convergence study [22]. Material properties, constitutive laws, and element types are described in the previous study [22].
Next, the implanted pre-op patient-specific scoliotic model was modified to simulate the surgical procedure and thus match the post-op scoliotic curve parameters. This correction was done to match the sagittal contour of rods to the post-op radiographs of the patient, and to obtain the stresses generated on the rods (simulation of rod attachment) [19]. An iterative correction process (model calibration) was done by controlling the magnitude of applied distraction in ABAQUS to each of the three FEA models to match post-operative parameters (Table 3). The final post-op scoliotic curve parameters for the spinal model were within ± 5° of the radiographic measurements except for the kyphosis parameter of patients 1 and 3 [24, 25]. For the aforementioned patients, there was a computational limitation where we could not achieve the necessary kyphosis angle without having an adverse effect on the coronal Cobb angles. However, the primary parameter for quantification of scoliotic deformity is the coronal Cobb angle which was within ± 5°. Furthermore, in patients 2 and 3, we have chosen to use pedicle screws as our proximal anchors in place of laminar or cranial hooks. This choice was based on two reasons. First, our framework did not model any ribcage required for simulation of these devices which would increase the complexity and computational expense. Furthermore, a few clinical studies investigated the effect of using pedicle screws versus proximal hooks, and have shown no significant differences, rather suggesting the use of pedicle screws where possible [26, 27].
Contact and loading conditions
For the final step, the patient-specific spinal instrumentation was implanted into the surgically corrected spinal model (Figs. 1, 2, 3). This was followed by application of bilateral longitudinal distraction, application of weight-specific follower load, and application of a 1 Nm flexion moment at the T1 vertebra to simulate worst-case bending motion [22, 23, 28]. The interactions between different interfaces of the FEA model are as listed in similar studies published previously [8, 9, 22]. The inferior endplate (base) of the S1 vertebra was fixed in all directions during all steps.
Qualitative validation
Spatial distribution of stresses was recorded on the rods after (1) surgical correction and application of distraction forces and follower loads, and (2) following 1 Nm of flexion bending. Next, areas of high-stress concentrations (first principal stress) were identified and compared directly to the corresponding rod-failure locations obtained from the patient-specific radiographs obtained from the retrieval database. The first principal stress was chosen as the preferred stress parameter as it helps in understanding the maximum tensile stress induced in the part due to the loading conditions [29]. This was based on evidence from Hill et al., which revealed that the failure mechanism was due to repeated flexion motion with crack initiation on the posterior side of the rod, causing stresses to be tensile in nature [23].
Quantitative data evaluation
The study also quantified the location and magnitude of highest principal stress markers that were extracted from the regions on the rods where the respective rod-failure locations were identified. The location of the maximum stress in the FEA model was calculated with respect to the bottom of the nearest tandem connector, as this was the landmark used to calculate the analogous failure location in the clinical radiographs. This location and stress magnitude were compared to the clinical data from radiographs and the global stress maxima, respectively.
Results
The spatial distribution of FEA stresses for each of the rods compared to the retrieval data was analyzed (Figs. 4–7). The posterior surface was examined, because the fracture initiation sites identified in previous retrieval analyses were on the posterior surface [23]. For Patient 1, the bottom left rod failed. The high-stress region in the mid-construct region observed on the bottom left rod in the FEA model matched the failure location from the retrieval data (Fig. 4). Patient 2 had two fracture locations: the top left and the bottom right rod. For the top left rod, the high-stress region observed adjacent to the tandem connector in the FEA model matched the retrieval data (Fig. 5). Furthermore, the fracture location for the bottom right rod matched the high-stress location in the FEA model (Fig. 6). Finally, the identified high-stress region in the FEA model matched the retrieval data for Patient 3 (adjacent to distal anchor, Fig. 7). The stress distribution data shown in Figs. 4–7 correspond to the post-operative radiographs at the last follow-up before rod breakage.
The magnitude of principal stress at the failure location in each of the localized rod regions (identified from the qualitative study) was within 30% of the maximum principal stress on the identified rod for all patients except patient 1 (Table 4). Similarly, the spatial location of these maximum principal stress regions on each of the rods showed a close correlation (within 10 mm—length of the growing rod that interfaces with one pedicle screw) when compared to the radiographically measured location of rod failure for all patients except patient 1 (Table 5).
Discussion
Rod fracture is a common complication with high incidence rates, especially in growing rod implants used for EOS patients [4,5,6,7,8,9]. A recent proof-of-concept study developed a patient-specific FEA framework to identify and validate the high-stress regions of traditional growing rods against their fracture location obtained from clinical registry data [22, 23]. This study helped establish that clinically observed fracture regions had high localized stress values; two out of the three high-stress regions matched the data from the retrieval analysis conducted by Hill et al. [23]. The results of the current study reinforce the validity of this framework, where the high-stress regions on the FEA growing rods match the fracture locations in the clinical data for all three patients.
Location-wise, the most frequently observed failure locations were adjacent to the distal anchor and adjacent to the tandem connector. Additionally, the results also suggest that the distal region might be more susceptible to fracture. The higher incidence of fractures at these specific locations matches the retrospective clinical observations of a larger study conducted by Hill et al. [23]. The quantitative data are largely consistent (exception in patient 3) in the magnitude and location of maximum stress regions when compared to the clinical data (Tables 4 and 5). As seen in patient 3, there is a significant difference in the stress magnitude in the localized high-stress region when compared to the maximum stress on the fractured rod (Fig. 4, difference of 58%). This actual location of the maximum stress is closer to the tandem connector, as opposed to being adjacent to the distal screw. This could be attributed to the fact that there were two high-stress regions at the proximal and distal ends of the long rod (bottom left rod, Fig. 4). The authors posit that this difference could be attributed to the intra-operative procedures such as notching on the rods, along with other biomechanical effects such as residual stresses due to spinal rod contouring and rod-screw interconnection assembly, which are not included in the scope of the study. [30,31,32].
Here, it is important to highlight that the authors used static simulations (i.e., no modeling of local damage or damage accumulation) for the purpose of this FEA study. This assumption of correlating stress concentrations from static simulations to fatigue fracture location has been confirmed to be reasonable in other biomechanical studies and employed in ASTM FEA standards as well (e.g., ASTM F2996, F3161, and F3334) [31, 33,34,35,36]. These fractures could also have occurred due to loading conditions specific to each patient such as trauma or common physical activities that cause significant flexion or extension in the spine. These simplified simulation methods, therefore, were able to successfully match the high-stress regions in the FEA to the clinical failure locations. This also follows previously published biomechanical data which showed factors like patient weight and distraction loading significantly contributed to generation of high stresses on the rods [37, 38]. Therefore, this patient-specific FEA framework has the potential to help identify high-risk areas and examine the underlying cause of potential failures, with the flexibility to be tailored to different loading regimes in different clinical applications. To our knowledge, the present study is the first in the field of early onset scoliosis to combine patient-specific FEA with clinical registry data to biomechanically evaluate potential rod failure in traditional growing rods.
The current study has some limitations, one being the limited cohort size of the study. This multi-patient study, however, builds on the previous study (limited to one patient compared to a pool of clinical data) and effectively validates the framework for future investigations of failure biomechanics in EOS. Another limitation is the potential error introduced due to a single operator using Surgimap software to obtain the radiographic measurements. Future efforts will quantify operator error through an uncertainty analysis for the multi-patient data. Similarly, another source of potential error is the use of the same material properties across all patients sourced from published literature. However, availability and extraction of patient-specific material properties for a retrospective study is very challenging. Given this limitation, validation studies focusing on patient-specific-FE models have a comparable sample size, in terms of datapoints use [29, 36, 38, 39]. Literature has shown this to mainly affect the range of motion of the spine, which would primarily affect the magnitude of the stresses recorded on the rods [40]. Furthermore, the magnitude of stresses reported in this study seems to be on the lower spectrum compared to the yield stress of the rod material. This is possibly due to the static loading assumption. However, given the scope of the current study, where the focus is on evaluating the distribution of stress on the rods to identify relatively high-stress concentration regions, evaluation of the magnitude of stress becomes less relevant. In addition, these stresses may further intensify with fatigue loading, vary among patients, and may even increase with consecutive distractions for each individual (due to changes in spinal flexibility and/or autofusion) [41].
Overall, this study aids in understanding implant failure from a biomechanical perspective. This patient-specific finite element modeling framework showed a decent correlation to clinical results when examining three EOS patients. Therefore, the computational modeling framework adopted in this study has clinical relevance, since we were successful in validation of this patient-specific in-silico modeling framework; ratified via retrospective clinical data to predict rod-failure locations. With further refinement, this approach may provide specific information that would be valuable to the surgeons/end users about the failure-prone regions on implants to supplement surgical planning. This refinement can be achieved in future studies by performing uncertainty quantification and incorporating more patients retrospectively and potentially prospective as well. This future work would also include analysis based on a variety of device characteristics. Additionally, the results strengthen the finding of the clinical retrieval analysis from a biomechanical perspective and may help device manufacturers optimize implant designs for future cases.
Conclusion
Overall, this study helps further validate the patient-specific finite element modeling framework. FEA stress concentrations spatially correlated to the analogous retrieval data for all patients. The authors believe that this framework may aid in predicting traditional growing rod failures by providing specific information that would be valuable to the surgeons/end users about the failure-prone regions on implants to supplement surgical planning. Furthermore, this biomechanical analysis may aid manufacturers in device design and development, to optimize the implant design for reducing rod failures. The study also elucidates the potential of utilizing clinical/patient registry data as a validation comparator to help the incumbent state of spinal treatment by providing biomechanical reasoning on traditional growing rod failure which remains a question in the treatment of early onset scoliosis.
References
Van Goethem JW, Van Campenhout A (2007) Scoliosis. Neuroimaging Clin N Am 17.1:105–115. https://doi.org/10.1007/978-3-540-68483-1_4
Blevins K, Battenberg A, Beck A (2018) Management of scoliosis. Adv Pediatr 65:249–266. https://doi.org/10.1016/j.yapd.2018.04.013
Schlösser TPC, Kruyt MC, Tsirikos AI (2021) Surgical management of early onset scoliosis: indications and currently available techniques. Orthop Trauma 35:336–346. https://doi.org/10.1016/j.mporth.2021.09.004
Sankar WN, Acevedo DC, Skaggs DL (2010) Comparison of complications among growing spinal implants. Spine 35:2091–2096. https://doi.org/10.1097/brs.0b013e3181c6edd7
Greggi T, Lolli F, Di Silvestre M et al (2012) Complications incidence in the treatment of early onset scoliosis with growing spinal implants. Stud Health Technol Inform 176:334–337. https://doi.org/10.3233/978-1-61499-067-3-334
Bess S, Akbarnia BA, Thompson GH et al (2010) Complications of growing-rod treatment for early onset scoliosis: analysis of one hundred and forty patients. J Bone Joint Surg 92:2533–2543. https://doi.org/10.2106/jbjs.i.01471
Cryar KA, Bumpass DB, McCullough L, McCarthy RE (2017) Rod breakage in Shilla growth guidance constructs: when, where, and why? Spine J. https://doi.org/10.1016/j.spinee.2017.07.091
Agarwal A, Kelkar A, Agarwal AG et al (2020) Device-related complications associated with Magec rod usage for distraction-based correction of scoliosis. Spine Surg Relat Res 4:148–151. https://doi.org/10.22603/ssrr.2019-0041
Lemans JVC, Kodigudla MK, Kelkar AV et al (2021) Finite element comparison of the spring distraction system and the traditional growing rod for the treatment of early onset scoliosis. Spine. https://doi.org/10.1097/brs.0000000000004297
Prendergast PJ (1997) Finite element models in tissue mechanics and orthopaedic implant design. Clin Biomech 12:343–366. https://doi.org/10.1016/s0268-0033(97)00018-1
Harrysson OLA, Hosni YA, Nayfeh JF (2007) Custom-designed orthopedic implants evaluated using finite element analysis of patient-specific computed tomography data: femoral-component case study. BMC Musculoskelet Disord. https://doi.org/10.1186/1471-2474-8-91
Kiapour A, Kiapour A, Palepu V, Demetropoulos C (2015) Finite element analyses (fea)finite element analysis. Dyn Reconstr Spine. https://doi.org/10.1055/b-0035-108522
Gefen A (2012) Patient-specific modeling in tomorrow’s medicine. Stud Mechanobiol Tissue Eng Biomater. https://doi.org/10.1007/978-3-642-24618-0
Henao J, Aubin C-É, Labelle H, Arnoux P-J (2015) Patient-specific finite element model of the spine and spinal cord to assess the neurological impact of scoliosis correction: preliminary application on two cases with and without intraoperative neurological complications. Comput Methods Biomech Biomed Engin 19:901–910. https://doi.org/10.1080/10255842.2015.1075010
Little JP, Adam CJ (2013) Geometric sensitivity of patient-specific finite element models of the spine to variability in user-selected anatomical landmarks. Comput Methods Biomech Biomed Engin 18:676–688. https://doi.org/10.1080/10255842.2013.843673
Balasubramanian S, D’Andrea CR, Viraraghavan G, Cahill PJ (2022) Development of a finite element model of the pediatric thoracic and lumbar spine, ribcage, and pelvis with orthotropic region-specific vertebral growth. J Biomech Eng. https://doi.org/10.1115/1.4054410
Le Navéaux F, Larson AN, Labelle H et al (2016) How does implant distribution affect 3D correction and bone-screw forces in thoracic adolescent idiopathic scoliosis spinal instrumentation? Clin Biomech 39:25–31. https://doi.org/10.1016/j.clinbiomech.2016.09.002
Wang X, Boyer L, Le Naveaux F et al (2016) How does differential rod contouring contribute to 3-dimensional correction and affect the bone-screw forces in adolescent idiopathic scoliosis instrumentation? Clin Biomech 39:115–121. https://doi.org/10.1016/j.clinbiomech.2016.10.002
Pei B, Lu D, Wu X et al (2022) Kinematic and biomechanical responses of spine to distraction surgery in children with early onset scoliosis: a 3D finite element analysis. Front Bioeng Biotechnol 10:933341. https://doi.org/10.3389/fbioe.2022.933341
Pan A, Ding H, Wang J et al (2022) The application of finite element analysis to determine the optimal UIV of growing-rod treatment in early onset scoliosis. Front Bioeng Biotechnol 10:978554. https://doi.org/10.3389/fbioe.2022.978554
Pei B, Lu D, Wu X et al (2022) Effects of growing rod technique with different surgical modes and growth phases on the treatment outcomes of early onset scoliosis: a 3D finite element analysis. Int J Environ Res Public Health 19(4):2057. https://doi.org/10.3390/ijerph19042057
Agarwal A, Kodigudla M, Kelkar A et al (2021) Towards a validated patient-specific computational modeling framework to identify failure regions in traditional growing rods in patients with early onset scoliosis. North Am Spine Soc J 5:100043. https://doi.org/10.1016/j.xnsj.2020.100043
Hill G, Nagaraja S, Akbarnia BA et al (2017) Retrieval and clinical analysis of distraction-based dual growing rod constructs for early onset scoliosis. Spine J 17:1506–1518. https://doi.org/10.1016/j.spinee.2017.04.020
Tauchi R, Tsuji T, Cahill PJ et al (2015) Reliability analysis of Cobb angle measurements of congenital scoliosis using X-ray and 3D-CT images. Eur J Orthop Surg Traumatol 26:53–57. https://doi.org/10.1007/s00590-015-1701-7
Tanure MC, Pinheiro AP, Oliveira AS (2010) Reliability assessment of Cobb angle measurements using manual and digital methods. Spine J 10:769–774. https://doi.org/10.1016/j.spinee.2010.02.020
Wilke HJ, Kaiser D, Volkheimer D et al (2016) A pedicle screw system and a lamina hook system provide similar primary and long-term stability: a biomechanical in vitro study with quasi-static and dynamic loading conditions. Eur Spine J 25(9):2919–2928. https://doi.org/10.1007/s00586-016-4679-x
Kim YJ, Lenke L, Cho SK et al (2004) Comparative analysis of pedicle screw versus hook instrumentation in posterior spinal fusion of adolescent idiopathic scoliosis. Spine 29(18):2040–2048. https://doi.org/10.1097/01.brs.0000138268.12324.1a
Schultz A, Andersson GB, Örtengren R et al (1982) Analysis and quantitative myoelectric measurements of loads on the lumbar spine when holding weights in standing postures. Spine 7:390–397. https://doi.org/10.1097/00007632-198207000-00009
Sutradhar A et al (2014) Experimental validation of 3D printed patient-specific implants using digital image correlation and finite element analysis. Comput Biol Med 52:8–17
Berti F, La Barbera L, Piovosen A et al (2018) Residual stresses in titanium spinal rods: effects of two contouring methods and material plastic properties. J Biomech Eng 140(11):111001. https://doi.org/10.1115/1.4040451
Piovosen A, Berti F, Villa T et al (2019) Computational and experimental fatigue analysis of contoured spinal rods. J Biomech Eng 140(4):044505. https://doi.org/10.1115/1.4042767
Ciriello L, Berti F, La Barbera L et al (2022) Global stiffness and residual stresses in spinal fixator systems: a validated finite element study on the interconnection mechanism. J Mech Behav Biomed Mater 135:105460. https://doi.org/10.1016/j.jmbbm.2022.105460
ASTM international (2020) ASTM F2996-20 Standard practice for finite element analysis (FEA) of non-modular metallic orthopaedic hip stems. https://doi.org/10.1520/F2996-20
ASTM international (2016) ASTM F3161-16 Standard test method for finite element analysis (FEA) of metallic total knee femoral components under closing conditions. https://doi.org/10.1520/F3161-16
ASTM international (2019) ASTM F3334-19 Standard practice for finite element analysis (FEA) of metallic total knee tibial components. https://doi.org/10.1520/F3334-19
Mansi T et al (2011) Towards patient-specific finite-element simulation of mitralclip procedure. In: Medical image computing and computer-assisted intervention–MICCAI 2011: 14th international conference, Toronto, Canada, September 18–22, 2011, Proceedings, Part I 14. Springer Berlin Heidelberg
Agarwal A, Agarwal AK, Jayaswal A, Goel V (2014) Smaller interval distractions may reduce chances of growth rod breakage without impeding desired spinal growth: a finite element study. Spine Deform 2:430–436. https://doi.org/10.1016/j.jspd.2014.08.004
Trabelsi N, Yosibash Z (2011) Patient-specific finite-element analyses of the proximal femur with orthotropic material properties validated by experiments. J Biomech Eng 133:061001
Shim VB et al (2008) Development and validation of patient-specific finite element models of the hemipelvis generated from a sparse CT data set. J Biomech Eng 130:051010
Schlager B, Niemeyer F, Galbusera F et al (2018) Uncertainty analysis of material properties and morphology parameters in numerical models regarding the motion of lumbar vertebral segments. Comput Methods Biomech Biomed Engin 21:673–683. https://doi.org/10.1080/10255842.2018.1508571
Agarwal A, Goswami A et al (2019) Quantitative characteristics of consecutive lengthening episodes in early onset scoliosis (EOS) patients with dual growth rods. Spine 44(6):397–403. https://doi.org/10.1097/BRS.0000000000002835
Acknowledgements
The authors would like to thank the entire Growing Spine Foundation for their collaboration via a Research Collaborative Agreement to use registry data for this study.
Funding
This study was funded by the U.S. Food and Drug Administration (FDA).
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design, material preparation, data collection and analysis, and manuscript writing. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Conflict of interest
All authors have no competing interests to declare that are relevant to the content of this article.
Ethical approval
This research study was conducted retrospectively from data obtained for clinical purposes. The data procured via the Growing Spine Foundation were de-identified.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Jayaswal, D., Kodigudla, M., Kelkar, A. et al. Validation of a patient-specific finite element analysis framework for identification of growing rod-failure regions in early onset scoliosis patients. Spine Deform (2024). https://doi.org/10.1007/s43390-024-00846-7
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s43390-024-00846-7