1 Introduction

The relative composition of an X-ray spectrum in terms of bremsstrahlung and radiation properties depends on anode material, filtration and kV which comprises a diverse region of wavelengths used in fluoroscopy. While the bandwidth is lower, a better image can be obtained with respect to resolution, contrast, noise and beam hardening artifacts [1,2,3]. Recently, numerous studies have led to the emergence of new technologies in the field of medical imaging, and have also incurred a huge cost considering expensive new equipment, which is due to the fact that the improvement of image quality in the diagnosis plays a vital role in diseases treatment via dose reduction [4,5,6,7]. Thus, in order to obtain the best possible X-ray band pass spectrum, the analysis of attenuation and absorption profiles of different elements has an essential position in computing the output intensities using selection of the finest as a filter. At present, X-ray low-band filters are widely used in medical centers [8,9,10,11]. With these filters, the passage of high-energy photons could be prevented without affecting or impairing the image quality [12,13,14,15,16].

Leschka et al. [17] found that applying tin filters in dual-energy CT leads to improved sensitivity and specificity for the differentiation of renal cysts from the improvement of the solid mass in a kidney phantom model. Also, Suntharalingam et al. [18] have shown in their phantom study that unimproved thoracic CT executed at 100 kVp or 150 kVp with tin filtration allows a reduction in the adult radiation dose, but not in children and neonates. Then, they have proposed a tube voltage between 70 and 110 kVp for the tested radiation dose. Here, a tube potential between 100 and 120 kVp is recommended for preschool children. Meanwhile, May et al. [19] have shown that spectral shaping by pre-filtration of tin plates allows for an average dose reduction of 17% without spoiling diagnostic image quality in low-dose parasinus CT with 100 kVp tube voltages. Moreover, Kim et al. [20] have demonstrated that the combination of tungsten, barium, silicon and molybdenum had a 0.3 mm lead tantamount dose and that the particle porosity and packing were 12–22% and 28–36%, respectively. Nonetheless, a satisfactory shield-ability could be achieved with a particle packing of 30% and a porosity > 20%. Consequently, it is a potential replacement for lead sheet and can be taken into account as a suitable medical radiation protection sheet with a fine economic feasibility. This can be achieved with a combination of milliampere (mA)s, kVp, collimation and additional copper filtration. Some recent studies [21,22,23] have shown that image quality remains the same or even better when adding a copper filter accompanied by lower dose. This is one reason to examine image quality and dose in pediatric pelvic tests. To prove that the image quality remains acceptable, it is important to perform a visual and physical assessment [24].

Here, the fundamental problem is that each filtration increases the load on the X-ray tube, and thus decreases its working life. It can be scrutinized impressiveness of band pass filters in enhancing image the quality and reducing the absorbed patient dose and staff, gives the filter is correctly chosen. Hence, this research considers how to compute parameters and how to distinguish the best filter for 100 keV, which is the conventional voltage in general diagnostic radiology [25, 26].

2 Methods

According to Fig. 1a, the X-ray absorption curves were used to settle the correct filter. When we want to employ X-ray band pass filters, low-energy and high-energy spectra are eliminated or attenuated so that a certain work area remains in the spectrum peak, which is the preferred area for medical imaging. In addition, reducing the absorbed dose is another important result of the removal of low-energy spectra. Based on the lead attenuation profiles in Fig. 1b and the investigation of the energies of the elements at the absorption edge (κ) and considering the X-ray spectrum peak, we can basically designate the appropriate elements and then the relevant filters. Therefore, by executing computations, the output X-ray spectra can be specified after the correct filters are applied, and then the improvement of the image specifications and the reduction of the output dose can be examined. The challenging problem of the stage is the selection of the correct filter thickness in terms of availability factors and cost effectiveness. It should be indicated that the absorption of X-ray raises with the third power of the atomic number. Thus, if the absorption edge of the desired element corresponds accurately to the peak energy of our spectrum and field of action, this leads to a strong attenuation of the high-energy photons and a thinning out of the X-ray spectrum besides a reduction of the extra absorbed patient dose through lessening low-energy photons.

Fig. 1
figure 1

a The X-ray source derived from tungsten rotating anode by focal point of 0.6 mm, 120 kVp, 28 mAs, and a 1.7 mm thick copper filter [25] (with permission from the author to publish the photo). b The attenuation profile for lead filter [26] (with permission from the author to publish the photo)

We utilized the X-ray tube with rotating tungsten anode and a focal point of 0.6 mm, 18 mA current, 125 kVp voltage, and a 1.7 mm thick copper filter. In this situation we want to take away photons with energy of more than 90 keV. Base on the attenuation profiles of the various elements, the K absorption edge of lead (Z = 82) is 88 keV. Therefore, choosing lead as a filter in the desired energy range is a positive choice [27, 28]. The attenuation coefficient of lead has been shown in Fig. 1b at the range of 10−3–102 MeV. The thickness, x, needed for the element is given using the following equation regarding the intensity attenuation in the spectrum peak from the Beer–Lambert law (Fig. 1a):

$$I = I_{0} e^{ - \mu x}$$
(1)

In this case, I0 is the initial intensity of the X-ray beam, I is the intensity of the X-ray beam passing through the thickness x of the materials examined, and μ is the linear absorption coefficient of the material under investigation.

The Monte Carlo N Particle version 4C (MCNP4C) was used for the simulation. MCNP4C is a universal Monte Carlo radiation transport code with three dimensional geometry and continuous energy transport of ions and particles. It includes flexible source and output options (tally), interactive graphics (Vised software), and upholds both sequential and multi-processing computer stages. Simulations were performed to obtain the output spectra before and after the assignment of the filters in front of the X-ray tube with different thicknesses by examining the results within the aforesaid energies. The code was run for ten defined volumes with distinct allocated material, and then the density (g/cm3) of each region was calculated. After that, a sphere with a radius of 40 cm was defined at the origin, and the body of the simulation was typified by a rectangular parallelepiped (rpp) of 200 to 900 cm, which is related to the definition of the cube-shaped from the coordinate axis. These cubes were positioned at a distance of 0.1 mm and represented the thickness of the filters utilized in each area. In addition, a 0.1 mm thick lead filter and a thin 0.1 mm thick tungsten target were chosen for the location where the photons pass and collide to be attenuated [29, 30]. The input electron energy was assigned to 100 keV and the direction of the particles incident on the target was in the vertical direction to the Z axis. The simulation was performed for 10 million particles ranging from 5 to 100 keV at an interval of 30.

The surgical C-arm (APELEM-DMS GROUP, APX HF III series, Parc Scientifique Georges Besse, France) was used in experimental method. A C-arm intensifies the image as a scanner. This name derives from the C-shaped arm, which connects the X-ray detector and the X-ray source. The C-arms have radiological capabilities, although they are mainly employed for fluoroscopic intra-operative imaging in surgical, emergency care, and orthopedic protocols. This device gives real-time and high-resolution X-ray images, allowing the clinician to pursue progress and make corrections immediately. The energy applied to the fluoroscopic part was 100 kVp beside a small focus of 0.1 to 3 mA, a constant voltage, and the machine current was raised following each exposure. In a radiographic style, an alpha phantom was located on the holder to measure the contrast, resolution, and quality. Following each exposure, the resolution and quality of the obtained images were examined in automatic mode, and the output dose was read in the third reading sector from the piranha detector, which was located in the palm rest. Then, utilizing a lead sheet thickness of 0.2 mm at a size of 10 cm × 10 cm at the beam exit, the dose ratio, quality of the image, resolution, and contrast were examined. Figure 2 shows the device and setting situations.

Fig. 2
figure 2

a The C-arm apparatus of fluoroscopy used. b The lead plate at the thickness of 0.2 mm by 10 cm × 10 cm size. c The settings of alpha phantom in experiment. d The phantom, piranha, and filter array in experiment

2.1 Piranha detector characteristics and calibration

The internal Piranha detector is very sensitive and can measure the peak tube voltage for as low outputs as 50 kV/0.050 mA at a distance of 50 cm. By calibration in estimation of total filtration, 1–9 mm Al and 60–120 kV range was used and the inaccuracy was ± 10% or ± 0.3 mm as well as ± 15% by 50 mm Al. Here, the obtained resolution was 0.1 or 1 mm by more than 10 μGy/s dose ratio.

After positioning the Piranha detector in front of the image intensifier, the Position Check should be used to confirm the position of the detector. When fluoroscopy is at play, the Piranha system automatically updates the display using the highest possible sensitivity.

On the other hand, the external Dose Probe is normally used to measure the lowest possible dose rate levels down to 0.1 µGy/s. Another reason for using the external Dose Probe is that the detector is much smaller than the Piranha, which facilitates the positioning in front of the image intensifier without affecting the measuring field for the mA feedback loop. If the image intensifier can manually control the mA and kV, then it can be used for measurement down to about 0.7 µGy/s. As a secondary parameter the total dose is accumulated, and after switching off the fluoroscopy, this value is used to calculate the average dose rate as the total dose divided by the exposure time. Figure 3 shows the physical specifications of the detector.

Fig. 3
figure 3

Physical dimensions of Piranha detector

The inaccuracy is here defined as the root of the square sum of systematic errors, which has not been eliminated, and random errors (dispersion around a mean value). The calculation of the inaccuracy is based on 15 different measurements with a confidence level of 95%. Of the total inaccuracy, random error is 20% and general inaccuracy is 80%. As a reference condition to the IEC61674 standard and inside the Piranha top panel, the calibration is done with a field size typically 5 mm less than the size of the top panel.

The standard calibration for the Piranha External Dose Probe is W/23 mm Al. This calibration was chosen because the main use of the detector is to measure the dose by the image intensifier after the phantom study. However, it can just be used for measurements of skin dose as well. The detector reacts very linearly to energy response and is not affected by different filtrations.

The dose rate was 4 nGy/s to 150 mGy/s with ± 5% or ± 1 nGy/s inaccuracy, and a resolution of ± 200 pGy/s by free-run or timed. Also, the irradiation time range was 0.1 ms to 34,000 s with ± 1% or ± 0.5 ms inaccuracy and 0.5 ms resolution.

2.2 Phantom study

The alpha phantom has been designed to test relative positions of the X-ray field, light beam diaphragm and vertical X-ray beam alignment, all in one exposure. According to Fig. 2c, it has 3 main sections: 1. Main section of the Alpha phantom for testing synchronicity of the radiation field to that of the Light Beam Diaphragm. 2. Center Tube fixes to the Alpha phantom to check the accuracy of the Vertical Beam Alignment. 3. Bucky Wall Stand Holder is an adjustable hanger that supports the alpha phantom on the face of a vertical Bucky.

By using the phantom, since the external dose probe (Piranha Dose Probe) is not sensitive to back scatter, a lower value compared to a transmission ion chamber is typically detected (typically in the range of 5–20%). Moreover, the beam correction factor can be used to make automatic corrections. This correction factor may also be stored permanently in a Favorite for easy access.

Here according to Fig. 2d, the radiation detector was placed in the middle of the imaged area and in the middle of the phantom field to obtain the phantom image and to investigate the filter impact. The image quality was appraised with a fairly standard phantom in high and low contrast resolution to compute the patient skin entrance dose.

3 Outcomes

Figure 4 shows the achieved X-ray spectra with and without utilizing 0.1 mm lead filter via attenuation formula and initial flux. For energies greater than 90 keV and to increase the attenuation, a thicker filter like 0.2 mm may be used, but this leads to a significant decrease in the intensity of the useful radiation. In order to approximate the thickness effect, the thickness of the lead was assumed to be 0.2 mm. Therefore, by re-computing eμx for different energies, the output intensities in the range of 45–120 keV can be resolved. If one compares the intensities for the case with the 0.2 mm thick lead filter and the case with 0.1 mm, it is resolved with increasing thickness that the damping ratio for high energies increases. For instance, the output intensity ratio (OIR) of 0.2 mm filter to 0.1 mm filter will be 0.75 and 0.48 in 80 keV and 95 keV, respectively from:

Fig. 4
figure 4

Comparison of two X-ray spectra before (top) and after (bottom) filter assignment

$${\text{OIR}}_{\text{lead}} = \left( {\frac{{I_{0.2} }}{{I_{0.1} }}} \right)_{E}$$
(2)

This means that as the thickness of the filter increases, the attenuation increases more at high energies than at low energies. It should be noted that increasing the thickness would decrease the intensity of the useful energy. We also found that, in accordance with the calculations, strong attenuation occurs at high and low energies immediately after the lead absorption peak by utilizing the filter at high energies.

By the MCNP4C code, the general geometry of the X-ray tube, the position of the filters in different thickness, and the photons and electrons generated by the characteristic and bremsstrahlung radiations were simulated for 10 million particles. The total simulation error was less than 1.5%, which is an indication of accurate results. The outcomes given in Table 1 demonstrate the particles counted by 0.1 to 0.7 mm thick per energy after assignment of the lead filter.

Table 1 Number of simulated output photons per starting particle at diverse thickness of the lead filter

Figure 5 shows the output X-ray spectra after assignment of the lead filter in diverse thicknesses. The outcomes showed that the desired results are achieved by setting the lead filter to a thickness of 0.1 mm and dismissing high and low energy photons. In the meantime, the useful part of the spectrum was retained and agreed with the lead absorption edge curve at 75 keV. Subsequently, a strong decrease of the particle numbers was followed. Meanwhile, increasing the filter thickness leads to more attenuation, that is undesirable to produce an image. Due to the consideration of the pure tungsten target and the lack of accurate information about the filters of the intrinsic tubes, there was no match between the X-ray output curves and the computed curves. Lastly, after executing the simulations under these circumstances, the desired results were obtained at a thickness of 0.1 mm, which was consistent with the computation outputs. The contrast and quality of the image were low due to the utilizing high voltage, and the absorbed dose was also on a large scale. Although the resolution of the image remained unchanged by using of the lead filter, the quality of the image and contrast improved significantly, and the absorbed dose decreased considerably, so that the lead filter with thickness of 0.2 mm lost its effects. Therefore, the thickness of the lead filter was raised to 0.4 mm to enhance the quality and lessen the absorbed dose. The use of 0.6 mm of the lead filter reduced the resolution, contrast, and image quality. In the other phase, 0.2 mm tinplate and 0.2 mm lead plate were employed, which did not influence the quality and contrast but significantly lessened the patient’s absorbed dose. By calculations, the selection of a 0.2 mm lead filter in this energy range was mathematically the right choice. So in order to get a high quality image, besides raising the mAs, the lead filter should be thicker. Figure 6 illustrates the acquired images of an alpha phantom with and without lead filter having a thickness of 0.2 mm at a voltage of 100 keV and 0.2, 1.2 and 2.7 mA. If the filter was not used, the image became whiter by increasing the current. For this energy range used, the HVL for the lead plates with a thickness of 0.2 mm and 0.4 mm corresponded to that of aluminum with a thickness of 10.1 mm and 11.3 mm, respectively. Table 2 demonstrates the dose ratio and the image resolution before and after the assignment of the tin filter. Without a filter, raising the mAs had not influence on the resolution, but on the image quality. Nevertheless the resolution preserved constant and the image quality enhanced well, the absorbed dose was significantly decreased. In addition, Fig. 7 shows the absorbed dose ratio in terms of intensity at 100 keV energy. While the tin filter was used, the dose ratio acquired was markedly diminished by 50 and 250 μGy/s for 0.5 and 3 mA, correspondingly.

Fig. 5
figure 5

The obtained X-ray spectra utilizing lead plates at diverse thicknesses

Fig. 6
figure 6

a The obtained alpha phantom image without using filter by 100 keV and 0.2 mA (Left). The obtained image using 0.2 mm lead plate (Right). b The obtained alpha phantom image without using filter by 100 keV and 1.2 mA (Left). The obtained image using 0.2 mm lead plate (Right). c The obtained alpha phantom image without using filter by 100 keV and 2.7 mA (Left). The obtained image using 0.2 mm lead plate (Right)

Table 2 Outcomes experimented using and without using tin filter
Fig. 7
figure 7

The acquired absorbed dose ratio in terms of intensity by 100 keV without and with the use of tin plate

Haga et al. [31] have used three copper attenuators by different thickness of 0.5, 1.5 and 3 mm in a rotatable panel to evaluate the image performance in a quality control test. They discovered the spatial resolution in the thin attenuation setting (0.5 mm Cu) was higher than that in the 3.0 mm Cu. In interventional radiology with several fluoroscopes, the stored fluoro loops are only saved after multiple activations of the fluoro pedal to achieve steady state dosimetric parameters (kVp, mA, spectral filter). Therefore, a method often is described as a “double clutch” so that a possible spectral filter or other parameters may change as needed due to increased or different phantom loading. Goode et al. [32] have been utilized 0.3 mm Cu spectral filter in a normal fluoro dose level and to reach a low dose setting with high beam quality, the copper spectral filter increased 0.6 mm besides lessening 5 mGy/min dose rate and growing kVp by 3. They indicated that the signal to noise ratio (SNR) is increased from normal to low dose level for a thickened phantom. Furthermore, Ngaile et al. [33] have reported that 0.35 mm thickness of copper can be used in fluoroscopy without significant impact on the image quality, and using 0.35 mm Cu plus 1 mm Al additional filtration preserves the contrast by 71% reduction in surface air Kerma.

The limitations of our research are that only one supplier of fluoroscopes and a small selection of models were available for these tests. Future work will include at least two more fluoroscopy suppliers and several other models to evaluate the filtration influence on image quality. Through a performance test with a simple patient-equivalent phantom, the quality control criteria can be further examined to design a distinct fluoroscope spectrum via diverse or complex filters.

4 Conclusions

In fact, extra filtration has the potential to change the X-ray spectrum, which often brings benefits in terms of patient dose and image quality. It is recommended to use the band pass filter to obtain high quality images, which should be determined by the attenuation profiles of the different materials. The work area in radiology, the X-ray spectrum peak, and the absorption edge are the most important factors in filter selection. For best outcomes, increasing the filter thickness also requires raising the mAs or exposure time. Because the spectrum peak is at 88 keV, utilizing the lead filter creates a narrow spectrum to enhance the image quality and decrease the absorbed patient dose. As the increase in thickness results in quality degradation, the mAs are raised as much as possible to obtain a high quality image. The results of experiments, simulations and calculations showed that lead submission at 100 kVp can lessen the tube output dose and a boost in mA can enhance the image quality while the dose is still lower than that of the unfiltered instance. By virtue of the outcomes, the 0.1 mm thick lead filter was preferred as the optimal filter.