An Approach Method to Calibrate the Autocollimator with Small Angle Measurement Range

  • Original Article
  • Published:
MAPAN Aims and scope Submit manuscript

Abstract

A new method to calibrate the autocollimator with small angle measurement range has been developed. This method used a system based on rotary encoder for angle measurements with self-calibration function (SelfA47). The SelfA47 has 18,000 graduation lines with angle interval of 0.02°, corresponds to 72 arcsec. Thereby this system cannot be used to calibrate the autocollimator with measurement range <72 arcsec. On the other hand, SelfA47 can be used to calibrate an electronic level. The deviation of electronic level shaped linear gradient. Therefore a small nominal gradient of electronic level can be approximated by interpolation. Since the gradient of a curve corresponds to the angle quantities, it is feasible to calibrate the autocollimator with nominal angle <72 arcsec using SelfA47 and an electronic level. In this proposed method, SelfA47 is placed horizontally on the same plane with electronic level and mirror reflector position. The electronic level value that have been corrected by SelfA47 is used as a reference. This reference value was converted to the angle, thus the angle deviation of autocollimator can be determined. The prominent uncertainty of angle measurement using this new technique derived from the uncertainty of electronic level and rotary encoder. Using this method, the expanded uncertainty of 0.36 arcsec is obtained from calibration of autocollimator with measuring range of ±20 arcsec and it is consistent statistically with the manufacturer calibration result shown by E n value is <1. In conclusion, this approach method can be applied to calibrate the autocollimator with a resolution smaller than SelfA47 as standard, but it is not recommended for the calibration of highly accurate autocollimator.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

References

  1. Walter Link, et al., Calibration Angles at the Level of Tenth of Seconds, XVIII IMEKO World Congress, Rio de Janeiro (2006).

    Google Scholar 

  2. T. Yandayan, et al., A novel technique for calibration of polygon angles with non-integer subdivision of indexing table, Precision Engineering, 26 (2002), 412–424.

    Article  Google Scholar 

  3. JCGM International vocabulary of metrology-Basic and general concepts and associated terms (VIM), 200, ISO (2008).

  4. Lung-Hen Chow, et al., Practical interpretation of metrological traceability on gauge block as to VIM 3, MAPAN-Journal of Metrology Society of India, 26 (2011), 79–86.

    Google Scholar 

  5. M. Arif Sanjid and K. P. Chaudhary, A novel multiple reflections technique to calibrate polygons and evaluation of its uncertainty of measurement, MAPAN-Journal of Metrology Society of India, 26 (2011), 29–35.

    Google Scholar 

  6. T. Watanabe, et al., Self-Calibratable Angle Device for Angular Standard, The 10th Int. Symp. of Meas. Tech. And Int. Inst. Daejeon, D3-1.1–4 (2011).

  7. M. Arif Sanjid, Improved direct comparison calibration of small angle blocks, Measurement, 46 (2013), 646–653.

    Article  Google Scholar 

  8. A. Just, et al., Calibration of high-resolution electronic autocollimators against an angle comparator, Metrologia, 40 (2003), 288–294.

    Article  ADS  Google Scholar 

  9. T. Watanabe, et al., High Resolution SelfA Rotary Table by the Interpolation Signal Calibration, ASPEN, Taipei (2013), 53-58.

    Google Scholar 

  10. T. Watanabe, et al., Research on the advancement of angle standard technology and the establishment of international comparison in ASEAN countries, NEDO (2011).

  11. JCGM Evaluation of measurement data—Guide to the expression of uncertainty in measurement, 100, ISO (2008).

  12. S. Yadav, et al., Least squares best fit line method for the evaluation of measurement uncertainty with electromechanical transducers (EMT) with Electrical Outputs (EO), MAPAN-Journal of Metrology Society of India, 25 (2010), 97-106.

    Google Scholar 

  13. Hewlett Packard, 5528A Laser Measurement System User’s Guide, 21-1–21-2, Hewlett Packard Co. (1988).

  14. O. Terra, Distance Scale Calibration of Optical Time Domain Reflectometer Using Active Intensity Modulation, MAPAN-Journal of Metrology Society of India, 30 (2015), 155-160.

    Google Scholar 

  15. Taylor Hobson, Calibration Sertificate of DA20 Autocollimator, 43930, UKAS (2008).

  16. Hala A. Mageed, et al., Interlaboratory Comparison in Measuring Low Levels of AC Voltage between Italy and Egypt, MAPAN-Journal of Metrology Society of India, 29 (2014), 175-181.

    Google Scholar 

  17. National Association for Proficiency Testing, Examining the Why’s and How’s of Proficiency Testing, (www.proficiency.org), Accessed on 18 Dec 2014.

Download references

Acknowledgments

The author would like to thank Dr. Tsukasa Watanabe and his team members who have provided SelfA47 in RCM-LIPI. Mr. A. Praba Drijarkara is kindly thanked for his advices from the beginning of this research work. Mrs. Eka Pratiwi and Mr. R. Rudi Anggoro Samodro are kindly thanked for all their help. The author would like to thank RCM-LIPI management for supporting this research work.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Okasatria Novyanto.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Novyanto, O., Alfiyati, N. An Approach Method to Calibrate the Autocollimator with Small Angle Measurement Range. MAPAN 31, 9–15 (2016). https://doi.org/10.1007/s12647-015-0153-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12647-015-0153-1

Keywords

Navigation