Abstract
This research was conducted to apply polyimide tape, which has the advantages of low price ans strong adhesive strength, to the neural electrode process. In addition, to maximize the low-cost characteristics, a fabrication process based on UV laser patterning rather than a photolithography process was introduced. The fabrication process started by attaching the gold sheet on the conductive double-sided tape without being torn or crushed. Then, the gold sheet and the double-sided tape were patterned together using UV laser. The patterned layer was transferred to the single-side polyimide tape. For insulation layer, electrode site opened single-sided polyimide tape was prepared. Polydimethylsiloxane was used as an adhesion layer, and alignment between electrode sites and opening sites was processed manually. The minimum line width achieved through the proposed fabrication process was approximately 100 \(\mu\)m, and the sheet resistance of the conductive layer was 0.635 \(\Omega\)/sq. Measured cathodal charge storage capacity was 0.72 mC/cm\(^2\) and impedance at 1 kHz was 4.07 k\(\Omega\)/cm\(^2\). Validation of fabricated electrode was confirmed by conducting 30 days accelerated soak test, flexibility test, adhesion test and ex vivo stimulation test. The novel flexible neural electrodes based on single-sided polyimide tape and UV laser patterned gold sheet was fabricated successfully. Conventional neural electrode fabrication processes based on polyimide substrate has a disadvantages such as long fabrication time, expensive costs, and probability of delamination between layers. However, the novel fabrication process which we introduced can overcome many shortcomings of existing processes, and offers great advantages such as simplicity of fabrication, inexpensiveness, flexibility and long-term reliability.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
In recent decades, various neural prosthetic devices have been developed to treat neurological diseases. Representative examples include cochlear implants for hearing restoration, deep brain stimulation for the treatment of brain disorders, cardiac implants for patients with electrical problems of heart, and retinal prostheses to restore the vision field [1,2,3,4,5,6]. To develop such a neural prosthetic device, it is important to design nerve electrodes that transmit electrical signals to the desired nerve cells. Furthermore, for the stable operation of the neural electrodes in vivo environment, an appropriate material should be used as a substrate. A variety of materials, including metal, silicon, glass, and polymer have been used in neural prosthetic devices, and polymers have recently received the most attention due to their flexibility, lightness, and ease of processing [7,8,9].
The first decision to make when designing a polymer based electrode is which polymer to use as the substrate. To achieve this, various factors that can affect the properties of neural electrodes, such as cytotoxicity, water absorption rate, density, Young’s modulus, and permeability must be considered. Past research has shown that polymers such as polyimide, parylene-C, and polydimethylsiloxane (PDMS) are biocompatible substrates for neural electrodes [10,11,12,13,14].
Among them, polyimide is the most widely used material for neural electrodes due to its negligible cytotoxicity and an appropriate level of Young’s modulus and density. Moreover, a thin polyimide multilayer film can be manufactured using spin coating, making it easy to achieve a desired thickness or apply existing MEMS (micro-electro-mechanical systems) processes [15, 16]. Due to these advantages, various neural electrodes using polyimide as a substrate have been manufactured.
However, delamination between metal-polyimide or polyimide-polyimide is a major disadvantage of polyimide-based neural electrodes, which occurs due to the relatively high water absorption rate [17,18,19]. It is a serious problem that affects the long-term reliability due to a significant decrease in the adhesion of the interface over time in an in vivo environment. In order to overcome this, various complementary methods have been proposed, such as controlling the curing temperature of the polyimide layer, increasing adhesion by forming a silicon carbide or titanium layer at the electrode area, or preventing delamination between layers due to moisture absorption. [20,21,22,23]. However, these methods make the fabrication process much more complex and increase manufacturing time and cost.
To overcome the delamination and complex fabrication of the existing polyimide-based neural electrodes, in the previous paper, the authors reported the fabrication process with laser patterned double-sided tape and gold sheet for the fabrication of the neural electrodes [24]. However, the uniformity and the stability of manufactured results were not at a satisfying level and proper level of validation had not proceeded. In this study, to overcome the shortcomings of the traditional polyimide-based neural electrode fabrication process and our previous research, a new polyimide tape-based process is proposed. Although several researchers have fabricated polyimide tape-based sensors in the past, the fabrication and verification of neural electrodes array are the first of their kind [25, 26]. Furthermore, the proposed process can fabricate neural electrodes much faster and cheaper than existing processes, while significantly suppressing delamination at the interface between layers. The electrode fabricated in this way can be used as a nerve stimulating electrode at various locations in the form of planner. Also it is expected that if an appropriate anchoring method is introduced, it can be used as a cuff electrode by wrapping the nerve such as vagus nerve or sciatic nerve due to its high flexibility.
In section 2, the materials and the equipment used in our fabrication process were introduced. Also, the detail of the updated fabrication process was described. The whole process of the fabrication process was introduced step by step. In section 3, the evaluation of the suggested fabrication process was presented. First of all, to estimate the performances of the fabrication process, minimum line width, line resistance, and sheet resistance were measured. Second, to evaluate the performance of fabricated electrodes, cathodal charge storage capacity (CSCc) and the impedance of fabricated electrodes were represented. Next, to show the durability and the resistance to the delamination of the electrodes, a 30-days accelerated soak test, flexibility test, and adhesion test proceeded. The adhesion test of the two interfaces, metal-polyimide and polyimide-polyimide was conducted respectively. Last, to confirm the feasibility of fabricated electrodes, ex-vivo stimulation test was conducted.
2 Materials and methods
2.1 Materials
Materials that are widely used and easy to get were selected to fabricate flexible neural electrodes. Double-sided tape with the conductive acrylic adhesive (YCNW50D, YoungJin Co, South Korea), Single-sided polyimide tape (Alphaflon Co, South Korea), 1 \(\mu\)m thickness gold sheet (DongYang Gold Silver Leaf & Powder Industry, South Korea) with 99.9 % purity and the Polydimethylsiloxane (PDMS, Dow Inc, USA) were used as materials. The structure of double-sided tape was introduced in Fig. 1b. The conductive substrate of double-sided tape is polyester with thin Ni-Cu film plated on both sides. The thickness of the substrate is 30 \(\mu\)m and the thickness of the acrylic adhesive is 10 \(\mu\)m for both sides. Single-sided polyimide tape consists of 25 \(\mu\)m thickness polyimide and 40 \(\mu\)m thickness silicone adhesives. For manufacturing, UV laser (Marcs laser Inc, South Korea), Spin coater (Dong-Ah trade Co, South Korea), O\(_{2}\) plasma treatment system (JP materials, South Korea), Centrifuge (Eppendorf, Germany), and the muffle furnace (SH SCIENTIFIC, South Korea) were used.
2.2 Fabrication process
The basic structure of flexible neural electrodes was represented in Fig. 1a. The whole fabrication process was introduced in Figs. 2 and 3. The overall fabrication time was less than 30 min except curing time of PDMS and the material cost was less than a few dollars.
2.2.1 Preparation of electrode layer
To fabricate flexible neural electrodes with the gold sheet and the double-sided tape, the first priority was to layer the gold sheet and conductive tape uniformly. Due to the very thin and fragile nature of the gold sheet, it was necessary to establish an appropriate adhesion method to attach it without breaking or lifting. Because the van der Waals force between the gold sheet and the tape unintentionally causes the two layers to stick together, the gold sheet must be firmly fixed before attaching with the tape. Therefore, the appropriate adhesion layer was needed and after testing several bio-compatible resins and curing agents, polydimethylsiloxane curing agent (Sylgard 184, silicone elastomer curing agent) was selected as the material for fixing the gold sheet. This worked well as an adhesion layer by firmly fixing the gold sheet to the glass slide without breaking it. The adhesion process was conducted as follows: First, the curing agent of PDMS was coated on the glass slide using spin coater. Since sufficient thickness should be ensured to provide appropriate adhesion, spin coating was performed at 500 rpm for 30 s. Next, the gold sheet was placed gently on top of the adhesion layer. The double-sided conductive tape was attached on the gold sheet evenly. Finally, by removing the release liner of the tape, preparation for laser patterning was ended.
2.2.2 Patterning and transfer
In this fabrication process, a UV laser was used for the electrode patterning. During the laser patterning process, over irradiation may cause problems such as pattern distortion or burning of the pattern edge due to physical damage. To prevent this problems, the condition that could transmit the minimum power to the conductive layer while cutting off at once was found. The patterning conditions were 5 W output power, 22 mm/s laser speed, 30 kHz output frequency, and 10 \(\mu\)s pulse width. After patterning, residue parts should be removed carefully. Single-sided polyimide tape was used to transfer the patterned layer without any distortion. Using the adhesive side of the tape, the gold layer of the pattern was revealed. Since some curing agent of PDMS may remain on the surface of the pattern, cleaning was performed with acetone for 10 s.
2.2.3 Preparation of insulation layer
To open the electrode sites in the insulation layer, they were patterned using the UV laser as Sect. 2.2.2. The pattern of the electrode sites was designed slightly larger than the metal pattern to prevent the overflowing of uncured PDMS over the electrode sites, which will be mentioned later. To clean up the polluted area after patterning, the insulation layer was rinsed in the acetone for 20 s.
2.2.4 Integration for insulation
The electrode layer and the insulation layer should be integrated well to provide uniform sealing. In the initial trial, single-sided polyimide tape was used as an insulation layer directly. However, since both thicknesses of the pattern and the single-sided polyimide tape were similar, precise and uniform sealing with direct packaging was almost impossible. Figure 1c showed the cross-section image of direct integration. The packaged result was not uniform, and a large air bubble existed near the pattern (red circle). This may cause low reliability, so to overcome this issue, the gap between the electrode layer and the insulation layer was filled with the PDMS as shown in Fig. 1d. 10: 1 ratio of the PDMS resin and the PDMS curing agent were mixed uniformly using a centrifuge for 60 s with 2000 rpm. O\(_{2}\) plasma treatment at the laboratory level was conducted to increase the adhesion between the polyimide tape and the PDMS as showun in Fig. 3b and e. The conditions of O\(_{2}\) plasma treatment were 150W power, 13.56 MHz frequency, 180 s, and 20 sccm gas flow rate. The uniformly mixed PDMS was spin-coated on the electrode layer and the insulation layer at 2000 rpm for 30 s and 4000 rpm for 30 s, respectively. Those conditions were selected to provide sufficient thickness to cover the pattern. The two layers coated with PDMS was manually aligned and attached very carefully to prevent the air bubbles from getting trapped between the two layers. Lastly, the curing process was proceeded for 4 h, at 80 \(^\circ\)C.
2.2.5 PDMS etching for site opening
PDMS etching solution for site opening was prepared by mixing Tetrabutylammonium fluoride solution (TBAF solution, Sigma-Aldrich, USA) in N-methyl-2-pyrrolidinone (NMP, Sigma-Aldrich, USA) by 1:3 ratio. The fabricated sample which was made in Sect. 2.2.4 was immersed into the etching solution and stirred at 70 rpm for 20 min. After etching process, the site-opened electrode was rinsed with deionized water. The fabricated results were shown in Fig. 4.
3 Results and discussion
In this section, the fabricated electrode is evaluated from various perspectives. First of all, to check out the distortion during the laser irradiation, the line widths of the pattern after laser patterning were measured. Also, the minimum line width achieved through this fabrication process was confirmed. To prove the performance, sheet resistance of the conductive layer and the line resistance after fabrication were gauged. In addition, cathodic charge storage capacity and impedance of fabricated electrodes were measured. In the case of impedance, measured results right after fabrication and 30 days after being stored in the 75 \(^\circ\)C phosphate-buffered saline (PBS) solution was compared. To show the reliability and the durability of the fabrication process, the accelerated soak test for 30 days and the flexibility test were conducted. Resistance to delamination between the layers was presented using tape test and blister test. Lastly, to make sure it works properly as a neural electrode, ex vivo stimulation test were conducted.
3.1 Line width of the pattern
In our suggested fabrication process, patterning was processed using the UV laser. Since the laser is irradiated in a circled area, rather than a dot, it induced the distortion between the intended laser pattern and the actual pattern. Therefore, differences between the expected pattern and the actual pattern after laser patterning should be measured.
In order to examine how much the laser distorts the intended pattern, the line width from 200\(\mu\)m to 500\(\mu\)m was patternes at intervals of 50 \(\mu\)m, and this was measured using a scanning electron microscope (SEM, HITACHI S-4800). The measurement was conducted 10 time respectively for each line width, and the averaged data were shown in Table 1. As we can see in the table, the difference between actual line width and intended line width was about 60 \(\mu\)m. Since laser engrave both sides of the patterned line, each side was peeled off about 30 \(\mu\)m. In addition, it was confirmed that the minimum line width was about 100 \(\mu\)m. Line widths less than this can also be obtained through laser patterning, but in this case, there is a high possibility of breakage during the transfer process. For convenience, the pattern width spoken from the next paragraph is the width of the expected pattern.
3.2 Line resistance and sheet resistance
The sheet resistance of the conductive layer (multi-layer with the gold sheet and the conductive adhesive) was measured using 4-point probe (FFP-5000, Changmin Inc. South Korea). Measured sheet resistance was 0.635 \(\Omega\)/sq and the uniformity was about 98 \(\%\). Also, line resistance after packaging was measured varying the line width from 200 \(\mu\)m to 500 \(\mu\)m with 100 \(\mu\)m steps. Measurement proceeded using a multimeter and the results of measurements were exploited in Table 2. Each line width was measured 10 times. As shown in the table, the smaller the line width causes the greater variation in resistance. In addition, it was also confirmed that the resistance value was inversely proportional to the line width according to Ohm’s law.
3.3 Charge storage capacity and impedance
The cathodic charge storage capacity and the impedance of fabricated electrodes after whole fabrication process were measured using impedance analyzer (Solatron SI 1260, AMETEK, USA) and electrochemical interface analyzer (Solatron SI 1287, AMETEK, USA). Cyclic voltammograms with 30 cycles, 20 mV/s scan rate were shown in the Fig. 5. In the Fig. 5a, red line represented the cyclic voltammogram of the last cycle. Figure 5b shows the current versus time graph. The CSC\(_{c}\) calculated from the last cycle of the cyclic voltammogram, which is represented with the red line was 0.72 mC/cm\(^2\). This value was slightly larger than other gold electrodes with hermetic sealing. The CSC\(_{c}\) of gold electrode fabricated on cyclic olefin copolymer (COC) substrate was 0.31 mC/cm\(^2\) and on the liquid crystal polymer (LCP) was 0.32 mC/cm\(^2\) [27, 28].
Impedance measurement was proceeded from 100 kHz to 1 Hz, 10 mV amplitude. Measured impedance at 1 kHz after fabrication was 4.07 k\(\Omega\)/cm\(^2\) magnitude and \(-81^\circ\) phase. The normalized results were described in the Fig. 5c and d. Also, after putting electrodes in a PBS solution of 75 \(^\circ\)C for 30 days, impedance of same electrode was measured again. Measured impedance at 1 kHz after 30 days were 0.470 k\(\Omega\)/cm2 magnitude and \(-77^\circ\) phase. Those results were better than the other gold electrodes with hermetic sealing [27, 28]. The superiority of CSC\(_{c}\) and impedance characteristics compared to the conventional fabrication processes using metal deposition comes from the high roughness of gold sheet. Compared to deposited gold, the relatively low flatness of the gold sheet and the adhesive had the effect of increasing the effective surface area. This increase made the CSC\(_{c}\) and impedance characteristics better.
3.4 Accelerated soak test
To estimate the long-term reliability of our fabrication process, 30-days accelerated soak test was conducted. Two interdigitated electrode patterns (IDE) with line width and line spacing of 400 \(\mu\)m were vertically arranged, and only one was immersed in phosphate-buffered saline solution (PBS) for use as a target. The other one was used as a reference without being immersed. The temperature of the PBS was maintained at 75 \(^\circ\)C, and the applied voltage was 5V. Measurement was proceeded using a data acquisition multi-meter system (DAQ6510, Keithley, USA) and picoammeter (6485 picoammeter, Keithley, USA) for 30 days. For the reliability of the results, the experiment was conducted five times and the averaged result was shown in Fig. 6 with standard deviation. We also showed a rough estimate of IDE soak test results based on different polymer substrates reported in previous studies in the same figure [15, 25] for comparison. Each experiment has a different starting point because the equipment and settings used in each experiment are different. However, comparing the tendency of each experimental results, the water absorption rate and long-term reliability of each polymer substrate could be analyzed. Our results were almost similar with cyclic olefin copolymer (COC), which is known to provide hermetical sealing and the results were taken by the same experiment environments. Also, the leakage current value after 30 days was almost the same as the liquid crystal polymer (LCP) which is known as providing hermetic sealing and was much better than parylene-C or polyimide, which has a significantly poor water absorption rate. By applying the 10-degree rule, 30 days in 75 \(^\circ\)C could be roughly estimated to 480 days in human body temperature (37 \(^\circ\)C).
We were able to demonstrate two facts from this experiment: the low moisture permeability of polyimide tape and the strong adhesion between PDMS and polyimide tape. Even though PDMS of several tens of micro-thickness has almost no resistance to moisture, polyimide tape, which has high moisture resistance prevented the leakage current from increasing during the test period. In addition, when PDMS is adhered to a polymer substance such as parylene-C and exposed to an in vitro environment, the interface is usually broken within a short period of time. However, the bond between the interfaces did not break throughout this experiment. This means that the bond between the PDMS and the polyimide tape is strong, and the blister test was conducted in Sect. 3.6 to confirm the adhesion strength between the interfaces.
3.5 Flexibility test
To verify the reliability of our fabricated results in the flexible application, two kinds of flexibility tests were performed. First, to represent the resistance variance during bending, the R-bending test was conducted using a 1-axis motion controller (STM-1-TS, ST1 corp. South Korea) in Fig.7a. Second, to check out the failure in a folding environment, the folding test proceeded using a 1-axis motion controller (STM-1-USB, STI corp, South Korea) in Fig.7b.
For R-bending test, the radius of bending (R) was varied from 5 mm to 10 mm with 1 mm step and the line width was 500 \(\mu\)m. For each radius, resistance was checked every 100 cycles, and measurements were conducted up to 1000 cycles. The speed of bending test was 100 mm/s. The ratio between the measured resistance of the first bending (Rf) and the measured resistance of the bending at every 100 cycles (Rc) were represented in Fig. 7c. No electrical failure was founded during the measurement and the increase of resistance was in the reasonable range. However, the increase in resistance did not return to its original state even if measured in a planar state after finishing the test. The reason for those increases was due to the non-uniformity of the PDMS layer while attaching the conductive layer to the substrate. Due to the imperfections during fabrication, the fabricated result is uneven in thickness. So, bending stress was collected at the thinnest point during bending, causing damage in that part. Those imperfections should be improved in future work.
For folding tests, the folding angle was set to 90 degrees in in-folding and 90 degrees in out-folding, and a total of 10,000 folding was performed. 30 mm long 10 line patterns with line width of 300 \(\mu\)m and 400 \(\mu\)m respectively, were prepared. The standard for electrical failure of line resistance was set over 100 \(\Omega\) so a resistance greater than 100 \(\Omega\) after folding was considered to be broken. For each degrees and line width, the experiment was conducted for 5 times respectively, and the average results were shown in Fig 7d. If the line width is thin, which is 300 \(\mu\)m, as shown in figure, the possibility of electrical failure increases. Also, there was a slightly higher probability of failure in the in-folding situation. This difference in probability of failure according to the folding angle seems to be because the conductive layer is not placed on the neutral plane. If the conductive layer can be accurately located on the neutral plane by improvement of fabrication process, this difference of failure probability seems reducible. When the line width was large at 400 \(\mu\)m, 10,000 times of folding did not significantly affect the electrical connection of the electrode.
3.6 Adhesion test
To show that the fabricated electrode is resistant to inter-layer delamination, it is necessary to confirm that the adhesive force between each layer is sufficient. To check out this, two tests were conducted. First, a tape test was performed to examine the adhesion between the conductive layer (combination of gold sheet and double-sided tape) and the polyimide tape. Next, a blister test was conducted to confirm that the adhesion between the substrate and packaging layer was large enough. For the tape test, a 1 cm square size conductive layer put on the polyimide tape was prepared. To show that the bonding was maintained in the in-vitro environment, the prepared sample was immersed in PBS solution at 75 \(^\circ\)C for 30 days. The tape test proceeded with scotch tape. (3 M, USA) During the tape test, no delamination or peeled-off particles were found. From this, it was confirmed that the adhesion between the conductive layer and the substrate was strong enough to withstand the internal body environment. For blister test, the sample with 3 mm size blister, the holder with 2 mm hole, and N\(_{2}\) gas injector with pressure controller were prepared. The blister of the sample was aligned to the center of the hole and the sample was fixed with instant glue (Loctite SG Easy Brush, Loctite, Germany). Then, the pressure applied to the blister was gradually increased until the destruction of bonding between layers was observed. Also, to demonstrate the delamination of packaging does not occur in the in-vitro environment, the samples placed in the PBS solution at 75 \(^\circ\)C for 30 days were tested. All samples showed the results that the instant glue layer and sample detached before the delamination between PDMS-polyimide tape occurred. The detachments between the instant glue and the samples occurred when a pressure of about 150 psi was applied. Through these results, it can be seen that the critical pressure at which adhesion between PDMS-polyimide tapes is destroyed was greater than 150 psi even after 4 weeks of exposure to a 75 \(^\circ\)C PBS solution. The corresponding value is significantly higher than 130 psi, which is the critical pressure of the polyimide-polyimide interface in the same environment [19]. Therefore, it was confirmed that the interfacial adhesion was significantly improved through our suggested process.
3.7 Ex-vivo stimulation test
To confirm the feasibility of the fabricated electrode, an ex-vivo stimulation test was conducted. The retina of mouse (C57BL/6N, male, 8 weeks, KOATECH, Pyeongtaek-si, Korea) was used for the stimulation test. All animal procedures were approved by the Institutional Animal Care and Use Committee (IACUC) at Ewha Womans University (IACUC 22-007). The setup for the ex-vivo test was described in Fig. 8a. Before the test, to harvest the fresh retina from the mouse was anesthetized through inhalation using isoflurane for 5 min. After the cervical dislocation, both eyes were enucleated and the whole layers of retina were harvested. The retinal patch was prepared under dim red light in artificial cerebrospinal fluid (aCSF) (124 mM NaCl, 5 mM KCl, 1.15 mM KH2PO4, 1.15 mM MgSO4, 10 mM Glucose, 25 mM NaHCO3, 5 mM CaCl2) bubbled with 95% O2 and 5% CO2 to maintain a pH of 7.3 \(\sim\) 7.4 at room temperature. Then, for the neural recording, prepared retinal patch was loaded on the 60 channel recording MEA (perforated microelectrode array; 60pMEA200/30iR-Ti, 30 \(\mu\)m of electrode site diameter and 200 \(\mu\)m of spacing; Multi Channel Systems MCS GmbH, Reutlingen, Germany). In order to perform subretinal stimulation, the ganglion cells of the retina were loaded to be positioned downwards. By putting the electrodes on the retina, like Fig. 8b, preparations for ex-vivo stimulation test was finished.
The results of stimulation test were plotted from Fig. 8c–e. For the stimulation, cathodic-first biphasic pulse was generated from stimulus generator (multi channel systems MCS GmbH, Reutlingen, Germany). The voltage stimulation was used and the stimulation parameters were set as follows: Phase amplitude: 3 V, Phase duration: 500 \(\mu\)s, inter-pulse interval: 999 ms, stimulation frequency: 1 Hz. In Fig. 8c, evoked response of retinal ganglion cell (RGC) after stimulation was described. Because the size of the stimulation electrode is larger than the size of the RGC, a single stimulation activates more than one RGC. Therefore, multiple RGCs respond to a single nerve stimulus, and these responses were spike-sorted using a threshold voltage of -32 \(\sim\) -12 \(\mu\)V, as represented in Fig. 8d. In addition, in Fig. 8e, it is shown in the post-stimulus time histogram (PSTH) that the response of the RGC exhibits a clear contrast before and after the stimulation moment (0 s). This confirms that the fabricated neural electrode can successfully stimulate all the nearby nerves in the site.
4 Conclusion
In this study, the fabrication process and validation of flexible neural electrodes using single-sided polyimide tape and gold sheet patterned with UV laser were introduced. The charge storage capacity and the impedance of the fabricated neural electrodes were at an appropriate level, and they showed outstanding performance in the flexibility test. To verify the long-term reliability of the electrode, the accelerated soak test to measure leakage current and the adhesion test to confirm the resistant to delamination were performed, and test results showed excellent reliability. Also, the feasibility of fabricated electrodes was investigated through ex-vivo stimulation test.
The proposed process only required less than 15 min of fabrication time except for the PDMS curing, and the material and process cost were very low. In addition, unlike the conventional MEMS process, the use of substances that can cause environmental pollution was minimized.
The introduced process method has several areas for improvement. First, the minimum line width over 100 \(\mu\)m should be improved. The minimum line width using direct patterning of UV laser seems to be close to about 50 \(\mu\)m, but there was considerable difficulty in achieving a minimum line width of 100 \(\mu\)m or less due to damage in the transfer process. Second, the aligning method that is currently being done by hand must be changed. It is almost impossible to perform alignment at line widths less than 100 \(\mu\)m by manual alignment. Finally, long term reliability over a longer period of time should be verified in an in vivo environment. These points will be reported in further studies.
References
Zeng F-G, Rebscher S, Harrison W, Sun X, Feng H. Cochlear implants: system design, integration, and evaluation. IEEE Rev Biomed Eng. 2008;1:115–42.
Wilson BS, Dorman MF. Cochlear implants: current designs and future possibilities. J Rehabil Res Dev. 2008;45(5):695–730.
Perlmutter J. Deep brain stimulation. Annu Rev Neurosci. 2006;29:229.
Dostrovsky JO, Lozano AM. Mechanisms of deep brain stimulation. Mov Disord Off J Mov Disord Soc. 2002;17(S3):63–8.
Weiland JD, Liu W, Humayun MS. Retinal prosthesis. Annu Rev Biomed Eng. 2005;7:361–401.
Margalit E, Maia M, Weiland JD, Greenberg RJ, Fujii GY, Torres G, Piyathaisere DV, O’Hearn TM, Liu W, Lazzi G, et al. Retinal prosthesis for the blind. Surv Ophthalmol. 2002;47(4):335–56.
Hong G, Lieber CM. Novel electrode technologies for neural recordings. Nat Rev Neurosci. 2019;20(6):330–45.
Cogan SF. Neural stimulation and recording electrodes. Annu Rev Biomed Eng. 2008;10:275–309.
Kotov NA, Winter JO, Clements IP, Jan E, Timko BP, Campidelli S, Pathak S, Mazzatenta A, Lieber CM, Prato M, et al. Nanomaterials for neural interfaces. Hoboken: Wiley; 2009.
Rousche PJ, Pellinen DS, Pivin DP, Williams JC, Vetter RJ, Kipke DR. Flexible polyimide-based intracortical electrode arrays with bioactive capability. IEEE: Piscataway; 2001.
Stieglitz T, Beutel Hr, Schuettler M, Meyer JU. Micromachined, polyimide-based devices for flexible neural interfaces. Biomed Microdevices. 2000;2, 283–294.
Hsu JM, Rieth L, Normann RA, Tathireddy P, Solzbacher F. Encapsulation of an integrated neural interface device with Parylene C, IEEE. 2008.
Rodger DC, Fong AJ, Li W, Ameri H, Ahuja AK, Gutierrez C, Lavrov I, Zhong H, Menon PR, Meng E, et al. Flexible parylene-based multielectrode array technology for high-density neural stimulation and recording. Sens Actuators, B Chem. 2008;132(2):449–60.
Guo L, Guvanasen GS, Liu X, Tuthill C, Nichols TR, DeWeerth SP. A PDMS-based integrated stretchable microelectrode array (isMEA) for neural and muscular surface interfacing. IEEE Trans Biomed Circuits Syst. 2012;7(1):1–10.
Richardson R Jr, Miller J, Reichert WM. Polyimides as biomaterials: preliminary biocompatibility testing. Biomaterials. 1993;14(8):627–35.
Seo J-M, Kim SJ, Chung H, Kim ET, Yu HG, Yu YS. Biocompatibility of polyimide microelectrode array for retinal stimulation. Mater Sci Eng, C. 2004;24(1–2):185–9.
Constantin CP, Aflori M, Damian RF, Rusu RD. Biocompatibility of polyimides: a mini-review. Materials. 2019;12(19):3166.
Kim E, Lee S, Seo J-M, Chung H, Kim S. Long-term reliability of polyimide electrode array in rabbit retina. Investig Ophthalmol Vis Sci. 2008;49(13):1773–1773.
Lee S, Kim E, Kim S. Application of the blister test to assess reliability of polyimide based retinal electrode. Investig Ophthalmol Vis Sci. 2007;48(13):664–664.
Kim ET, Kim C, Lee SW, Seo J-M, Chung H, Kim SJ. Feasibility of microelectrode array (MEA) based on silicone-polyimide hybrid for retina prosthesis. Investig Ophthalmol Vis Sci. 2009;50(9):4337–41.
Schander A, Gancz JM, Tintelott M, Lang W. Towards long-term stable polyimide-based flexible electrical insulation for chronically implanted neural electrodes. Micromachines. 2021;12(11):1279.
Ordonez JS, Boehler C, Schuettler M, Stieglitz T. Improved polyimide thin-film electrodes for neural implants, 2012;5134–5137. IEEE.
Ceyssens F, Puers R. Insulation lifetime improvement of polyimide thin film neural implants. J Neural Eng. 2015;12(5):054001.
Jeong H, Seo J. Electrode design on plastic substrates using laser patterned double-sided tape and gold leaf. In: 2019 41st annual international conference of the IEEE engineering in medicine and biology society (EMBC), 2019; pp 1286–1289. IEEE.
Santos MSF, Ameku WA, Gutz IGR, Paixão TRLC. Gold leaf: From gilding to the fabrication of disposable, wearable and low-cost electrodes. Talanta. 2018;179:507–11.
Alam AU, Clyne D, Lush W, Deen MJ. A reusable, reagent-less free chlorine sensor using gold thin film electrode. Analyst. 2021;146(8):2626–31.
Baek C, Kim J, Lee Y, Seo J. Fabrication and evaluation of cyclic olefin copolymer based implantable neural electrode. IEEE Trans Biomed Eng. 2020;67(9):2542–51.
Lee SW, Min KS, Jeong J, Kim J, Kim SJ. Monolithic encapsulation of implantable neuroprosthetic devices using liquid crystal polymers. IEEE Trans Biomed Eng. 2011;58(8):2255–63.
Acknowledgements
This research was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science and ICT(NRF-2017R1A5A1015596), Ministry of Education (NRF-2022R1I1A4063209), and the Korea Medical Device Development Fund grant funded by the Korea government (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health and Welfare, the Ministry of Food and Drug Safety (Project Number: 1711139110, KMDF-PR-20210527-0006).
Funding
Open Access funding enabled and organized by Seoul National University. The authors have not disclosed any funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Ethics approval
All animal procedures were approved by the Institutional Animal Care and Use Committee (IACUC) at Ewha Womans University (IACUC 22-007).
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Jeong, H., Lee, T., Kim, J. et al. Fabrication and validation of flexible neural electrodes based on polyimide tape and gold sheet. Biomed. Eng. Lett. 14, 267–278 (2024). https://doi.org/10.1007/s13534-023-00345-0
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13534-023-00345-0