Skip to main content
Log in

A compact-sized surface EMG sensor for myoelectric hand prosthesis

  • Original Article
  • Published:
Biomedical Engineering Letters Aims and scope Submit manuscript

Abstract

Myoelectric prosthesis requires a sensor that can reliably capture surface electromyography (sEMG) signal from amputees for its controlled operation. The main problems with the presently available EMG devices are their extremely high cost, large response time, noise susceptibility, less amplitude sensitivity, and larger size. This paper proposes a compact and affordable EMG sensor for the prosthetic application. The sensor consists of an electrode interface, signal conditioning unit, and power supply unit all encased in a single package. The performance of dry electrodes employed in the skin interface was compared with the conventional Ag/AgCl electrodes, and the results were found satisfactory. The envelope detection technique in the sensor based on the tuned RC parameters enables the generation of smooth, faster, and repeatable EMG envelope irrespective of signal strength and subject variability. The output performance of the developed sensor was compared with commercial EMG sensor regarding signal-to-noise ratio, sensitivity, and response time. To perform this, EMG data with both devices were recorded for 10 subjects (3 amputees and 7 healthy subjects). The results showed 1.4 times greater SNR values and 45% higher sensitivity of the developed sensor than the commercial EMG sensor. Also, the proposed sensor was 57% faster than the commercial sensor in producing the output response. The sEMG sensor was further tested on amputees to control the operation of a self-designed 3D printed prosthetic hand. With proportional control scheme, the myoelectric hand setup was able to provide quicker and delicate grasping of objects as per the strength of the EMG signal.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
Fig. 11
Fig. 12
Fig. 13

Similar content being viewed by others

References

  1. Day S. Important factors in surface EMG measurement. Calgary: Bortec Biomedical Ltd Publishers; 2002. p. 1–7.

    Google Scholar 

  2. Tavakoli M, Benussi C, Lourenco JL. Single channel surface EMG control of advanced prosthetic hands. Expert Syst Appl. 2017;79:322–32.

    Article  Google Scholar 

  3. Liu J, Zhou P. A novel myoelectric pattern recognition strategy for hand function restoration after incomplete cervical spinal cord injury. IEEE Trans Neural Syst Rehabil Eng. 2013;21:96–103.

    Article  Google Scholar 

  4. Pancholi S, Joshi AM. Portable EMG data acquisition module for upper limb prosthesis application. IEEE Sens J. 2018;18:3436–43.

    Article  Google Scholar 

  5. Phinyomark A, Phukpattaranont P, Limsakul C. Fractal analysis features for weak and single-channel upper-limb EMG signals. Expert Syst Appl. 2012;39:11156–63.

    Article  Google Scholar 

  6. Baek J-Y, An J-H, Choi J-M, Park K-S, Lee S-H. Flexible polymeric dry electrodes for the long-term monitoring of ECG. Sens Actuators A. 2008;143:423–9.

    Article  Google Scholar 

  7. Pylatiuk C, Muller-Riederer M, Kargov A, Schulz S, Schill O, Reischl M, et al. Comparison of surface EMG monitoring electrodes for long-term use in rehabilitation device control. In: 2009 IEEE international conference on rehabilitation robotics [Internet]. Kyoto, Japan: IEEE; 2009 [cited 2018 Dec 3]. p. 300–4.

  8. Searle A, Kirkup L. A direct comparison of wet, dry and insulating bioelectric recording electrodes. Physiol Meas. 2000;21:271.

    Article  Google Scholar 

  9. Laferriere P, Lemaire ED, Chan ADC. Surface electromyographic signals using dry electrodes. IEEE Trans Instrum Meas. 2011;60:3259–68.

    Article  Google Scholar 

  10. Jamal MZ, Kim K-S. A finely machined toothed silver electrode surface for improved acquisition of EMG signals. In: 2018 IEEE Sensors Applications Symposium (SAS) [Internet]. Seoul: IEEE; 2018 [cited 2018 Dec 3]. p. 1–5.

  11. Parker P, Englehart K, Hudgins B. Myoelectric signal processing for control of powered limb prostheses. J Electromyogr Kinesiol. 2006;16:541–8.

    Article  Google Scholar 

  12. Asghari Oskoei M, Hu H. Myoelectric control systems—a survey. Biomed Signal Process Control. 2007;2:275–94.

    Article  Google Scholar 

  13. Khushaba RN, Al-Timemy A, Kodagoda S, Nazarpour K. Combined influence of forearm orientation and muscular contraction on EMG pattern recognition. Expert Syst Appl. 2016;61:154–61.

    Article  Google Scholar 

  14. Lenzi T, De Rossi SMM, Vitiello N, Carrozza MC. Proportional EMG control for upper-limb powered exoskeletons. In: 2011 annual international conference of the IEEE Engineering in Medicine and Biology Society [Internet]. Boston: IEEE; 2011 [cited 2018 Dec 3]. p. 628–31.

  15. Fougner A, Stavdahl O, Kyberd PJ, Losier YG, Parker PA. Control of upper limb prostheses: terminology and proportional myoelectric control—a review. IEEE Trans Neural Syst Rehabil Eng. 2012;20:663–77.

    Article  Google Scholar 

  16. Herle S, Man S, Lazea G, Raica P. Myoelectric control strategies for a human upper limb prosthesis. J Control Eng Appl Inf. 2012;14(1):58–66.

    Google Scholar 

  17. Hudgins B, Parker P, Scott RN. A new strategy for multifunction myoelectric control. IEEE Trans Biomed Eng. 1993;40:82–94.

    Article  Google Scholar 

  18. Chowdhury RH, Reaz MBI, Ali MABM, Bakar AAA, Chellappan K, Chang TG. Surface electromyography signal processing and classification techniques. Sensors (Basel). 2013;13:12431–66.

    Article  Google Scholar 

  19. Farina D, Merletti R, Indino B, Graven-Nielsen T. Surface EMG crosstalk evaluated from experimental recordings and simulated signals. Reflections on crosstalk interpretation, quantification and reduction. Methods Inf Med. 2004;43:30–5.

    Article  Google Scholar 

  20. Andrade AO, Soares AB, Nasuto SJ, Kyberd PJ. EMG decomposition and artefact removal. In: Computational intelligence in electromyography analysis—a perspective on current applications and future challenges. IntechOpen; 2012.

  21. Shobaki MM, Malik NA, Khan S, Nurashikin A, Haider S, Larbani S, et al. High quality acquisition of surface electromyography—conditioning circuit design. In: IOP conference series: materials science and engineering, vol. 53. 2013. p. 012027.

    Article  Google Scholar 

  22. Agostini V, Knaflitz M. An algorithm for the estimation of the signal-to-noise ratio in surface myoelectric signals generated during cyclic movements. IEEE Trans Biomed Eng. 2012;59:219–25.

    Article  Google Scholar 

  23. Supuk TG, Skelin AK, Cic M. Design, development and testing of a low-cost sEMG system and its use in recording muscle activity in human gait. Sensors (Basel). 2014;14:8235–58.

    Article  Google Scholar 

  24. Imtiaz U, Bartolomeo L, Lin Z, Sessa S, Ishii H, Saito K, et al. Design of a wireless miniature low cost EMG sensor using gold plated dry electrodes for biomechanics research. In: 2013 IEEE international conference on mechatronics and automation. 2013. p. 957–62.

  25. Drost G, Stegeman DF, van Engelen BGM, Zwarts MJ. Clinical applications of high-density surface EMG: a systematic review. J Electromyogr Kinesiol. 2006;16:586–602.

    Article  Google Scholar 

  26. Farina D, Jiang N, Rehbaum H, Holobar A, Graimann B, Dietl H, et al. The extraction of neural information from the surface EMG for the control of upper-limb prostheses: emerging avenues and challenges. IEEE Trans Neural Syst Rehabil Eng. 2014;22:797–809.

    Article  Google Scholar 

  27. Farrell TR, Weir RF. The optimal controller delay for myoelectric prostheses. IEEE Trans Neural Syst Rehabil Eng. 2007;15:111–8.

    Article  Google Scholar 

  28. Milosevic B, Benatti S, Farella E. Design challenges for wearable EMG applications. In: Design, Automation Test in Europe Conference Exhibition (DATE), 2017; 2017. p. 1432–7.

  29. Electrode|Electrodes|Myo Control Elements|Myo Hands and Components|Upper Limb Prosthetics| Prosthetics|Ottobock US Healthcare [Internet]. [cited 2018 Dec 4]. https://professionals.ottobockus.com/Prosthetics/Upper-Limb-Prosthetics/Myo-Hands-and-Components/Myo-Control-Elements/Electrodes/Electrode/p/13E200~560.

  30. Pizzolato S, Tagliapietra L, Cognolato M, Reggiani M, Müller H, Atzori M. Comparison of six electromyography acquisition setups on hand movement classification tasks. PLoS ONE. 2017;12:e0186132.

    Article  Google Scholar 

  31. MyoWare Muscle Sensor—SEN-13723—SparkFun Electronics [Internet]. [cited 2019 Jan 28]. https://www.sparkfun.com/products/13723.

  32. Gerdle B, Karlsson S, Day S, Djupsjöbacka M. Acquisition, processing and analysis of the surface electromyogram. In: Windhorst U, Johansson H, editors. Modern techniques in neuroscience research. Berlin: Springer; 1999. p. 705–55. https://doi.org/10.1007/978-3-642-58552-4_26.

    Chapter  Google Scholar 

  33. Wang J, Tang L, Bronlund JE. Surface EMG signal amplification and filtering. Int J Comput Appl. 2013;82:15–22.

    Google Scholar 

  34. Rice DA, Venkatachalam V, Wegmann MJ. A simple envelope detector. IEEE Trans Instrum Meas. 1988;37:223–6.

    Article  Google Scholar 

  35. Patla AE. Some characteristics of EMG patterns during locomotion: implications for the locomotor control process. J Mot Behav. 1985;17:443–61.

    Article  Google Scholar 

  36. Medved V, Tomkovic S. Locomotion diagnostics: some neuromuscular and robotic aspects. IEEE Eng Med Biol Mag. 1991;10:23–8.

    Article  Google Scholar 

  37. D’Alessio T, Conforto S. Extraction of the envelope from surface EMG signals. IEEE Eng Med Biol Mag. 2001;20:55–61.

    Article  Google Scholar 

  38. Balbinot A, Favieiro G. A neuro-fuzzy system for characterization of arm movements. Sensors (Basel). 2013;13:2613–30.

    Article  Google Scholar 

  39. FSR Integration Guide—Interlink Electronics|DigiKey [Internet]. [cited 2019 Jun 15]. https://www.digikey.com/en/pdf/i/interlink-electronics/interlink-electronics-fsr-force-sensing-resistors-integration-guide.

  40. Konrad P. A practical introduction to kinesiological electromyography. 61.

  41. Grimnes S. Impedance measurement of individual skin surface electrodes. Med Biol Eng Comput. 1983;21:750–5.

    Article  Google Scholar 

  42. Sinderby C, Lindström L, Grassino AE. Automatic assessment of electromyogram quality. J Appl Physiol. 1995;79:1803–15.

    Article  Google Scholar 

  43. Thuau D, Abbas M, Chambon S, Tardy P, Wantz G, Poulin P, et al. Sensitivity enhancement of a flexible MEMS strain sensor by a field effect transistor in an all organic approach. Org Electron. 2014;15:3096–100.

    Article  Google Scholar 

  44. Rodrigues DMC, Lopes RN, Franco MAR, Werneck MM, Allil RCSB. Sensitivity analysis of different shapes of a plastic optical fiber-based immunosensor for Escherichia coli: simulation and experimental results. Sensors. 2017;17:2944.

    Article  Google Scholar 

  45. Greene EJ, Lo PH. Method for measuring RF pulse rise time, fall time and pulse width. United States patent US 5,805,460. 1998.

  46. Englehart K, Hudgin B, Parker PA. A wavelet-based continuous classification scheme for multifunction myoelectric control. IEEE Trans Biomed Eng. 2001;48:302–11.

    Article  Google Scholar 

  47. Kargov A, Pylatiuk C, Martin J, Schulz S, Döderlein L. A comparison of the grip force distribution in natural hands and in prosthetic hands. Disabil Rehabil. 2004;26:705–11.

    Article  Google Scholar 

  48. Geethanjali P. Myoelectric control of prosthetic hands: state-of-the-art review. Med Devices (Auckl). 2016;9:247–55.

    Google Scholar 

  49. Belter JT, Segil JL, Dollar AM, Weir RF. Mechanical design and performance specifications of anthropomorphic prosthetic hands: a review. J Rehabil Res Dev. 2013;50:599.

    Article  Google Scholar 

Download references

Acknowledgements

The authors would like to thank the Design Innovation Centre, Indian Institute of Technology (BHU) for funding this project.

Funding

This research work was funded by Design Innovation Centre, Indian Institute of Technology (BHU).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Alok Prakash.

Ethics declarations

Conflict of interest

The authors declare that there are no conflicts of interest.

Ethical approval (involvement of animals)

This article does not contain any studies with animals performed by any of the authors.

Ethical approval (involvement of human subjects)

This article involves surface EMG data acquisition from various human subjects. Ethical approval was taken from the Ethical committee, Institute of medical sciences, BHU, Varanasi before performing this experiment. All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Informed consent

Informed consent was obtained from all individual participants included in the study.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Prakash, A., Sharma, S. & Sharma, N. A compact-sized surface EMG sensor for myoelectric hand prosthesis. Biomed. Eng. Lett. 9, 467–479 (2019). https://doi.org/10.1007/s13534-019-00130-y

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13534-019-00130-y

Keywords

Navigation