Abstract
Quality assurance of radiotherapy linear accelerator beams demands the use of equipment with high resolution and reliability of reproducible results over time. This study examines the suitability of a commercially available two-dimensional ionization chamber array—the StarCheck array (PTW, Frieburg, Germany) to measure symmetry and flatness for both photon and electron beams. The study was conducted over a period of 4 years whereby the reliability of the array could be established. The reproducibility, uniformity of chamber response, and comparison of both photon and electron profiles acquired with the StarCheck array to that of the water-tank were examined. The most significant result was that across all profiles acquired using the StarCheck array, a defective chamber was detected, manifested by a ‘dip’ in the cross-plane profile at the same position. Assuming all detectors are functional, StarCheck was shown to be within 1% of the water-tank results for both flatness and symmetry. The detector array exhibited many calibration and detector issues over the period in which it has been used in the department. Furthermore, PTW recommends that recalibration of the array should to be performed every 2 years at either PTW, Freiburg, or at a qualified calibration laboratory. Therefore, the department requires a backup device for the recalibration time, approximately 4 months as worst case.
References
Huq MS, Fraass BA, Dunscombe PB, Gibbons JP et al (2016) AAPM TG 100 report: application of risk analysis methods to radiation therapy quality management. Med Phys 43(7):4209–4262
Xu L, McEwen M (2009) Commissioning of a PTW StarCheck 2-D ion chamber array. Med Phys 36(6):15. https://doi.org/10.1118/1.3181866
Severgnini M, Aslian H, Denaro MD (2016) PTW StarCheck 2D array for quality control in IOERT: an evaluation of accuracy and dose consumption. Radiother Oncol 119(supplement 1):s733–s734. https://doi.org/10.1016/S0167-8140(16)32831-6
Simon TA, Kozelka J, Simon WE, Kahler D, Li J, Liu C (2010) Characterization of a multi-axis ion chamber array. Med Phys 37(11):6101–6111. https://doi.org/10.1118/1.3505452
Ritter TA, Gallagher I, Roberson PL (2014) Using a 2D detector array for meaningful and efficient linear accelerator beam property validations. J Appl Clin Med Phys 15(6):46–58
Klein EE, Hanley J, Bayouth J, Yin F, Simon W et al (2009) AAPM TG 142 report: quality assurance of medical accelerators. Med Phys 36(9):4197–4212
Xu L, McEwen M, Cojocaru C, Faddegon B (2009) Measurement of lateral dose distributions using GafChromic EBT films and PTW StarCheck 2-D array. Med Phys 36(6Part16):2624. https://doi.org/10.1118/1.3181925
International Electrotechnical Commission (1989) IEC publication 976: Medical electron accelerators - functional performance characteristics
Fogliata A, Garcia R, Knӧӧs T, Nicolini G et al (2012) Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy. Med Phys 39(10):6455–6464
Khan FM, Doppke KP, Hogstrom KR, Kutcher GJ et al (1991) Clinical electron-beam dosimetry: Report of AAPM Radiation Therapy Committee Task Group No. 25. Med Phys 18(1):73–109
Acknowledgements
The authors would like to thank Eva Sau Fan Liu and Patrick O’Connor for their help in acquiring some of the water-tank measurements for this study.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors have no conflicts of interest to declare.
Ethical approval
This article does not contain any studies with human participants or animals performed by any of the authors.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Ibrahim, S., Seshadri, V. & Charles, P. Assessing the suitability of a two-dimensional array for routine quality assurance checks of flatness and symmetry. Phys Eng Sci Med 43, 1451–1460 (2020). https://doi.org/10.1007/s13246-020-00932-w
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13246-020-00932-w