To the Editor,

Figure
figure 1

Example of a size 2.5 (left) and 3.0 (right) Shiley™ uncuffed tracheal tube (TT) (Covidien™, Mansfield, MA, USA). Each TT has insertion depth marks printed (in cm) along the length of the TT. In addition, each TT has two glottic depth marks printed on the distal end. The package inserts indicate the location (in mm) of the glottic depth marks A and B from the tip of the TT

Table Distance in airway length between the vocal cords and carina in infants and children measured from computed tomography and bronchoscopy

The purpose of this letter is to highlight a new marking convention appearing on the unit packages of uncuffed tracheal tubes (TTs) (Figure). In 1974, Loew and Thibeault1 described two marking conventions for uncuffed TTs, namely the black safety line (to be placed at the level of the vocal cords) and the insertion depth marks (to be documented at the lips or nares). In a study of 28 infants the TT tip was positioned in the middle third of the trachea, with this method preventing endobronchial intubation.

Subsequently, the International Organization for Standardization (ISO) mandated the marking of insertion depth marks at 1-cm intervals from the TT tip. Nevertheless, it was silent on the marking of the black safety line(s). Voluntarily, manufacturers marked the black safety line(s), but they did not standardize the marking convention. Worldwide, both neonatologists and anesthesiologists have identified this lack of standardization as a source of imprecision, confusion, and morbidity. In its 2016 revision of the medical device standard for TTs and connectors, ISO considered these concerns, renaming the black safety line as the glottic depth mark.Footnote 1 As superiority of one marking convention over another has not been proven, the revised ISO standard stipulates that the distance between the glottic depth mark and the TT tip be marked on the unit package, as shown in the Figure.

The correct use of the glottic depth mark requires knowledge of the changes in airway length during development. In 12 infants (aged zero to two years), Griscom and Wohl2 reported the airway length between the vocal cords and carina (LengthVC-C) to be 5.4 cm. In 56 infants and toddlers, Sirisopana et al.3 used their data to derive the following formula: LengthVC-C (cm) = 5.3 + 0.05 age (months); these values were consistent with those measured during bronchoscopy4,5 (Table).

In preterm infants, Loew and Thibeault1 marked the black safety line 2.2 cm from the tip of the smallest (internal diameter 2.5 mm) uncuffed TT. In a study of 114 fetal autopsies, Rigo and Fayoux6 measured the length of the airway between the vocal cords and the mid trachea and reported this length, in millimeters, to be 2.8 + 0.62 gestational age (weeks); estimates that were 4.3 mm shorter than those reported by Loew and Thibeault.

Flexion and extension of the neck can move the TT tip 0.5 to 1 cm in the caudad and cranial direction, respectively. In 2% of intubations, Mariano et al.7 reported that positioning the double black safety line of SheridanTM uncuffed TTs (Hudson RCI©; Temecula, CA, USA) at the level of the vocal cords, located the TT tip within 0.5 cm of the carina (observed via fluoroscopy). During neck flexion, such a low insertion depth risks endobronchial intubation. In 45% of the intubations, the TT tip was above the sternoclavicular junction (observed via fluoroscopy). During neck extension, such a high insertion depth risks accidental extubation.

Thus, the correct use of the glottic depth mark requires: 1) knowledge of the developmental changes in the airway length, 2) expert judgment to determine the optimal insertion depth, 3) technical expertise to optimally position the glottic depth mark, and 4) knowledge of the specific manufacturer’s TT markings on the unit package, all while documenting the corresponding insertion depth mark at the lip or nares.