Abstract
Introduction
The aim of this paper is to show how to improve diagnostic accuracy using CDUS and twinkling artifact in patients experiencing discomfort due to the presence of small FBs in the soft tissues not clearly visible at US grayscale examination.
Materials and methods
We enrolled 7 adult patients presenting with small (2–4 mm) superficial FBs located in the subcutaneous and muscle tissues, barely or not detectable on US grayscale. All patients underwent US grayscale and CDUS examinations.
Results
We identified superficial FB with twinkling artifact in all 7 patients. All of these were confirmed to represent foreign bodies after surgical excision.
Conclusion
TA is useful in the evaluation of subcutaneous and muscular FBs and provides information on their location, depth and shape, which is useful if surgical excision is required.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Twinkling artifact (TA) is a color Doppler ultrasound (CDUS) artifact consisting of a fluctuating artefactual color mix of rapidly changing red and blue vertical bands behind a highly reflective, granular structure composed of several reflecting elements. On the contrary, TA does not occur behind smooth reflective structures [1].
TA occurs due to erroneous estimates of the speed of Doppler signals as the initial wave is transformed into several waves with different directions, amplitudes and frequencies. These waves are incorrectly interpreted by US equipment thus creating TA.
In the presence of large reflecting surfaces, such as calculi, the phenomenon acquires iconographic value, however adding nothing to diagnostic specificity in the presence of a rear shadow cone [2].
The usefulness of TA is enhanced in the presence of small reflective surfaces of about 2–4 mm without posterior shadow cones which make foreign bodies (FBs) difficult to detect at US grayscale imaging, whereas they can be better characterized in the presence of TA.
FBs may be introduced into the skin/muscles through lacerations and soft tissue wounds. Long-term complications of retained FBs include chronic pain and neurovascular impairment, and diagnosing and locating are therefore important in patient management and treatment.
The aim of this paper is to show how to improve diagnostic accuracy using CDUS and TA in patients experiencing discomfort due to the presence of small FBs in the soft tissues not clearly visible at US grayscale examination.
Materials and methods
From August to October 2022 we enrolled 7 adult patients presenting with small (2–4 mm) superficial FBs located in the subcutaneous and muscle tissues, barely detectable on US grayscale. US examination was carried out on the most painful site indicated by the patient. All patients underwent US grayscale and CDUS examinations. Informed written consent, for processing personal data, was obtained from all patients. IRB approval is not required for our study.
Two radiologists, each with more than 5 years of experience in ultrasound examination, performed US Gray Scale and color Doppler ultrasound on each patient, looking for the artifacts, using a Samsung RS80A with Prestige (with a 4–15 MHz linear probe).
First, the presence of foreign bodies was investigated on the affected site using US Gray Scale, then, CD examination was performed searching the TA or the RA, that was used to verify the presence and localize the FB.
During the scanning process, the focus was placed slightly deeper than the foreign body, the gain setting was controlled, and the pulse repetition frequency (PRF) was decreased.
Patients who presented symptoms that limited daily or work activities, underwent surgery to remove the foreign body.
Discussion
In a study reported by Mitchell DG, TA appeared as a rapidly changing mixture of red and blue behind presumed calcifications. The color change was not in any identifiable order, and the phenomenon was inexhaustible [2]. The spectrum was composed of close vertical bands with no outer wrapping, and the audio mode displayed treble squeaks. The color TA was found on 42 out of 140 color Doppler US images of presumed calcifications.
In vitro experiments reported by Aytac SK and Ozcan H showed that TA was generated by granular structures including sodium chloride, iron filings, emery paper, and ground chalk [3].
TA is due to a signal of slow movement of the tissue underneath the reflecting surface caused by the positioning of the probe on the skin. This signal is picked up by the probe and imperfectly eliminated by the noise elimination filter thus creating the complex beam of multiple waves that bounce between the various components of the interface and deepen in an area shielded by the reflectance of the interface itself.
In addition, erroneous estimates of the speed of Doppler signals may contribute to the genesis of TA [2, 3].
TA is an inconstant phenomenon varying in relation to two main factors, i.e. the anatomical and instrumental factors. The anatomical factor because TA is increasingly evident the more irregular the impact surface of the US beams is. The instrumental factor occurs in new devices with latest generation CDUS systems, where TA is diagnostic in over 95% of urinary stones [4].
In the presence of large reflecting surfaces, the phenomenon takes on an iconographic value due to the suggestiveness of the images but it adds nothing or little to the diagnostic specificity linked to the presence of the posterior shadow cone that is a US pathognomonic semeiological sign of highly reflective calcified structures [5].
TA is useful in the presence of small reflective surfaces (2–4 mm) without a posterior shadow cone (microcalcifications), making it possible to characterize their calcified nature, and in the presence of small FBs, which are poorly or not visible at grayscale US imaging.
TA was present regardless of the velocity (range, 1.0 cm/s–1.5 m/s), wall filter setting (100–800 Hz), probe frequency, focal depth, probe-to-interface distance (up to 21 cm) and color Doppler system [4, 6]. However, TA can be enhanced by placing the fire below the reflecting granular surface and decreasing PRF [6].
Thanks to TA, CDUS may be a reliable and non-invasive technique for evaluating and confirming the presence of small irregular or smooth reflective surfaces of 2–4 mm, not or barely visible at grayscale US, thereby providing useful information on the localization of FBs and increasing the detection of calcifications [7].
In conclusion TA is useful in the evaluation of subcutaneous and muscular FBs and provides information on their location, depth and shape, which is useful if surgical excision is required.
TA is also useful to confirm the presence of small calcifications in case of diagnostic doubt, when grayscale US does not allow a reliable visualization.
During our study, we observed that some materials generated a better visualization of TA than others. Glass proved to be a strong determinant of TA whereas wood exhibited TA to a lesser extent.
References
Rahmouni A, Bargoin R, Herment A et al (1996) Color Doppler twinkling artifact in hyperechoic regions. Radiology 199:269–271
Mitchell DG (1990) Color Doppler imaging: principles, limitations, and artifacts. Radiology 177:1–10
Aytac SK, Ozcan H (1999) Effect of color Doppler system on the twinkling sign associated with urinary tract calculi. J Clin Ultasound 27:433–439
Kamaya A, Tuthill T, Rubin JM (2003) Twinkling artifact on color Doppler sonography: dependence on machine parameters and underlying cause. AJR Am J Roentgenol 180:215–222
Granata A, Fiorini F, D’Amelio A, et al. L’ecocolordoppler nella pratica clinica nefrologica. Vol. 1. Milano: Ed. AccademiaNazionale di Medicina 2009, pp 163–5
Herment A, Guglielmi JP (1994) Principles of colour imaging of blood flows. EurUltrasound 1:197–204
Glica ML, Chirila CN, Podeanu DM et al (2017) Twinkle, twinkle little stone: an artifact improves the ultrasound performance. MedUltrason 19:272–275
Funding
Open access funding provided by Università degli Studi di Roma La Sapienza within the CRUI-CARE Agreement. The authors have not disclosed any funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Consent to publish
Informed consent for publication of the images was obtained from all individual participants included in the study.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Drudi, F.M., Maroncelli, R., Angelini, F. et al. Unusual application of twinkling artifact. J Ultrasound (2024). https://doi.org/10.1007/s40477-023-00861-w
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s40477-023-00861-w