INTRODUCTION

Animal models are frequently used in the study of various central nervous system (CNS) disorders, such as vascular, inflammatory, neurodegenerative, and neoplastic diseases. These disease models may represent either selected aspects or, occasionally, the full spectrum of the corresponding human disease. Mouse colonies are relatively easy to maintain, and the universal availability of transgenic animals coupled with the capability of generating new, genetically modified mice from well-characterized “standard” strains represents a major advantage in identifying and exploring pathogenic mechanisms relevant to CNS diseases.

One of the most common approaches in biomedical research is to visualize tissue compartments or cellular interactions in various tissues. In this context, standard optical microscopy has two major limitations: 1) only ex vivo tissues can be studied at high resolution and 2) the degree of specificity varies depending on the staining method used. Newer intravital microscopy techniques are also emerging and have become very useful tools for in vivo monitoring of cellular motion and cell–cell interactions. The main advantages of intravital microscopy are high microscopic resolution, multiplex and real-time imaging with good image quality, and the capability of tracking of various labeled cell types. However, there are also limitations, including the required surgical window, smaller field of view, and limited tissue penetration.

Because most disease processes are dynamic, noninvasive in vivo imaging modalities are of great advantage as they allow for investigations to be performed at multiple different time points. Small animal magnetic resonance imaging (MRI) studies are usually preclinical: they establish or study new aspects of a disease process that have not yet been clarified in the human disease. Small animal MRI also allows for translational projects in which the pathomechanism of human MRI findings are studied. The first report of MR imaging of a rat was published over 30 years ago [1]. MRI has the capability of studying live organisms without exposing them to potentially harmful ionizing radiation. Besides anatomical imaging, MRI is also capable of providing physiological information about several important aspects of biological processes, including circulation and cerebrospinal fluid flow, cerebral blood flow and volume, activity mapping with functional MRI (fMRI) or Mn++-based techniques, metabolite distribution with chemical shift imaging, diffusion or perfusion properties of the studied tissue, or in vivo pH measurement via phosphorus MR spectrometry (MRS) [2, 3].

TECHNICAL ASPECTS OF SMALL ANIMAL MRI

Many universities and research institutions have acquired small animal MR imaging systems, most commonly as part of a core facility. Most research facilities use 4.7– to 11.7–Tesla (T) narrow-bore magnets, of which the 7T magnet is the most commonly utilized. The two leading manufacturers of these systems are Bruker Biospin (Ettlingen, Germany) and Varian Medical Systems (Palo Alto, CA). Both vertical and horizontal bore systems are available. Vertical bore neutron magnetic resonance spectrometers equipped with gradient coils and imaging probes are the most common systems used in biochemistry-based core facilities, whereas in radiology research facilities, horizontal bore magnets are more common. Obviously, the physical principles remain the same regardless of the orientation of the bore. However, the horizontal bore system often allows for easier monitoring of the mice and larger flexibility in customizing the probes and coils. Most facilities employ their own engineers and technicians. Larger facilities often develop not only their own pulse sequences and post-processing software tools, but also design hardware, including radio frequency (RF) coils and rodent head stabilizing probes. Most commonly, only the field strength is mentioned when the differences between imaging systems is discussed. However, the utilized RF and gradient coils are equally important and can literally determine the success of an experimental MRI research program. Custom-made coils specific to the studied research question often have a significantly better signal-to-noise ratio (SNR) than stock coils.

While MR image acquisition is a very complex process, image analysis is often far more time consuming and computation intense. Image processing and visualization tools are included with the software provided by the scanner manufacturer, but they are generally insufficient to provide the desired detailed answers. Several commercially available biomedical image post-processing software packages are available, and many research institutions develop their own software applications. Notable examples of image analysis software are Analyze [4, 5], Statistical Parametric Mapping [6], the freely downloadable National Institutes of Health image [7, 8], and Analysis of Functional NeuroImages [9]. Most of these provide basic analysis methods, including intensity and region of interest-based area or volume measurements. They also include tools for slice extraction from 3–dimensional (3D) volumes and for a variety of 3D visualizations. Several offer segmentation, coregistration, and other tools for complex analyses requiring image algebra methods, including diffusion tensor imaging (DTI) visualization, perfusion calculations, or fMRI visualization tools.

SPECIAL REQUIREMENTS FOR SMALL RODENT MRI

MRI requires the study subjects to be completely immobilized, and the importance of this cannot be stressed enough. In order to achieve this goal, general anesthesia (in most cases, inhalational) must be used in small animal experiments [10]. Unlike injectable agents, inhalational anesthesia is administered continuously and can be calibrated to the specific requirements posed by the animal and the experiment. The anesthesia may in some cases alter the studied biological processes and/or the desired imaging outcome. This is especially true for cerebral blood flow/blood volume and fMRI studies [11]. Animal monitoring under anesthesia is also an important requirement and includes monitoring of core temperature, respiratory rate, and electrocardiogram. Oxygen saturation and CO2 level monitors are also available. Monitoring is also important from the standpoint of gated acquisition, which is often needed to eliminate respiration or cardiac motion-related artifacts during acquisition. The most commonly used systems are manufactured by SA Instruments (Stony Brook, NY; http://www.i4sa.com/).

The core temperature inside narrow-bore scanners can be in the low 50°F range (15–20°C), and various heating systems are utilized to maintain the normal core temperature of the animals while in the scanner. These are usually supplied by the imaging systems vendors. Most utilize either warm airflow or circulating water beds.

In addition to the anesthesia and maintaining a suitable core temperature, another important factor in successful small rodent MRI is that the probe should keep the animal stable during the imaging session [12]. Most small animal systems are shipped with mouse and rat probes, but many laboratories develop custom-made holders for specific applications and/or for added stability. A good probe should achieve maximal stabilization without any trauma to the animal. Ideal animal holders should also allow for easily reproducible stereotactic placement.

DIFFERENCES BETWEEN SMALL ANIMAL IMAGING AND HUMAN MRI: ADVANTAGES AND DISADVANTAGES OF THE HIGHER FIELD STRENGTH

The study of experimental animals with MRI differs in more ways than just the smaller scale that rodent imaging requires. However, the smaller scale (of a mouse CNS) itself is a major problem to overcome (FIG. 1) Most small animal magnets operate at a high field strength, namely, in the range of 4.7–11.7T, as opposed to the standard clinical range of 1.5–3T. The main advantage of high field scanners is the increased SNR (FIG. 2); at the same time, increased image artifacts and other limitations make higher field strength scanning technically more challenging. In living organisms, which are comprised of multiple compartments with different physico-chemical properties, artifacts can be generated due to susceptibility and different relaxation properties at the higher field strength [13]. Why certain imaging artifacts are more prominent at the very high field strengths utilized can be best understood in the context of the well-known scaling relationships that describe how SNR, chemical shift, susceptibility variation, specific absorption rate, and RF wavelength vary with increased field strength (Table 1). Each artifact that is seen at 4.7–11.7T can also be seen at 1.5–3T—but not as prominently. Some artifacts are related to the hardware, including spike-noise artifacts, RF interference artifacts (crosstalk between the RF transmit and receive channels, RF penetration due to magnet room shielding issues, especially for vertical bore magnets), and central dot artifacts. Artifacts that are typically more prominent due to increased SNR include aliasing (wraparound) artifacts (also prominent at lower field strengths), Gibbs ringing (truncation artifact), fine line artifact [particularly common in rapid acquisition relaxation enhancement (RARE) sequences], and half field-of-view ghosting (especially with parallel acquisition). Chemical shift- and susceptibility-related artifacts are especially common at high field strengths. Relaxation time effects also need to be considered: the T1 relaxation time of many tissues increases by 20–50% between 1.5 and 3T, while the T2 relaxation time is generally unchanged, although the apparent T2* relaxation time decreases in most tissues. It is good practice to conduct relaxometry experiments in the tissue compartments of interest to sort out these issues in the given system. The specific absorption rate is a key limiting factor at higher field strengths for in vivo experiments. Novel pulse sequences and/or changes in acquisition parameters in an effort to minimize RF power levels are needed to allow for image acquisition without potentially dangerous energy absorption and related heating of the tissue.

FIG. 1
figure 1

Comparison of magnetic resonance imaging (MRI) scans of a human (left) versus mouse (right) brain. Parameters of the clinical MR image: Field of view (FOV), 22 × 22 cm2; matrix size, 256 × 256; slice thickness, 4 mm. The mouse MR image was extracted from a 3–dimensional dataset with parameters: FOV, 3.2 × 1.92 ×  1.92 cm3; matrix size, 256 × 128 × 128. The entire mouse brain is about the size of the human caudate nucleus head. (High resolution version of this image is available in the electronic supplementary material.)

FIG. 2
figure 2

7 Teslar (T) (300 MHz) (a, c) versus 16.4T (700 MHz) (b, d) images of the same live mouse. Spin echo [turbo-rapid acquisition relaxation enhancement (RARE)] images of 1-mm-thick slices were obtained using transmission time (TR) = 4 s, echo time (TE) = 14.7 ms, RARE factor = 4, number of averages = 8, total scanning time = 34 min. The in-plane isotropic image resolution is 59 μm/pixel. Note the much higher signal-to-noise ratio (SNR) in the 16.4T MR images (b, d), as well as stronger susceptibility artifacts in (b). (High resolution version of this image is available in the electronic supplementary material.)

Table 1 The Scaling of Key Physical Parameters with the Strength of the Main Magnetic Field B 0.

Most investigators would like to obtain image sets of excellent resolution and high SNR within a very short space of time. Each of the above three components (SNR, resolution, imaging time) are interdependent. For example, the price one pays for a good SNR is frequently a lower resolution and/or longer acquisition time. Or, if both good SNR and resolution are the goals, then the imaging time usually is prolonged. Finding an acceptable balance may be a difficult task. Several parameters of the studied model should be considered when the goal is to optimize image acquisition.

Living organisms can be considered inhomogeneous samples, with compartmentalized organs containing solid, gaseous, and various consistencies of liquid (from fluid to gelatinous) substances. These different consistencies have different magnetic susceptibilities, and whereas this in general does not represent a major problem at lower field strength, it becomes a very important problem at the higher field strength. Magnetic susceptibility is a known source of artifacts. One of the best examples for susceptibility artifacts is the loss of brain signal near air-filled sinuses. Gradient echo-based sequences are especially prone to these artifacts. Several methods have been proposed to reduce these artifacts, including the GESEPI sequence [14, 15]. In gradient echo imaging, slice selection correction methods have also been tried and successfully used [16, 17]. For multislice studies, the use of phase–phase encoding is helpful in reducing susceptibility-related artifacts. Sequences that use RF refocused echoes (fast spin echo, or rapid acquisition with refocused echo) [18], which are variants of spin echo sequences, eliminate most of the susceptibility-related artifacts. However, in some applications, susceptibility-related artifacts might be of benefit. The blood oxygen level-dependent (BOLD) signal (described later in this review)—the basis of fMRI studies—and iron oxide-based negative contrast agents [superparamagnetic iron oxides (SPIOs)] are good examples for this.

Susceptibility-weighted imaging (SWI) is another novel method that takes advantage of different magnetic susceptibilities of tissues, and represents a new tool to enhance tissue contrast [1921]. This method is based on relative differences between the magnetic susceptibility in one tissue and the surrounding tissue or a background. At long echo times (TE), signals from two tissues that have different magnetic susceptibilities will be out of phase and obtaining the phase images is the crucial point that defines the contrast in these experiments [21]. SWI is a very convenient tool in assessing the iron content or any other substance that distorts the local field [22]. SWI was initially utilized mostly as a method for MR venography procedures [2326], but more recently this method has found its place in studying arterial venous malformations [27, 28], hemorrhagic lesions [29], brain tumors [30, 31], and multiple sclerosis (MS) [32].

COMMON MRI METHODS IN STUDIES OF CNS DISEASE MODELS

Anatomical/structural imaging

Several neurological disease models have been studied using small animal MRI systems. The greatest advantage of the numerous MRI methods utilized in small animal research is that they are performed noninvasively and allow anatomical, pathological, and functional information to be obtained repeatedly over time. One of the most basic applications is to generate 3D anatomical atlases of rodent brains. This has been accomplished by several groups [3336]. Similarly, MRI atlases describing mouse embryonic development have also been generated and reported [37]. A web-based application assembled by the California Institute of Technology allows insight into one of these databases [38, 39]. An example of high-resolution ex vivo anatomical imaging is shown in FIG. 3.

FIG. 3
figure 3

16.4T (700 MHz) MR microscopy of ex vivo mouse brain (right) compared to histological section stained with hematoxylin/eosin (left). Note the excellent anatomical resolution of the CA1 cell layer in the hippocampus and dentate gyrus on the right image extracted from an MRI 3D dataset. The high SNR and high isotropic resolution (55 μm/pixel) are achieved at the expense of long scanning time (5 h and 27 min). Other parameters: FOV, 14 × 14 ×  14 mm3; matrix size, 256 × 256 ×  256; TR = 400 ms; TE = 7 ms. (High resolution version of this image is available in the electronic supplementary material.)

MRI-based tractography

In basic neuroscience research, tract tracing via MRI is an important and still under-utilized application. While DTI-based techniques allow for fiber visualization based on physical properties of the studied CNS tissue, manganese-enhanced MRI (MEMRI) allows for investigations of actual tracts and their synaptic connections. This technique utilizes paramagnetic Mn++ ions. In MRI, Mn++ ions cause a shortening of the water T1 relaxation time with strong contrast enhancement in T1-weighted MRI [40]. Having similar chemical properties as Ca++, Mn++ ions are actively taken up by neurons via voltage-gated Ca++ channels [41]. Once inside the neurons, they can pass the Mn++ ions to each other via the synaptic junctures, thus allowing visualization of neural pathways by MRI [4245]. It is thought that Mn++ ions enter the cells due to the opening of receptor-activated calcium channels. Once inside the cell, Mn++ ions utilize axonal microtubule transport mechanisms to reach the axon terminal. During synaptic transmission, Mn++ is released from the presynaptic neuron. It is able to enter the postsynaptic neuron via calcium channels [43]. Thus, Mn++ ions will accumulate in activated chains of neurons, leading to increased signal intensity in activated parts of the CNS. If applied topically to areas of interest from the standpoint of their connectivity, such as cortical areas and olfactory neurons, or to the retina, then Mn++ can be used for tract tracing [4548]. Bilgen et al. [46] used MEMRI to visualize injected Mn++ ions in normal and hemisected spinal cords in rats. They also suggested MEMRI as a method for indirect assessment of axonal integrity and connectivity of corticospinal tract in injured rat spinal cords [47]. An example of MEMRI is shown in FIG. 4 [49].

FIG. 4
figure 4

Typical gray matter lesion in a rat model of hypoxic-ischemic injury (arrows in C and F) was observed in late phase by manganese-enhanced MRI (MEMRI) after Mn++ injection on day 7, which was not visible in the T2WI, T1WI, and ADC map (Reproduced with permission from J. Yang and E.X. Wu [49]). T2WI = T2-weighted MR image, T2WI = T2-weighted MR image, ADC = apparent diffusion coefficient. Reproduced with permission from the Institute of Electrical and Electronics Engineers (IEEE). (High resolution version of this image is available in the electronic supplementary material.)

Functional MRI studies

Functional MRI studies are also conducted in small rodents. fMRI studies are based on the BOLD principle. Blood flow and oxy/deoxyhemoglobin-related changes in activated areas of the cortex produce mild hypointensity on T2* images [5053]. Images obtained in the activated and nonactivated state are used to generate activity maps, including sensory or visual cortex mapping in animals [50, 54]. Mild hypocapnia may be helpful in establishing BOLD contrast [55]. Some rodent fMRI studies have also utilized SPIO/ultrasmall (US)PIO contrast materials to enhance signal loss at the activated areas [51]. These techniques also allow the measurement of cerebral blood volume [51, 56, 57]. One recent fMRI study was the first such study to successfully utilize thermal stimuli to distinguish the noxious and innocuous pathways in the rat spinal cord [58]. Another group demonstrated an fMRI method with simultaneous acquisition of rat brain and spinal cord images [59]. These researchers proposed that the simultaneous brain and spinal cord fMRI can be used not only in animal models, but also in several neurological disorders in humans, such as MS, movement disorders, brain trauma, and spinal cord injuries, among others.

Cell-specific and molecular MRI

Cell-specific and molecular MRI is now used in the study of several disease models [6062]. These techniques are important tools for establishing a better correlation between conventional histology and MRI methods. Many groups have reported success with immune cell-specific imaging, utilizing superparamagnetic contrast materials (SPIOs) that are either internalized by cells after in vitro incubation or bound to the cells by specific antibodies (FIG. 5) [60, 6266]. SPIOs are detectable through their property of generating strong susceptibility effects (shortening of T2* relaxation) as well as shortening of the T1 and T2 relaxation time. Since susceptibility artifacts are stronger at high field strengths, these contrast materials are especially suitable for small animal imaging. The SPIOs include standard SPIOs (size, usually 50–150 nm), USPIOs (size, 30–50 nm) [67] and micron-sized paramagnetic iron oxide (MPIO) (size, approaching or more than 1 μm) [68]. The first use of SPIOs as a contrast agent was in liver and splenic tumor imaging [69, 70]. As a cellular label, SPIOs are used for MR cell tracking, as in detecting the specific locations of inflammation [71] or macrophage infiltration [72] in experimental autoimmune encephalomyelitis (EAE), the most widely used murine model of MS. Subsequently, USPIOs with a longer blood half-life were developed and utilized for the imaging of macrophage infiltration in the brains of animals with EAE [7375]. Several groups have investigated the effects of particle size and iron concentration on SPIO and USPIO uptake rates. In comparison with USPIOs, macrophages and monocytes appear to have much a higher uptake preference for SPIOs and are more effectively labeled by the latter [76, 77]. Thorek et al. [78] tested the uptake rate of different sized SPIOs in non-phagocytic T-cells and showed that particles <300 nm were most efficient in cell labeling. Even though USPIOs provide less contrast enhancement and SPIOs may be more efficient and suitable for macrophage labeling, USPIOs are overall preferred due to 1) longer blood half life (>24 h) and 2) larger number of particles loaded per cell [79]. One potential application of similar USPIO-based labeling techniques is to follow the distribution of biologically active proteins [60]. More recently, McAteer et al. [80] reported a molecular imaging methodology utilizing MPIOs. By utilizing antibody conjugated MPIOs, they showed clear upregulation and activation of Vascular Cell Adhesion Molecule 1 in a model of acute brain inflammation. In addition to the above labeling methods, “smart” molecular imaging probes and “sensing” contrast agents that are activated at specific sites of interest by enzymatic modification represent an interesting avenue for future development [81, 82].

FIG. 5
figure 5

CD8+ T-cell labeling with ultrasmall superparamagnetic iron oxide (USPIO)-conjugated antibodies in Theiler’s murine encephalomyelitis virus infection. Left image: T2*-weighted image shows diffuse hypointensities in several areas of the cerebrum and cerebellum. Middle image: T1-weighted study shows faint hyperintense areas. Right image: composite image, using a mask generated by dividing the T1-weighted image matrix with the T2* weighted. Reproduced with permission from I. Pirko et al. [150]. (High resolution version of this image is available in the electronic supplementary material.)

MRI STUDIES OF COMMON CNS DISEASE CATEGORIES

Inflammatory diseases of the CNS

Inflammatory and demyelinating diseases of the CNS are frequently studied in animal models. The most common models include EAE, Theiler’s murine encephalitis virus infection TMEV), and toxic demyelination models. A successful in vivo DTI of the spinal cord in mice with EAE has been reported [83] and proposed as a potential surrogate marker of axonal and myelin damage in this model. The authors demonstrated that relative anisotropy was a very sensitive biomarker to detect white matter destruction in EAE-affected mice, whereas the ADC was not. Similar to human diseases, lesion formation can be monitored by T2–weighted sequences (FIG. 6). Moreover, classical high resolution T2–weighted MRI may be used to investigate volumetric changes in brains of EAE mice as well as to correlate with subsequent histopathological analysis [84]. The presence of deep gray matter T2 hypointensity was recently demonstrated in the TMEV model with strong functional correlations [85]. A combination of gadolinium (Gd)-enhanced T1– with T2– and T2*-weighted imaging was used to characterize a new animal model of an acute hemorrhagic leuko-encephalomyelitis, a rare and fatal neurological disorder in humans [86]. Toxin-induced demyelination is less frequently studied by MRI. T2W imaging together with diffusion-weighted imaging (DWI) has been demonstrated to be useful in studying patterns of corpus callosum demyelination in cuprizone fed mice, a toxic model of MS [87]. In the same toxic model, both the magnetization transfer ratio [88] and DTI [89] were proposed to serve as sensitive and reproducible surrogate markers in quantifying myelin loss and repair.

FIG. 6
figure 6

Demyelinating lesions in interferon-γ receptor knockout mice following TMEV infection. Note the high signal intensity areas (red arrows) in the brainstem and near the thalamus on these T2-weighted in vivo images. Reproduced with permission from I. Pirko et al. [150]. (High resolution version of this image is available in the electronic supplementary material.)

Gd enhancement is used to monitor the blood–brain barrier permeability associated with new lesion formation [90]. Measurements of brain atrophy are important for assessing the chronic aspects of many neurological diseases. Volumetric MRI techniques based on 3D acquisition sequences [90, 91] are the key components in the study of brain atrophy. DWI is also important in these models because this technique may show lesion formation even earlier than Gd contrast imaging [92]. Through the use of DTI methods, tract integrity is also studied [93]. Voxel-based MRS or chemical shift imaging studies are also used, mainly to assess axonal integrity and membrane turnover [91, 94].

Optic neuritis (ON) is an inflammatory demyelinating syndrome that may be seen as a standalone disease or as a typical component of MS and neuromyelitis optica. Due to the uniformity of the optic nerve fibers, ON models represent an obvious subject for DTI studies (FIG. 7) [95]. One research group demonstrated that in a mouse model of EAE-associated ON, autoimmune axonal injury correlated with a decrease in axial diffusivity, whereas demyelination correlated with changes in radial diffusivity, as measured by DWI [96]. This noninvasive method provides a means to test the efficacy of potential therapeutic agents.

FIG. 7
figure 7

Optic nerves (indicated by arrows) from the experimental autoimmune encephalomyelitis (EAE)-affected mouse appear hypointense (B) in relative anisotropy (RA) maps compared with the control (A). Decreased axial diffusivity (D) and increased radial diffusivity (F) in optic nerves from EAE-affected mice were observed in the expanded views of axial (C and D) and radial diffusivity (E and F) corresponding to the rectangles in (A) and (B). Reproduced with permission from Xu et al. [95]. (High resolution version of this image is available in the electronic supplementary material.)

Neurodegenerative disease models

In the study of neurodegenerative disease models, morphometric studies are used for atrophy measurements. Both global brain atrophy measurements as well as focal volume loss of brain structures are assessed using MRI [97100]. One example is the recently reported cerebellar cortical atrophy in an EAE model [85].

The pathology of Alzheimer’s disease (AD) has been very difficult to study with MRI. However, through the use of novel contrast agents, one group has reported good success in imaging AD plaques ex vivo [101, 102]. More recently, even without contrast agents, plaque imaging in small rodent models has become possible both ex vivo [103] and in vivo (FIG. 8) [104106]. Because amyloid plaque deposits contain iron, it was considered that T2* (GE) sequences would be best in visualizing the plaques, which in general are considered to be more sensitive to iron deposits than to spin echo-based T2–weighted sequences. However in MS-related iron deposition of deep gray nuclei, T2–weighted—and not T2*-weighted sequences—have been successful in capturing the phenomenon in both the human disease and one of its animal models [107, 108]. A recent comprehensive ex vivo study of a transgenic mouse model of AD provides an excellent comparative review of several MRI sequences, including T1W, T2W, T2*W, proton density, and SWI, and establishes the foundation for future in vivo studies [109].

FIG. 8
figure 8

12–month-old Alzheimer’s disease (AD) mouse. In vivo (a, b) and ex vivo (c, d) T2SE-weighted MR images, thioflavin-S-stained (e, f) and iron-stained (g, h) histological brain sections have been precisely spatially registered over a circumscribed area of the cortex, indicated by the box. The boxes in the right column (scale bar, 100 μm) represent 3× magnified portions of the adjacent parent image in the left column (scale bar, 1.0 mm). The numbered arrows indicate individual plaques visualized in each of the four different image types that matched with the linked-cursor system. Reproduced with permission from C.R. Jack [106]. (High resolution version of this image is available in the electronic supplementary material.)

Animal models for amyotrophic lateral sclerosis (ALS) are available, and MRI is becoming an attractive tool to study this disease. The most commonly studied animal model of ALS is a transgenic mouse that overexpresses a human G93A-SOD1 mutation [110]. Using this mouse model, two different groups have successfully used T2–weighted MRI to visualize the neuronal degeneration of brain stem motor nuclei as well as to evaluate the disease progression [111113]. More recently, DWI, specifically ADC studies have been suggested as an additional marker for the time-course evaluation of disease progression in ALS mice [114]. Another group used a model in which a cycad toxin introduced via the diet causes a neurodegeneration which resembles a human ALS–Parkinsonism dementia complex [115]. This group demonstrated that volumetric MR microscopy is a sensitive tool for detecting early neurodegeneration, and the result correlated well with histopathology findings and behavioral deficits.

Iron deposition as a component in the pathology of neurodegenerative diseases is also studied by MRI. Changes suggestive of iron deposition are also seen in certain MS models, as recently reported by our team of investigators (FIG. 9) [85]. Since iron is known to cause susceptibility-related artifacts and high field strength imaging is more prone to these artifacts (see above), small animal MRI is very suitable to study this phenomenon [99].

FIG. 9
figure 9

Example of progressive thalamic T2 hypointensity in a virally induced multiple sclerosis (MS) model at 1 (A), 4 (B), 6 (C), and 12 (D) months postdisease induction. A–D represent coronal slices extracted from the original 3D datasets. Note the hypointensity of the mediodorsal thalamic nucleus and the increasing hypointensity at the later time points (arrows). (High resolution version of this image is available in the electronic supplementary material.)

Stroke models

Experimental stroke models in small rodents are frequently studied with MRI methods. These generally require classic MR imaging modalities, including T1, T2, and proton density imaging. Newer techniques include diffusion and perfusion imaging and MRS for determining certain metabolites [116, 117]. Many groups have used experimental middle cerebral artery ligation combined with hypoxia in rats or mice as models for human stroke (FIG. 10) [118, 119]. DWI sequences are especially sensitive to early ischemia, and many studies have utilized this resource in animal imaging, along with perfusion-weighted imaging, which is capable of visualizing the penumbra—the potentially salvageable area at risk [117, 120, 121]. Most recently, a novel mouse model of transient ischemic attack has been developed, as confirmed by DWI and T2–weighted MRI [122].

FIG. 10
figure 10

ADC, T2, and T1 maps show the severity and growth of the lesion in the mouse stroke model induced by a transient middle cerebral artery occlusion (reproduced with permission from P.A. Barber et al. [119]). Images on the top row were obtained 50–60 min after reperfusion, and images on the bottom row are of similar slices 24 h later. (High resolution version of this image is available in the electronic supplementary material.)

When DWI is performed using a fast imaging sequence like echo-planar imaging, the temporal resolution of scanning can be as low as every 20–30 s/entire brain scan, allowing close monitoring of stroke development. MRS studies have been conducted with the aim of looking at the lactate peak, which becomes prominent when anaerobic glycolysis predominates [116, 117, 123]. Another important aspect is the N-acetyl-aspartate peak that may show a decrease in completed strokes, but may remain normal if the animals still can recover from the ischemic event [124]. Magnetic resonance angiography [125] and perfusion-weighted imaging [57] have been performed and evaluated in several stroke models. These studies accurately characterize the cerebral blood flow and blood volume during ischemia.

Experimental intracerebral hemorrhages in animal models have been developed and only recently studied by MRI. In a warfarin-induced intracerebral hemorrhage model, T2* MRI was shown to almost perfectly correlate with the hematoma size observed from cryo-sections [126]. Another group used direct infusion of autologous blood in a rat striatum as a model and demonstrated that MRI also correlated with histopathological features [127]. Taken overall, these results suggest that MRI is very useful as a tool to evaluate functional outcomes after treatments as well as to elucidate their mechanism of action.

Traumatic injury models of the brain and spinal cord

Several models of traumatic brain injury have been developed in experimental animals. These range from diffuse injuries to controlled cortical impacts, chemical insults, and nonimpact models. MRI allows monitoring of the damaged tissue, also the tissue at risk [12, 128, 129]. For this purpose, DWI/DTI/magnetization transfer (MT) techniques are often used, similarly to the stroke studies described above. One group has provided evidence that DTI signal abnormalities correlate with electron microscopic and histological features of pericontusional injury. Based on these findings, these researchers propose DTI as an important tool to validate these injuries and predict approximate time since trauma [130, 131] (FIG. 11). Another group utilized a rat model and demonstrated that DWI was useful for the in vivo examination of brain edema following traumatic brain injury, with or without secondary hypoxia [132]. MRS may also be utilized for the study of tissue injury in the CNS [133, 134].

FIG. 11
figure 11

Diffusion tensor imaging of normal mouse brain (top row) and mouse brain 24 h following TBI (bottom row) (Reproduced from C.L. Mac Donald et al. [131]. Relative anisotropy and axial diffusivity show a gradient of signal changes within the red outlined area on post-TBI as compared to normal brain images. Other images showing radial diffusivity, trace diffusion and T2-weighted images are similar in the normal and injured brain. CTL = Control, TBI = Traumatic brain injury. (High resolution version of this image is available in the electronic supplementary material.)

Spinal cord injury is most commonly studied in rats. Traditionally, spinal cord integrity and neural circuitry either under normal conditions or after injury are studied by using histology and immunohistochemistry [135137]. During the last decade several MRI techniques have been utilized for in vivo research in spinal cord injury models. Diffusion anisotropy MRI has been suggested as an approach for quantitative assessment of secondary degeneration and recovery after spinal cord injury in rats [138]. Moreover, a good correlation between functional MRI activation in the rat spinal cord with neuronal activity determined by immunohistochemistry has been reported [139]. In addition, MEMRI may be a sensitive in vivo method for assessing neuronal functionality after spinal cord injury and eventual treatment outcomes. The main limitation in using MEMRI in vivo is the dose-dependent neurotoxicity of manganese [140].

Brain tumor models

Animal models of brain tumors are also frequently studied by MRI and MRS (FIG. 12) [141143]. The basic T1– and T2–weighted techniques may not always allow an obvious delineation between healthy and tumor infiltrate. Thus, advanced MRI methods and cell-labeling techniques are needed for the identification of transplanted tumor cells and growth monitoring. A combination of several MRI methods may enable visualization of human neural stem cells targeting the brain tumor [144].

FIG. 12
figure 12

MR imaging in a mouse model of a brain tumor showing a different outcome after two treatments (Reproduced with permission from L.K. Phuong et al. [143]. Panels (A) and (D) are brain MRI scans before the treatment, whereas (B) and (E) are the MRI scans 18 days after the different treatments. (B) Note the complete regression of a tumor treated with measles virus genetically engineered to produce carcinoembryonic antigen (MV-CEA) with a visible needle trace. (C) A stained (hematoxylin/eosin; H&E) brain section from the same mouse shows no residual tumor cells except for the glial scar and macrophages (arrows) (×200). (E) Mouse treated with UV-inactivated MV-CEA shows a profound tumor growth. (F) H&E stain from the same mouse’s brain shows a hypercellular U87 tumor with frequent mitotic figures (arrows) (×200). (High resolution version of this image is available in the electronic supplementary material.)

The use of intravenous contrast materials shows a blood–brain barrier breakdown in certain tumor models. MRS studies (proton and phosphorous) are also used to study biochemical differences between cancerous and normal tissue in vivo. Diffusion and perfusion studies together with blood flow and BOLD studies may provide useful information because tumor cells generally show increased metabolism [145, 146]. A recent study described a new radiation necrosis mouse model [147]. This important study provided milestones in the development of MRI tools to distinguish radiation changes from tumor recurrence. Chemical shift imaging is a multi-voxel MRS method that may be particularly useful in studying the various brain diseases [90]. In addition, a novel method of heteronuclear imaging has emerged. Kato et al. [148, 149] utilized 1H/13C magnetic resonance spectroscopic imaging to noninvasively analyze the delivery pattern and intratumoral distribution of the 13C-labeled anti-cancer drug temozolomide, which is commonly used in treatment of several brain tumors.

CONCLUSIONS

Over the last several years, microscopic resolution in vivo MRI has become a popular and versatile tool in the study of CNS diseases. MR image resolution of tens of microns in all three dimensions can now be achieved for in vivo studies, although for most practical applications, an isometric resolution of approximately 100 μm is sufficient. Because of the obstacles encountered with physiological microscopic movements of tissues due to circulation, breathing, and fluid flow, significantly higher resolutions are probably unrealistic for in vivo applications; the numerous inherent artifacts related to high field strength imaging have also to be kept in mind. New developments from the fields of neutron magnetic resonance and clinical MRI research will find their way to small animal experimental MRI research. Since a growing number of human CNS diseases have MRI-based diagnostic criteria, translational MRI research utilizing rodent models allows insight into important substrates of tissue dysfunction related to these clinically relevant MRI findings. The increased versatility that microscopic resolution MRI provides will result in enhanced understanding of critically important biological processes and in the development of new diagnostic and treatment approaches in a variety of human CNS diseases.