Introduction

Cholesterol has aroused considerable interest in recent years on account of its being an important parameter in clinical diagnosis. Almost all membranes contain varying level of cholesterol and the level changes during different physiological conditions [1]. There is a strong positive correlation between high serum cholesterol level and arteriosclerosis, hypertension, and myocardial infarction. As biosensors have the ability to measure constantly the presence, absence, or concentration of specific organic or inorganic substances in desired test specimens, the thrust to monitor cholesterol in blood is driving the need for the development of efficient cholesterol biosensors to improve the life quality [2, 3].

Enzyme-based electrochemical biosensors have shown high selectivity and excellent sensitivity [46], but the enzyme is easily denatured during its immobilization procedure and its activity is also affected by temperature, pH, and toxic chemicals. Besides, the reproducibility of enzyme-based sensors is not very good [7] which further restrict the application of cholesterol biosensor. In recent years, nonenzymatic cholesterol sensors have been designed and fabricated to overcome the disadvantages of enzymatic cholesterol sensors.

The cholesterol biosensors based on carbon nanotube (CNT) have been continuously reported [817]. It has been demonstrated that carbon nanotubes could promote electron transfer with various redox active proteins, ranging from cytochrome c to glucose oxidase with a deeply embedded redox center and horseradish peroxidase with surface redox centers. To make use of carbon nanotubes in biosensing applications, the ability to immobilize biomolecules on the carbon nanotube structure, without diminishing their bioactivity is indispensable. Therefore, different types of physicochemical approaches have recently been devised for functionalization of carbon nanotubes with biomolecules. Many biological species, such as DNA, proteins and enzymes, have been immobilized on to carbon nanotubes either on their sidewalls or at the end caps. Lee et al. [8] developed a layer-by-layer assembled CNT network as a matrix for loading platinum nanoparticles to accomplish the nonenzymatic determination of cholesterol. The sensor with 24 bilayer of CNT shows very low detection limit and a wide linear range from 0.005 to 10 mM with a sensitivity of 8.7 μA mM−1 cm−2. Guo et al. [11] fabricated an amperometric cholesterol biosensor based on layer-by-layer deposition of a cationic polyelectrolyte poly (diallyl dimethyl ammonium chloride) and cholesterol oxidase (ChOx) on multiwalled carbon nanotubes modified gold electrode. A linear range up to 6.0 mM has been observed for the biosensor with a detection limit of 0.2 mM.

Li et al. [12] studied the effect of electron transfer and response current of carbon nanotube modified screen printed electrode for monitoring total cholesterol in blood using cholesterol esterase, cholesterol oxidase, peroxidase, and potassium ferrocyanide. Tan et al. [13] designed an amperometric cholesterol biosensor based on sol–gel chitosan/silica and multiwalled carbon nanotubes, which was used to estimate free cholesterol concentration in real human blood samples. However, these reported biosensors were witnessed by the narrow linear range, low sensitivity and unsatisfactory stability.

Since we have synthesized carbon nanotubes (CCNT) from coconut oil by our own indigenous method and it has also worked as glucose biosensor [18], it was felt to fabricate a new cholesterol biosensor based on this CCNT to check its sensitivity and detection limit towards cholesterol. Hence, in continuation of our earlier work on the synthesis and applications of nanomaterials [1822], we have now reported a nonenzymatic cholesterol biosensor at carbon nanotube (CCNT) modified electrode. Interestingly, the fabricated sensor showed a high sensitivity of ~15.31 ± 0.01 μA μM−1 cm−2 with the lower detection limit of 0.01708 μM in case of cholesterol. In literature many cholesterol sensors based on CNT electrodes have been reported, but synthesis of CNT from a readily available, cheap raw material and its use in cholesterol sensing is quite interesting and it reveals a new direction in research on biosensors.

Experimental

Reagents and apparatus

Coconut oil of edible quality was used for the synthesis. Cholesterol was purchased from E. Merck. All solutions were prepared with deionized water. The structure and morphology of CCNT were characterized by a scanning electron microscope (Zeiss EVO 50, India) and XRD were recorded with JSO ISO DEBYEFLEX 2002 Model X-ray powder diffractometer. Raman spectra was recorded using a Raman spectrometer; WITEC MODEL with 514 nm excitation. Electrochemical studies were performed using a mini potentiostat (Dropsens μstat 100, Spain).

Synthesis of CCNT

CCNT was synthesized by a low cost and easy method from coconut oil of edible quality as reported in our earlier paper [18]. It was functionalized using 2 M HNO3, washed and dried. Excess nitrate was removed and complete removal of nitrate was confirmed by Griess’s test [23]. The powdered CCNT was then characterized by SEM and XRD.

Fabrication of CCNT electrode

New modified electrode was designed like the commercially available standard electrode DS110CNT. In case of standard DS110CNT electrode, working electrode was made of functionalized carbon nanotubes, counter electrode was made of carbon, whereas reference electrode and electric contacts were made of silver. In our case, both working and counter electrode were made of CCNT and it was fabricated on an insulating teflon material containing three silver wires, following our reported method [18, 19]. For this, CCNT was first stirred with polystyrene solution, prepared in chloroform (9:1 ratio) followed by sonication. A drop of the slurry was then deposited as a very fine thin film on the teflon substrate covering two silver wires, serving as working and counter electrodes. The third silver wire was used as a reference electrode. The electrode has a dimension of 3.5 × 1.0 × 0.5 cm (length × width × height) and it was ideal for working with 50 μL volume like the standard electrode.

Nonenzymatic detection of cholesterol with fabricated CCNT electrode

For nonenzymatic detection of cholesterol, CV and DPV studies with various concentrations of cholesterol solutions were performed using 0.001 M H2SO4 as supporting electrolyte, at modified CCNT electrode. To carry out the electrochemical detection, different concentrations of cholesterol solutions were prepared by dissolving required amount of cholesterol directly into the solution of 0.001 M H2SO4, maintaining the pH ~ 5.3. Then, 50 μL of the cholesterol solution was taken by micropipette and dropped on the surface of electrode. The measurement conditions maintained for CV were E step = 0.005 V, E cond = −0.6 V, E dep = −0.5 V, whereas E step = 0.005 V, E pulse = 0.007 V, E cond = −0.6 V, E dep = −0.5 V, t pulse = 0.07 V for DPV studies.

Results and discussions

Characterizations of CCNT

The SEM micrograph of CCNT was shown in Fig. 1 which clearly shows the formation of the tubular structure CCNT having diameter below 100 nm. Figure 2 shows the XRD patterns of CCNT. The XRD pattern indicated the presence of two predominant peaks at 25.2° and 43.8°, which corresponds to (002) and (001) planes of CCNT, respectively. Raman spectra of CCNT, shown in Fig. 3, showed two prominent peaks at around 1350 and 1600 cm−1. The less intensed peak was known as D band and peak having more intensity was known as G band.

Fig. 1
figure 1

a Image of carbon nanotube (CCNT) modified electrode. b Image of commercially available screen printed carbon nanotube electrode (DS110CNT)

Fig. 2
figure 2

SEM images of carbon nanotube (CCNT)

Fig. 3
figure 3

a XRD pattern of carbon nanotube (CCNT). b Raman spectra of carbon nanotube (CCNT)

Electrochemical characterization of CCNT electrode

The active surface area of CCNT electrode was estimated by CV using 5 × 10−4 M K4[Fe(CN)6] in 0.1 M H2SO4 as a redox probe at various scan rate, by following Randles–Sevcik equation (Eq. 1) [24].

I P = ( 2 . 69 × 1 0 5 ) n 3 / 2 AD 1 / 2 C ν 1 / 2
(1)

where n is the number of electrons participating in the redox reaction, A is the electroactive surface area (cm2), D is the diffusion coefficient (cm2 s−1), C is the concentration of the redox probe molecule (mol cm−3), and ν is the scan rate (mVs−1).

Furthermore, IP increases with ν1/2 (Fig. 4) and is directly proportional to concentration and the ratio of IPaIPc is close to one. The [Fe(CN)6]3−/4−is one of the most extensively studied redox couples in electrochemistry, which exhibits a heterogeneous one electron transfer (n = 1).The calculated electroactive surface area of CCNT electrode was found to be 0.0765 cm2, where C = 10−4 M, D = 6.5 × 10−6 cm2/s for [Fe(CN)6]3−/4− system.

Fig. 4
figure 4

CV plot of 5 × 10−4 M [K4Fe(CN)6] in 0.1 M H2SO4 on CCNT electrode. Inset the plot of peak current (IP/μA) versus square root of scan rate (ν1/2/mV s−1)

Mechanism of nonenzymatic detection of cholesterol at CCNT electrode

For nonenzymatic detection of cholesterol, CV and DPV studies with various concentrations of cholesterol solutions were performed using 0.001 M H2SO4 as supporting electrolyte. CV and DPV plots clearly showed the presence of two peaks which suggests the generation of two electro active compounds during the reaction of cholesterol and H2SO4. The probable mechanism may be attributed to the fact that when cholesterol reacts with H2SO4, initially one product is formed which was unstable and therefore decomposed to another stable product in second step. This mechanism is also supported by Liebermann–Burchard method for cholesterol [25]. Therefore, the probable reaction may be represented as:

Cholesterol + H 2 SO 4 bis - cholestadienyl monosulfonic acid
(2)
Bis - cholestadienyl monosulfonic acid + H 2 SO 4 bis - cholestadienyl disulfonic acid
(3)

From CV and DPV studies, it can be concluded that the low intensity of the peaks was due to the formation of bis-cholestadienyl monosulfonic acid which decomposes to bis-cholestadienyl disulfonic acid resulting in a high intense peak.

Electrochemical response of cholesterol on different electrodes

The electrochemical behavior of 10−5 M (10 μM) cholesterol in 0.001 M H2SO4 (pH ~ 5.3) at a scan rate of 100 mVs−1 on bare silver, standard screen printed carbon nanotube electrode (DS110CNT) and CCNT modified electrode was studied by CV. Figure 5 showed that for the first peak, the peak potentials on DS110CNT electrode and CCNT modified electrode were almost same, i.e., −0.27 and −0.26 V, respectively. No peak current was observed at bare silver electrode which indicated that it cannot undergo the reaction in this potential range but for the second peak, a little rise in the peak current was observed at—0.056 V. DS110CNT electrode and our CCNT electrode again showed very close proximity in the peak potential values for the second peak, i.e., −0.092 and 0.003 V, respectively. Although the peak potentials were in very good agreement with each other, but heights of the peaks were different for each electrode. The order of the heights of the peaks was found to be I/μA (DS110CNT electrode) > I/μA (CCNT electrode) > I/μA (bare silver electrode). Thus, it may further be concluded that the unique shape and dimension of nanostructured CCNT electrode offered good electrical conductivity almost like standard CNT electrode, resulting in the increased peak current than bare silver electrode and almost same peak potential like DS110CNT electrode.

Effect of pH

Figure 6a displayed the effect of different pH on the electrochemical response of 10−5 M cholesterol in 0.001 M H2SO4 at a scan rate of 100 mVs−1 at CCNT electrode. It was observed that when the pH changed from 4.7 to 6.7, the anodic peak in CV moved towards the negative direction and the current response decreased. Figure 6b showed the linear relationship between the anodic peak current and the pH value in the range of 5.3 and 6.7 following the equation I/μA = 16.24791 − 1.19302 pH (R2 = −0.99636). I/μA vs. pH graph showed that for cholesterol, maximum peak current was observed at pH 5.3 and hence this pH was chosen for the subsequent analytical experiments for cholesterol.

Fig. 5
figure 5

CV plots of 10−5 M cholesterol in 0.001 M H2SO4 at 100 mV s−1scan rates on bare Ag electrode, screen printed carbon electrode (DS110), and CCNT modified electrode at pH ~ 5.3

Effect of scan rate

CV technique was employed to find out the effect of the scan rate on the electrochemical detection of cholesterol at the CCNT electrode. Figure 7a displayed CV plots showing the overlapping of voltagrams of 10−5 M cholesterol in 0.001 M H2SO4 (pH ~ 5.3) at various scan rates on CCNT electrode. The current versus scan rate plot shown in the inset exhibited a linear relationship following the equation I/μA = 6.28481 + 0.39227 ν/mV s−1, R2 = 0.99724. This confirmed that the electrochemical kinetics reaction was adsorption-controlled [26] as the adsorption-controlled process should result in a linear plot in case of I vs. ν. The linearity was observed over the entire range (50–112 mV s−1) of scan rates studied with standard deviation value of 0.6189, where N = 13. A plot of logarithm of peak current vs. logarithm of scan rate gave a straight line with a slope of 1.2 (Fig. 7b) closed to the theoretical value of 1 for a purely adsorption-controlled process [24] which in turn further confirms that the process was adsorption-controlled. The linear regression equation log I/μA = −1.00262 + 1.26127 log ν/mV s−1 gave the value R2 = 0.99811.

Fig. 6
figure 6

a The effect of different pH on the electrochemical response of 10−5 M cholesterol in 0.001 M H2SO4 at a scan rate of 100 mV s−1 at CCNT modified electrode. b The linear relationship between the anodic peak current and the pH value from 5.3 to 6.7

In an adsorption-controlled system, the analyte does not have to travel to the electrode. Thus, when the potential required for reaction is reached, the current increases and decreases much more rapidly which results in a sharp peak with a high current density since all the analyte can react at once.

Effect of accumulation time

The effect of accumulation time on peak current was also investigated with 10−5 M cholesterol in 0.001 M H2SO4 (pH ~ 5.3) at CCNT electrode. Figure 8 showed the plots of CV of 10−5 M cholesterol in 0.001 M H2SO4 (pH ~ 5.3) at 100 mV s−1 scan rate and at various time intervals (1–6 min). Calibration plot (in inset) showed a linear dependence of the anodic peak current on the entire range of time interval studied in the both cases giving the correlation coefficient of 0.99456 (I/μA = 6.26 + 3.39714 min) [27, 28]. The standard deviation value was found to be 0.74412, where N = 6. However, if the study was conducted over a more extended time interval range then there was a breakdown in linearity relationship. This suggests that with the increase in accumulation time, the peak current increases gradually but after 6 min the peak current reached the maximum and becomes stable. Therefore, the optimal accumulation time of 6 min was employed in further experiments.

Fig. 7
figure 7

a CVs of 10−5 M cholesterol in 0.001 M H2SO4 at various scan rates (50, 55, 60, 65, 70, 75, 80, 85, 90, 95, 100, 110, 112 mVs−1) on CCNT modified electrode at pH ~ 5.3. Inset: the plot of peak current (I/μA) versus scan rate (ν mV s−1). b Linear relation between logarithm of peak current (log I/μA) and logarithm of scan rate (log ν mV s−1)

Linearity, sensitivity, and limit of detection of the nonenzymatic cholesterol sensor

To determine the quantitative analysis of cholesterol at the CCNT electrode, DPV technique was employed because the peaks were sharper and better defined at lower concentration of cholesterol than those obtained by CV studies. 0.001 M H2SO4 (pH ~ 5.3) was selected as the supporting electrolyte for the quantification of cholesterol as it gave maximum peak current intensity at pH = 5.3. Figure 9 showed that peak current increases linearly with the increasing concentration of cholesterol. A good linear relationship was observed between anodic current and cholesterol concentrations in the range of 1–50 μM. The linear regression equation I μA = 6.96911 + 0.11713 gave the R2 value as 0.99856. However, if the study was conducted with more concentrated solution then there was a breakdown in linearity relationship which may be attributed to adsorption of oxidized products of cholesterol on the electrode surface. The sensitivity was calculated using the slope of the current versus concentration calibration plot (Fig. 9, inset) divided by the active surface area of CCNT according to the following equation [29]:

Sensitivity = slope of the plot / active surface area of the electrode
(4)

where the active surface area of CCNT was 0.0765 cm2 and the slope was 1.1713 μA μM−1. The sensitivity of the modified CCNT nonenzymatic cholesterol sensor was calculated to be ~15.31 ± 0.01 μA μM−1 cm−2.

Fig. 8
figure 8

CVs of 10−5 M cholesterol at CCNT modified electrode in 0.001 M H2SO4 in various time interval (1, 2, 3, 4, 5, 6 min) at 100 mV s−1 scan rate at pH ~5.3. Inset: the plot of peak current (I/μA) versus time interval (t/min)

Fig. 9
figure 9

CV plots at CCNT modified electrode in 0.001 M H2SO4 with different concentrations of cholesterol (0–50 μM) at 100 mV s−1 scan rate at pH ~5.3. Inset: calibration curve of response current (I/μA) versus concentrations of cholesterol (conc. μM)

The limit of detection (LOD) and limit of quantification (LOQ) were calculated using the following equation:

LOD = 3 s/m; LOQ = 10 s/m, where s is the standard deviation of the peak currents of the blank (five runs), and m is the slope of the calibration curve [30]. The values were found to be 0.017 and 0.056 μM, respectively. Statistical data of the calibration curve are obtained from five different determinations were shown in Table 1.

Table 1 Characteristics of calibration plot of nonenzymatic cholesterol sensor using DPV at CCNT modified electrode

The detection limit, sensitivity, and response time of CCNT electrode was compared to that of other reported electrodes used for cholesterol sensors [813] and it was found that these results were in good agreement with literature values (Table 2).

Table 2 A comparison of CCNT electrode with other modified electrode, using as cholesterol sensor

Reproducibility and stability of the nonenzymatic cholesterol sensor

Reproducibility and stability are two important characteristics for the any modified electrode. The reproducibility of the proposed electrochemical sensor was evaluated by determining the voltammetric response of 10−5 M cholesterol (five separate samples) in 0.001 M H2SO4 at the same pH at CCNT electrode (five times) by DPV studies. It was found that the average peak current was 12.5 μA, and the relative standard deviation (RSD) was 3.5 % which showed acceptable reproducibility of the proposed cholesterol sensor (Table 3). The stability of the nonenzymatic cholesterol sensor was also explored by storing the modified electrode in air for 10 days and the current response of 10−5 M cholesterol was found to be stable, keeping ~97 % of its initial intensity.

Table 3 Influence of interfering species on the voltammetric response of 10−5 M cholesterol

Interference

Under optimized experimental conditions mentioned above, some species can possibly interfere with cholesterol determination. So, the effects of the presence of some interfering substances on the current responses of 10−5 M cholesterol have been evaluated by DPV studies to ensure the selectivity of the reported cholesterol biosensor. Voltammetric studies were carried out by analyzing the sample solutions of 10−5 M cholesterol, spiked with various amount of urea, uric acid, glucose, vitamin C, l-alanine, glycine, l-serine, l-phenylalanine, tryptophan, tyrosine under the same experimental conditions. The experimental results (Table 3) indicated that ten times of each substance and five times of glucose showed almost no interference with the determination of cholesterol. Hence, it can be further concluded that the proposed method was able to assay cholesterol in the presence of interfering substances also. The average signal change was found to be 1.364 %.

Conclusions

A nonenzymatic cholesterol biosensor has been fabricated at the surface of CCNT electrode. The most attractive features of this biosensor were mainly its selectivity and sensitivity. A higher producible sensitivity of ~15.31 ± 0.01 μA μM−1 cm−2, response time of 6 s and detection limit of 0.01708 μM were obtained from the fabricated biosensor. The long-term storage stability of the sensor was tested for 10 days. The sensitivity retained ~97 % of initial sensitivity up to 10 days. However, the detailed studies of CCNT cholesterol biosensor for estimation of total cholesterol in serum and blood samples by nonenzymatic method, as well as the effect of various interfering substances with their increasing concentration, are in progress.