Introduction

Respiratory diseases such as asthma and chronic obstructive pulmonary disease (COPD) are routinely managed using inhaled bronchodilators and corticosteroids [1, 2]. However, the medication must reach the receptors in the lower airways to exert its effect. To ensure this happens, correct use of the inhaler device is critical for optimal drug delivery [3].

Patient-related factors such as poor adherence and inhaler handling errors have a negative impact on asthma control [4]. Studies have shown that poor inhaler technique is common and inhaler use is often highly suboptimal [5]. Estimates of improper inhaler use range from 20% to 82% of patients [4, 69]. There is increasing evidence in support of the claim that correct inhaler technique is fundamental for effective therapy and asthma management [5, 10, 11]. Errors in the use of inhalers have been shown to have serious effects in terms of the management of the disease [6, 8, 9]. Poor inhaler technique in patients with asthma is associated with increased hospitalization, more emergency department visits, increased use of oral corticosteroids and antimicrobials, and poorer asthma control [6, 8, 9]. Some errors are device-independent and include not breathing out before actuation of the device [12].

Although correct inhaler technique involves some common steps for all devices, the optimal inhalation pattern differs between devices [3, 13]. Inhalers that are perceived as difficult to use are usually associated with poor inhalation technique and low treatment adherence, leading to worsening in asthma control [14]. As such, the technical characteristics of an inhaler device can determine how well it is handled, and consequently, how often it is correctly used. An inhaler that is easy to use, intuitive, and preferred by the patient could ultimately improve adherence [1416].

In patients with inadequately controlled asthma, the combination therapy involving an inhaled corticosteroid (ICS) and a long-acting β2 agonist (LABA) is a recommended therapy [17]. The fixed-dose combination of the ICS/LABA, budesonide/formoterol (BF), has proven to be an effective therapy by improving pulmonary function and asthma control [1, 2] in patients with asthma and COPD. DuoResp® (BF) Spiromax® (Teva Pharmaceuticals) is a BF dry powder inhaler designed to provide dose equivalence with enhanced user-friendliness compared to Symbicort® (BF) Turbuhaler® (AstraZeneca) [18]. Both inhalers are approved in Europe for the treatment of patients with asthma and COPD (for whom an ICS/LABA combination is indicated).

To determine which of these two inhalers patients might find easiest to use correctly, it is necessary to establish objective criteria by which these inhaler devices are compared with regards to their ease-of-use and potential handling errors. In this study, practicing clinicians with a specialty in device handling in asthma or COPD were asked to identify potential handling errors for the Spiromax and the Turbuhaler device. Subsequently, a Delphi process was used to reach consensus on the errors most likely to be critical for effective drug delivery. A Delphi process was chosen because it is an established tool for reaching a consensus on a complex problem in a highly objective way [1921]. In the Delphi process, a peer group of experts work together anonymously to reach consensus on a complex problem as objectively as possible.

The aims of this Delphi process study were to independently establish what is considered an error (for optimal drug delivery) when handling empty versions of DuoResp Spiromax or Symbicort Turbuhaler dry powder inhalers and to assign a level of severity to each of these errors.

Methods

Study Procedures

This study was conducted by InterPhase Consult, an independent strategic consulting firm, on behalf of Teva Pharmaceuticals Inc. InterPhase Consult ensured that the methodology and conduct of the study were free from bias and Teva Pharmaceuticals had no influence on the choice of advisors, design, and administration of the Round 1–4 questionnaires, or the data analysis.

This Delphi process was initiated with selection of the advisors and development of the questionnaires, followed by four rounds of the Delphi process and then data analysis (Fig. 1). Per Office for Human Research Protections (OHRP) approved assurance created under 45 CFR 46.103, the protocol was exempt from informed written consent because (1) the research presented no more than minimal risk of harm to subjects, (2) the exemption did not adversely affect the rights and welfare of the subjects, (3) the research could not practicably be carried out without the exemption and (4) whenever appropriate, the subjects were provided with additional pertinent information after participation. (http://www.hhs.gov/ohrp/humansubjects/guidance/45cfr46.html).

Fig. 1
figure 1

Outline of the study processes

Advisor Selection

Fifteen European advisors were identified. Advisors were invited to participate until 10 had accepted, per the recommendations of Delbecq et al. (1975) for panel selection when performing a Delphi methodology study [22]. Please refer to the Appendix in the supplementary material for additional details on the advisors selection.

Delphi Process (Rounds 1–4)

Using the Delphi process, advisors were asked via questionnaires to agree on a list of potential inhaler usage errors and then rate them according to the perceived impact each error would have on effective drug delivery. Advisors remained anonymous to each other until the teleconferences. Questionnaires were developed before the study by InterPhase analysts, but the wording could be adapted during the course of the study, if necessary, to ensure instructions to the advisors were clear and the Delphi process was able to achieve consensus. Advisors responded to the questionnaires online using the website SurveyMonkey® (Palo Alto, CA, USA). The Delphi process aimed to achieve consensus (see Data Analysis below) on ≥90% of all errors.

Round 1

Participating advisors were sent empty Spiromax and Turbuhaler devices along with the respective patient instruction leaflets. Once they were familiar with the devices and their instructions, the advisors were asked to list all of the potential handling errors they could think of for each device. Handling errors were defined as any way in which the inhaler could be used incorrectly during the inhalation process. The advisors were randomly divided into two groups; one group reviewed the Spiromax device first and the other reviewed the Turbuhaler device first. The resultant lists from the 10 advisors were consolidated by InterPhase Consult into one list of errors for each device.

Round 2

Advisors received the consolidated list of errors (ordered randomly) and were asked to score each error with respect to its severity using an integer scale of 0–10, (0, least critical; 10, most critical). The severity levels were then separated into three categories: ‘error', ‘potentially critical', and ‘critical' as shown in Fig. 2. Based on the advisors’ expert opinion with regards to the impact on treatment outcomes and/or inhaler device functionality, ‘error' indicated an error that would have minor or no impact, ‘potentially critical' indicated an error would have a mild-to-moderately negative impact, and ‘critical' indicated an error would have a negative impact. An error could be removed from the list if seven or more advisors provided approval. Advisors were also asked to give their reasoning behind the score they assigned for each error.

Fig. 2
figure 2

Integer scale used to rate error severity

Round 3

Advisors were given an individualized report showing their responses relative to the consolidated group severity scores and were given the opportunity to re-evaluate their scores and revise them if they wished. If a consensus was not achieved for ≥90% of errors for each device, a fourth round was to be conducted.

Round 4

Advisors were provided with a list of all errors grouped within their corresponding categories as determined by the median score from Round 3. They were then asked to agree or disagree with the error category for each error based on the median error severity ranking from Round 3. The outcome from Round 3 was accepted as consensus group opinion if ≥90% of the advisors agreed in Round 4.

Post-hoc Study Amendment

In a post hoc study amendment, advisors were asked to participate in a series of teleconferences to discuss and refine the output from the Delphi process. A summary of the consensus from the calls was subsequently approved by the participating advisors.

Data Analysis

Scores obtained for each error in Rounds 2 and 3 were analyzed using descriptive statistics and median scores were calculated using Microsoft® Excel 2010® software (Microsoft, Redmond, USA). Errors were ranked according to median scores within each of the three severity categories. Interquartile ranges (IQRs) were calculated to give a measure of dispersion and, therefore, consensus. Consensus for each error was deemed to have been reached when the IQR was ≤2, or if ≥80% of advisors put the error into the same severity category.

Results

Conduct of the Study

Ten advisors participated in this study (Table S1). Full consensus was not reached for ≥90% of the errors after Round 3; a fourth round was conducted. Some changes were made to the wording and design of the questionnaire following Round 1 to emphasize the intended meaning and definition of a handling error.

Identification of Errors Using the Delphi Process (Rounds 1–4)

Round 1

A total of 29 potential errors were identified for Spiromax in Round 1 and 31 were identified for Turbuhaler (data on file, Teva Pharmaceuticals and InterPhase Consult).

Rounds 2 and 3

None of the errors was deemed to be invalid by seven or more advisors in Round 2 and all were retained.

Median severity scores were assigned to each of the errors and did not change significantly between Rounds 2 and 3 (Table 1). A critical severity score was given on fewer occasions for errors in using Spiromax compared with Turbuhaler.

Table 1 Final error categorization and Round 2 and 3 severity scoresa for (A) Spiromax and (B) Turbuhaler

The errors considered by the advisors to be most critical (median severity score of ≥9) when using Turbuhaler were ‘not inhaling through the mouthpiece (e.g., nose)’, ‘difficulty/failure to remove the cap prior to usage’, ‘forgetting to load the device prior to each dosage’, and ‘exhaling into the device’. The error ‘forgetting to load the device prior to each dosage’ was assigned to the critical error category by all ten advisors; nine advisors put ‘exhaling into the device’, ‘not inhaling through the mouthpiece (e.g., nose)’, ‘red grip is only rotated fully in one direction prior to using the device’, and ‘holding the mouthpiece instead of the barrel when twisting the red grip’ in the critical error category.

For Spiromax, the errors considered to be most critical were ‘not inhaling through the mouthpiece (e.g., nose)’ and ‘failure to read the dose indicator correctly, potentially leading to using an empty device’. None of the errors were classified as being critical by all 10 advisors; eight advisors classed ‘not inhaling through the mouthpiece (e.g., nose)’ as a critical error.

After Round 3, consensus was reached for 38% of the errors for Spiromax and 68% of the errors for Turbuhaler.

Round 4

In Round 4, the median severity scores from Round 3 were accepted as the consensus for an additional nine errors for Spiromax and eight errors for Turbuhaler, resulting in 69% consensus for Spiromax and 94% for Turbuhaler.

Post-hoc Review and Analysis

A post-hoc review (by InterPhase Consult and Teva Pharmaceuticals) of the errors identified for Spiromax and Turbuhaler found some anomalies, such as: handling errors that had been scored differently for the two devices, but which may reasonably have been expected to be scored consistently; handling errors that had been scored consistently, but which may reasonably have been expected to have been scored differently; and handling errors that may be invalid based on the definition of a handling error. Following this review, all advisors were invited to attend a teleconference to discuss the results. Six advisors agreed to participate and two teleconferences were held to discuss the output of the Delphi process and the anomalies, and to give the advisors an opportunity to revise their scores. After the teleconferences, a summary of the consensus from the calls was sent to the participating advisors for their approval.

A total of 16 errors were identified as being anomalous and were discussed. After the calls, four of these retained their original classification and 12 were reclassified. For Spiromax, six errors were removed, as they were not thought to be valid handling errors for the device, and two potentially critical errors were reclassified as non-critical errors (not holding the breath for long enough following inhalation; device not held upright during loading/inhalation). For Turbuhaler, five errors were removed, as they were not thought to be valid handling errors for the device, and two errors that had originally only been identified for Spiromax were added (not holding the breath for long enough following inhalation; unable to/forgetting to breathe out fully prior to inhalation).

Final Outcome

Following the teleconferences, the number of potential handling errors was finalized as 22 for Spiromax and 27 for Turbuhaler (Table 1; Fig. 3). Based on the group median scores from Round 3 and the subsequent error categorization during the teleconferences, the 22 potential handling errors for Spiromax comprised four critical errors, 12 potentially critical errors, and six non-critical errors (Table 1; Fig. 3). In comparison, the 27 potential handling errors for Turbuhaler comprised nine critical errors, 14 potentially critical errors, and four non-critical errors (Table 1; Fig. 3).

Fig. 3
figure 3

Number and severity of identified potential error for Spiromax and Turbuhaler (final results, following Round 4)

Discussion

Despite the availability of a wide variety of medications, a significant portion of patients with asthma and COPD remain uncontrolled [1416]. The correct use of an inhaler device is crucial for optimal drug delivery. As such, the inhaler device itself plays an important and active role in asthma management. It is therefore important to have devices which are intuitive to use and to train patients.

Using the Delphi process, advisors participating in this study independently identified potential handling errors for Spiromax and Turbuhaler and ranked them according to their severity. Spiromax was associated with fewer errors overall (22 vs. 27) and fewer errors that were classified as critical (four vs. nine) or potentially critical (12 vs. 14) than Turbuhaler. This suggests that there may be less potential for critical handling errors when using the Spiromax device compared with Turbuhaler.

Findings from this study are in agreement with outcomes reported in two independent device mastery studies among healthy Finnish volunteers and healthcare practitioners (HCPs) in Australia. In healthy adult Finnish volunteers [23] and Australian HCPs [24], Spiromax was associated with higher levels of device mastery and fewer errors by intuitive use/no instructions and after reading the patient information leaflet compared with Turbuhaler.

Training patients in inhaler use is a core component of good clinical practice [25]. Consequently, providing HCPs with an inhaler that is less prone to device handling errors and easy to use and teach patients could potentially reduce the number of device handling errors and ultimately improve asthma control. Results from this study highlight specific errors that could potentially be committed using Spiromax or Turbuhaler. Two errors for Turbuhaler received the maximum median severity score of 10; these were ‘not inhaling through the mouthpiece (e.g., nose)' and ‘difficulty/failure to remove the cap prior to usage'. A number of errors regarding priming of the device were classified as critical for Turbuhaler (which were not identified for Spiromax). These included, ‘forgetting to load the device prior to each dosage', ‘failing to fully rotate the red grip', ‘failure to read the dose indicator correctly, potentially leading to using an empty device', and ‘red grip is only rotated fully in one direction prior to using the device'. The highest median severity score for a Spiromax error was 9.5 for ‘not inhaling through the mouthpiece (e.g., nose)'. Thus, HCPs should focus on these potential errors when training patients to use Spiromax or Turbuhaler.

Although training in inhaler technique is important and assessing technique is recommended at every visit [25], in everyday life this might not prove practical or possible. Additionally, training alone might not be sufficient to ensure proper technique mastery. Some patients fail to achieve correct inhaler technique despite repetitive training [3]. The patient’s opinion on a particular device, psychosocial factors such as preference and willingness to use a particular device, and the device choice are also important aspects to consider [26]. The ability to use the inhaler correctly can influence adherence [3]. An intuitive device that is easy to use and preferred by the patient might also improve asthma control by minimizing the risk of handling error committed by the patient and improving adherence. However, it is important to note that a device that is shown to potentially be associated with fewer errors in an experimental setting does not necessarily translate to less potential errors and better inhalation technique in real life. Also, a device with less potential errors is not automatically the preferred device by the patient. Real-life scenarios can be different and several patient-related factors such as comorbidities, dexterity, familiarity with inhaler devices, and other lifestyle factors may play a role [8, 27]. Matching the correct patient with the correct inhaler is now thought to be essential when prescribing an inhaler that a patient can and will use correctly at every use [28, 29].

The Delphi process methodology was deemed most suitable for this study because it is an established tool for reaching consensus on a complex problem as objectively as possible. It has been proven to be effective for reaching a consensus when factual data do not exist or cannot easily be obtained, or when the problem is so subjective that an individual may give an opinion that conforms to those of other participants rather than maintaining their own critical opinion [19, 30]. Among the strengths of this study are the key characteristics of the Delphi process: the involvement of an expert panel, the anonymity of the advisors and their responses to other participants, the use of a series of questionnaires administered in rounds, the summarization of information from each round via a moderator and sharing of anonymized group opinion, the progressive convergence of opinions with administration of each questionnaire, and quantification of the extent of agreement using simple statistics.

This study has a number of limitations, one of which was the requirement for the teleconferences after Round 4 of the Delphi process to finalize the lists of potential handling errors. It was felt necessary to conduct the teleconferences for a number of reasons: there was < 90% consensus for the errors for Spiromax; some of the listed errors were not thought to be valid handling errors while others were thought to unfairly disadvantage one device over another; and some errors were listed for one device that should be applicable to both. Examples of invalid handling errors that were removed include: ‘washing the device using liquids/getting the device wet', ‘chewing the mouthpiece of the device', ‘information leaflet inaccurate/difficult to understand', and ‘attempting to fit a spacer device'. The need for the teleconferences may be interpreted as interference in the Delphi process; however, neither device was unfairly prejudiced by the teleconferences—both devices had some errors removed by the advisors. The aim of the teleconferences was to ensure the final error lists were accurate and representative of handling characteristics of the devices in the real world.

Other limitations include that the respiratory device experts chosen as advisors may not have been able to identify all of the handling errors that may occur in daily usage of the devices owing to their inherent expertise. As such, the findings from this study rely on the judgment of a select group who are experts in the field and may not be representative of the wider healthcare community. Additionally, the Delphi process employs a consensus approach. Extreme positions are usually eliminated and the outcome may be a diluted version of the overall expressed opinions.

Conclusions

Through the Delphi process, experts in respiratory devices independently identified and reached consensus on potential handling errors for Spiromax and Turbuhaler and their severity. ‘Not inhaling through the mouthpiece (e.g., nose)' was one of the most highly rated critical errors for both devices. Critical errors in priming the device were identified for Turbuhaler, but not for Spiromax, as might be expected given the difference in operation of the two devices. There were fewer errors identified overall for Spiromax compared with Turbuhaler, and fewer errors with Spiromax were classified as critical or potentially critical than with Turbuhaler. Altogether, this indicates that there may be less potential for critical errors with Spiromax than Turbuhaler.