Introduction

Due to the highly reactive nature of titanium, the synthesis of NiTi shape memory alloy (also known as Nitinol) requires non-contaminating processing methods. The gathering of C, N, and O on melting causes the formation of inclusions such as TiC, TiO2, and Ti2Ni x O y on bulk level and hot-working oxidation may leave distributed inclusions at surface level of semi-finished products. A recent observation, the surface particulates of NiTi and in biomedical devices, is reported by Shabalovskaya et al. [1]. The inclusions, and their sizes, could affect both the NiTi functional properties and the processing aspects such as the workability of low dimension semi-finished products. Nowadays, the role and the amount of inclusions affecting specific properties such as the fatigue life, biocompatibility, or corrosion resistance of devices made in NiTi are not univocally explained. Regarding TiC inclusions, which are found typically in VIM processed Nitinol due to carbon crucibles employed in this kind of melting plant, their presence is particularly detrimental to pitting corrosion resistance, as asserted in Kimura and Sohmura [2], Shabalovskaya et al. [3]. Also, regarding oxide-type inclusions (i.e., Ti2NiO x ) Liang and Huang [4] have found that their presence is the main reason for pitting. Coda et al. [5] characterized the main types of inclusions of commercial Nitinol produced by VIM and vacuum arc melting (VAR) routes; by means of extreme value statistics, Urbano et al. [6] showed that the maximum inclusion dimension in the highly strained volume of a fatigue test contributes to a decrease in the fatigue performance of 0.3-mm-diameter superelastic wires. Morgan et al. [7] reported about extra low inclusions content (ELI) alloys appeared to have improved fatigue performances. Rahim et al. [8] investigated the effect of C and O impurity levels on fatigue lives, of pseudoelastic NiTi, reporting that the fatigue resistance is affected mainly by surface aspects and that lower C and O contents will not result in greater fatigue lives. However, cleaner processing decreases the probability for fatigue crack initiation due to inclusions, but it cannot suppress it completely. It is a matter of fact that non-contaminating melting route should be used for the production of high-quality NiTi alloys for biomedical applications, according to the ASTM standard specification F 2063-12 [9] that limits both the impurity contents as well as the amount of particulates. Among non-contaminating melting methods, PAM represents one of the most important achievements in the field of high-quality alloy production over the last three decades. The elevated working pressure of PAM prevents the selective evaporation of the high-vapor pressure elements avoiding chemical alterations of final alloys composition. Moreover, inert operating conditions of PAM, in conjunction with the transfer of a high energy density, permit alloying of a great number of metallic elements. For over 20 years, industrial PAM furnaces have been employed for the production of reactive, high melting points, refractory metals, and superalloys, as reported in [1015]. Thus, PAM seems particularly suitable for melting “cleaner,” near equiatomic NiTi shape memory alloys (SMA), not only on laboratory scale NiTi-based button preparations [16, 17], but also for Nitinol industrial production.

The aim of this work is the evaluation of the corrosion resistance of a Ni50.8Ti49.2 alloy produced by a pilot scale PAM furnace in comparison to one prepared through a conventional VIM melting technique by using the same starting Ni and Ti metals. Microstructures of PAM and VIM NiTi alloys were investigated by FEG-SEM, and the corrosion resistance was determined by potentiodynamic as well as potentiostatic tests in a simulated body fluid (SBF) solution.

Experimental

NiTi Melting and Processing

Commercially pure Ti rods, 3 mm in diameter (ASTM grade 1—Titania srl) with typical impurities content of N 100; C 200; H 26; Fe 500; O 600 ppm; and electrolytic Ni (Falconbridge ™ >99.99 %: Cu, Co, Fe, Pb, S, Zn <5 ppm) were chosen as starting metals. It is worth noting that the ASTM F67 standard specification for unalloyed titanium for surgical implant accepts greater oxygen (up to 1800 ppm) and carbon (up to 800 ppm) contents for grade 1 Ti. So the chosen starting metals assure a detectable amount of endogenous forced inclusions into the NiTi alloys produced by PAM and VIM plants of this work. A Ni-rich nominal composition of Ni50.8Ti49.2 (at.%) was selected for guaranteeing the austenitic state at body temperature, required for developing superelastic medical devices.

PAM NiTi was produced by a laboratory scale plasma arc furnace (Leybold—Plasma Lab 3000 ™) equipped with a transferred plasma torch of 150 kW (PEC Plasma Electric Company®). Feeding materials were drip melted into a cylindrical (60 mm dia.) water-cooled, copper crucible, under pure He (purity >99,995) atmosphere. An ingot of 400 mm in length was processed by continuous casting. The PAM ingot was then re-melted 5 times by drip melting for increasing compositional homogeneity.

As comparison, a low-purity NiTi ingot of the same Ni50.8Ti49.2 (at.%) nominal composition was melted in a VIM furnace (Balzers VSG10 ™) in a dense graphite crucible using the same starting Ni and Ti metals. Then, the VIM material was cast in a water-cooled copper mold. Details of both PAM and VIM laboratory plants employed for small productions were reported elsewhere [18, 19]. After melting, the two PAM and VIM NiTi ingots were hot forged at 900 °C in air to rods 20 mm in diameter. Then the rods were fully annealed at 900 °C for 2 h and water quenched.

Material Characterization Techniques

The non-metallic C, O, and N impurities of the fully annealed produced NiTi rods were measured with LECO analyzers. Transformation temperatures of PAM and VIM NiTi specimens taken from fully annealed NiTi rods were checked by a differential scanning calorimeter (DSC—TA Instruments Q100™) at heating/cooling rate of 10 °C min−1. Cross and longitudinal sections of the PAM and VIM rods were cut with a metallographic saw and then were prepared with standard metallographic polishing to 1 μm diamond cloth. Scanning electron microscope investigation was conducted using a FEG-SEM (Hitachi SU-70™) equipped with SE, BSE and EDS detectors.

Corrosion Resistance Analysis

For the corrosion resistance tests, cylindrical specimens 15 mm in dia. and 6 mm high were machined from the PAM and VIM rods. Due to the well-known effect of surface roughness on the corrosion resistance, the surfaces of the samples were prepared using emery paper mechanical polishing (MP), diamond clothes polishing (1 μm MP) as well as electro-polishing (EP) in a perchloric acid 10 % butylene glycol 90 % solution at applied voltage of 40 V for 120 s. Three surface finish conditions were investigated:

  1. (a)

    #600 Grit emery paper (#600 MP).

  2. (b)

    1 μm diamond clothes polishing (1 μm MP).

  3. (c)

    1 μm diamond clothes + electro-polishing (1 μm EP).

All the corrosion tests were performed in a SBF solution with the following composition: NaCl 8.44 g/L, NaHCO3 0.35 g/L, NaHPO4 0.06 g/L, and NaH2PO4 0.06 g/L. The solution was maintained at constant temperature of 37 °C and protected by an extremely high purity nitrogen gas flow.

Cycling potentiodynamic tests were performed using the experimental apparatus suggested in ASTM G5 Standard practice [20]. A specimen holder for flat specimens according to ASTM G61 Standard practice [21] was used (exposed area 1 cm2). The specimen was immersed in the test solution for 1 h before potentiodynamic scanning. Initial potential was 600 mV versus saturated calomel electrode (SCE), and scan rate was 600 mV/h. Reversal of the scanning direction was done when the anodic current density reached 100μA/cm2. The reference electrode was a home-made Ag–AgCl electrode, and the potential was then expressed as SCE values. The tests were re-done at least in triplicate.

Potentiostatic scratch tests were performed also using the same electrode assembly for flat specimen as described above. In the scratch method originally proposed by Pessal and Liu [22], the metal surface is slightly damaged using a diamond tip while the potential is maintained constant. If the specimen reaches the passivation state under the applied potential value, the scratching test is repeated by applying an increased (in 10 mV steps) potential value so that the potential at which the material does not re-passivate can be determined.

Results and Discussion

Non-metallic Impurities and DSC

Oxygen, carbon, and nitrogen contents were analyzed in different regions of the fully annealed rods used for corrosion resistance tests. C, O, and N average values of five measurements for each non-metallic impurity are summarized in Table 1.

Table 1 Average O, C, and N contents of PAM and VIM NiTi rods

As expected, lower non-metallic impurity contents are found in PAM NiTi. These values are consistent with the impurity level of the starting Ti and confirm the extraordinary non-contaminating characteristic of PAM melting route. On the contrary, VIM NiTi shows greater C and O contents than commercially available Nitinol resulting in an elevated impurities gathering effects. That is due to some unavoidable features related to the VIM melting method: (i) a direct pickup of C by liquid alloy contained in the graphite crucible; (ii) the reaction between the graphite crucible and the surrounding Al2O3-based concrete used as thermal insulation and crucible holder that can produce CO and CO2; these products could be accumulated into the alloy; (iii) the residual humidity of both the graphite crucible and the insulating concrete that can be absorbed by the reactive Ti upon heating and by liquid NiTi during soaking and casting times. For both the alloys, the total N content is negligible.

It is worth noting that the detected impurity values of VIM NiTi exceed the limits of the active standard ASTM F2063 that, for biomedical applications, reasonably, restricts the amount of both carbon and oxygen + nitrogen impurities of wrought NiTi up to 500 ppm [9]. In any case for this work, suitable standard purity PAM and a low purity VIM NiTi alloys were realized for studying the effect of different inclusions levels on Nitinol corrosion resistance characteristics.

DSC results, of specimens taken from fully annealed rods of PAM and VIM materials, show typical reversible martensitic transformation curves with austenite finish temperature, Af, below 0 °C (see Fig. 1). The broader DSC peaks of PAM alloy compare to those of the VIM NiTi are indicative of a minor compositional homogeneity of the NiTi alloy. This is due to the laboratory scale, single torch, PAM furnace used in this work that cannot guarantee the same chemical homogeneity of VIM melting alloys. The latter is quite good thanks to the magnetic stirring effect during VIM melting. Similar homogeneity degree can be reached by multiple torches PAM furnaces, as employed in the industrial production of Ti alloys.

Fig. 1
figure 1

DSC curves (heating/cooling rate 10 °C/min) of PAM and VIM Ni50.8Ti49.2 fully annealed specimens

Microstructure and Particles Statistical Analysis

The nature, distribution, and size of particulates and secondary phases in PAM and VIM alloys were analyzed through SEM observation with backscattered electron detector. Selected images are reported in Fig. 2. PAM NiTi revealed two populations of secondary phases with different sizes. The predominant phase, of very small size, was localized at grain boundaries. EDS analyses revealed Ti/Ni ratios of 2:1. The oxygen content was not easily assessed by EDX, but varied from one particle to another. Hence, Ti2Ni intermetallic phase and Ti2NixOy oxides comprise this population. The same compositional results were associated with the larger inclusions distributed throughout the NiTi matrix, with dimension <5 μm, as indicated by the arrows of Fig. 2a. EDX microanalysis results of particulates (see Fig. 4a) do not show any C signal according to the low C content measured in PAM alloy reported in Table 1.

Fig. 2
figure 2

SEM micrographs of PAM (a) and VIM (b) of Ni50.8Ti49.2 rods after hot rolling and fully annealing. The arrows show typical inclusions

A different particulates distribution and type of inclusions were observed in the VIM alloy as depicted in Fig. 2b. A uniform distribution of small particulates, mainly oxides, with the presence of several larger TiC inclusions, with size ranging from 5 to 40 μm, can be observed. The latter TiC inclusions are usually fractured throughout NiTi cold working steps as also reported in [23, 24].

From microstructure results, it is confirmed that the non-contaminating nature of the PAM melting route is suitable for the production of NiTi with an adequate microstructure and refined particle distribution. EDX spectra of typical inclusions and the NiTi matrix of both PAM and VIM alloys are reported in Fig. 3.

Fig. 3
figure 3

EDX spectra of a typical inclusions of PAM NiTi; b typical inclusion of VIM NiTi; c NiTi matrix of both NiTi PAM and VIM showing near equiatomic NiTi composition

The micro-cleanliness of the purest PAM NiTi alloy was also evaluated according to ASTM F2063-12: the samples were viewed with backscattered electrons at 500× magnification and several images were collected, inspecting a total area of more than 2 mm2. In order to evaluate the number and total area of inclusions, selected fields of view were analyzed with image analysis software (X-Plus, Alexasoft™). The statistical outcomes for PAM are in fully compliance with the ASTM F2063-12 and the particles size distributions are presented in Table 2 and Fig. 4, respectively. In all fields of view, no particle with size greater than 39 μm was found, and 1.73 % (area %) of total particles was evaluated. These results are in agreement with the industrial NiTi alloys statistical inclusion distributions reported in Coda et al. [5] confirming that PAM is a suitable melting route for the production of Nitinol for medical devices and surgical tools.

Table 2 Statistical analysis results of standard purity PAM NiTi particulates
Fig. 4
figure 4

Particle size distribution of PAM NiTi. Most of inclusions are characterized by a small size (<3 μm), and particle larger than 19 μm was not detected (see subview). The ASTM F2063-12 size limit of 39 μm (dashed line) was depicted also

Corrosion tests

Potentiodynamic polarizations curves of different surface preparations of PAM NiTi specimens are plotted in Fig. 5. As shown, the 1 μm MP and the 1 μm EP samples show the same polarization behavior with pitting at high potential values (>1100 mV vs SCE). This happens in metallic materials completely passivized by an un-cracked protecting layer. It should be remarked that for these specimens the increasing current, is due to oxygen evolution reaction rather than pitting phenomena, as verified by the fact that the specimens do not show the presence of attacks at the end of the tests. From the polarization curves shown in Fig. 5, the role of the surface condition on pitting potential is also highlighted. In fact, Epit increases from 600 to 1100 mV vs SCE as the surface finish changes from #600 MP to 1 μm MP (or after surface EP). The latter results on PAM alloy are in agreement to the Burstein et al. results [25, 26] asserting that rougher surfaces may create more vulnerable sites to corrosion compared to smoother surfaces. Furthermore, the high Epit values measured for the 1 μm MP and 1 μm EP samples confirmed that a protective TiO2-enriched thin layer is created on EP surfaces, as reported by several authors [2732]. It has been asserted that electro-polishing of Nitinol assures, on one hand, very smooth, defect-free surfaces and on the other hand promotes the formation of a very thin protective TiO2 layer (≈2 nm in thickness).

Fig. 5
figure 5

Potentiodynamic tests in SBF 37 °C of MP and EP polished PAM NiTi alloy

In Fig. 6, a comparison of potentiodynamic polarization scans of PAM and VIM alloys are reported. From these plotted curves, it is unmistakable revealed that the NiTi alloys characterized by two different amount of inclusions show the same corrosion resistance behavior. Thus, from potentiodynamic tests, the correlation between the amount of inclusions, related to the purity degree of a NiTi alloy, and the corrosion resistance does not represent a reasonable approach for evaluating the acceptance of Nitinol purity degree. These results agreed the ones by Bai et al. [27] where any correlation between Ti2NiO x inclusions and the resistance to pitting was not found.

Fig. 6
figure 6

Potentiodynamic tests in SBF at 37 °C of MP polished PAM and VIM NiTi alloys

In Table 3, the pitting values of the potentiostatic scratch tests conducted on PAM and VIM NiTi alloys are collected together with the potentiodynamic results and with similar testing results on AISI 316 and Ti6Al4 V reported elsewhere [33]. The potentiostatic tests of PAM and VIM alloys show comparable Epit values with the PAM alloy slightly greater (300 mV vs. SCE) than VIM alloy. These values are comparable to those for AISI 316L, while Ti alloys show greater potentiostatic pitting potential (>800 mV vs. SCE) indicating a superior corrosion resistance. This result is in agreement with those of Vengupoland and Trepanier [34].

Table 3 Pitting potentials (mV vs. SCE) of 1 μm MP specimens

It must be emphasized that there is still an open question on acceptable criteria for corrosion resistance of implantable medical metallic devices. This matter is subjected to the ASTM F2129-08 Standard specifications [35]. Presently, this method for conducting cyclic potentiodynamic polarization measurements to determine the corrosion susceptibility of small implant devices does not specify any “acceptance criteria” for such devices. Rosenbloom and Colbert [36] proposed first an acceptance criterion based on results of potentiodynamic polarization tests as described in the ASTM F2129-08. In Tab. 4, their proposed classification is summarized (Table 4).

Table 4 Acceptance criteria based on pitting values of potentiodynamic polarization tests of implantable medical devices

Following potentiodynamic tests required by this proposed criteria the results of the present work are re-plotted in Fig. 7. As seen, both the standard purity PAM alloy as well as low purity VIM NiTi could be accepted as implantable medical devices.

Fig. 7
figure 7

Results of potentiodynamic tests re-plotted according to the classification of Rosenbloom and Colbert [36]. The value referring to Ti6Al4 V and AISI 316L stainless steel is taken from a previous work [33]

Conclusion

PAM was successfully employed to produce NiTi shape memory alloys. Non-metallic impurities in PAM alloy are consistent with the raw materials thus demonstrating the non-contaminating NiTi alloying process.

In vitro corrosion resistance tests of Nitinol with different inclusions content produced by PAM and VIM were reported. Potentiodynamic test did not show any effect of inclusions on the polarization curves, while the expected strong outcome related to sample surface preparation was highlighted.

The potentiodynamic results demonstrated that for EP and 1 μm MP surface preparations, high pitting values >800 mV versus SCE were reached independently from the inclusions content. In this case, both the alloys show excellent corrosion resistance.

For rough # 600 MP surface, Nitinol pitting value of about 600 mV versus SCE was detected for both NiTi inclusions contents. That it could be considered as borderline value for acceptance of NiTi as implantable material as proposed classification.

For PAM and VIM, Nitinol potentiostatic polarization results showed lower pitting potentials 250–300 mV vs SCE. In this case, the purer NiTi alloy showed slightly better corrosion resistance compared with the lower quality, but comparable to AISI 316L.