Introduction

Recently, non-thermal atmospheric pressure plasma has attracted considerable attention in biomedical applications due to its simplicity and efficiency [1]. Conventionally, UV emissions, heat (more than 120 \(^{\circ }\)C), ethanol and strong chemicals [2, 3] are effective sterilizing methods for inactivation of microorganisms. These methods have recently raised general disagreements about their environmental effects. Therefore, new methods of sterilization should be sought which have some advantages such as safety, convenience and lack of residual toxicity [4]. Non-thermal plasma is an alternative to conventional sterilization methods [5] and is considered as a new sterilization technique in the conservation of materials from bacteria [4]. This method has several advantages such as the possibility of inactivating the bacteria at low temperatures (near room temperature), being economical, absence of residual toxicity, and appropriateness for applications when product conservation is required [6, 7]. They have become very attractive for biomedical applications such as bacteria inactivation [8, 9], wound healing [10], decontamination of medical equipment [11], and blood coagulation [12]. The non-thermal plasma consists of neutral gas atoms or molecules, ions, electrons and reactive species [13]. It is reported that the reactive species and excited neutrals in the non-thermal plasma are responsible for the microorganism inactivation since they are able to modify the DNAs, proteins, and cell membranes [14, 15].

Nishime et al. [16] evaluated the sterilizing efficacy of a DBD non-thermal helium plasma jet toward Gram-positive and Gram-negative bacteria, and demonstrated that the sterilizing efficacy depends critically on the active species such as ozone. Maisch et al. [17] evaluated the antimicrobial effects and the mechanism of cold atmospheric plasma treatment on MRSA, Staphylococcus aureus and E. coli bacteria and showed that the bacteria inactivation increased with longer treatment time.

The fundamental objective of this work is to develop an understanding of the effects of non-thermal plasma jet on bacterial decontamination. Although there have been numerous studies that deal with decontamination of bacteria using atmospheric pressure plasma jet (APPJ), less attention has been paid to using double-ring electrodes in enhancing the performance of plasma jets system, which can lead to more efficient sterilization of bacteria. In addition, the efficiency of sterilization has been shown clearly when treating with plasma jets in spite of the lower applied voltage.

Experimental part

Experimental facility

The plasma jet used in this study is a dielectric barrier discharge (DBD) type APPJ. The schematic diagram of the DBD plasma jet is shown in Fig. 1. It consists of a Pyrex tube representing a dielectric barrier discharge. The length of the tube was 95 mm with wall thickness of 0.925 mm, while the internal and external diameters of the tube were 2 mm and 3.85 mm, respectively. Helium gas with commercial grade of 99.998% was fed at the upper part of the Pyrex tube and the flow rate was controlled by flow meters 11420 (Mathesen, USA) and needle valve 288 01 B2 (Leybold, Germany). The DBD plasma jet device was based on a double-ring structure with two aluminum ring electrodes covering the external Pyrex tube. The thickness and width of the aluminum ring were 0.1 and 12 mm, respectively. The distance between two electrodes was 13 mm and the distance between the downstream electrode and the nozzle of the Pyrex tube was 4 mm. The upstream electrode was connected to a power supply and the downstream electrode was grounded.

Fig. 1
figure 1

Schematic representation of the DBD plasma jet system

Instrumentation for plasma characterization

The BDB plasma jet was driven by a homemade high-voltage AC power supply under fixed frequency of 12 kHz with variable voltage from 0 to 7.5 kV\(_{p-p}\). The electrical properties were measured by a high-voltage probe P6015A (Tektronix, USA) and a current probe AT-C202 (Generic, China). The waveforms of the applied voltage and discharge current were recorded using PC USB oscilloscope 6022BE (Hantek, China) with a 20 MHz bandwidth and a 48 MS/s sampling rate. The plasma gas temperature was measured by alcohol thermometer using different flow rates at a distance of 15 mm from the nozzle of the Pyrex tube. The emission spectra of the DBD plasma jet was determined by ultraviolet visible (UV–Vis) spectrometer device SV2100 (Kmac, Korea) with wavelength range from 200 to 1100 nm. The spectrometer device was connected with an optical fiber cable M92L01 (Thorlabs, USA) to record the spectral emissions. To avoid the dispersion of plasma radiation, a collimator was used to assemble the radiation emitted from the plasma. The collimator was located at 20 mm from the edge of the Pyrex tube (axial direction) and 20 mm from the wall of the Pyrex tube (radical direction).

Sample preparation

A bacterial suspension was prepared for S. aureus bacteria with specific concentrations. This concentrations were measured by UV–Vis spectrophotometer device SP-3000 Plus (Optima, Japan) at 625 nm and was 0.402, which is equivalent to \(6.5 \times 10^{8}\) bacteria cell-forming units per ml (CFU/ml). After the suspension was prepared, 0.1 ml of this suspension, containing \(6.5 \times 10^{8}\) CFU/ml bacteria, was spread on the nutrient agar via the spreading method by a sterile swab in a standard Petri dish. The dishes were stored in an aseptic location to dry for 10 min and then exposed to the plasma jet; see Fig. 2.

Fig. 2
figure 2

Photograph of the DBD plasma jet interacting with Staphylococcus aureus bacteria

Treatment conditions

For all treatments, the DBD plasma jet was operated under a frequency of 12 kHz with 7.5 kV\(_{p-p}\) of applied voltage and 4 slm of helium flow rate. The distance between the nozzle of the Pyrex tube and the sample was fixed at 15 mm. The bacteria were exposed to the DBD plasma jet for different time intervals (1, 3, 5, 10 and 15 min). After treatment, all samples were incubated for 24 h at 37 \(^\circ\)C.

Measuring the inhibition zones

To determine the effect of plasma treatment on bacterial samples, the approximate diameter of inhibition zones which formed on agar was measured using Vernier calipers. To improve the measurements precision of irregular zones, advanced image software was used to calculate the diameter of the inhibition zones.

Measuring the Reynold number

The state of the helium flow is achieved by calculating the Reynolds number (\(R_{n}\)), which describes the stability of flow such as the stable laminar flow and the turbulent flow [18]. The Reynolds number can be calculated by the following formula [19]:

$$\begin{aligned} R_{n} = 2.12 \times 10^{-2} \left( \frac{\rho }{\mu }\right) \times \left( \frac{Q}{r_{D}}\right) , \end{aligned}$$
(1)

where \(\rho\) is the fluid density, \(\mu\) is the viscosity, Q is the flow rate and \(r_\mathrm{{D}}\) is the inner diameter of the tube.

Results and discussion

The typical waveforms of current discharge produced in the DBD plasma jet is shown in Fig. 3. The current in the first cycle increases rapidly with increased applied voltage, but decreases dramatically before the applied voltage reaches the peak value. This is attributed to the shielding effect (memory effect), which is caused by the accumulation of negative charge on the Pyrex tube [20]. This effect is a well-known mechanism of the DBD plasma which avoids arcing and makes the plasma plume harmless. Also, at higher voltage the values of the current pulses increase and their amplitude decreases. Further, the discharge current has many pulses for each cycle of the applied voltage which can be attributed to a glow-like discharge (multi pulse glow discharge) [21]. Therefore, the discharge current produced from the DBD plasma jet is considered a uniform breakdown [22].

Fig. 3
figure 3

Waveform of the applied voltage and discharge current under a helium flow rate of 4 slm

Figure 4 shows the effect of helium flow rate on the length of the plasma plume with two values of applied voltage. As can be seen in this figure, the length of the plume rapidly increases with the gas flow rate from 1 to 4 slm. The plasma plume reached the maximum length of 44 mm when the applied voltage was 7.5 kV\(_{p-p}\) and 41 mm when the applied voltage was 4.7 kV\(_{p-p}\). This indicates that the effect of increased applied voltage leads to a slight increase in the length of the plasma plume because of the increase of electron mobility and electric field intensity [23].

If the gas flow rate increases over 4 slm, the length of the plasma plume decreases dramatically at both values of applied voltage. This is because of the effect of Reynolds number. If the \(R_{n}\) is relatively small, such as \(R_{n}\) < 2000, the flow is laminar. If it is large, i.e., \(R_{n}\) > 4000, the flow is turbulent [24]. At 5 slm, the \(R_{n}\) is 4527, and this causes the flow rate to become turbulent and results in instabilities and deformations of the gas flow rate [25].

Fig. 4
figure 4

Plasma plume length as a function of helium flow rate

Figure 5 shows the gas temperature as a function of the irradiation time with various gas flow rates at the plasma conditions of 12 kHz and 7.5 kV\(_{p-p}\). As observed in this figure, the highest temperature reached 39 \(^{\circ }\)C at 1.5 min with a gas flow rate of 2 slm and the lowest temperature reached 29 \(^{\circ }\)C at 0.5 min with a flow rate of 5 slm. This indicates that the gas temperature increased with irradiation time and decreased with gas flow rate. Further, the gas temperature at all flow rates reached thermal equilibrium after 1.5 min of plasma exposure and remained the same up to 15 min.

Fig. 5
figure 5

Gas temperature as a function of the time for various helium flow rates

The effect of the distance from plasma jet exit on the gas temperature is shown in Fig. 6. It is observed that the gas temperature at 4 slm decreased from 38 to 34 \(^{\circ }\)C and from 42 to 38.5 \(^{\circ }\)C at 2 slm when the distance increased from 5 to 15 mm. It is worth mentioning that the effect of reducing the distance between the plasma jet exit and the target has an important role in the inactivation efficiency of bacteria. This is because the decrease in the distance leads to the generation of more active species, which play an important role in the inactivation process of bacteria. On the other hand, reducing the distance between the plasma jet exit and the target increases the temperature. Therefore the distance should be optimized so as not to exceed the plasma temperature above 43 \(^{\circ }\)C [26].

Fig. 6
figure 6

Gas temperature as a function of the distance from the nozzle of the Pyrex tube

The DBD plasma jet propagation to the surrounding air leads to interaction with air molecules, forming active species such as reactive oxygen and nitrogen species [27]. Figure 7 shows the spectral lines of the DBD plasma jet emitted in the UV–Vis regions. The band of excited molecular nitrogen is detected in the UV region at 357 nm, while the ionized nitrogen can be found in the Vis region at 427 nm. In addition, the spectral lines of ionic oxygen are detected in the UV–Vis region between 337 and 469 nm. The spectral lines of hydrogen and atomic oxygen can be found in the Vis regions at 656 nm for hydrogen, with 645.3 and 777 nm for atomic oxygen. Several spectral lines of atomic helium can be observed in UV–Vis regions between 388 and 728 nm.

Fig. 7
figure 7

Typical spectral lines of the DBD plasma jet

The spectral line intensities of the N\(_{2}^{+}\), N\(_{2}\), H, O II and O I as a function of helium flow rate are shown in Fig. 8. As observed in this figure, increasing the helium flow rate leads to higher intensities of spectral lines. This is because the increase in gas flow rate results in a lower mole fraction of air which surrounds the plasma plume. The lower mole fraction of air can lead to higher local electron concentration. This explains the higher emission intensity from excited species [28, 29]. Some authors showed that the increase of gas flow rate is very significant to increase the amount of reactive oxygen and nitrogen species (RONS), which in turn contribute to cell death by damage of the DNA molecules [30].

Fig. 8
figure 8

Emission intensities of a N\(_{2}^{+}\) and N\(_{2}\), b H, c O II and d O II as a function of helium flow rate

Figure 9 shows the images of bacteria after treatment by DBD plasma jet for 3, 5 and 15 min. It is observed that the inhibition zone diameter extends with increased plasma exposure time. Some authors have reported that the increase of plasma treatment time leads to increasing effectively the inactivation of microorganisms [31].

Fig. 9
figure 9

Photographs of S. aureus samples after treatment by DBD plasma jet. a Sample not exposed to plasma. Sample exposed to bd plasma for 3, 5 and 15 min

The diameter of the inhibition zones for S. aureus bacteria was determined and it is presented in Fig. 10 under different times of treatment. It is observed that the inhibition zone diameter exhibits a linear increase with plasma exposure time, and the inhibition zone diameter reached the maximum value at 53.37 ± 0.8 mm under 15 min of plasma exposure.

Fig. 10
figure 10

Inhibition zone diameter of the S. aureus bacteria as a function of the plasma treatment time

Figure 11 shows the relationship of the survival amounts versus helium plasma jet treating times. The curve drops sharply in about 1 min, which means the amounts of alive cells decrease rapidly in the sample. After an extension of the treating time from 3 to 15 min, the amounts of alive cells decrease relatively slowly. Therefore, it may be concluded that the killing rate is dependent on the treating times.

Fig. 11
figure 11

Live cell amounts as a function of helium plasma exposure times

The mechanism of bacteria inactivation in plasma jet can occur by two known factors: physical and chemical factors. The physical factors include the heat, UV irradiation and charged particles, while the chemical factors include the active species.

At the operating plasma conditions of 12 kHz, 7.5 kV\(_{p-p}\) and 4 slm used to treat the bacteria, the gas temperature reached 34 \(^{\circ }\)C. The effect of heat at this degree does not contribute to the process of bacteria inactivation. The charged particles produced from DBD plasma can be accumulated on the external surface of the cell membrane. These charges form an electrical force which can overcome the tensile strength of the cell membrane and cause its rupture. This process occurs if the plasma is directly exposed to the bacteria [32]. The concentration of charged particles decreases if the plasma indirectly exposes the bacteria (plasma afterglow) and this is due the quick recombination of the electrons and ions [33]. Therefore, it can be concluded that the charged particles play a secondary role in the process of bacteria inactivation, because all bacteria samples in this study were exposed to the plasma plume indirectly. Furthermore, the UV-C radiation intensity in the wavelength region of 180–280 nm may cause lethal harm to cells. Thus, the inactivation effect on microorganisms by ultraviolet irradiation in this range is mostly associated with DNA damage [34]. The intensities of the spectral lines in the range 180–280 nm were not observed when the plasma spectrum was diagnosed through optical emission spectroscopy (OES). Thus, it can be concluded that the UV irradiations do not play a major role in the process of bacteria inactivation.

As shown in Fig. 9, the number of colonies decreases significantly with increase in irradiation time, as a large number of colonies are killed even at the ends of the inhibition zone diameter. On the other hand, it is worth mentioning that the diameter of inhibition zone was always larger than the plasma plume diameter (equal to 1.5 mm), which can be attributed to the fact that the inactivation occurred mostly due to the action of RONS. Studies have reported that reactive species in DBD plasma jet can contribute significantly to the inactivation process of bacteria [35]. Reactive species such as ROS generated by DBD plasma jet have damaging effects on cells, since they target proteins, DNA, cell wall and membrane [36]. For example, ROS, such as O and O\(_{3},\) can cause leakage of the cell membrane and induce DNA deterioration, which in turn causes the inactivation of bacteria [37].

Conclusion

In this study, the DBD non-thermal plasma jet was well designed to be an effective bactericidal when tested against S. aureus bacteria. With increased helium flow rate, the gas temperature decreased, while plasma plume length increased. According to OES, there are no peaks in the range 180–280 nm. The absence of these peaks suggests that the UV radiation did not contribute to the inactivation process and therefore no lethal harm will be caused to the cells. The bacteria exposed to plasma showed an increase in the inhibition zone diameter with longer irradiation time. In addition, the diameter of inhibition zone was always larger than the plasma plume diameter, suggesting that the RONS played a major role in the inactivation process.