Abstract
Purpose
Frequent glucose monitoring is necessary for optimal glycaemic control. Second-generation intermittently scanned glucose monitoring (isCGM) systems inform users of out-of-target glucose levels and may reduce monitoring burden. We aim to compare FreeStyle Libre 2 (Abbott Diabetes Care, Witney, U.K.) to self-monitoring of blood glucose in children with type 1 diabetes and sub-optimal glycaemic control.
Methods
This open-label randomised controlled trial will enrol 100 children (4–13 years inclusive, diagnosis of type 1 diabetes ≥ 6 months, HbA1c 58–110 mmol/mol [7.5–12.2%]), from 5 New Zealand diabetes centres. Following 2 weeks of blinded sensor wear, children will be randomised 1:1 to control or intervention arms. The intervention (duration 12 weeks) includes second-generation isCGM (FreeStyle Libre 2) and education on using interstitial glucose data to manage diabetes. The control group will continue self-monitoring blood glucose. The primary outcome is the difference in glycaemic control (measured as HbA1c) between groups at 12 weeks. Pre-specified secondary outcomes include change in glucose monitoring frequency, glycaemic control metrics and psychosocial outcomes at 12 weeks as well as isCGM acceptability.
Discussion
This research will investigate the effectiveness of the second-generation isCGM to promote recommended glycaemic control. The results of this trial may have important implications for including this new technology in the management of children with type 1 diabetes.
Trial registration
This trial was prospectively registered with the Australian New Zealand Clinical Trials Registry on 19 February 2020 (ACTRN12620000190909p) and the World Health Organization International Clinical Trials Registry Platform (Universal Trial Number U1111-1237-0090).
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Background
In New Zealand, there are an estimated 2,500 youth aged 0–18 years living with type 1 diabetes [1,2,3]. New Zealand has one of the highest rates of paediatric diabetes in the world, with the incidence growing annually [4]. Internationally, only one in four children with diabetes achieve international standards of glycaemic control (HbA1c < 58 mmol/mol [< 7.5%]) [5,6,7]. This increases their risk for short and long-term diabetes complications as shown by the Diabetes Care and Control Trial [8,9,10].
Frequent and timely self-monitoring of blood glucose (SMBG) is essential for guiding diabetes management decisions and keeping glucose levels in a safe range. Conventional SMBG involves finger-stick blood tests six or more times each day [11]. Children may infrequently perform SMBG because of social pressure to not be seen as ‘different’ [12], physical discomfort from pricking their fingers, and the technology is not user friendly (requires multiple steps to obtain a reading) [13].
Real-time continuous glucose monitoring (rtCGM) and intermittently scanned CGM have significant advantages over SMBG [14]. rtCGM systems use a subcutaneous glucose sensor to transmit and display a continuous stream of real-time interstitial glucose data to a pump/reader. Despite rtCGM systems being an accurate and effective glucose monitoring tool, like other diabetes technologies they are costly which can limit, or lead to inequity in uptake, and alarms can contribute to alarm fatigue and subsequent discontinuation of rtCGM use [15,16,17]. An alternative to rtCGM is intermittently scanned continuous glucose monitoring (isCGM) technology. isCGM involves applying a small factory-calibrated sensor to the back of the upper arm to detect interstitial glucose levels and then scanning the sensor with a reader to immediately display the glucose level. As with newer versions of rtCGM, isCGM technology provides accurate glucose information for up to 2 weeks [18]. Randomised controlled studies and real-world data based on first-generation isCGM use have found evidence of better glycaemic control with use over a sustained period of time [19, 20].
First-generation isCGM is highly acceptable to children and young people with diabetes and their caregivers [21, 22]. The second-generation isCGM system (FreeStyle Libre 2) is more accurate than the previous generation and additionally provides personalisable hypoglycaemia and hyperglycaemia alarms [23]. First-generation isCGM has been associated with improved quality of life and improved glycaemic control over 3 months in children ages 5–18 years [24]. The optional alarm feature in the second-generation system may particularly benefit families of children with above recommended HbA1c given the alarms prompt action to treat above target glucose levels and provide peace of mind that below target glucose levels will be detected. There is one randomised controlled trial currently being conducted in adult patients with type 1 diabetes in the UK [25]. However, there are no randomised controlled trials of second-generation isCGM in paediatric patient populations. In adolescents and young adults with type 1 diabetes, the first-generation of isCGM was found to increase glucose monitoring compared to SMBG, but this did not translate into significant differences in glycaemic control (as measured by HbA1c) between groups at 6 months [26]. Given the ease of being able to scan (even through clothing), the reduction in SMBG testing and both hypoglycaemia and hyperglycaemia alarms, second-generation isCGM may provide an important opportunity to help children and their families improve self-management behaviours [26].
The proposed trial aims to investigate the effectiveness of the second-generation isCGM for reducing HbA1c in children above the recommended glycaemic control target compared to SMBG.
Methods
Study design
This research is comprised of a multisite 12-week randomised, controlled, parallel-group trial. As shown in Fig. 1, 100 children with type 1 diabetes will be randomised to 12 weeks of standard care (control group) or standard care plus isCGM (intervention group). The study was approved by the Northern A Health and Disability Ethics Committee (ethics reference: 20/NTA/12) and Māori (indigenous New Zealanders) Research Consultation Committees in each region. Recruitment began in October 2020 and the study is expected to be completed by December 2022.
Study procedures
Study population and recruitment
The study will be conducted at 5 diabetes centres across New Zealand. Participants will be paediatric patients receiving standard diabetes care through district health board (DHB) diabetes services (Auckland DHB, Southern DHB, Capital Coast DHB, Bay of Plenty DHB, and Northland DHB). These diabetes services provide care for approximately 500 + children in the study age range. During routine clinical visits, eligible children will be identified by their usual paediatric endocrinologist/diabetologist/paediatrician and invited to participate. Inclusion criteria are: diagnosis of type 1 diabetes ≥ 6 months; age 4 to 13 years inclusive; on > 0.5 units of insulin/kg/day; no regular use of isCGM or CGM in the previous 3 months; and current HbA1c ≥ 58 mmol/mol and ≤ 110 mmol/mol, on day of consent. Children will not be included if they are diagnosed with any severe chronic diabetes related complications or severe medical or psychiatric co-morbidity/severe mental illness requiring ongoing treatment (e.g., diagnosed eating disorder); are participating in another study that could affect glucose measurements; or have plans to leave study site regions prior to study completion. Written informed consent will be obtained from parents/guardians, written informed assent will be obtained from participants aged 7 to 13 years, and verbal assent will be obtained from participants aged 4 to 6 years. Any participant can withdraw (or be withdrawn by their parent or guardian) from the study at any point.
Randomisation
Prior to study commencement, a randomisation table was generated by a biostatistician using Stata 15.1 software and pre-defined parameters (pre-study HbA1c [58 to 74 mmol/mol, or ≥ 75 mmol/mol; 7.5 to 8.9%, or ≥ 9.0%], study site), and imported into the REDCap randomisation module. REDCap is a secure, web-based application designed to support data capture for research studies [27]. Participants will be randomised in a 1:1 ratio to either the control (SMBG) group or the intervention (isCGM) group at Visit 2 by research staff using the randomisation module. Participants, investigators, and study staff will not be masked to group allocation.
Control group
All participants will continue standard diabetes care from their usual paediatric diabetes care provider. Routine diabetes clinics are attended regularly (at least every 3 months) to provide diabetes care by a multi-disciplinary team (paediatric endocrinologist/diabetologist/paediatrician, diabetes nurse specialist, dietitian, psychologist). Between scheduled study visits, participants will have the usual ability to contact their clinical team as is routine for all patients. Control group participants will continue SMBG using conventional finger stick BG testing with a glucometer and be fitted with a blinded isCGM, sensor, which they will wear for the first and final 2 weeks of the RCT.
Intervention group
The intervention consists of a FreeStyle Libre 2 isCGM system (sensors, reader, USB cable, power adapter, user’s manual, and quick start guide) and structured education from trained research staff. Education will include sensor insertion, interpreting the readings, and optimisation of insulin dosages, if appropriate. The first sensor will be applied by research staff. Participants will insert the next sensor 14 days later under supervision (Visit 3) and for the remainder of the study. Participants will be instructed to scan a minimum of 6–10 times each day with no longer than 8 h between two scans. Research staff will set the initial recommended reader settings to 3.9 mmol/L (70 mg/dL) and 15.0 mmol/L (270 mg/dL). Research staff will access glucose data online through LibreView, a secure, cloud-based system, to generate a report of participants’ average interstitial glucose level, time above/in/below range, and scans per day at 2-, 4-, 8- and 12-weeks from isCGM commencement. If the report shows time spent 'low' is > 4% or time spent 'very low' is > 1% then the report will be forwarded to the participant's clinical team for follow-up.
As a safety precaution, participants will be instructed to perform SMBG to confirm their glucose level before therapeutic interventions or corrective action are taken if hypo- or hyperglycaemic levels (≤ 4.0 or ≥ 14.0 mmol/l) or symptoms occur.
To prevent sensor loss prior to the end of the 14-day sensor session, participants will be provided with either Rockadex (pre-cut sports tape), Hypafix® (BSN medical GmbH, Hamburg, Germany) or cohesive tape to be used to attach the sensor securely in the event the adhesive becomes loose.
Procedures
At screening and enrolment (Visit 1, beginning of Week 1) a point of care HbA1c will be measured to confirm eligibility. Date of diabetes diagnosis for subsequent calculation of duration of diabetes (month and year will be recorded when the exact date is unknown), current insulin regimen, insulin dosing, HbA1c measurements in previous 6 months, and co-morbidities will be recorded from electronic medical records. Diabetic ketoacidosis (DKA) [28] and severe hypoglycaemia events (defined as a blood glucose value ≤ 3.9 mmol/L and resulting in loss of consciousness, a call for an ambulance and/or admission to hospital, or use of glucagon) in the past 6 months will also be recorded from electronic medical records to provide baseline estimates of frequency for these events. All participants will start blinded CGM (FreeStyle Libre Pro, Abbott) to continually measure and store glucose level data for up to 14 days [29]. This glucose monitoring system uses similar sensor technology to the FreeStyle Libre 2 system in the intervention; however, the Pro system masks all glucose data until it is downloaded at Visit 2. Participants with sensor data for at least 50% of the blinded wear period will be randomised at Visit 2. Questionnaires for the participant-reported outcomes will be administered before randomisation and at the end of the 12-week RCT.
Outcome assessments
The primary outcome is the between group change in HbA1c at 12-weeks (i.e., end of week 14 of study). The timing of all assessments is presented in Table 1. Trained research staff will be responsible for completing assessments. Visit 2 measurements will be taken before randomisation.
* Pediatric Quality of Life Inventory (PedsQL) 3.2 Diabetes Module, Hypoglycaemia Fear Survey (HFS), Self-Efficacy for Diabetes Self-Management (SEDM). ** Diabetic ketoacidosis, moderate and severe hypoglycaemia, issues related to glucose monitoring device use.
Demographics
A self-administered questionnaire will collect demographic information including age, gender, ethnicity, address, and education level. Participants may choose to select more than one ethnicity; however, each person will be allocated to a single ethnic group for the purposes of statistical analyses that will be prioritised in the order of Māori, Pacific, Asian and European/Other [30]. The address where the participant lives more than 50% of the time will be used to assess their New Zealand deprivation score, which is a validated index of the relative socioeconomic deprivation of the area in which an individual lives [31].
Anthropometry
Weight and height will be measured using standard procedures and calibrated instruments. Weight will be measured with a fixed scale (DigiTol, Toledo, Switzerland or similar) or portable scale (Tanita Corporation, Japan or similar) to the nearest 0.1 kg, with shoes and heavy clothing removed. Height will be measured to the nearest 0.1 cm, by wall-fixed stadiometer (Harpenden stadiometer, Holtain Limited, Pembs, UK or similar) or a portable stadiometer (Leicester Height Measure, Invicta Plastics Ltd., Oadby, England). Height and weight will be used to calculate body mass index (BMI)-z-scores using Centers for Disease Control and Prevention growth standards [32].
HbA1c
Glycated haemoglobin (HbA1c) will be measured by traditionally calibrated point-of-care instrument (DCA Vantage Analyzer, Siemens Healthcare Diagnostics, Ireland) at all sites, which meets acceptance criteria of having a total CV < 3% in the clinically relevant HbA1c range [33]. In the event a value is > 130 mmol/mol (> 14%, maximum reading possible) the value will be recorded as 130.
isCGM glucose metrics
During all follow-up visits, all retrospective glucose readings from the previous 2 weeks will be downloaded from the isCGM reader or LibreView. Hypoglycaemia (time below target) will be recorded as percentage of time below target (< 3.9 mmol/L). Time in range will be recorded as the percentage of time in the range (3.9–10.0 mmol/L) [34, 35]. Hyperglycaemia (time above target) will be recorded as percentage of time above target (> 10 mmol/L). Glucose levels < 3.9 mmol/L between 10 pm and 7am (nocturnal hypoglycaemia) will be reported to the appropriate diabetes care provider for follow-up.
Glucose monitoring behaviour
Glucose monitoring behaviour will be defined as scanning (intervention group) or SMBG (intervention and control group), which will be determined by device downloads of glucose monitoring device data.
isCGM acceptability
isCGM acceptability will be evaluated using a non-validated instrument adapted from previous similar research [36]. On an ordinal scale from 0 (strongly disagree) to 5 (strongly agree), participants will rate their opinion regarding the following areas: acceptability of sensor application, wear/use of the device and comparison to SMBG.
Psychosocial assessments
Psychosocial data and overall diabetes treatment acceptance will be collected through validated self-report questionnaires completed online using (REDCap Research Electronic Data Capture) software and the order of administration will be standardised to increase reliability. Together the questionnaires will take between 30 and 45 min to complete at each time point. All questionnaires will be administered in English. Clinical care teams will be notified if participants report physical or mental health problems necessitating follow-up.
The 33-item Pediatric Quality of Life Inventory (PedsQL) 3.2 Diabetes Module is a measure of diabetes-specific health-related quality of life that assesses participant’s and parent’s/guardian’s perceptions of the participant’s diabetes-specific symptoms and management challenges during the past month [37]. The PedsQL 3.2 Diabetes Module measures five domains: Diabetes Symptoms, Treatment Barriers, Treatment Adherence, Worry and Communication. Participant self-report forms are specific for ages 5–7, (young child), 8–12 (child), and 13–14 (adolescent). The parent proxy form is specific to ages 2–4 (toddler), 5–7 (young child), 8–12 (child), 13–14 (adolescent). The PedsQL 3.2 Diabetes Module Diabetes Symptoms and Diabetes Management Summary scores have demonstrated excellent measurement properties and are recommended as useful standardised patient-reported outcomes of diabetes symptoms and diabetes management in clinical research in children with type 1 diabetes [37]. Items are rated from 0 (never a problem) to 4 (almost always a problem). Item ratings are then reverse scored and linearly transformed to a 0–100 scale, with higher scores reflecting a better quality of life.
The Hypoglycaemia Fear Survey for Children (HFSC) is a 25-item instrument adapted from the adult HFS [38]. The HFSC will be completed by children aged 6 years and older. Overall, higher scores reflect greater fear of hypoglycaemia, a higher score on the Behaviour Subscale reflects a greater tendency to avoid hypoglycaemia and/or its negative consequences, and a higher score on the Worry Subscale indicates more worry concerning episodes of hypoglycaemia and its consequences. The CHFS has shown adequate internal consistency (HFSC behaviour subscale alpha = 0.70; CHFS worry subscale alpha = 0.89; and CHFS-Total alpha = 0.85) [38]. HFSC worry subscale and total scores have been shown to correlate significantly with other measures of anxiety [38]. Total scores and subscale scores will be calculated as z-scores standardised to the instrument-specific and baseline means and standard deviations.
The Self-Efficacy for Diabetes Self-Management (SEDM) is a 10-item self-report questionnaire for youth aged 10–16 years that examines confidence to carry out self-care behaviours and covers all the key areas of diabetes self-management [39]. The SEDM will be completed by participants who are 10 years and older. Participants are asked “How sure are you that you can do each of the following, almost all the time” and items are rated from 1 (not at all sure) to 10 (completely sure) and averaged. Higher scores indicate higher self-efficacy. The SEDM has demonstrated good validity and reliability (Cronbach’s alpha 0.9) [39].
At Visit 1, parents/guardians of enrolled participants who provide written consent for their own participation in the study will complete a short questionnaire collecting demographic characteristics (e.g., age, gender, education level, and ethnicity). At the baseline and follow-up visits parents/guardians will complete questionnaires to assess their perceptions of their own fear of their child experiencing hypoglycaemia using the parent version of the scale [38].
Statistical analysis
A sample size of 88 (44 participants in each group) would provide 80% power to detect a difference in changes in HbA1c of 7 mmol/mol (0.75%) between the intervention and control group using standard deviation of 15 mmol/mol and correlation of 0.7 between repeated observations on the same person and a two-sided test at the 0.05 level [26, 40]. This is a clinically important difference and similar to other proven technologies such as insulin pumps or CGM. To account for a small amount of missing data and loss to follow-up, we will recruit a sample size of 100 (50 participants per group) at baseline.
The statistician will be blinded to allocation arm and will use non-informative group codes until all planned analyses are completed. Descriptive statistics will be calculated for all variables. The primary analysis will follow the intention-to-treat principle with all participants analysed in the group to which they were randomised, regardless of actual sensor wear. Additional analyses include: HbA1c, glucose monitoring frequency and adherence, episodes of moderate and severe hypoglycaemia (as defined in Safety section below), episodes of DKA, and psychosocial variables using Poisson and linear mixed models as appropriate. Statistical analysis will be performed using Stata software with two-sided p < 0.05 considered significant.
Safety
For safety monitoring purposes, LibreView reports will be produced at 2-, 4-, 8-, and 12-weeks from isCGM commencement and checked for episodes of moderate (blood glucose values ≤ 3.9 mmol/L) and severe (child is having altered mental status and unable to assist in their care or is semiconscious or unconscious) hypoglycaemia. In the event the proportion of ‘low’ values is > 4% or ‘very low’ values is > 1% the report will be forwarded to the participant’s usual diabetes care provider for follow-up. Sensor failure rates and cutaneous adverse events (e.g., pain, itching, redness, subcutaneous haemorrhage, infection) will be self-reported to research staff at each visit or by phone call every four weeks throughout the study. All adverse events will be recorded in an Adverse Event form.
Participants will be asked to contact research staff immediately (by sending a photo of their affected skin site, if possible) if they notice a cutaneous issue associated with wearing the sensor. Clinical research staff will then advise if medical treatment is necessary. Participants will be referred to their general practitioner or emergency department, as appropriate, for management of medical events.
For more significant or persistent adverse events involving skin, a barrier product will be offered (e.g., Cavilon spray, SkinTac™) or drug therapy (e.g., zinc ointment, Fenistil gel, or hydrocortisone cream) prescribed, and the participant’s caregiver will be instructed to relocate the sensor to another area of the skin such that the effects are maintained at a tolerable level. Ultimately, the decision to continue or discontinue the use of the FreeStyle Libre 2 when localised skin symptoms occur will be made in consultation with the participant.
An internal Safety Monitoring Committee will be notified of severe adverse events (e.g., severe hypoglycaemia [BG value ≤ 3.9 mmol/L and resulting in loss of consciousness, a call for an ambulance and/or admission to hospital, or use of glucagon], DKA [being unwell due to hyperglycaemia and high ketones, and requiring a visit to the doctor, emergency room, or admission to hospital]) immediately after being reported to research staff. The Committee will then discuss any necessary action. Non-urgent events (moderate events) will be reported to the lead investigator after being reported to research staff. The internal Safety Monitoring Committee will be comprised of clinical investigators (CJ, BW, EW, AL, VC).
Data management
All study participants will be assigned a non-informative study identification number. Only research staff and investigators will have access to the electronic study records for the purposes of recording data and checking completeness of data. Data will be recorded and stored electronically in REDCap, which is securely hosted at the University of Otago. Identifiable information (e.g., date of diagnosis, address, date of birth) will not be stored in REDCap. Instead, age in whole numbers and duration of diabetes in whole numbers will be recorded in REDCap. Local sites will, however, hold in locked Excel sheets their own participants with address and contact details (phone number and emails), so that if the local sites need to contact participants (for replacement Libre 2 devices etc.) they can do so.
REDCap features (e.g., calendar and colour-coding forms to indicate complete or missing data) will help ensure adherence to timeframes, compliance to measurement procedures, and completeness of data. Data will be routinely checked for missing and/or erroneous values by the study coordinator. At the end of the study, original data collection sheets and written informed consent will be stored securely at the lead site along with copies of all data collected electronically. The lead investigator will retain an anonymised electronic copy of the cleaned data set, with all identifying information removed. The data set may be shared as part of the scientific peer-review process or shared to conduct a meta-analysis (e.g., impacts of flash glucose monitoring on glycaemic control). The electronic dataset will be destroyed 10 years from the end of the study.
Discussion
isCGM technology has the potential to significantly improve diabetes control in children, and limited data is available especially for the second-generation isCGM system. Increasing time in range, reducing HbA1c, reducing burden, and improving quality of life for children with this lifelong chronic disease is important and improving glycaemic control reduces the risk of acute and chronic diabetes complications. If next generation isCGM is effective in an RCT, it will then increase our ability to have this device available and funded for children worldwide.
Availability of data and material
De-identified data related to the primary and secondary outcomes will be available to those involved in the peer review process for publication in a scientific journal, upon request.
Code availability
Not applicable.
Abbreviations
- BG:
-
Blood glucose
- BMI:
-
Body mass index
- CGM:
-
Continuous glucose monitoring
- DHB:
-
District health board
- DKA:
-
Diabetic ketoacidosis
- HbA1c:
-
Glycated haemoglobin
- isCGM :
-
Intermittently scanned continuous glucose monitoring
- RCT:
-
Randomised controlled trial
- SMBG:
-
Self-monitoring blood glucose
References
Karvonen M, Viik-Kajander M, Moltchanova E, Libman I, LaPorte R, Tuomilehto J. Incidence of childhood type 1 diabetes worldwide. Diabetes Mondiale (DiaMond) Project Group. Diabetes Care. 2000;23(10):1516–26.
Campbell-Stokes PL, Taylor BJ. New Zealand Children’s Diabetes Working G. Prospective incidence study of diabetes mellitus in New Zealand children aged 0 to 14 years. Diabetologia. 2005;48(4):643–8.
You WP, Henneberg M. Type 1 diabetes prevalence increasing globally and regionally: the role of natural selection and life expectancy at birth. BMJ Open Diabetes Res Care. 2016;4(1):000161.
Derraik JG, Reed PW, Jefferies C, Cutfield SW, Hofman PL, Cutfield WS. Increasing incidence and age at diagnosis among children with type 1 diabetes mellitus over a 20-year period in Auckland (New Zealand). PLoS ONE. 2012;7(2):32640.
de Bock M, Jones TW, Fairchild J, Mouat F, Jefferies C. Children and adolescents with type 1 diabetes in Australasia: an online survey of model of care, workforce and outcomes. J Paediatr Child Health. 2019;55(1):82–6.
Foster NC, Beck RW, Miller KM, Clements MA, Rickels MR, DiMeglio LA, et al. State of type 1 diabetes management and outcomes from the T1D exchange in 2016–2018. Diabetes Technol Ther. 2019;21(2):66–72.
Phelan H, Clapin H, Bruns L, Cameron FJ, Cotterill AM, Couper JJ, et al. The Australasian Diabetes Data Network: first national audit of children and adolescents with type 1 diabetes. Med J Aust. 2017;206(3):121–5.
Gubitosi-Klug RA, Braffett BH, White NH, Sherwin RS, Service FJ, Lachin JM, et al. Risk of severe hypoglycemia in type 1 diabetes over 30 years of follow-up in the DCCT/EDIC study. Diabetes Care. 2017;40(8):1010–6.
Diabetes C, Complications Trial/Epidemiology of Diabetes I, Complications Research G, Lachin JM, White NH, Hainsworth DP, et al. Effect of intensive diabetes therapy on the progression of diabetic retinopathy in patients with type 1 diabetes: 18 years of follow-up in the DCCT/EDIC. Diabetes. 2015;64(2):631–42.
Tamborlane WV, Attia N, Saif R, Sakati N, Al AA. Impact of the diabetes control and complications trial (DCCT) on management of insulin-dependent diabetes mellitus: a pediatric perspective. Ann Saudi Med. 1996;16(1):64–8.
Miller KM, Beck RW, Bergenstal RM, Goland RS, Haller MJ, McGill JB, et al. Evidence of a strong association between frequency of self-monitoring of blood glucose and hemoglobin A1c levels in T1D exchange clinic registry participants. Diabetes Care. 2013;36(7):2009–14.
Borus JS, Blood E, Volkening LK, Laffel L, Shrier LA. Momentary assessment of social context and glucose monitoring adherence in adolescents with type 1 diabetes. J Adolesc Health. 2013;52(5):578–83.
Blackwell M, Wheeler BJ. Clinical review: the misreporting of logbook, download, and verbal self-measured blood glucose in adults and children with type I diabetes. Acta Diabetol. 2017;54(1):1–8.
Edelman SV, Argento NB, Pettus J, Hirsch IB. Clinical implications of real-time and intermittently scanned continuous glucose monitoring. Diabetes Care. 2018;41(11):2265–74.
Addala A, Auzanneau M, Miller K, Maier W, Foster N, Kapellen T, et al. A decade of disparities in diabetes technology use and HbA1c in pediatric type 1 diabetes: a transatlantic comparison. Diabetes Care. 2021;44(1):133–40.
Hennessy LD, De Lange M, Wiltshire EJ, Jefferies C, Wheeler BJ. Youth and non-European ethnicity are associated with increased loss of publicly funded insulin pump access in New Zealand people with type 1 diabetes. Diabet Med. 2021;38(1):14450.
Farfel A, Liberman A, Yackobovitch-Gavan M, Phillip M, Nimri R. Executive functions and adherence to continuous glucose monitoring in children and adolescents with type 1 diabetes. Diabetes Technol Ther. 2020;22(4):265–70.
Ajjan RA, Cummings MH, Jennings P, Leelarathna L, Rayman G, Wilmot EG. Accuracy of flash glucose monitoring and continuous glucose monitoring technologies: Implications for clinical practice. Diab Vasc Dis Res. 2018;15(3):175–84.
Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Kroger J, Weitgasser R. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet. 2016;388(10057):2254–63.
Jangam S, Lang J, Dunn T, Xu Y, Hayter G. Sustained improvement in glycaemia following flash glucose monitoring: an expanded worldwide analysis. Am Diabet Assoc Clin Diabet. 2019;68:1.
Boucher S, Blackwell M, Galland B, de Bock M, Crocket H, Wiltshire E, et al. Initial experiences of adolescents and young adults with type 1 diabetes and high-risk glycemic control after starting flash glucose monitoring - a qualitative study. J Diabetes Metab Disord. 2020;19(1):37–46.
Boucher SE, Aum SH, Crocket HR, Wiltshire EJ, Tomlinson PA, de Bock MI, et al. Exploring parental perspectives after commencement of flash glucose monitoring for type 1 diabetes in adolescents and young adults not meeting glycaemic targets: a qualitative study. Diabet Med. 2020;37(4):657–64.
Alva S, Bailey T, Brazg R, Budiman ES, Castorino K, Christiansen MP, et al. Accuracy of a 14-day factory-calibrated continuous glucose monitoring system with advanced algorithm in pediatric and adult population with diabetes. J Diabetes Sci Technol. 2020:1932296820958754.
Pintus D, Ng SM. Freestyle libre flash glucose monitoring improves patient quality of life measures in children with Type 1 diabetes mellitus (T1DM) with appropriate provision of education and support by healthcare professionals. Diabetes Metab Syndr. 2019;13(5):2923–6.
Wilmot EG, Evans M, Barnard-Kelly K, Burns M, Cranston I, Elliott RA, et al. Flash glucose monitoring with the FreeStyle Libre 2 compared with self-monitoring of blood glucose in suboptimally controlled type 1 diabetes: the FLASH-UK randomised controlled trial protocol. BMJ Open. 2021;11(7):e050713.
Boucher SE, Gray AR, de Bock M, Wiltshire EJ, Galland BC, Tomlinson PA, et al. Effect of 6 months’ flash glucose monitoring in adolescents and young adults with type 1 diabetes and suboptimal glycaemic control: managing diabetes in a “flash” randomised controlled trial protocol. BMC Endocr Disord. 2019;19(1):50.
Harris PA, Taylor R, Thielke R, Payne J, Gonzalez N, Conde JG. Research electronic data capture (REDCap): a metadata-driven methodology and workflow process for providing translational research informatics support. J Biomed Inform. 2009;42(2):377–81.
Wolfsdorf JI, Glaser N, Agus M, Fritsch M, Hanas R, Rewers A, et al. ISPAD Clinical Practice Consensus Guidelines 2018: diabetic ketoacidosis and the hyperglycemic hyperosmolar state. Pediatr Diabetes. 2018;19(Suppl 27):155–77.
Distiller LA, Cranston I, Mazze R. First clinical experience with retrospective flash glucose monitoring (FGM) analysis in South Africa: characterizing glycemic control with ambulatory glucose profile. J Diabetes Sci Technol. 2016;10(6):1294–302.
Carter PJ, Cutfield WS, Hofman PL, Gunn AJ, Wilson DA, Reed PW, et al. Ethnicity and social deprivation independently influence metabolic control in children with type 1 diabetes. Diabetologia. 2008;51(10):1835–42.
Atkinson J, Salmond C, Crampton P. NZDep2013 index of deprivation. New Zealand, Ministry of Health. 2014.
Ogden CL, Kuczmarski RJ, Flegal KM, Mei Z, Guo S, Wei R, et al. Centers for Disease Control and Prevention 2000 growth charts for the United States: improvements to the 1977 National Center for Health Statistics version. Pediatrics. 2002;109(1):45–60.
Lenters-Westra E, Slingerland RJ. Six of eight hemoglobin A1c point-of-care instruments do not meet the general accepted analytical performance criteria. Clin Chem. 2010;56(1):44–52.
Beck RW, Bergenstal RM, Riddlesworth TD, Kollman C, Li Z, Brown AS, et al. Validation of time in range as an outcome measure for diabetes clinical trials. Diabetes Care. 2019;42(3):400–5.
Beck RW, Bergenstal RM, Cheng P, Kollman C, Carlson AL, Johnson ML, et al. The relationships between time in range, hyperglycemia metrics, and HbA1c. J Diabetes Sci Technol. 2019;13(4):614–26.
Edge J, Acerini C, Campbell F, Hamilton-Shield J, Moudiotis C, Rahman S, et al. An alternative sensor-based method for glucose monitoring in children and young people with diabetes. Arch Dis Child. 2017;102(6):543–9.
Varni JW, Delamater AM, Hood KK, Raymond JK, Chang NT, Driscoll KA, et al. PedsQL 3.2 diabetes module for children, adolescents, and young adults: reliability and validity in type 1 diabetes. Diabetes Care. 2018;41(10):2064–71.
Gonder-Frederick L, Nyer M, Shepard JA, Vajda K, Clarke W. Assessing fear of hypoglycemia in children with Type 1 diabetes and their parents. Diabetes Manag (Lond). 2011;1(6):627–39.
Iannotti RJ, Schneider S, Nansel TR, Haynie DL, Plotnick LP, Clark LM, et al. Self-efficacy, outcome expectations, and diabetes self-management in adolescents with type 1 diabetes. J Dev Behav Pediatr. 2006;27(2):98–105.
Boucher SE, Gray AR, Wiltshire EJ, de Bock MI, Galland BC, Tomlinson PA, et al. Effect of 6 months of flash glucose monitoring in youth with type 1 diabetes and high-risk glycemic control: a randomized contraolled trial. Diabet Care. 2020;43(10):2388–95.
Funding
The study funder is The Starship Foundation (Auckland, New Zealand). The funder and the isCGM manufacturer have no roles or responsibilities in study design, conduct, data analysis and interpretation, or manuscript writing. Intervention supplies and blinded glucose sensors were purchased commercially from the isCGM manufacturer.
Author information
Authors and Affiliations
Contributions
CJ and BW were responsible for the study concept, oversight during protocol development, and will be responsible for the conduct of the study. VC, AL, MdB, MdL contributed to the study design and provided expert opinion during protocol development and funding application support. CJ, BW, EW, VC, and AL will facilitate recruitment of patients as lead regional investigators and will serve as the internal Safety Advisory Committee. WC will assist in study recruitment. CJ, SS, BW, HC, and DS were responsible for providing expert opinion during protocol development and preparing the detailed protocol with AB. MdL will be responsible for randomisation, statistical analyses, and the statistical interpretation of results. AS provided expert opinion during protocol development and funding application support, as well as to all aspects related to mental health. All authors contributed to refinement of the study protocol and approved the final manuscript.
Corresponding authors
Ethics declarations
Conflict of interest
The authors have no conflicts of interest to declare that are relevant to the content of this article.
Ethics approval
The protocol underwent Māori (indigenous New Zealanders) consultation, which fostered input into this study. The study protocol was approved by the Northern A Health and Disability Ethics Committee (ethics reference: 20/NTA/12). All district health boards approved recruitment and conduct of the study at their site.
Consent to participate
Written informed consent will be obtained from parents/guardians, written informed assent will be obtained from participants aged 7 to 13 years, and verbal assent will be obtained from participants aged 4 to 6 years.
Consent for publication
Not applicable.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Styles, S., Wheeler, B., Boucsein, A. et al. A comparison of FreeStyle Libre 2 to self-monitoring of blood glucose in children with type 1 diabetes and sub-optimal glycaemic control: a 12-week randomised controlled trial protocol. J Diabetes Metab Disord 20, 2093–2101 (2021). https://doi.org/10.1007/s40200-021-00907-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s40200-021-00907-y