Evaluation of a Gafchromic EBT2 film dosimetry system for radiotherapy quality assurance

  • Scientific Paper
  • Published:
Australasian Physical & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

This study examines the dosimetric accuracy of Gafchromic EBT2 model radiochromic film for use in radiotherapy quality assurance. In this study, film was scanned using an Epson Perfection V700 flatbed scanner in transmission mode at 75 DPI with the subsequent analysis performed using the red and blue colour channels and ImageJ software. Results of this study suggest that the conversion of film optical density to measured dose should, at present, utilise red channel data only, without application of a blue channel correction to the data. For the batch of film examined here, film uniformity and reproducibility appear to have improved compared with published results using older batches. The orientation of the film on the scanner and the side of the film facing the light source were found to have substantial effects on results. Based on the results of this study, it is possible to recommend the use of EBT2 film in routine quality assurance testing for radiotherapy, in situations where a dose uncertainty of up to 2.8% is acceptable.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Fuss M, Sturtewagen E, De Wagter C, Georg D (2007) Dosimetric evaluation of GafChromic EBT film and its implication on film dosimetry quality assurance. Phys Med Biol 52:4211–4225

    Article  PubMed  Google Scholar 

  2. Garcia-Garduno OA, Celis MA, Larraga-Gutierrez JM, Moreno-Jimenez S, Martinez-Davalos A, Rodruiguez-Villafuerte M (2008) Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film. J Appl Clin Med Phys 9(3):90–98

    Google Scholar 

  3. Cosica G, Vaccara E, Corvisiero R, Cavazzani P, Ruggieri FG, Taccini G (2009) Fractionated stereotactic radiotherapy: a method to evaluate geometric and dosimetric uncertainties using radiochromic films. Med Phys 36(7):2870–2880

    Article  Google Scholar 

  4. Ho AK, Gibbs IC, Chang DD, Main B, Adler JR (2007) The use of TLD and gafchromic film to assure submillimeter accuracy for image-guided radiosurgery. Med Dosimetry 33(1):36–41

    Article  Google Scholar 

  5. Wilcox EE, Daskalov GM (2007) Evaluation of Gafchromic EBT film for CyberKnife dosimetry. Med Phys 34(6):1967–1974

    Article  PubMed  CAS  Google Scholar 

  6. Lagerwaard FJ, Van Der Hoorn EAP, Verbakel WFAR, Haasbeek CJA, Slotman BJ, Senan S (2009) Whole-brain radiotherapy with simultaneous integrated boost to multiple brain metastases using volumetric modulated arc therapy. Int J Rad Oncol Biol Phys 75(1):253–259

    Article  Google Scholar 

  7. Zeidan OA, Stephenson SAL, Meeks SL, Wagner TH, Willoughby TR, Kupelian PA, Langen KM (2006) Characterisation and use of EBT radiochromic film for IMRT dose verification. Med Phys 33(11):4064–4072

    Article  PubMed  Google Scholar 

  8. Fiandra C, Ricardi U, Ragona R, Anglesio S, Giglioli FR, Calamia E, Lucio F (2006) Clinical use of EBT model Gafchomric film in radiotherapy. Med Phys 33(11):4314–4319

    Article  PubMed  Google Scholar 

  9. Butson MJ, Cheung T, Yu PKN (2005) Absorption spectra variations of EBT radiochromic film from radiation exposure. Phys Med Biol 50:N135–N140

    Article  PubMed  CAS  Google Scholar 

  10. Schneider F, Polednik M, Wolff D, Stell V, Delana A, Wenz F, Menegotti L (2008) Optimisation of the Gafchromic EBT protocol for IMRT QA. Zeitschrift fur Medizinische Physik 19:29–37

    Article  Google Scholar 

  11. Battum LJ, Hoffmans D, Piersma H, Heukelom S (2008) Accurate dosimetry with Gafchromic EBT film of a 6 MV photon beam in water: what level is achievable? Med Phys 35(2):704–716

    Article  PubMed  Google Scholar 

  12. Saur S, Frengen J (2008) Gafchromic EBT film dosimetry with flatbed CCD scanner: a novel background correction method and full dose uncertainty analysis. Med Phys 35(7):3094–3101

    Article  PubMed  Google Scholar 

  13. Devic S, Seuntjens J, Sham F, Podgorsak EB, Schmidtlein CR, Kirov A, Soares CG (2005) Precise radiochromic film dosimetry using a flat-bed document scanner. Med Phys 32(7):2245–2253

    Article  PubMed  Google Scholar 

  14. Lynch BD, Kozelka J, Ranade MK, Li JG, Simon WE, Dempsey JF (2006) Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT GAFCHROMIC film. Med Phys 33(12):4551–4556

    Article  PubMed  Google Scholar 

  15. Martisikova M, Ackermann B, Jakel O (2008) Analysis of uncertainties in Gafchromic EBT film dosimetry of photon beams. Phys Med Biol 53:7013–7027

    Article  PubMed  Google Scholar 

  16. Bouchard H, Lacroix F, Beaudoin G, Carrier J-F, Kawrakow I (2009) On the characterisation and uncertainty analysis of radiochromic film dosimetry. Med Phys 36(6):1931–1946

    Article  PubMed  Google Scholar 

  17. ISP (2010) Gafchromic EBT2 self-developing film for radiotherapy dosimetry. http://online1.ispcorp.com/_layouts/Gafchromic/content/products/ebt2/pdfs/GAFCHROMICEBT2TechnicalBrief-Rev1.pdf (Accessible from http://gafchromic.com)

  18. ISP (2010) Gafchromic EBT2 scan handling guide. http://online1.ispcorp.com/_layouts/Gafchromic/content/products/ebt2/pdfs/GAFCHROMICEBT2ScanHandlingGuide1.pdf (Accessible from http://gafchromic.com)

  19. Butson MJ, Cheung T, Yu PKN, Alnawaf H (2009) Dose and aborption spectra response of EBT2 Gafchromic film to high energy x-rays. Aust Phys Eng Sci Med 32(4):196–202

    Article  CAS  Google Scholar 

  20. Butson MJ, Cheung T, Yu PKN (2008) Measuring energy response for RTQA radiochromic film to improve quality assurance procedures. Aust Phys Eng Sci Med 31(3):203–206

    Article  CAS  Google Scholar 

  21. Rink A, Vitkin AI, Jaffray DA (2007) Intra-irradiation changes in the signal of polymer-based dosimeter (Gafchromic EBT) due to dose rate variations. Phys Med Biol 52:N523–N529

    Article  PubMed  Google Scholar 

  22. Cheung T, Butson MJ, Yu PKN (2006) Measurement of high energy x-ray beam penumbra with Gafchromic EBT radiochromic film. Med Phys 33(8):2912–2914

    Article  PubMed  Google Scholar 

  23. Arjomandy B, Tailor R, Anand A, Sahoo N, Gillin M, Prado K, Vicic M (2010) Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies. Med Phys 37(5):1942–1947

    Article  PubMed  CAS  Google Scholar 

  24. Butson MJ, Yu PKN, Cheung T, Alnawaf H (2010) Energy response of the new EBT2 radiochromic film to x-ray radiation. Rad Meas 45:836–839

    Article  CAS  Google Scholar 

  25. Sutherland JGH, Rogers DWO (2010) Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film. Med Phys 37(3):1110–1116

    Article  PubMed  CAS  Google Scholar 

  26. Richley L, John AC, Coomber H, Fletcher S (2010) Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapy. Phys Med Biol 55:2601–2617

    Article  PubMed  CAS  Google Scholar 

  27. Paelinck L, De Neve W, De Wagter C (2007) Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry. Phys Med Biol 52:231–242

    Article  PubMed  CAS  Google Scholar 

  28. Kairn T, Aland T, Kenny J (2010) Local heterogeneities in early batches of EBT2 film: a suggested solution. Phys Med Biol 55:L37–L42

    Article  PubMed  CAS  Google Scholar 

  29. ISP (2010) Correction Protocol for Gafchromic EBT2 Dosimetry Film. http://online1.ispcorp.com/_layouts/Gafchromic/content/products/ebt2/pdfs/GAFCHROMICEBT2CorrectionProtocol-Rev1.pdf (Accessible from http://gafchromic.com)

  30. Kairn T, Kenny J, Crowe SB, Fielding AL, Franich RD, Johnston PN, Knight RT, Langton CM, Schlect D, Trapp JV (2010) Technical note: modelling a complex micro-multileaf collimator using the standard BEAMnrc distribution. Med Phys 37(4):1761–1767

    Article  PubMed  CAS  Google Scholar 

  31. Johnson RA, Bhattacharyya GK (2006) Statistics: principles and methods. Wiley, New York

    Google Scholar 

  32. Butson MJ, Cheung T, Yu PKN (2009) Evaluation of the magnitude of EBT Gafchromic film polarisation effects. Aust Phys Eng Sci Med 32(1):21–25

    Article  CAS  Google Scholar 

  33. Cheung T, Butson MJ, Yu PKN (2005) Post-irradiation colouration of Gafchromic EBT radiochromic film. Phys Med Biol 50:N281–N285

    Article  PubMed  Google Scholar 

  34. Hartmann B, Martisikova M, Jakel O (2010) Technical note: homogeneity of Gafchromic EBT2 film. Med Phys 37(4):1753–1756

    Article  PubMed  Google Scholar 

  35. Desroches J, Bouchard H, Lacroix F (2010) Technical note: potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry. Med Phys 37(4):1565–1570

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Dr Kairn’s contribution to this work was funded by the Wesley Research Institute, Australia (grant number 2008/11).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to T. Aland.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Aland, T., Kairn, T. & Kenny, J. Evaluation of a Gafchromic EBT2 film dosimetry system for radiotherapy quality assurance. Australas Phys Eng Sci Med 34, 251–260 (2011). https://doi.org/10.1007/s13246-011-0072-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13246-011-0072-6

Keywords

Navigation